K150672 · Durr Dental AG · NBL · Jan 27, 2016 · Dental
Device Facts
Record ID
K150672
Device Name
VistaCam iX Proof
Applicant
Durr Dental AG
Product Code
NBL · Dental
Decision Date
Jan 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1745
Device Class
Class 2
Indications for Use
The VistaCam iX "Proof" is intended to be used as an aid in the detection and diagnosis of dental caries.
Device Story
VistaCam iX "Proof" is an intraoral camera system used by dental professionals to detect dental caries. The device consists of a toothbrush-sized handpiece with a "Proof" head, connected via USB to a PC running DBSWIN imaging software. Operation involves placing a camera cover and autoclave-able spacer over the distal end, then positioning the handpiece over the tooth. The device illuminates the tooth surface with violet LEDs (495 nm), causing bacteria associated with caries to fluoresce. The system captures this fluorescence, converts it into an electrical signal, and processes it into a color-coded image displayed on a monitor. Colors (green, blue, red, orange, yellow) and numerical values represent the fluorescence ratio, indicating the likelihood and magnitude of tooth decay. This visual output assists clinicians in diagnosing caries and assessing tooth health.
Clinical Evidence
Bench testing only. Performance was validated through comparative testing against the predicate device, focusing on LED illumination, image quality, and color separation. Safety testing was conducted per IEC 60601-1, IEC 60601-1-2, and IEC 80601-2-60. Biocompatibility was assessed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Fluorescence-based caries detection using 4 violet LEDs (495 nm) and a light-collecting lens. Enclosure: ASA+PC. Power: USB-powered via PC. Connectivity: USB to PC. Software: DBSWIN imaging software. Patient-contacting materials: Ethylene methyl acrylate copolymer, low-density polyethylene, and Septon with UV stabilizer. Sterilization: Autoclave-able spacers (moist heat).
Indications for Use
Indicated for use as an aid in the detection and diagnosis of dental caries in patients.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
Predicate Devices
Spectra Fluorescence Caries Detection Aid Device (K090169)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection. The profiles are simple and abstract, with no facial features, and are rendered in a dark color against a light background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 8, 2016
Durr Dental Ag % Ms. Suzanne Lucas Air Techniques, Inc 1296 Walt Techniques, Inc Melville, New York 11747
Re: K150672
Trade/Device Name: VistaCam iX "Proof" Regulation Number: 21 CFR 872.1745 Regulation Name: Laser Fluorescence Caries Detection Device Regulatory Class: II Product Code: NBL Dated: April 24, 2015 Received: May 26, 2015
Dear Ms. Lucas:
This letter corrects our substantially equivalent letter of January 27, 2016
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
Susan Runno DDS, mA
Tina Kiang, Ph.D Acting Division Director Science and Policy Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K150672
Device Name VistaCam iX "Proof"
Indications for Use (Describe)
The VistaCam iX "Proof" is intended to be used as an aid in the detection and diagnosis of dental caries.
Type of Use (Select one or both, as applicable)
| <span> <b> \[X] Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> |
|------------------------------------------------------------------------------------|
| <span> \[ ] Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) is being submitted in accordance with the requirements of 21 CFR §807.92.
### 1. Date Summary Prepared:
January 21, 2016
### 2. Submitter's Identification:
DÜRR DENTAL AG Höpfigheimer Strasse 17 74321 Bietigheim-Bissingen Deutschland/Germany Telefon: (0 71 42) 70 5-0 Telefax: (0 71 42) 705-500 E-Mail info@duerr.de Internet: www.duerrdental.com
# Contact:
Oliver Lange Email: lange.o@@duerr.de Head of Quality Management DÜRR Dental AG Höpfigheimer Strasse 17 D-74321 Bietigheim-Bissingen Deutschland / Germany Phone: + 49 (0) 7142 / 705-0 Fax: + 49 (0) 7142 / 705-500 Email: lange.o@duerr.de
# U.S. Contact:
Suzanne Lucas Air Techniques, Inc. 1295 Walt Whitman Road Melville, NY 11747 Tel: 516-214-5514 Email: slucas@airtechniques.com
### 3. Device:
| Trade /Proprietary Name: | VistaCam iX "Proof" |
|--------------------------|------------------------------------------------------|
| Common Name: | Intraoral Camera with Fluorescence Caries Detection |
| | Aid |
| Classification: | 21 CFR 872.1745; Laser Fluorescence Caries Detection |
| | Aid |
| Product Code: | NBL |
{4}------------------------------------------------
### Predicate Device: 4.
Spectra Fluorescence Caries Detection Aid Device by Air Techniques, Inc. 510K# K090169
### 5. Device Description:
The VistaCam iX "Proof" aids in the detection and diagnosis of caries. It consists of a toothbrush-sized handpiece and a "Proof" head. A USB cable connects the handpiece to a personal computer with PACS software such as DBSWIN to enable communication between a PC computer and the handpiece.
