Exceed Tx

K150702 · Great Lakes Orthodontics, Ltd. · DYW · Mar 24, 2016 · Dental

Device Facts

Record IDK150702
Device NameExceed Tx
ApplicantGreat Lakes Orthodontics, Ltd.
Product CodeDYW · Dental
Decision DateMar 24, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

eXceed Computerized Precision Bracket Placement Solution is a software system intended for use as an aid in orthodontic treatment planning to correct Malocclusions in Orthodontic Patients. For use by dental professionals trained in orthodontic treatment, including radiographic analyses and treatment planning. eXceed Computerized Precision Bracket Placement Solution is intended for use with commercially available brackets currently used in standard orthodontic treatment. The end product is an indirect bonding tray for use by the Dental professional to place multiple brackets at the same time.

Device Story

Software system for orthodontic treatment planning; inputs include dental impressions or 3D models, panoramic X-rays, and facial/intra-oral images. Software generates 3D model and calculates ideal bracket placement; clinician reviews and approves/adjusts plan. Manufacturing facility prints 3D model featuring physical/visual landmarks for bracket positioning; technician bonds brackets to model based on landmarks. Vacuum forming creates indirect bonding tray with brackets in place. Tray shipped to dental professional for clinical application using standard adhesives. Facilitates precise, simultaneous placement of multiple brackets; improves treatment planning accuracy; benefits patient through optimized orthodontic outcomes.

Clinical Evidence

No clinical data. Performance verified via bench testing comparing the digital model generation and bracket placement accuracy against the clinician-approved treatment plan.

Technological Characteristics

Proprietary software for 3D modeling and bracket placement calculation. Inputs: scanned analog impressions or intra-oral scans, X-rays, and images. Output: 3D printed working model with physical/visual landmarks for bracket positioning and vacuum-formed indirect bonding trays. Operates via clinician-approved digital treatment plans. Compatible with commercially available orthodontic brackets.

