K150668 · Cook Biotech Incorprated · FTM · Dec 9, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150668
Device Name
Biodesign Enterocutaneous Fistula Plug
Applicant
Cook Biotech Incorprated
Product Code
FTM · General, Plastic Surgery
Decision Date
Dec 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biodesign® Enterocutaneous Fistula Plug is intended for implantation to reinforce soft tissue for repair of enterocutaneous fistulas. The device is supplied sterile and is intended for one-time use.
Device Story
Bioabsorbable, collagen-derived device for enterocutaneous fistula repair. Consists of implanted SIS (small intestinal submucosa) plug, flange/gasket assembly, suture tether, and Molnar Disc; includes delivery system (pusher, dilator, sheath, transfer tube). SIS plug fills fistula tract, remodeling into patient tissue; flange/gasket assembly seals internal opening to prevent fluid ingress; suture connects internal assembly to external Molnar Disc to maintain tension. Used by clinicians in surgical settings. Device provides mechanical reinforcement and biological scaffold for tissue healing; temporary components pass naturally or fall off skin.
Clinical Evidence
Bench testing included leak resistance, pushability, deployment, gasket expansion, and tensile strength. GLP animal study in domestic swine model demonstrated fistula closure and complete plug incorporation at five weeks with no negative clinical sequelae.
Indicated for implantation to reinforce soft tissue for repair of enterocutaneous fistulas.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9, 2015
Cook Biotech Incorprated Mr. Nick Wang Regulatory Affairs Scientist 1425 Innovation Place West Lafayette, Indiana 47906
Re: K150668
Trade/Device Name: Biodesign Enterocutaneous Fistula Plug Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM Dated: March 13, 2015 Received: November 10, 2015
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150688
Device Name Biodesign® Enterocutaneous Fistula Plug
#### Indications for Use (Describe)
The Biodesign® Enterocutaneous Fistula Plug is intended for implantation to reinforce soft tissue for repair of enterocutaneous fistulas. The device is supplied sterile and is intended for one-time use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## 13 March 2015
### Cook Biotech Incorporated
# Biodesign® Enterocutaneous Fistula Plug
| Manufacturer Name: | Cook Biotech Incorporated<br>1425 Innovation Place<br>West Lafayette, Indiana 47906<br>Telephone: +1 (765) 497-3355<br>Fax: +1 (765) 807-7709 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Perry W. Guinn, VP of Quality Assurance and<br>Regulatory Affairs |
| Device Name and Classification: | |
# e Name and Classification:
| Trade Name: | Biodesign® Enterocutaneous Fistula Plug |
|-------------------------------|----------------------------------------------------|
| Common/Usual Name: | Surgical Mesh |
| Proposed classification name: | Surgical Mesh<br>21 CFR 878.3300 (FTM)<br>Class II |
# Indications for Use:
The Biodesign® Enterocutaneous Fistula Plug is intended for implantation to reinforce soft tissue for repair of enterocutaneous fistulas. The device is supplied sterile and is intended for one-time use.
### Predicate Device:
The predicate device is the Surgisis® Biodesign® Enterocutaneous Fistula Plug (K082682), also manufactured by Cook Biotech Incorporated.
# Device Description:
The Biodesign® Enterocutaneous Fistula Plug is a bioabsorbable, collagen-derived, minimally invasive treatment option for the repair of enterocutaneous fistulas. The device consists of two parts - an implanted device and a delivery system. The implanted device
{4}------------------------------------------------
has four components - the small intestinal submucosa (SIS) plug, the flange/gasket assembly, the suture tether and the Molnar Disc. The SIS plug is manufactured from porcine small intestine that has been stripped of its serosal, mucosal, and muscle layers and virally inactivated. The resulting acellular collagenous layer, termed Small Intestinal Submucosa, is manufactured into a rolled, freeze dried plug. The function of the SIS plug is to fill and aid in the healing of the tract. The plug fully remodels into patient tissue over time. The other three components, the flange/gasket assembly, the suture tether and the Molnar Disc are all temporary implants designed to either naturally pass out of the body or fall off the skin. The function of the flange/gasket assembly is to seal the internal opening of the fistula tract and prevent the ingress of any gastric/intestinal fluid from entering the fistula tract. The suture connects the flange assembly at the internal opening of the fistula to the Molnar Disc at the external opening. Tethering the suture to the Molnar Disc provides tension to keep the flange/gasket assembly securely in place. In addition to the implanted device, a delivery system is provided to ensure proper device delivery and deployment.
