For the administration of Nitroglycerin solution from a container into the patient's vascular system through a vascular access device.
Device Story
Single-use disposable IV administration set; designed for delivery of Nitroglycerin solution. Features vented spike (DUO-VENT), drip chamber, PVC pump segment, and trilayer tubing with inner polyethylene (PE) layer to prevent drug absorption/loss associated with standard PVC. Includes roller clamp, slide clamp, and optional injection sites (Interlink, Clearlink, or needle-accessed). Operated by clinicians in clinical settings to connect drug containers to patient vascular access devices. Modifications include updated PE material in tubing inner layer and transition to latex-free needle-accessed y-sites. Device ensures accurate drug delivery by maintaining Nitroglycerin concentration during infusion.
Clinical Evidence
Bench testing only. Evaluated functional performance (clear passage, roller clamp force/shut-off/leak, bond strength, air pressure, drug compatibility) and biocompatibility (ISO-10993-1 for prolonged duration, external communicating, indirect blood path). Tests included cytotoxicity, systemic toxicity, irritation, sensitization, hemocompatibility, and USP physiochemical testing. All results met acceptance criteria.
Technological Characteristics
Single-use IV administration set. Materials: non-DEHP drip chamber, PVC pump segment, trilayer tubing (inner PE layer), luer lock. Connectivity: None. Sterilization: Not specified. Software: None. Biocompatibility per ISO-10993-1.
Indications for Use
Indicated for the administration of Nitroglycerin solution from a container into the patient's vascular system through a vascular access device.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Nitroglycerine Administration Sets (Solution Admin. Set 2C9096 w/10 PVC) (K832284)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing to the right, stacked on top of each other. The text reads 'DEPARTMENT OF HEALTH & HUMAN SERVICES - USA'.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 16, 2015
Baxter Health Corporation c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K150860
Trade/Device Name: Nitroglycerin Set with DUO-VENT Spike Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Sets Regulatory Class: II Product Code: FPA Dated: March 30, 2015 Received: April 1, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) TBD
K150860
Device Name
Nitroglycerin Sets with DUO-VENT Spike
Indications for Use (Describe)
For the administration of Nitroglycerin solution from a container into the patient's vascular system through a vascular access device.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 5 - 510(k) Summary
April 14, 2015
### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
# CONTACT PERSON:
Dhiraj Bizzul Specialist, Global Regulatory Affairs 32650 N. Wilson Road Round Lake, IL 60073 Telephone: (224) 270-2177 Fax: (224) 270-4119
### DEVICE NAME:
Trade Name: Nitroglycerin Sets with DUO-VENT Spike
Common name: Intravascular (IV) Administration Sets
Classification Panel: 80 General Hospital
Classification: 21 CFR 880.5440
Class: Class II
Product Code: FPA
## Table 1. Representative Product Codes for Nitroglycerin Sets with DUO-VENT Spike
| Code Number | Name |
|-------------|------------------------------------------------------------------------|
| 1C8043 | Nitroglycerin Set with DUO-VENT Spike |
| 2C7551 | Nitroglycerin Set with DUO-VENT Spike (INTERLINK System) |
| 2C7552 | Nitroglycerin Set with DUO-VENT Spike (Needle-Accessed Injection Site) |
| 2C8851 | Nitroglycerin Set with DUO-VENT Spike (CLEARLINK System) |
{4}------------------------------------------------
# PREDICATE DEVICE:
| Device | Company | Predicate 510(k) | Clearance date |
|---------------------------------------------------------------------------------|----------------------------------|------------------|-------------------|
| Nitroglycerine Administration<br>Sets(Solution Admin. Set 2C9096 w/10<br>PVC).a | Travenol<br>Laboratories, S.A.b | K832284 | December 29, 1983 |
| Solution Sets/Extension Sets | Baxter Healthcare<br>Corporation | K142011 | August 18, 2014 |
### Table 2. Predicate 510(k)
4 Nitroglycerine Administration Sets have been renamed Nitroglycerin Sets with DUO-VENT Spike. K832284 was submitted to the FDA under the device name of Nitroglycerine Administration Sets but was cleared under the device name of Solution Admin. Set 2C9096 w/10 PVC.
