K150855 · Lima Corporate S.P.A. · MEH · Oct 21, 2015 · Orthopedic
Device Facts
Record ID
K150855
Device Name
H-MAX S Femoral Hip System
Applicant
Lima Corporate S.P.A.
Product Code
MEH · Orthopedic
Decision Date
Oct 21, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup. Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia; rheumatoid arthritis; osteoarthritis after femoral heads fractures; revisions in cases of good remaining femoral bone stock. The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction. The Limacorporate Delta TT Cup is intended for uncemented use.
Device Story
H-MAX S Femoral Hip System is a modular hip prosthesis for total hip arthroplasty. Components include a monolithic Ti6Al4V cementless femoral stem (macro-roughened surface with hydroxyapatite coating), modular femoral heads (CoCrMo or Biolox Delta ceramic), and acetabular cups (cemented UHMWPE or uncemented Delta TT titanium alloy). The system replaces the femoral head and acetabular socket. Surgeons select components based on patient anatomy and clinical needs. The device is implanted in an OR setting. The prosthesis restores hip joint function, reduces pain, and facilitates mobility. The modular design allows for various head diameters and neck lengths to optimize joint stability and biomechanics.
Clinical Evidence
No clinical data was necessary to demonstrate substantial equivalence. The submission relies on bench testing and design comparisons to previously cleared predicate devices.
Technological Characteristics
Femoral stem: Ti6Al4V (ISO 5832-3, ASTM F1472) with hydroxyapatite coating. Modular heads: CoCrMo (ASTM F1537) or Biolox Delta ceramic. Cemented cups: UHMWPE (ASTM F648, ISO 5834-2) with AISI 316-L stainless steel radiographic marker. Delta TT cups: Ti6Al4V (ASTM F1472, ISO 5832-3) manufactured via Electron Beam Melting (EBM) with Trabecular Titanium porous structure. Coupling: 12/14 Morse taper. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty for pain relief or improved function due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, hip dysplasia), rheumatoid arthritis, post-fracture osteoarthritis, or revisions with adequate femoral bone stock.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, representing the department's focus on health and human well-being. The profiles are depicted in a simple, abstract manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2015
Limacorporate S.P.A. Stephen Peoples Principal Consultant Via Nazionale 52 33038 Villanova Di San Daniele Udine-Italy
Re: K150855
Trade/Device Name: H-MAX S Femoral Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: MEH, JDI, KWY, LPH Dated: September 15, 2015 Received: September 22, 2015
Dear Mr. Peoples:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
#### Page 2 - Ms. Jayana Kenana
(21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# H-MAX S Femoral Hip System Indications for Use
The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.
Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis ●
- and hip dysplasia; ●
- rheumatoid arthritis; ●
- osteoarthritis after femoral heads fractures;
- revisions in cases of good remaining femoral bone stock. ●
The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction. The Limacorporate Delta TT Cup is intended for uncemented use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
Date: October 19, 2015
| <b>Manufacturer:</b> | <b>U.S. Contact Person:</b> |
|----------------------------------|-----------------------------|
| Limacorporate S.p.A. | Dr. Stephen J. Peoples |
| Via Nazionale, 52 | Principal Consultant |
| 33038 – Villanova di San Daniele | Phone: 260-645-0327 |
| Udine - Italy | FAX: +39 0432945512 |
| Product | Product Code | Regulation and Classification Name |
|-------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------|
| H-Max S femoral hip<br>system | MEH | Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented<br>prosthesis per 21 CFR 888.3353 |
| | JDI | Hip joint metal/polymer semi-constrained<br>cemented prosthesis per 21 CFR 888.3350 |
| | KWY | Hip joint femoral (hemi-hip) metal/polymer<br>cemented or uncemented prosthesis per 21 CFR<br>888.3390 |
| | LPH | Prosthesis, Hip, Semi-Constrained,<br>Metal/Polymer, Porous Uncemented per 21<br>CFR 888.3358 |
#### Description
The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.
The H-MAX S femoral stem is a monolithic, cementless stem made from Ti6Al4V (ISO 5832-3. ASTM F1472). The external surface has a macro-roughened surface and a layer of hydroxyapatite is applied along the length of the stem. The H-MAX S stem is manufactured in 11 sizes for each of two configurations, standard and lateralizing, that vary in the CCD angle.
