K150680 · Maxx Orthopedics, Inc. · MBH · Jun 26, 2015 · Orthopedic
Device Facts
Record ID
K150680
Device Name
Freedom Cementless Femoral Component
Applicant
Maxx Orthopedics, Inc.
Product Code
MBH · Orthopedic
Decision Date
Jun 26, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Freedom® Cementless Femoral Components are designed to be used as a part of the Freedom® Total Knee System, and is indicated for the following: - Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Correction of functional deformities. - Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy. - Moderate valgus, varus, or flexion trauma. - Knee fractures untreatable by other methods - Revision surgery where sufficient bone stock and soft tissue integrity are present. The Freedom Total Knee System, Freedom Stemmed Tibial Components and Freedom PCK Components are indicated for cemented fixation. Only Cementless Femoral (CR and PS) components with porous coating are additionally indicated for cementless biological fixation application.
Device Story
Freedom® Cementless Femoral Components are implantable orthopedic devices used in total knee replacement (TKR) surgery. Designed for use within the Freedom® Total Knee System, these components replace damaged knee joint surfaces. The device is implanted in an operating theater by a qualified surgeon. The components feature a porous coating, allowing for cementless biological fixation as an alternative to traditional cemented fixation. By mimicking normal knee geometry, the device aims to restore joint function and mobility. The system is available in various sizes and configurations to accommodate patient anatomy. Clinical benefit is derived from the restoration of joint mechanics and reduction of pain associated with degenerative or traumatic knee conditions.
Clinical Evidence
No clinical data was required or provided to support the substantial equivalence determination. Evidence is based on bench testing, including stereological evaluation of porosity, pore size, and coating thickness, as well as static shear, tensile, and shear fatigue testing of CoCr coupons.
Technological Characteristics
Implantable femoral components made of biocompatible materials (CoCr). Features a porous coating (sintered CoCr powder, -40/+60 mesh) for biological fixation. Designed to mimic human knee geometry; available in multiple sizes for left/right configurations. Mechanical testing performed per ASTM F75 standards. No software or electronic components.
Indications for Use
Indicated for patients with severe knee pain, mobility loss, or disability due to rheumatoid, osteoarthritis, traumatic arthritis, or polyarthritis; functional deformities; post-traumatic joint contour loss; moderate valgus, varus, or flexion trauma; fractures untreatable by other methods; or revision surgery with sufficient bone stock/soft tissue.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three heads, representing health, services, and people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.
June 26, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Maxx Orthopedics, Incorporated Ms. Gabriela Gottlieb Regulatory Affairs Manager 531 Plymouth Rd., Suite 526 Plymouth Meeting, Pennsylvania 19462
Re: K150680 Trade/Device Name: Freedom® Cementless Femoral Components Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH Dated: March 16. 2015 Received: March 18, 2015
Dear Gabriela Gottlieb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Freedom® Cementless Femoral Components Device Name:
Indications for Use:
The Freedom® Cementless Femoral Components are designed to be used as a part of the Freedom® Total Knee System, and is indicated for the following:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities.
