The Vyntus/SentrySuite Product Line is intended to be used for measurements, data collection and analysis of lung function (PFT) and cardio-pulmonary (CPET) parameters, aiding in the diagnosis of related conditions. The results of the test can be viewed online with the help of a computer screen and can be printed after the test. The test results can be saved for future reference or report generation purposes. The products can be utilized with patients age 4 years and older as long as they can cooperate in the performance - no special limit to patient's sex or height exists. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar setting (professional healthcare facilities). A qualified physician has to reassess all Vyntus/SentrySuite measurements. An interpretation by Vyntus/SentrySuite is only significant if it is considered in connection with other clinical findings. The Vyntus ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on the screen or printed on paper. 12-lead ECGs are analyzed automatically and suggestions for the interpretation of the resting ECG can be made by the software. ECG interpretation statements made by the Vyntus/SentrySuite represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements. The Vyntus ECG can be used for non-interpretive applications in patients age 4 years and older and a weight of 20 kg or higher. The Vyntus ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital). The Vyntus ECG is not intended for intracranial use. The Vyntus ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The Vyntus ECG is not intended for use in home healthcare environments. Automatic interpretation of the ECG is not possible for pediatric and adolescent patients below 16 years of age and for patients with pacemakers.
Device Story
Vyntus ECG is a 12-lead PC-ECG extension for Vyntus CPX; records electrical heart signals via surface electrodes; transmits data via Bluetooth to SentrySuite platform; used in hospitals/clinics by trained professionals under physician direction. Device performs A/D conversion, filtering, and automated analysis using Hannover ECG System (HES). Outputs include rhythm, trace, and full disclosure views; HR, ST-values, and ST-slope metrics; and automated interpretation suggestions. Healthcare providers use output to support cardiovascular assessment; requires physician reassessment; cannot be sole basis for therapy/drugs. Benefits include integrated lab workflow and routine ECG monitoring.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
PC-ABS housing; 12-lead ECG; Bluetooth connectivity; 1.5V AA or 1.2V NiMH battery; 500 Hz transmission rate; 4000 Hz pacemaker detection; 0.05-150 Hz bandwidth; <2.5 µV/bit resolution; CF applied parts, defibrillator proof. Software: SentrySuite platform with HES algorithm.
Indications for Use
Indicated for patients age 4 years and older (ECG non-interpretive: weight 20kg+) for lung function, cardiopulmonary, and 12-lead ECG measurements. Contraindicated for intracranial use, EMS environments, home healthcare, and automatic ECG interpretation for patients <16 years or with pacemakers.
Regulatory Classification
Identification
A pulmonary-function data calculator is a device used to calculate pulmonary-function values based on actual physical data obtained during pulmonary-function testing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an emblem that resembles an abstract representation of an eagle or a similar bird, with stylized feathers or wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 14, 2015
CareFusion Germany 234 GmbH Elmar Niedermeyer Consult, Regulatory Affairs Leibnizstrasse 7 Hoechberg, Bavaria 97204 Germany
Re: K150810
Trade/Device Name: Vyntus / SentrySuite Product Line Regulation Number: 21 CFR 868.1880 Regulation Name: Pulmonary Function Data Calculator Regulatory Class: Class II Product Code: BZC, DPS Dated: July 14, 2015 Received: July 17, 2015
Dear Elmar Niedermeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150810
Page 1 of 1
Device Name
Vyntus/SentrySuite Product Line
### Indications for Use (Describe)
The Vyntus/SentrySuite Product Line is intended to be used for measurements, data collection and analysis of lung function (PFT) and cardio-pulmonary (CPET) parameters, aiding in the diagnosis of related conditions. The results of the test can be viewed online with the help of a computer screen and can be printed after the test results can be saved for future reference or report generation purposes.
The products can be utilized with patients age 4 years and older as long as they can cooperate in the performance - no special limit to patient's sex or height exists. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar setting (professional healthcare facilities).
A qualified physician has to reassess all Vyntus/SentrySuite measurements. An interpretation by Vyntus/SentrySuite is only significant if it is considered in connection with other clinical findings.