After a camera cover is placed over the distal end, and an autoclave-able spacer is installed, the Handpiece is positioned over the tooth to be examined. The camera functions by illuminating the tooth surface with a light that causes the bacteria resident in carries to fluoresce. The fluoresced light is then converted into an electrical signal, sent to a computer, converted into an image (by imaging software) and presented on a monitor in multiple colors to illustrate suspected areas of decay.
### 6. Indications for use:
The VistaCam iX "Proof" is intended to be used as an aid in the detection and diagnosis of dental caries.
### 7. Summary of the technological characteristics of the device compared to the predicate device:
Dürr Dental's VistaCam iX "Proof" is identical in terms of indications for use and the technology to the Spectra caries detection aid manufactured by Air Techniques'. Whose predicate dental devices are currently in commercial distribution and US-FDA cleared under 510K #K090169.
Table 1 below, summarizes the technological characteristics of VistaCam iX "Proof" vs. the predicate device.
{5}------------------------------------------------
# Table 1: Comparison to the Predicate device:
| Comparison criteria | New device<br>VistaCam iX "Proof"<br>510K # K150672 | Predicate device<br>Air Technique's Spectra<br>Fluorescence Caries<br>Detection Device<br>510K #K090169 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Owner/ Operator | Dürr Dental AG | Dürr Dental AG |
| Manufacturer | Dürr Dental AG | Air Techniques, Inc |
| Similarities | | |
| Intended use | Caries detection aid | Caries detection aid |
| Indications for use | The VistaCam iX "Proof" is<br>intended to be used as an aid in<br>the detection and diagnosis of<br>dental caries. | The Spectra is indicated as an<br>aid in the detection and<br>diagnosis of dental caries. |
| Technology | Fluorescence technology to aid<br>in the detection of carious<br>lesions | Fluorescence technology to aid<br>in the detection of carious<br>lesions |
| Mode of operation | Excites bacteria to fluoresce | Excites bacteria to fluoresce |
| Image evaluation | Same as predicate | Image Color<br>Relative level<br>Green: 0 - 1.0<br>Blue: 1.0 - 1.5<br>Red 1.5 - 2.0<br>Orange 2.0 - 2.5<br>Yellow 2.5 - 3.0<br>Lighter colors and higher<br>numbers indicate an increasing<br>magnitude of the red / green<br>fluorescence ratio and the<br>likelihood of poorer tooth<br>health. |
| Detection wavelength | 495 nm | 495 nm |
| Power Supply | Uses power from PC USB port | Uses power from PC USB port |
| Returned light | Fluorescence | Fluorescence |
| Differences: | | |
| Enclosure | ASA+PC (Acrylonitrile Styrene<br>Acrylate + Polycarbonate) | ABS (Acrylonitrile Butadiene<br>Styrene) Plastic |
| Intensity | 3 mW/cm squared at 8mm | 3.5 mW/cm squared at 10mm |
| Light source | Utilizes 4 Violet LEDs with light<br>collecting lens. | Utilizes 6 Violet LEDs with light<br>collecting lens. |
| Comparison criteria | New device<br>VistaCam iX "Proof"<br>510K # K150672 | Predicate device<br>Air Technique's Spectra<br>Fluorescence Caries<br>Detection Device<br>510K #K090169 |
| Software: | VistaCam iX "Proof" driver<br>software and DBSWIN.<br>VistaCam iX "Proof" can only<br>be used with DBSWIN imaging<br>software | ATI Spectra driver software and<br>Visix. |
| Operating | 10-40 C | 10-40 C |
| environment | 20 to 75% RH non-condensing | 10 to 90% RH non-condensing |
| Length | 7.48 inches | 8.6 inches |
| Width | 1.03 inches | 0.87 inches |
| Height | 1.03 inches | 1.15 inches |
| Spacers | 8 mm | 10 mm |
{6}------------------------------------------------
### Discussion of Differences: 8.
- i. Enclosure: The material of composition for VistaCam iX "Proof" is substantially equivalent to that of Spectra enclosure as both these materials are non-carcinogenic and poses no health hazards as per their MSDSs. The enclosure of the VistaCam iX "Proof" was safety tested by Intertek to IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007), Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. The enclosure of the Spectra was safety tested by Underwriters Laboratory to IEC 60601-1:1988 +A1:1991 +A2:1995, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- ii. Intensity: The intensity for VistaCam iX is 3 mW/cm squared at 8mm and for Spectra is 3.5 mW/cm squared at 10mm. The difference in intensity of these devices does not impact VistaCam iX "Proof's" performance, safety and effectiveness. VistaCam iX's performance is demonstrated in the document - Executed validation report.
- Light source and Lens: VistaCam iX "Proof" uses 4 violet LEDs and 2 iii. lenses whereas Spectra use 6 violet LEDs and 4 lenses. The difference in the number of lenses and LED's has geometrical and constructional/ design reasons.