Indications for Use

Indicated for orthodontic patients requiring treatment for malocclusions. For use by dental professionals trained in orthodontic treatment and planning.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a wing-like shape above them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 24, 2016 Great Lakes Orthodontics, Ltd. Mr. Dave Graver Director of Logistics 200 Cooper Ave. Tonawanda, New York 14150 Re: K150702 Trade/Device Name: eXceed Computerized Precision Bracket Placement Solution Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: DYW, NJM, EJF Dated: February 19, 2016 Received: February 23, 2016 Dear Mr. Graver: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | |-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number (if known) | K150702 | | Device Name | eXceed Computerized Precision Bracket Placement Solution | | Indications for Use (Describe) | eXceed Computerized Precision Bracket Placement Solution is a software system intended for use as an aid in orthodontic treatment planning to correct Malocclusions in Orthodontic Patients. For use by dental professionals trained in orthodontic treatment, including radiographic analyses and treatment planning. eXceed Computerized Precision Bracket Placement Solution is intended for use with commercially available brackets currently used in standard orthodontic treatment. The end product is an indirect bonding tray for use by the Dental professional to place multiple brackets at the same time. | | Type of Use (Select one or both, as applicable) | | | | <div> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (8/14) | Page 1 of 1 | |----------------------|-------------| |----------------------|-------------| {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Great Lakes Orthodontics, LTD. The logo features a stylized map of the Great Lakes region of the United States, surrounded by the words "Great Lakes Orthodontics." Below the logo, the company's name is written in a larger font, followed by the tagline "An Employee Owned Company." Our Vision "Delight our customers. Respect and help our co-workers" #### K150702 510K Summary #### Submitter's name, address, phone, and contact person: Great Lakes Orthodontics, Ltd. (Registration# 1316408) 200 Cooper Ave. Tonawanda, NY 14150 Phone 716-319-1250 (direct) Fax 716-871-0550 Contact person: Dave Graver Email: dgraver@greatlakesortho.com Date summary prepared: March 24, 2016 Trade / Proprietary Name of Device: Exceed Computerized Precision Bracket Placement Solution #### Common name of device: Accessory to Orthodontic Brackets #### Device classification name: Orthodontic Plastic Bracket (Class II) Classification regulation: 21CFR 872.5470 Product codes: DYW (Orthodontic plastic bracket), NJM (Orthodontic Ceramic Bracket), and ELF (Orthodontic Metal Bracket) Page 1/4 200CooperAvenue · P.O. Box5111 · Tonawanda, New York 14151-5111 716-871-1161 • 800-828-7626 • Fax:(716) 871-0550 e-mail info@greatlakesortho.com • Website: www.greatlakesortho.com {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Great Lakes Orthodontics, LTD. The logo includes a map of the Great Lakes region of the United States, surrounded by the words "Great Lakes Orthodontics." Below the logo, the words "Great Lakes Orthodontics, LTD." are written in a larger font. Underneath that, it says "An Employee Owned Company". Our Vision "Delight our customers. Respect and help our co-workers" ## Predicate Device: Ortho CADiQ (510k # K082207) ### Intended use/Indications for use eXceed Computerized Precision Bracket Placement Solution is a software system intended for use as an aid in orthodontic treatment planning to correct Malocclusions in Orthodontic Patients. For use by dental professionals trained in orthodontic treatment, including radiographic analyses and treatment planning. eXceed Computerized Precision Bracket Placement Solution is intended for use with commercially available brackets currently used in standard orthodontic treatment. The end product is an indirect bonding tray for use by the Dental professional to place multiple brackets at the same time. ### Technological Characteristics The device consists of proprietary software that calculates the ideal position of the dental brackets based on the dental impressions and or 3D models supplied by the patients Orthodontist. Commercially available brackets are used as part of the system. A side by side comparison table is shown at the end of this summary. ## Principals of operation Using the images provided by the Dental Professional, the software creates a 3D model and identifies the ideal placement of the brackets. The file is sent to the Dental Professional for review and approval. The Dental Professional may adjust the final position of the bracket if desired. A 3D model is printed, and the brackets are placed on the model in the prescribed location, approved by the Orthodontist. An indirect bonding tray is fabricated with the brackets in place. The tray and brackets are sent to the Dental professional. The Dental Professional places the indirect tray using their chosen commercially available bracket adhesive. While the subject and predicate devices do not have identical Indications for Use statements, they share the same intended use. Both are intended for use as aids in orthodontic treatment planning for use by dental professionals trained in orthodontic treatment, including radiographic analyses and treatment planning. Both are intended for use with commercially-available brackets. Page 2/4 200CooperAvenue · P.O.Box5111 · Tonawanda, New York 14151-5111 716-871-1161 · 800-828-7626 · Fax:{716}671-0550 Email: info@greatlakesortho.com Website: www.greatlakesortho.com {5}------------------------------------------------ Image /page/5/Picture/0 description: The image is a logo for Great Lakes Orthodontics. The logo features a map of the United States with the Great Lakes highlighted in white. The words "GREAT LAKES" are written in a circular fashion above the map, and the word "ORTHODONTICS" is written in a circular fashion below the map. The logo is simple and clean, and it effectively communicates the company's name and location. ## Great Lakes Orthodontics,LTD. An Employee