### Additional for Vyntus ECG:
The Vyntus ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on the screen or printed on paper. 12-lead ECGs are analyzed automatically and suggestions for the interpretation of the resting ECG can be made by the software. ECG interpretation statements made by the Vyntus/ SentrySuite represent partial qualitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements.
The Vyntus ECG can be used for non-interpretive applications in patients age 4 years and older and a weight of 20 kg or higher. The Vyntus ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital). The Vyntus ECG is not intended for intracranial use. The Vyntus ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The Vyntus ECC is not intended for use in home healthcare environments. Automatic interpretation of the ECG is not possible for pediatric and adolescent patients below 16 years of age and for patients with pacemakers.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the CareFusion logo. The logo consists of an orange circle with a white shield inside, with a portion of the shield colored yellow. To the right of the circle is the text "CareFusion" in a dark gray sans-serif font. The logo is simple and modern, with a focus on the company name.
# 510(k) Summary
# GENERAL INFORMATION
#### Type of Submission 5.1
Traditional 510(k) Submission
Submission date: 16/03/2015
#### 5.2 Submitter
Name: CareFusion Germany 234 GmbH
- Address: Leibnizstrasse 7 D-97204 Hoechberg Germany
## Contact person in Germany:
| (Official Correspondent) | Elmar Niedermeyer |
|--------------------------|-----------------------------------|
| Address: | CareFusion Germany 234 GmbH |
| | Leibnizstrasse 7, 97204 Hoechberg |
| | Germany |
| Phone: | +49 931 49 72 - 361 |
| FAX: | +49 931 49 72 - 62361 |
| E-mail | elmar.niedermeyer@carefusion.com |
## Contact person in the U.S.:
| (U.S. Agent) | Donald Sherratt |
|--------------|--------------------------------|
| Address | CareFusion |
| | 22745 Savi Ranch Parkway |
| | Yorba Linda, CA 92887 |
| Phone: | 714-919-3349 |
| Fax: | 714-283-8420 |
| E-mail: | donald.sherratt@carefusion.com |
{4}------------------------------------------------
### 5.3 Establishment Registration Number 9615102
#### 5.4 Common Name or Classification Name
Pulmonary-function data calculator (CFR 868.1880, Product Code BZC) Electrocardiograph (CFR 870.2340, Product Code DPS)
### 5.5 Trade Name Vyntus/SentrySuite Product Line
#### 5.6 Device Classification
This is a Class II device
#### 5.7 Classification Panel
73 Anesthesiology Part 868 Code BZC (primary) 74 Cardiovascular Part 870 Code DPS (secondary)
#### 5.8 Reason for Premarket Notification
- new medical device .
- additional cart 3.0 to available cart 2.0
#### 5.9 Legally predicate marketed devices
- MasterScope ECG ● K082539 Code BTY, DPS
- Cor12+ K091505 Code DSH ●
- Vyntus / SentrySuite Product line K133925 Code BZC .
#### 5.10 Predicate Device Company
- Cardinal Health Germany 234 GmbH (now CareFusion Germany 234 GmbH)
- . Cardinal Health Germany 234 GmbH (now CareFusion Germany 234 GmbH)
- . CareFusion Germany 234 GmbH
#### 5.11 Device Description
Description & function:
The Vyntus ECG is the perfect 12-Lead PC-ECG extension for the Vyntus CPX via secure Bluetooth® communication. One integrated solution through the SentrySuite platform helps laboratories ease procedures and integration to reduce costs.
The Vyntus ECG records the full 12-lead resting and stress ECG via its wireless, battery operated ECG amplifier. The evaluation and interpretation is using the proven Hannover ECG System (HES).
{5}------------------------------------------------
## Scientific Concept:
The surface electrodes of the Vyntus ECG record the electrical signals that are mainly generated by the heart muscle activity and repolarization. The small voltages are amplified and A/D converted. A low frequency high-pass filter is applied and the recorded signals are transmitted to the PC for recording, storing, evaluation, interpretation, display and reporting.