6 Travenol Laboratories, S.A. became Baxter International Inc. in 1988
### DEVICE DESCRIPTION:
The proposed devices, which are the subject of this Traditional 510(k) premarket notification, consist of Nitroglycerin Sets with DUO-VENT Spike. They are single use disposable devices intended for the administration of Nitroglycerin solution from a container into the patient's vascular system through a vascular access device. These devices are the same as the current marketed devices, previously cleared under 510(k) premarket notification K832284 (cleared on December 29, 1983).
Baxter's Nitroglycerin Sets with DUO-VENT Spike are comprised of non-DEHP drip chamber with a vented spike. PVC tubing pump segment, polyethylene (PE) lined trilayer tubing, and a luer lock. On all sets there is a fixture slide clamp and an on-off roller clamp. Configurations of these sets differ in overall length, type of injection site (Interlink, Clearlink, needle-accessed, or none), and type of drip chamber (10 or 60 drops per minute).
Polyethylene lined tubing is used to administer Nitroglycerin due to the compatibility issues between the drug and polyvinyl chloride (PVC). Nitroglycerin concentration can be significantly reduced when infused using standard PVC administration sets, due to absorption into the tubing. For this reason, trilayer tubing with an inner polyethylene layer is used in Baxter's line of Nitroglycerin Sets with DUO-VENT Spike.
The basis for this premarket notification is a modification to the PE lined trilayer tubing used in Baxter's Nitroglycerin Sets with DUO-VENT Spike. The modification consists of a change to the Polyethylene material used in the inner layer of the PE lined trilayer tubing.
{5}------------------------------------------------
Along with this modification, the needle-accessed injection site (y-site) is changing from a latex-containing y-site to one that is not manufactured with latex. This proposed needle-accessed y-site is currently used in Baxter's Intravascular (IV) Administration Sets with y-site(s) and has been previously cleared under 510(k) premarket notification K142011 (cleared on August 18, 2014).
These modifications do not impact the intended use or the fundamental scientific technology of the device. No other materials of construction are being introduced into this device as part of this update. The product labels are also being updated to add the indications for use statement of the device, revise statements regarding latex and references to pump devices, and make any other changes to comply with Baxter's labeling standards.
## STATEMENT OF INTENDED USE:
For the administration of Nitroglycerin solution from a container into the patient's vascular system through a vascular access device.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed devices are equivalent to Baxter's current legally marketed device cleared under 510(k) premarket notification K832284 (cleared on December 29, 1983). The modification consists of a change to the Polyethylene material used in the inner layer of the PE lined trilayer tubing. In addition to this, the needle-accessed y-sites will be changing, which has been previously cleared under 510(k) premarket notification K142011 (cleared on August 18, 2014). The intended use, the basic design, and function for the proposed device are equivalent to the predicate device.
### DISCUSSION OF NONCLINICAL TESTS:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed devices are appropriately designed for their intended use.
{6}------------------------------------------------
# Performance Data:
The following bench tests were conducted to evaluate the effect of the PE lined trilayer tubing modification on the functional performance of the Nitroglycerin Sets with DUO-VENT Spike:
- Clear Passage Test .
- Roller Clamp Force Test ●
- Roller Clamp Shut-Off Test ●
- Roller Clamp Tubing Leak Test ●
- . Solvent Bond Tensile Strength Test
- Solvent Bond Air Pressure Test
- . Drug Compatibility Test
The following bench tests were previously conducted in K142011 (cleared on August 18, 2014) to evaluate the effect of the design modification on the functional performance of the IV Administration Sets with a needle-accessed y-site:
- Puncture Test .
- 7-Day Indwell Test ●
- Burst Test ●
- Vacuum Leakage Test ●
- Solvent Bond Tensile Strength Test ●
- . Solvent Bond Air Pressure Test
- Coring Test ●
- Insertion Force Test .
All tests met the acceptance criteria.
{7}------------------------------------------------
# Biocompatibility:
Biocompatibility assessments of the Nitroglycerin Sets with DUO-VENT Spike were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for prolonged duration, external communicating device, indirect blood path, and Blue Book Memorandum G95-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", as recommended in the IV Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]". The following tests were conducted:
- Cytotoxicity ●
- Systemic Toxicity ●
- Irritation/Intracutaneous .
- . Sensitization
- Hemocompatibility ●
- USP Physiochemical
Based upon the results of this prolonged duration, external communicating, indirect blood path testing, the Nitroglycerin Sets with DUO-VENT Spike have been shown to be biocompatible and appropriate for its intended use.
# CONCLUSION:
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in Table 2.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.