← Product Code [DXF](/productcode/DXF) · K150709

# ProTrack RF Anchor Wire (K150709)

_Baylis Medical Company, Inc. · DXF · Jun 17, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150709

## Device Facts

- **Applicant:** Baylis Medical Company, Inc.
- **Product Code:** [DXF](/productcode/DXF.md)
- **Decision Date:** Jun 17, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5175
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ProTrack RF Anchor Wire is indicated for the creation of an atrial septal defect in the heart.

## Device Story

ProTrack RF Anchor Wire is a sterile, single-use device used to create atrial septal defects. It functions as a guidewire that delivers monopolar radiofrequency (RF) energy to a distal electrode to puncture tissue. The device connects to a Baylis Medical Company RF Puncture Generator via a connector cable. It features a stainless steel core with an insulating coating, an atraumatic curved distal end to prevent slippage, and a radiopaque marker coil for fluoroscopic visualization. The wire serves as a stiff rail for advancing ancillary devices into the left atrium. Used in clinical settings by physicians; the device facilitates access for subsequent cardiac procedures. Benefits include controlled tissue puncture and stable positioning for device delivery.

## Clinical Evidence

Bench testing only. Verification and validation included biocompatibility (ISO 10993-1, FDA G95-1), mechanical testing (ISO 11070), electrical safety (IEC 60601-2-2), arc integrity/thermal testing, and benchtop validation of radiopacity, cutting performance, and accessory compatibility.

## Technological Characteristics

Stainless steel core wire with insulating coating; monopolar RF energy delivery; curved atraumatic distal tip with radiopaque marker coil; compatible with BMC RF Puncture Generator; sterile, single-use; dimensions and materials similar to predicates; electrical safety per IEC 60601-2-2; mechanical compliance per ISO 11070.

## Regulatory Identification

A septostomy catheter is a special balloon catheter that is used to create or enlarge the atrial septal defect found in the heart of certain infants.

## Predicate Devices

- Nykanen Radiofrequency Wire ([K990284](/device/K990284.md))
- NRG Transseptal Needle ([K073326](/device/K073326.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2015

Baylis Medical Company Inc. Ms. Meghal Khakhar Director, Regulatory & Scientific Affairs 2645 Matheson Blvd. East Mississauga, ON Canada L4W 5S4

Re: K150709

Trade/Device Name: Protrack Rf Anchor Wire Regulation Number: 21 CFR 870.5175 Regulation Name: Septostomy Catheter Regulatory Class: Class II Product Code: DXF Dated: March 18, 2015 Received: March 19, 2015

Dear Ms. Khakhar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours

Mitchell Stein

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known):

Device Name: ProTrack RF Anchor Wire

Indications for Use:

The ProTrack RF Anchor Wire is indicated for the creation of an atrial septal defect in the heart.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

{3}------------------------------------------------

# 7. 510(k) Summary

### Submitter Information

- A. Company Name: Baylis Medical Company Inc.
- B. Company Address: 2645 Matheson Blvd. East Mississauga, Ontario L4W 5S4 Canada
- C. Company Phone: (905) 602-4875
- D. Company Facsimile: (905) 602-5671
- E. Contact Person: Meghal Khakhar
- F. Summary Prepared on: 18-Mar-2015

# Device Identification

- A. Device Trade Name: ProTrack RF Anchor Wire
- B. Device Common Name: Septostomy catheter
- C. Classification Name: 21 CFR 870.5175 Septostomy catheter
- D. Product Code: DXF
- E. Device Class: Class II

#### Identification of Predicate Devices

#### Table 1: Predicate Devices

| Predicate Device            | Manufacturer                | 510(k) No. |
|-----------------------------|-----------------------------|------------|
| Nykanen Radiofrequency Wire | Baylis Medical Company Inc. | K990284    |
| NRG Transseptal Needle      | Baylis Medical Company Inc. | K073326    |

#### Indications for Use

The ProTrack RF Anchor Wire is indicated for the creation of an atrial septal defect in the heart.

{4}------------------------------------------------

# Device Description

The ProTrack RF Anchor Wire is a sterile, single-use device that delivers radiofrequency (RF) power in a monopolar mode to its distal electrode. The device connects to a separately cleared Baylis Medical Company (BMC) RF Puncture Generator at its proximal end through a compatible BMC Connector Cable.

The ProTrack RF Anchor Wire is comprised of a core stainless steel wire. The body of the wire is coated with an insulating material to facilitate smooth advancement of the device and to provide electrical insulation. The wire's distal end is atraumatic and curved to prevent the wire from slipping back into the right atrium through the atrial septal defect when devices are being retracted or advanced over it. A marker coil is positioned on the distal curve for visualization under fluoroscopy. The main body of the RF Anchor Wire provides a stiff rail for advancing ancillary devices into the left atrium following creation of an atrial septal defect. The proximal end is curved and is bare metal to connect with the included BMC Connector Cable for connection to the compatible BMC RF Puncture Generator.

# Comparison to Predicate Devices

Table 2 provides a comparison of the main characteristics of the ProTrack RF Anchor Wire to those of the predicate devices.

Table 2: Comparison of ProTrack RF Anchor Wire Characteristics to Predicate Devices

| Characteristic                                        | Comment   |
|-------------------------------------------------------|-----------|
| Intended Use                                          | Identical |
| Indications for Use                                   | Similar   |
| Fundamental scientific technology                     | Identical |
| Operating principles                                  | Identical |
| Mechanism of action                                   | Identical |
| Materials                                             | Similar   |
| Dimensions                                            | Similar   |
| Accessory devices                                     | Identical |
| Packaging configuration                               | Similar   |
| Sterilization method and Sterility Assurance<br>Level | Identical |

# Performance Testing

Performance testing was completed for the ProTrack RF Anchor Wire to support its safety and effectiveness for its intended use and its substantial equivalence to the predicate devices. The following bench top verification and validation was completed:

{5}------------------------------------------------

- 1. Biocompatibility Verification: The biological safety of the device was verified as per the requirements of ISO 10993-1 and FDA's modified ISO guidelines in accordance with FDA's blue book memorandum #G95-1 on biocompatibility.
- 2. Mechanical Verification: Mechanical testing was performed to verify compliance of the device with ISO 11070 and Baylis self-enforced requirements.
- 3. Electrical Verification: Electrical testing was performed to verify compliance of the device with IEC 60601-2-2 and Baylis self-enforced requirements.
- 4. Arc Integrity Verification: Thermal testing was performed to ensure compatibility of the device with the BMC RF Puncture Generator.
- 5. Benchtop Validation: Validation testing was performed to demonstrate compatibility with accessory devices, radiopacity, and cutting performance as compared to the predicate devices.

The ProTrack RF Anchor Wire met all requirements as specified by the test protocols.

# Conclusions

The proposed and predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action. Differences in design and technological characteristics between the proposed and predicate devices do not raise any new concerns of safety and effectiveness. The results of verification and validation provide reasonable assurance of the safety and effectiveness of the ProTrack RF Anchor Wire for its intended use and its substantial equivalence determination to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K150709](https://fda-staging.innolitics.com/device/K150709)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