The Biolox Delta ceramic head is a modular femoral ceramic head with a female 12/14 taper for coupling to the male 12/14 taper of the H-Max S femoral stem. The Biolox Delta ceramic heads are available in diameters of 28, 32, and 36 mm in three (3) neck lengths (S, M, L); the 36 mm
{4}------------------------------------------------
diameter head is also available with a XL neck length.
The Delta TT Cup is a hemi-spherical, 2-piece modular cup consisting of a metal shell and a polyethylene liner. The metal shell is manufactured using an EBM (Electron Beam Melting) process with titanium alloy powder (Ti6Al4V, ASTM F1472 - ISO 5832-3) producing a non-porous bulk interior surface with a Trabecular Titanium porous structure on the external surface. A polar threaded hole is used for attachment of an introducer instrument; the hole mates with the polar peg of the polyethylene liner when the liner is inserted into the cup. Three threaded holes in lateral positions allow additional fixation of the cup using bone screws; polyethylene plugs are used to fill the holes when bone screws are not required. The Delta TT Cup metal shell is coupled with a liner made of either conventional or cross-linked ultra high molecular weight polyethylene (UHMWPE) by means of a taper coupling. The Delta TT Cup metal shell is available in external diameters of 44, 46 and 48 mm, for use with Small liners; 50 and 52 mm, for use with Medium liners; and 54, 56, 58, 60, 62, 64, 66, 68, 70, 72, 74 and 76mm mm for use with Large liners. The polyethylene liners are available in three (3) sizes, Small, Medium and Large, with inner diameters of 28, 32, and 36mm.
#### CoCrMo Femoral heads
Femoral head devices are used by surgeons to replace the head of the femur during total hip surgery. They are characterized by a spherical shape and are coupled with the acetabular cup inserted in the acetabulum.
The femoral heads are coupled with the H-Max S stems by means of a 12/14 Morse taper. Modular heads are made of CoCrMo (ASTM F1537) and are available in diameters of 28, 32 and 36 mm and in S. M. L and XL offsets.
#### Cemented cups
Cemented cups are used to completely replace the acetabular cavity, using cement as the fixation agent. Cemented acetabular cups are manufactured entirely from UHMWPE (ASTM F648, ISO 5834-2). An AISI 316-L stainless steel ring is inserted in the equatorial region to allow the evaluation of the device position through radiography.
The acetabular cups are available in two different versions:
- · standard
- · protruded
The standard version is completely hemispherical and is designed for use in deep acetabula. The protruded version has a lateral portion which provides greater coverage of the femoral head to resist dislocation. The cemented cups are available with articular diameters of 28 and 32 mm.
The following table shows the availability of the cemented cups diameters for each type of cup.
{5}------------------------------------------------
| Cup Type | Inner diameter | Range of Outer Diameter |
|----------------|----------------|----------------------------------|
| Standard Cups | 28 mm | 44-58 mm with increments of 2 mm |
| | 32 mm | 46-58 mm with increments of 2 mm |
| Protruded Cups | 28 mm | 44-58 mm with increments of 2 mm |
| | 32 mm | 46-58 mm with increments of 2 mm |
### Intended Use/Indications
The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.
Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis
- and hip dysplasia;
- rheumatoid arthritis;
- osteoarthritis after femoral heads fractures; ●
- revisions in cases of good remaining femoral bone stock. ●
The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction. The Limacorporate Delta TT Cup is intended for uncemented use.
#### Predicate Devices
- H-MAX S Femoral Hip System (Limacorporate, K112091)
- Minima S System (Limacorporate, K141327)
- Master SL Femoral Stem (Limacorporate, K143509) 0
- Modulus Femoral Stem (Limacorporate, K112158) ●
- Delta TT Acetabular System (Limacorporate, K112898, K141395). 0
# Comparable Features to Predicate Device(s)
The subject H-MAX S Femoral Hip Stems are identical to the H-MAX S Femoral Hip Stems cleared in K112091. In addition:
- The subject CoCrMo modular heads (36 mm) are identical to CoCrMo modular heads (36 ●
{6}------------------------------------------------
mm) approved for use with Master SL femoral stem (K143509) and with Minima S femoral stem (K141327) in combination with Delta TT Acetabular cups;
- . The subject Biolox Delta heads (36 mm) are identical to the Biolox Delta Heads cleared in K141327 (Minima S) and K143509 (Master SL) in combination with Delta TT Acetabular cups;
- The subject CoCrMo heads (28, 32 mm) are identical to CoCrMo heads (28, 32 mm) ● cleared in K141327 (Minima S) and K143509 (Master SL) in combination with Delta TT Acetabular cups;
- The subject Biolox Delta heads (28, 32 mm) are identical to the Biolox Delta Heads ● cleared in K141327 (Minima S) and K143509 (Master SL) in combination with Cemented cups or Delta TT Acetabular cups.