Significant performance characteristics:
- Full 12 lead ECG O
- Small, low weight (<300 g) o
- Data transfer by blue-tooth O
- Single AA battery operation O
- Pacemaker detection with 4000 Hz sample rate O
- 500 Hz per channel transmission rate O
- Bandwidth: 0.05-150 Hz o
- Resolution: < 2.5 µV/bit O
- Classification applied parts; CF, defibrillator proof O
- Automatic analysis by HES stress/resting algorithm package O
- Online evaluation and display of HR, ST-values and ST-slope O
- O Arrhythmia detection
- Different views O
- Rhythm view
- Trace view ●
- Full disclosure view (result view) .
- Different filters selectable
- 50/60 Hz
- · Baseline
- Muscle (20, 40, 100 Hz) .
- Continuous storage of raw ECG data for stress application O
#### 5.12 Intended Use Statement
O
The Vyntus/SentrySuite Product Line is intended to be used for measurements, data collection and analysis of lung function (PFT) and cardiopulmonary (CPET) parameters, aiding in the diagnosis of related conditions. The results of the test can be viewed online with the help of a computer screen and can be printed after the test. The test results can be saved for future reference or report generation purposes.
The products can be utilized with patients age 4 years and older as long as they can cooperate in the performance - no special limit to patient's sex or height exists. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar setting (professional healthcare facilities). A qualified physician has to reassess all Vyntus/SentrySuite measurements. An interpretation by Vyntus/SentrySuite is only significant if it is considered in connection with other clinical findings.
### Additional for Vyntus ECG:
The Vyntus ECG is intended for measuring the surface electrocardiogram (ECG) of a patient. The acquired ECG can be recorded and displayed on the
{6}------------------------------------------------
screen or printed on paper. 12-lead ECGs are analyzed automatically and suggestions for the interpretation of the resting ECG can be made by the software. ECG interpretation statements made by the Vyntus/SentrySuite represent partial qualitative and quantitative information on the patient's cardiovascular condition and no therapy or drugs can be administered based solely on the interpretation statements.
The Vyntus ECG can be used for non-interpretive applications in patients age 4 years and older and a weight of 20 kg or higher. The Vyntus ECG is intended to be used for routine ECG collection, recording both under resting and stress conditions. The measurement is performed by trained healthcare professionals under the direction of a physician in healthcare facilities (e.g. the doctor's office or hospital). The Vyntus ECG is not intended for intracranial use. The Vyntus ECG is not intended for use in an EMS environment (Emergency Medical Services Environment). The Vyntus ECG is not intended for use in home healthcare environments. Automatic interpretation of the ECG is not possible for pediatric and adolescent patients below 16 years of age and for patients with pacemakers.
#### 5.13 Required Components
- PC or Notebook with Bluetooth interface o
- SentrySuite Software O
- Vyntus ECG amplifier O
- Bag for the Vyntus ECG amplifier o
- Instruction for Use o
- NiMH battery charger O
- o Accessories
- Disposable electrodes (Rest) ●
- Disposable electrodes (Stress)
- NiMH rechargeable battery 1,2V AA ≥ 2500 mAh ●
#### 5.14 Summary Table of Comparison
| Comparison with MasterScope ECG K082539 | | |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MasterSope ECG<br>(K082539) | Vyntus ECG<br>(proposed device) |
| indication<br>for Use | The MasterScope / MasterScope ECG is intended to<br>be used for measurement and data collection of lung<br>function parameters. The system performs<br>cooperation-dependent flow volume measurements.<br>Mostly it will be used for COPD and Asthma patients.<br>In addition it is intended for measuring a 3/6- or 12-<br>channel surface electrocardiogram (ECG) of a patient.<br>The acquired ECG can be recorded and displayed on<br>the screen or printed on paper. 12-channel ECG's are<br>analysed automatically and suggestions for the<br>interpretation of the 12-channel ECG can be made by<br>the software.<br>MasterScope / MasterScope ECG can be used for non<br>interpretive applications for patients with an age of 4<br>years and older and a weight of 20 kg or higher.