The Flexipet® Denuding Pipettes are intended to be used for blastomere or polar body manipulation, oocyte and embryo manipulation, or denuding. The Flexipet® Manipulation Pipettes are intended to be used for blastomere or polar body manipulation, oocyte and embryo manipulation, blastocyst handling, and manipulation of the oocyte-cumulus complex.
Device Story
Flexipet® Denuding and Manipulation Pipettes are single-use, gamma-sterilized polycarbonate tubes used in assisted reproduction. The devices function as microtools for the physical handling of oocytes, embryos, and blastocysts. They are operated by clinicians in a laboratory setting using an aspiration tool (e.g., Cook Flexipet® Adjustable Handle) to create suction. The pipette tip diameter varies (80–600 µm) to accommodate different cell sizes and manipulation tasks. By providing a controlled means to aspirate and transfer delicate biological samples, the device facilitates embryological procedures, potentially improving the efficiency and safety of assisted reproductive technology workflows.
Clinical Evidence
Bench testing only. Testing included Mouse Embryo Assay (≥80% development to blastocyst), LAL endotoxin testing (<20 EU/device), cytotoxicity (grade <2), accelerated aging, pull-out force (≥0.4 lbf), and functional aspiration testing.
Technological Characteristics
Polycarbonate tubes, 3.5 inches in length. Proximal inner diameter 600 µm; distal inner diameter range 80–600 µm. Gamma sterilized. Single-use. Manual operation via aspiration handle.
Indications for Use
Indicated for use in assisted reproduction procedures including blastomere or polar body manipulation, oocyte and embryo manipulation, denuding, blastocyst handling, and manipulation of the oocyte-cumulus complex.
Regulatory Classification
Identification
Assisted reproduction microtools are pipettes or other devices used in the laboratory to denude, micromanipulate, hold, or transfer human gametes or embryos for assisted hatching, intracytoplasmic sperm injection (ICSI), or other assisted reproduction methods.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when the assisted reproduction microtools (pipettes) are manufactured from glass, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
THE STRIPPER® Micropipetter and Micropipetter Tips (K993699)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2015
Cook Incorporated Naomi Funkhouser Regulatory Affairs Specialist 750 Daniels Way P.O. Box 489 Bloomington, IN 47402
Re: K150748
> Trade/Device Name: Flexipet® Denuding Pipette, Flexipet® Manipulation Pipette Regulation Number: 21 CFR 884.6130 Regulation Name: Assisted reproduction microtools Regulatory Class: Class II Product Code: MOH Dated: November 18, 2015 Received: November 18, 2015
Dear Naomi Funkhouser,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K150748
Device Name
Flexipet® Denuding Pipette and Flexipet® Manipulation Pipette
Indications for Use (Describe)
The Flexipet® Denuding Pipettes are intended to be used for blastomere or polar body manipulation, occyte and embryo manipulation, or denuding.
The Flexipet® Manipulation Pipettes are intended to be used for blastomere or polar body manipulation, oocyte and embryo manipulation, blastocyst handling, and manipulation of the oocyte-cumulus complex.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font, and the word "MEDICAL" is in a smaller, white, sans-serif font below it.
# COOK INCORPORA NIELS WAY, P.O.
### 5.0 510(k) Summary
## Flexipet® Denuding Pipette Flexipet® Manipulation Pipette 510(k) Summary 21 CFR §884.6130 Date Prepared: December 11, 2015
Cook Incorporated
750 Daniels Way
P.O. Box 489
#### Submitted By:
| Applicant: | |
|------------|--|
| Address: | |
Phone Number: Fax Number:
Contact: Contact Address:
Contact Phone Number: Contact Fax Number:
(812) 335-3575 x104371 (812) 332-0281 Naomi Funkhouser Cook Incorporated 750 Daniels Way
Bloomington, IN 47402
P.O. Box 489 Bloomington, IN 47402 (812) 335-3575 x104371 (812) 332-0281
#### Device Information:
Trade name: Common name:
Classification Name:
Regulation: Product Code: Flexipet® Denuding Pipette Flexipet® Manipulation Pipette Microtools, Assisted Reproduction (Pipettes) Assisted Reproduction Microtools. 21 CFR §884.6130 MOH
#### Predicate Device:
The predicate device is "THE STRIPPER®" Micropipetter and Micropipetter Tips (Mid-Atlantic Diagnostics, Inc., K993699).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol to the right of the word. Below the word "COOK" is the word "MEDICAL" in white, and both words are set against a red background.