# Non-Clinical Testing:
The materials, design, femoral head sizes and acetabular components remain the same as described in previously cleared submissions and no new issues of safety and effectiveness are introduced by adding:
- CoCrMo 0 36 mm modular heads coupled with Delta TT system for use with the Ho MAX S Femoral Hip System;
- Biolox Delta Ø 36 mm modular heads coupled with Delta TT system for use with the Ho MAX S Femoral Hip System;
- CoCrMo 0 28, 32 mm modular heads coupled with Delta TT system for use with the H-. MAX S Femoral Hip System.
- Biolox Delta Ø 28, 32 mm modular heads coupled with Delta TT system and Cemented ● cups for use with the H-MAX S Femoral Hip System.
The tests provided in previously cleared submissions are therefore applicable to the new couplings proposed to be used with the H-MAX S Femoral Hip Stem.
In detail:
- Coupling of H-MAX S Femoral Hip system with CoCrMo modular heads (36 mm) and ● Delta TT acetabular cups : the design, material, and sizes of the CoCrMo modular heads (36 mm) articulating with compatible Delta TT acetabular system are identical to the design, material, and sizes of CoCrMo modular heads (36 mm) cleared for use with the Minima S femoral stem (K141327) and the Master SL femoral stem (K143509) articulating with Delta TT acetabular system.
- Coupling of H-MAX S Femoral Hip system with Biolox Delta heads (36 mm) and Delta TT acetabular cups : the design, material, and sizes of the Biolox Delta Femoral Heads for use with the H-MAX S Femoral Hip System Stems are identical to the design, material, and
{7}------------------------------------------------
sizes of the Biolox Delta Femoral Heads cleared for use with the Minima S Hip System (K141327) and Master SL femoral stem (K143509) articulating with Delta TT acetabular system. The coupling between the Biolox Delta Head and the H-MAX S Femoral Hip Stem is achieved through the same taper used to couple the Biolox Delta ceramic head to the Minima Hip Stem (K141327) and Master SL femoral stem (K143509). The H-MAX S Femoral Hip Stem (K112091) is manufactured from the same material (Ti6Al4V allov) as the Minima S Hip Stem and Master SL femoral stem cleared for use with Biolox Delta ceramic heads ( in K141327 and K143509 respectively).
- Coupling of H-MAX S Femoral Hip system with CoCrMo heads (28, 32 mm) and Delta TT ● acetabular cups : CoCrMo modular heads sizes 28 and 32 in offset S, M, L, XL are identical in design, material and sizes to the ones cleared in the original 510(k) K112091, for the use with H-MAX S stem and cemented cups. These devices are coupled with cemented cups in the original submission of H-MAX S (K112091): the coupling of these heads with Delta TT cups has already been cleared with Master SL (K140975) and Minima S (K141327) stems.
- Coupling of H-MAX S Femoral Hip system with Biolox Delta heads (28, 32 mm) and Cemented cups / Delta TT acetabular cups : the design, material, and sizes of the Biolox Delta Femoral Heads for use with the H-MAX S Femoral Hip System Stems are identical to the design, material, and sizes of the Biolox Delta Femoral Heads cleared for use with the Minima S Hip System (K141327) and Master SL femoral stem (K143509) articulating with Cemented cups / Delta TT acetabular system. The coupling between the Biolox Delta Head and the H-MAX S Femoral Hip Stem is achieved through the same taper used to couple the Biolox Delta ceramic head to the Minima Hip Stem (K141327) and Master SL femoral stem (K143509). The H-MAX S Femoral Hip Stem (K112091) is manufactured from the same material (Ti6Al4V alloy) as the Minima S Hip Stem and Master SL femoral stem cleared for use with Biolox Delta ceramic heads ( in K141327 and K143509 respectively).
Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.