<br>MasterScope / MasterScope ECG is intended for use<br>in routine ECG recording by trained physicians in the<br>office or hospital. MasterScope / MasterScope ECG is | The Vyntus/SentrySuite product line is intended to be<br>used for measurements, data collection and analysis<br>of lung function (PFT) and cardio-pulmonary (CPET)<br>parameters, aiding in the diagnosis of related<br>conditions. The results of the test can be viewed<br>online with the help of a computer screen and can be<br>printed after the test. The test results can be saved for<br>future reference or report generation purposes.<br>The products can be utilized with patients age 4 years<br>and older as long as they can cooperate in the<br>performance - no special limit to patient's sex or height<br>exists. Measurements will be performed under the<br>direction of a physician in a hospital environment,<br>physician's office or similar setting (professional<br>healthcare facilities). A qualified physician has to<br>reassess all Vyntus/SentrySuite measurements. An<br>interpretation by Vyntus/SentrySuite is only significant<br>if it is considered in connection with other clinical |
| | not intended for intracardial use. Automatic<br>interpretation of the ECG is not possible for pediatric<br>patients with an age below 16 years and for<br>pacemaker patients.<br>MasterScope CT (Clinical Trial version) incorporates<br>the identical measurements. In addition it offers<br>workflow control elements to restrict the use of the<br>equipment (e.g. individual access rights are defined for<br>different user roles like investigator, doctor, study<br>nurse, trainer and service personnel).<br>The interpretation software is intended to support the<br>physician in evaluation the ECG in terms of<br>morphology and rhythm.<br>A qualified physician has to reassess all MasterScope<br>/ MasterScope ECG measurements. An interpretation<br>by MasterScope / MasterScope ECG is only significant<br>if it is considered in connection with other clinical<br>findings. ECG interpretation statements made by the<br>MasterScope / MasterScope ECG represent partial<br>qualitative and quantitative information on the patient's<br>cardiovascular condition and no therapy or drugs can<br>be administered based solely on the interpretation<br>statements.<br>The MasterScope / MasterScope ECG / MasterScope<br>CT is powered from 100 - 240V / 50 - 60Hz wall<br>outlets. No energy is transferred to the patient. | findings.<br>Additional for Vyntus ECG:<br>The Vyntus ECG is intended for measuring the<br>surface electrocardiogram (ECG) of a patient. The<br>acquired ECG can be recorded and displayed on the<br>screen or printed on paper. 12-lead ECGs are<br>analyzed automatically and suggestions for the<br>interpretation of the resting ECG can be made by the<br>software. ECG interpretation statements made by the<br>Vyntus/SentrySuite represent partial qualitative and<br>quantitative information on the patient's cardiovascular<br>condition and no therapy or drugs can be administered<br>based solely on the interpretation statements.<br>The Vyntus ECG can be used for non-interpretive<br>applications in patients age 4 years and older and a<br>weight of 20 kg or higher. The Vyntus ECG is intended<br>to be used for routine ECG collection, recording both<br>under resting and stress conditions. The measurement<br>is performed by trained healthcare professionals under<br>the direction of a physician in healthcare facilities (e.g.<br>the doctor's office or hospital). The Vyntus ECG is not<br>intended for intracranial use. The Vyntus ECG is not<br>intended for use in an EMS environment (Emergency<br>Medical Services Environment). The Vyntus ECG is<br>not intended for use in home healthcare environments.<br>Automatic interpretation of the ECG is not possible for<br>pediatric and adolescent patients below 16 years of<br>age and for patients with pacemakers. |
| intended<br>Use | 12-channel<br>Surface ECG recording device | similar |
| target<br>population | 4 years and older and a weight of 20 kg or<br>higher | similar |
| application | ECG recording | similar |
| bandwidth | 0 - 150 Hz digital | similar |
| ECG Leads | 12 Standard | similar |
| A/D<br>resolution | 2,6 µV/bit | similar |
| pacemaker<br>detection<br>sample rate | 4000 Hz | similar |
| connection<br>to<br>electrodes | 4 mm snap connector, gold plated | similar |
| electrode<br>impedance | Impedance measurement for the electrode | similar |
| patient<br>contacting<br>accessory<br>(biocompati<br>bility) | Single use electrode (Ambu Blue<br>Sensor)Electrode cable | similar |
| material<br>ECG<br>amplifier | ABS/PC<br>(no patient contacting part) | ABS/PC white RAL 9003<br>(no patient contacting part) |
| enclosure | | |
| sterility | The device along with its accessories is<br>neither supplied sterile nor intended to be<br>sterilized | Similar |