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods
- Revision surgery where sufficient bone stock and soft tissue integrity are present. ●
The Freedom Total Knee System, Freedom Stemmed Tibial Components and Freedom PCK Components are indicated for cemented fixation. Only Cementless Femoral (CR and PS) components with porous coating are additionally indicated for cementless biological fixation application.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## Section 13
## 510(k) Summary
#### Submitter:
#### Date of Summary: June 25, 2015
Maxx Orthopedics, Inc. 531 Plymouth Rd #526 Plymouth Meeting, PA 19462
#### Establishment Registration Number: 3007311878
| Contact Person: | Alternate Contact: |
|------------------------------------------|-------------------------------------------|
| Gabriela Gottlieb | Nach Dave |
| Manager, Regulatory and Clinical Affairs | Director, Regulatory and Clinical Affairs |
| Phone: 215-606-7418 | Phone: 732-718-1385 |
| Email: gabriela.gottlieb@maxxortho.com | Email: nach.dave@maxxortho.com |
#### Device Information:
| Proprietary Name: | Freedom® Cementless Femoral Components |
|----------------------|-------------------------------------------------------------------------------------|
| Common/Usual Name: | Cementless Femoral Components |
| Classification Name: | Knee Joint Patellofemorotibial Metal/Polymer porous-coated<br>Cementless prosthesis |
| Regulation Number: | 21 CFR 888.3565 |
| Regulatory Class: | Class II |
| Product Code: | MBH |
#### Identification of Predicate Device(s):
The Freedom® Cementless Femoral Components are similar to the following commercially available devices in regards to the fundamental scientific technology, material composition and intended use:
| Device Name | Type | Device<br>Class | Product<br>Code | Regulation<br>Number | 510(k)<br>number |
|---------------------------------------------|-----------|-----------------|-----------------|----------------------|-------------------------------|
| Freedom Total Knee system | Primary | II | JWH | 21 CFR §<br>888.3560 | K082019<br>K091280<br>K131481 |
| Optetrak Logic® Porous Femoral<br>Component | Reference | II | MBH | 21 CFR §<br>888.3565 | K140302 |
{4}------------------------------------------------
## Device Description:
The Cementless Femoral Components ("Components") are intended for use with the currently available Freedom Total Knee System as part of total knee replacement (TKR) surgery. Total knee replacement surgery is a procedure where worn, diseased, or damaged surfaces of the knee joint are removed and replaced with artificial surfaces.
The Components will be prescription products consisting of single use only, implantable devices for implanting into patients in an operating theatre by a qualified surgeon.
## Intended Use:
The Freedom® Cementless Femoral Components are designed to be used as a part of the Freedom® Total Knee System, and is indicated for the following:
- . Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- . Correction of functional deformities.
- . Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, and/or prior patellectomy.
- . Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods
- Revision surgery where sufficient bone stock and soft tissue integrity are present.
The Freedom Total Knee System, Freedom Stemmed Tibial Components and Freedom PCK Components are indicated for cemented fixation. Only Cementless Femoral (CR and PS) components with porous coating are additionally indicated for cementless biological fixation application.
# Substantial Equivalence for Technological Characteristics:
The proposed Maxx Orthopedics' Freedom® Cementless Femoral Component and the primary predicate device are identical in that they consist of same geometric design, material of composition, and intended use. The proposed components are also the same in fundamental scientific technology as the referenced predicate device (porous coating). The only difference is the device's operating mechanism, which is expanded to give the Surgeon the option to apply implant without cement.
Both the proposed Freedom® Cementless Femoral Components and the predicate devices have been designed to mimic the normal knee geometry. Both the proposed and predicate devices are compatible with left and right configurations and available in a variety of sizes that are intended to mimic normal human anatomy. Both the proposed and predicate devices are made of biocompatible materials and are identical in technological design and materials.
{5}------------------------------------------------
## Performance Testing:
Mechanical and functional testing described in Section 8 of this 510(k) submission demonstrates that the Freedom® Cementless Components are mechanically and functionally equivalent to the parent Freedom® Total Knee System. The evaluations were performed in accordance with the Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Cementless Prostheses; Guidance for Industry and FDA. The results were found to be equivalent to the legally marketed predicate device.
| Test Name | Test # | Objective |
|-----------------------------|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Stereological<br>Evaluation | TR-15-01 | The purpose of this study was to determine (1) percent<br>porosity, (2) pore size (line length intercept),<br>and (3) coating thickness of CoCr coupons coated and<br>sintered with three layers of -40/+60 mesh<br>CoCr powder. The interconnection of porosity was also<br>examined on all coupons. The coating<br>technique utilized in this study is identical to the coating<br>technique used for specified medical<br>implants. |
| | Static Shear and<br>Tensile Testing | TR-15-02 |
| Abrasion Testing | | TR-15-03 |
| Shear fatigue<br>testing | TR-15-04 | The purpose of this study was to determine shear fatigue<br>strength of ASTM F75 CoCr coupons coated and sintered with<br>three layers of -40/+60 mesh CoCr powder. The coating<br>technique utilized in this study is identical to the coating<br>technique used for specified medical implants. |
# Clinical Testing:
Clinical Testing was not required for these components to support substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.