Owned Company Our Vision " Delight our customers. Respect and Help our co-worlters" The operating principle for Exceed Computerized Precision Bracket Placement Solution and OrthoCAD iQ is identical, with the exception of the guidance method used by the technician in the manufacturing facility when pasting the brackets to the working model, prior to tray fabrication. The difference between the Subject and predicate devices is that the subject device employs a different guidance method for placing the brackets on the working model prior to trays' fabrication. Exceed Computerized Precision Bracket Placement Solution includes visual/physical frames on its models, while OrthoCAD provides virtual visual quidance only. The difference in operating principle is minor, and does not render the subject device substantially different from the predicate device for the following reasons: - · The intended use of the different guidance method is the same as in the predicate device, providing the technician at the manufacturing facility clear and unequivocal guidance regarding where the brackets are to be placed on the working model, based on the approved placement plan. - As confirmed in comparative testing, the visual guidance system used by the subject ● device is accurate, in the sense that it corresponds exactly with the virtual coordinates dictated by the approved placement plan, thus enabling the technician to place the brackets on the working model precisely according to that plan. As such, it is as safe and effective as the predicate and does not raise any new concerns. #### Performance Data The software is verified and validated by the digital model generated, matching the 3D model and bracket placement locations prescribed and approved by the Orthodontist. #### Substantial Equivalence Great Lakes Orthodontics Ltd has demonstrated that, for the purposes of FDA's regulation of medical devices, the Exceed Computerized Precision Bracket Placement Solution is substantially equivalent to the predicate device in terms of intended use, indications, technical characteristics, and principles of operation. Page 3/4 200CooperAvenue · P.O.Box5111 · Tonawanda,New York 14151-5111 716-871-1161 · 800-828-7626 · Fax:{716}671-0550 Email: info@greatlakesortho.com Website: www.greatlakesortho.com {6}------------------------------------------------ Image /page/6/Picture/0 description: The image is a logo for Great Lakes Orthodontics. The logo features a map of the United States with the Great Lakes highlighted in white. The words "GREAT LAKES" are written in a semi-circle above the map, and the word "ORTHODONTICS" is written in a semi-circle below the map. The logo is simple and clean, and it effectively communicates the company's name and focus. # Great Lakes Orthodontics, LTD. An Employee Owned Company Our Vision " Delight our customers. Respect and Help our co-worlters" ## TECHNOLOGICAL CHARACTERISTICS (Similarities & Differences) | | Exceed Computerized Precision Bracket<br>Placement Solution | Ortho CAD iQ | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | eXceed Computerized Precision Bracket<br>Placement Solution is a software system intended<br>for use as an aid in orthodontic treatment planning<br>to correct Malocclusions in Orthodontic Patients.<br>For use by dental professionals trained in<br>orthodontic treatment, including radiographic<br>analyses and treatment planning. eXceed<br>Computerized Precision Bracket Placement<br>Solution is intended for use with commercially<br>available brackets currently used in standard<br>orthodontic treatment. The end product is an<br>indirect bonding tray for use by the Dental<br>professional to place multiple brackets at the same<br>time. | Ortho CAD iQ is a computer-guided system<br>intended for use as an aid in orthodontic<br>treatment planning for use by dental<br>professionals trained in orthodontic treatment<br>including radiographic analyses and<br>treatment planning. OrthoCAD iQ is intended<br>for use with commercially-available brackets<br>and wires that apply continuous gentle force<br>to reposition the teeth. It also uses indirect<br>bonding trays to affix the brackets in position. | | Key Records | A 3D model is generated from scanned analog<br>impressions or<br>directly from an intra-oral scan | A 3D model is generated from<br>scanned analog impressions or directly from<br>an intra-oral scan | | Additional<br>Records | A panoramic X-ray, facial and intra- oral images | A panoramic X-ray, facial and intra-oral<br>images | | Treatment Plan | A detailed treatment plan is provided by the<br>clinician as part of case submission. | A detailed treatment plan is provided by the<br>clinician as part of case submission. | | Virtual Bracket<br>Placement<br>Plan | Using the submitted records and the prescribed<br>treatment plan, a suggested placement plan is<br>generated and sent for review by the clinician. The<br>Clinician utilizes the eXceed software to adjust the<br>plan and later have it approved. | Using the submitted records and the<br>prescribed treatment plan, a suggested<br>placement plan is generated and sent for<br>review by the clinician. The Clinician utilizes<br>the OrthoCAD software to adjust<br>the plan and later have it approved. | | Manufacturing | A working pre-treatment model, which includes<br>tooth and bracket-specific landmarks depicting the<br>position of the brackets based on the approved<br>plan, is printed. Brackets are bonded by a<br>technician within the landmarks. | A working pre-treatment model is printed. A<br>technician bonds the brackets to the working<br>model using a pen-like device for steering<br>the appliances on the surface of the tooth and<br>a computerized silhouette screened on the<br>tooth depicting the approved bracket position. | | Finished<br>Product | Vacuum forming then follows to produce patient-<br>specific bracket transfer trays. | Vacuum forming then follows to produce<br>patient-specific bracket transfer trays. | Page 4/4 200CooperAvenue · P.O.Box5111 · Tonawanda,New York 14151-5111 716-871-1161 • 800-828-7626 • Fax:{716) 671-0550 Email: info@greatlakesortho.com _Website= www.greatlakesortho.com
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...