| Evaluation<br>algorithm | Hannover ECG System (HES) | Similar |
| Interpretati<br>on<br>algorithm<br>(resting ECG) | Hannover ECG System (HES) | Similar |
| Comparison with Vyntus / SentrySuite Product Line K133925 | | |
{7}------------------------------------------------
{8}------------------------------------------------
| | Vyntus / SentrySuite Product Line<br>(K133925) | Vyntus ECG<br>(proposed device) |
|--------------------------------|---------------------------------------------------------------------------|----------------------------------------------------------------------|
| Software | · Software platform SentrySuite<br>SentrySuite Software version 2.11<br>● | · Software platform SentrySuite<br>SentrySuite Software version 2.13 |
| Comparison with Cor12+ K091505 | | |
| | Cor12+ | Vvntus ECG |
| | Cor12+<br>(K091505) | Vyntus ECG<br>(proposed device) |
|----------------|-----------------------------|---------------------------------------------------------|
| Interface | Bluetooth & SD card | Bluetooth |
| energy<br>used | 2 x 1,5V (AA size, Lithium) | 1,5V AA battery or 1,2V AA NiMH<br>rechargeable battery |
# Summary of technological characteristics compared to the predicate device to the tables above:
- The ECG part of the indications for use for the predicate MasterScope ECG ● and the indication for use of the Vyntus ECG are similar. Both devices record ECG's for the similar patient population.
- . The material from the housing in both devices is PC-ABS. Different is that the housing of the Vyntus ECG is in addition with the colour additive white RAL 9003.
- . The software platform used for the Vyntus ECG is similar to the predicate device Vyntus / SentrySuite Product line K133925. The software which is used is the SentrySuite software but here with a different version.
- . The interface communication of the Vyntus ECG and the predicate Cor12+ is realized over Bluetooth connection. Different is that the Vyntus ECG uses Bluetooth connection only whereby the Cor12+ can also be used with a SD card for data transfer. Both devices work as intended and the proposed device is substantial equivalent to the predicate.
- The energy source of the Vyntus ECG is similar to the predicate Cor12+. Both ● devices use a battery whereby the Vyntus ECG can be used with a rechargeable battery too.
{9}------------------------------------------------
#### 5.15 Summary of Device Testing
### 1. Non-clinical tests conducted for determination of substantial equivalence:
| Characteristic | Standard/Test | Results Summary |
|------------------------|---------------------------------|---------------------------------------------------------------|
| 1. Risk Management | ISO 14971 | The proposed device passes the applicable tests and standards |
| 2. Usability | EN 62366 | The proposed device passes the applicable tests and standards |
| 3. Software life cycle | ISO 62304 | The proposed device passes the applicable tests and standards |
| 4. Basic Safety | IEC 60601-1 &<br>IEC 60601-2-25 | The proposed device passes the applicable tests and standards |
| 5. EMC Compatibility | IEC 60601-2 | The proposed device passes the applicable tests and standards |
| 6. Biocompatibility | ISO 10993-1 | The proposed device passes the applicable tests and standards |
### Summary Discussion of Bench Performance Data
The CareFusion Vyntus ECG passed all specified test requirements.
The validation and verification testing confirmed this device meets user needs and design inputs for an electrocardiograph system.
Testing also confirmed physical attributes and device performance meet the requirements of the standards listed in the performance testing summary above. These standards address risk management, usability, software life cycle, electrical safety, emc and biocompatibility. All testing which have been performed demonstrate substantial equivalence to the predicate devices.
### 2. Clinical tests conducted for determination of substantial equivalence and/or of clinical information:
Clinical Performance Data/Information:
Clinical testing was not performed with this device.
### 3. Conclusion drawn from non-clinical and clinical data:
The Carefusion Vyntus ECG meets the functional claims and intended use as described in the product labeling. The Vynuts ECG is substantially equivalent to the predicate devices described in the submission.
#### 5.16 Conclusion
Based on the above, CareFusion concludes that the Vyntus ECG is substantially equivalent to the legally marketed predicate devices and as safe as effective as the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.