#### Device Description:
The Flexipet® Denuding Pipettes are polycarbonate tubes with inner diameters measuring 600 um (microns) at the proximal end which taper in diameter in a range from 120 µm to 170 µm on the distal end. The micropipettes are all 3.5 inches in length and constructed from polycarbonate. The devices are supplied gamma sterilized and are intended for one-time use.
The Flexipet® Manipulation Pipettes are polycarbonate tubes with inner diameters measuring 600 um (microns) at the proximal end which taper in diameter in a range from 80 um to 600 um on the distal end. The micropipettes are all 3.5 inches in length and constructed from polycarbonate. The devices are supplied gamma sterilized and are intended for one-time use.
#### Intended Use:
The Flexipet® Denuding Pipettes are intended to be used for blastomere or polar body manipulation, oocyte and embryo manipulation, or denuding.
The Flexipet® Manipulation Pipettes are intended to be used for blastomere or polar body manipulation, oocyte and embryo manipulation, blastocyst handling, and manipulation of the oocyte-cumulus complex.
#### Comparison to Predicates:
The Flexipet® Denuding Pipettes and Flexipet® Manipulation Pipettes are substantially equivalent to the predicate device, "THE STRIPPER®" Micropipetter and Micropipetter Tips (Mid-Atlantic Diagnostics, Inc., K993699), in that these devices have similar designs, methods of construction and operation, and indications for use.
#### Technological Characteristics:
The Flexipet® Denuding Pipettes and Flexipet® Manipulation Pipettes were subjected to testing as to assure validation of design and performance.
- 1. MEA Testing The Flexipet® Denuding Pipettes and Flexipet® Manipulation Pipettes underwent lot-release Mouse Embryo Assay testing. The acceptance criterion was that ≥80% 1-cell embryos developed to blastocyst within 96 hours. The predetermined acceptance criterion was met.
- 2. LAL Testing The Flexipet® Denuding Pipettes and Flexipet® Manipulation Pipettes were lot-release tested using the Limulus Amebocyte Lysate (LAL) test. The devices must demonstrate < 20 EU/device. The predetermined acceptance criterion was met.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below that, the word "MEDICAL" is written in white letters on a black background.
- 3. Cytotoxicity Testing Testing was performed with the requirement that all test extracts must have a reaction grade of less than 2 (mild reactivity). The results showed that the predetermined acceptance criterion was met.
- 4. Accelerated Aging After sterilization and accelerated aging, the devices remained flexible and did not become brittle; therefore, the acceptance criterion was met.
- 5. Pull-Out Force Testing was performed with the requirement that the force needed to pull the pipette out of the aspiration tool must be a minimum of 0.4 lbf. The results showed that the predetermined acceptance criterion was met.
- 6. Aspiration Test Fluid was aspirated through the Flexipet® Denuding Pipettes and Flexipet® Manipulation Pipettes using the Cook Flexipet® Adjustable Handle. The acceptance criterion was met.
#### Conclusion:
The results of these tests support a conclusion that the Flexipet® Denuding Pipettes and Flexipet® Manipulation Pipettes met the design input requirements based on the intended use and support the conclusion that these devices do not raise new questions of safety or effectiveness as compared to, and are substantially equivalent to, the predicate device, "THE STRIPPER® Micropipetter and Micropipetter Tips, manufactured by Mid-Atlantic Diagnostics, Inc., cleared under 510(k) Premarket Notification Number K993699 on May 16, 2000.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.