K150718 · Agency For Medical Innovations GmbH · FHN · Jun 12, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K150718
Device Name
Wi-3 HAL-RAR System
Applicant
Agency For Medical Innovations GmbH
Product Code
FHN · Gastroenterology, Urology
Decision Date
Jun 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Wi-3 HAL-RAR System is a Bluetooth-enabled, pulsed Doppler, ultrasound system intended to be used for detecting hemorrhoidal arteries, manually ligating these arteries, and subsequently carrying out a mucopexy of hemorrhoidal prolapse if necessary.
Device Story
Wi-3 HAL-RAR System uses pulsed Doppler ultrasound to detect hemorrhoidal arteries; assists surgeons in operating room during hemorrhoidal artery ligation (HAL) and recto-anal repair (RAR). System consists of Bluetooth-enabled probe, arm, and fixation components. Surgeon operates device to locate arteries via Doppler signal; manually ligates arteries; performs mucopexy if needed. Bluetooth connectivity replaces traditional cabling; battery integrated into handpiece. Output provides real-time arterial detection feedback to surgeon, facilitating precise ligation. Benefits include minimally invasive approach to hemorrhoid treatment.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), software verification (IEC 60601-1-4), electrical safety, and electromagnetic compatibility (IEC 60601-1-2, IEC 60601-2-37).
Indicated for patients requiring detection of hemorrhoidal arteries, manual ligation of said arteries, and mucopexy of hemorrhoidal prolapse.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2015
Agency for Medical Innovations GmbH % Dr. Allison C. Komiyama, Ph. D., R.A.C. AcKnowledge Regulatory Strategies 2834 Hawthorn Street San Diego, California 92104
Re: K150718
Trade/Device Name: Wi-3 HAL-RAR System Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: Class II Product Code: FHN, IYN Dated: March 18, 2015 Received: March 19, 2015
Dear Dr. Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150718
Device Name Wi-3 HAL-RAR System
#### Indications for Use (Describe)
The Wi-3 HAL-RAR System is a Bluetooth-enabled, pulsed Doppler, ultrasound system intended to be used for detecting hemorrhoidal arteries, manually ligating these arteries, and subsequently carrying out a mucopexy of hemorrhoidal prolapse if necessary.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the text "K150718 Page 1 of 3" at the top. Below that is a large blue logo that says "A.M.I.". The logo has a small registered trademark symbol next to the I.
# 510(k) Summary K150718
#### DATE PREPARED
March 18, 2015
#### MANUFACTURER AND 510(k) OWNER
Agency for Medical Innovations GmbH Im Letten 1, 6800 Feldkirch, Vorarlberg, Austria +43 5522 90505-0 Telephone: Fax: +43 5522 90505-4006 Official Contact: Stefanie Hoellger, Quality Assurance & Regulatory Affairs
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies Telephone: akomiyama@acknowledge-rs.com Email:
#### PROPRIETARY NAME OF SUBJECT DEVICE
Wi-3 HAL-RAR System
#### COMMON NAME
Hemorrhoidal ligator
#### DEVICE CLASSIFICATION
Hemorrhoidal ligator (21 CFR 876.4400, Product Code FHN, Class II) Ultrasonic pulsed Doppler imaging system (21 CFR 892.1550, Product Code IYN, Class II)
#### INDICATIONS FOR USE
The Wi-3 HAL-RAR System is a Bluetooth-enabled, pulsed Doppler, ultrasound system intended to be used for detecting hemorrhoidal arteries, manually ligating these arteries, and subsequently carrying out a mucopexy of hemorrhoidal prolapse if necessary.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for A.M.I. in blue. Above the logo, the text "K151532 Page 2 of 3" is visible. The logo is in a bold, sans-serif font and has a modern, corporate look.
## DEVICE DESCRIPTION
The Wi-3 HAL-RAR System is a Bluetooth-enabled, pulsed Doppler, ultrasound system intended to be used for detecting hemorrhoidal arteries, manually ligating these arteries, and subsequently carrying out a mucopexy of hemorrhoidal prolapse if necessary. "HAL" is an acronym referring to "Hemorrhoidal Arteries Ligation", and "RAR" is an acronym referring to "Recto Anal Repair." The Wi-3 HAL-RAR System is designed for use in a surgical suite or operating room.
The Wi-3 HAL-RAR System includes components that come in contact with the patient during use. The RAR Probes (made of acrylic-based multipolymer) and Sleeve (PBT) are external communicating devices having limited contact duration (< 24 hours) with tissue. The Wi-3 Arm (made of Aluminum Alloy) and Wi-3 Fixation Nut (made of polyphenylsulfone and acetal) are surface devices having limited contact duration (< 24 hours) with skin.
## PREDICATE DEVICE IDENTIFICATION
The Wi-3 HAL-RAR System is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|------------------------------------------|-------------------|
| K061831 | MedChannel HTS System* / MedChannel, LLC | ✓ |
| K081429 | THD Slide / THD S.p.A | |
| K131282 | Nexus™ Hemorrhoid Ligator / inx Medical | |
* The MedChannel HTS System was purchased by A.M.I and rebranded as the HAL-RAR System.
## SUMMARY OF NON-CLINICAL TESTING
Biocompatibility: Patient contacting material was subjected to biocompatibility testing in compliance to ISO 10993-1 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process: cytotoxicity, sensitization, and irritation.
Software Verification: The software development and testing was executed in compliance to IEC 60601-1 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance and IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems.
Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests and IEC 60601-2-37 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
{5}------------------------------------------------
#### 510(k) Summary
Image /page/5/Picture/1 description: The image shows the text "K150718" and "Page 3 of 3" at the top. Below that is the logo "A.M.I." in a large, bold, blue font. There are gray dots below each letter in the logo. The logo appears to be a company or organization's branding.
#### EQUIVALENCE TO PREDICATE DEVICES
The subject device has a similar design and dimensions, and uses similar or identical materials as the device cleared in K061831. The new material (PBT) used in the Sleeve component of the subject device has undergone testing to ensure it is biocompatible for its contact duration (< 24 hours) and contact type (tissue). The subject device has the same intended use and similar technological characteristics (i.e. the Doppler is connected via a Bluetooth module instead of a cable, and the batteries are now incorporated into the handpiece of the device instead of the console) to the devices cleared in K061831, K081429, and K131282. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicates.
## CONCLUSION
Based on the testing performed, including biocompatibility, software verification testing, electrical safety, and electromagnetic compatibility testing, it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed Wi-3 HAL-RAR System are assessed to be substantially equivalent to the predicate devices. The device is considered safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.