The function of lens is to focus the image of an object (surface of a tooth) onto the image sensor (CCD). As the "Proof" spacers are "shorter" (8mm), the camera and thus the image sensor is closer to the surface
{7}------------------------------------------------
during image capturing. Therefore less lenses and LED's are needed for gaining a similar picture section and image quality. Most importantly, for both cameras, the recorded object area is nearly the same for the given distance, which is defined by the spacer.This difference in number of LEDs and lenses does not impact device's performance. VistaCam iX "Proof's" performance is demonstrated in the document - Executed validation report.
- iv. Software: VistaCam iX "Proof" works with VistaCam iX "Proof" driver software and Spectra device works with ATI Spectra driver software. Both VistaCam iX "Proof" and Spectra uses same algorithm developed for VistaCam iX by Duerr. VistaCam iX "Proof" can only be used with DBSWIN imaging software as opposed to Visix for Spectra. The compatibility of VistaCam iX "Proof" and DBSWIN is demonstrated by the DBSWIN software documents that are provided.
- Operating environment: The difference in the limits for relative humidity V. for VistaCam iX and Spectra is very minor and has no impact on the substantial equivalence of VistaCam iX.
- vi. Spacers: The 8 mm spacers for VistaCam iX "Proof" are exactly the same as the 10 mm spacers for Spectra in terms of shape and material* of construction. Moreover these are purchased from the same supplier. The 2 mm difference in the size of these spacers is not a significant change in their dimensions and thus has no impact on their performance and/or sterilization characteristics. The Sterilization validation study was performed on the 10 mm Spectra spacers in 2009 per ISO 11134:1994. Sterilization of health care products -- Requirements for validation and routine control -- Industrial moist heat sterilization, by Nelson laboratories. Nelson has provided their rationale dated October 27, 2015 to support the validity of the Sterilization for Spectra Spacers performed to ISO 11134:1994 in 2009 and it's conformance to the requirements of current FDA recognized standard, ISO 17665-1. Sterilization of health care products moist heat Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
*Septon plus a UV stabilizer (2-5-chloro-2H-benzotriazole-2-yl)-6-(1,1dimethylethyl)- 4-methylphenol (CAS-No. 3896-11-5) with the restriction SML(T) =30 mg/kg.
{8}------------------------------------------------
### Non-Clinical Data 9.
The purpose of the evaluation was to demonstrate substantial equivalence based on similar performance between the new device, the VistaCam iX "Proof"and the predicate device, Spectra Fluorescence Caries Detection Aid. Kev performance attributes tested and compared include:
- a. LED Illumination and output
- b. Image Quality
- c. Color Separation
Validation and verification test results showed that new device and the predicate device are equivalent, and that illumination and fluorescence of potential caries detection products are similar for both Dürr Dental AG's VistaCam iX "Proof" and Air Techniques' Spectra camera.
## 10. Performance Testing
Dürr Dental AG's VistaCam iX "Proof" complies with IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (30 edition) - Medical electrical equipment -Part 1: General requirements for basic safety and essential performance, IEC 60601-1-2:2014 (4" edition) - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests, and IEC 80601-2-60:2012 (first edition) Medical electrical equipment - Part 2-60: Particular requirements for basic safety and essential performance of dental equipment.
## 11. Biocompatibility Assessment
There is no biocompatibility issues known to be associated with any of the materials used to manufacture the VistaCam iX "Proof". Patient contact Distance Spacer component was tested and complies with ISO 10993-10:2002 Standard and Amendment 1, "Biological Evaluation of Medical Devices, Part 10-Tests for Irritation and Delayed-Type Hypersensitivity", and ISO 10993-5. Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
By design, many of the components of the VistaCam iX "Proof" are isolated from patient contact by a camera cover or through a distance spacer. Components that have no patient contact are:
{9}------------------------------------------------
- Hand piece housing ASA+PC (Acrylonitrile Styrene Acrylate + ● Polycarbonate)
- Lens window- Glass* .
- Handpiece control buttons Polyester ●
- Umbilical Silicone .
Components that contact the patient are:
- Camera cover* Ethylene methyl acrylate copolymer and low density polyethylene
- 8 mm Spacer* ●
- Septon plus a UV stabilizer (2-5-chloro-2H-benzotriazole-2-yl)-6-(1,1-dimethylethyl)- 4-methylphenol (CAS-No. 3896-11-5) with the restriction SML(T) =30 mg/kg
*These materials are currently in use and market distribution with the predicate device "Spectra Fluorescence Caries Detection Aid System" device, 510K# K090169.
## 12. Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate device in terms of technology, performance and indications for use, Dürr Dental AG concludes that the VistaCam iX "Proof" is substantially equivalent to the predicate device as described herein.
The differences between the new device and the predicate device shown in the comparison table above do not raise any new questions about safety and effectiveness and so we consider it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.