### Equivalence of Market Device:
The Biodesign® Enterocutaneous Fistula Plug, the subject device, is a design improvement of the predicate device, Surgisis® Biodesign® Enterocutaneous Fistula Plug (K082682). The modifications described in this 510(k) are made based on user feedback and clinical experience resulting from use of the predicate device. Analysis and comparison of the intended use, and material and technological characteristics, in conjunction with relevant testing, support the determination that the subject and predicate devices are substantially equivalent. A brief summary of the relevant testing and the substantial equivalence comparison are provided below in this 510(k) summary.
### Biocompatibility
The biocompatibility assessment of the Biodesign® Enterocutaneous Fistula Plug was conducted in accordance with FDA's biocompatibility testing guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' (May 1, 1995). Based on the standard, the plug, the flange/gasket assembly, the delivery system and the Molnar Disc each have different classifications. The classifications are summarized in Table 5-1.
{5}------------------------------------------------
| Device | Biocompatibility Classification | | |
|---------------------------|------------------------------------|--------------------------------|-----------------------------------------|
| Component | Device Type | Contact | Duration |
| SIS Plug | Implant device | Tissue contacting | Permanent (>30 days) |
| Flange/gasket<br>assembly | Surface device | Mucosal membrane<br>contacting | Prolonged contact (24<br>hrs - 30 days) |
| Delivery<br>device | Externally<br>communicating device | Tissue contacting | Limited (<24 hrs) |
| Molnar Disc | Surface device | Skin contacting | Prolonged contact (24<br>hrs - 30 days) |
Table 5-1. Biocompatibility Classification
Relevant biocompatibility testing was performed based the biocompatibility classification of each of the components.
For example, the following biocompatibility tests were performed on portions classified to be permanent tissue implant (e.g. the SIS material):
- Genotoxicity
- Cytotoxicity ●
- Muscle implantation ●
- Acute intracutaneous reactivity ●
- Skin irritation ●
- . ISO sensitization
- Acute systemic toxicity ●
- . Subchronic systemic toxicity
The results of the biocompatibility assessment provided evidence that the Biodesign® Enterocutaneous Fistula Plug meets the biocompatibility requirements of the ISO-10993 standard.
# Mechanical Testing
Mechanical testing was conducted to ensure the device design is appropriate and the device is able to function as intended. Tests performed include:
- . Leak resistance
- Pushability
- Deployment ●
- Gasket expansion ●
- Leak resistance .
{6}------------------------------------------------
- Tensile strength
- Two-plug deployment ●
The same types of testing were used to support the clearance of the predicate device (K082682). All testing were performed on terminally sterilized devices.
# Animal Testing
A GLP animal study was performed using the Biodesign® Enterocutaneous Fistula Plug (subject device) in a domestic swine model. The purpose of the animal study was to characterize the safety and biological response of the Biodesign® Enterocutaneous Fistula Plug under a simulated use animal model. Endpoints of the study included evaluation of local, regional responses of the intestine and adjacent organs, as well as the time over which the flange portion of the device remained in the luminal surface of the intestine. The study showed closure of the surgically created fistulas, complete incorporation of enterocutaneous fistula plugs at five weeks, with no negative clinical sequelae.
# Conclusions drawn from tests
Evaluation of the subject device, Biodesign® Enterocutaneous Fistula Plug, provides evidence of its suitability for use in soft tissue repair of enterocutaneous fistulas and substantial equivalency to the predicate device in terms of intended use, material, and technological characteristics.
# Substantial Equivalence
All differences between the predicate and the subject device were analyzed; the differences did not change the intended use, fundamental mode of action, or introduce new types of questions in risks or effectiveness, thus supporting a determination that the subject and predicate devices are substantially equivalent. See Table 5-2 below for a comparison of the subject and predicate devices.
| | Biodesign® Enterocutaneous<br>Fistula Plug<br>(subject device) | Surgisis® Biodesign®<br>Enterocutaneous Fistula Plug<br>(predicate device) |
|---------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Manufacturer | Cook Biotech Incorporated | Cook Biotech Incorporated |
| 510(k) number | Unassigned | K082682 |
| Indication | Implantation to reinforce soft<br>tissue for repair of<br>enterocutaneous fistulas | Implantation to reinforce soft<br>tissue for repair of<br>enterocutaneous fistulas |
{7}------------------------------------------------
| Mode of action | Enterocutaneous fistulas are repaired by sealing the internal opening of the fistula and filling the fistula tract by using cylindrical SIS plugs | Enterocutaneous fistulas are repaired by sealing the internal opening of the fistula and filling the fistula tract by using cylindrical SIS plugs |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Components | Implanted Device<br>- Plug<br>- Suture<br>- Molnar Disc<br>- Temporary flange<br>- Expandable gasket<br>Delivery System<br>- Pusher<br>- Dilator<br>- Sheath with Captor valve<br>- Transfer tube | Implanted Device<br>- Plug<br>- Suture<br>- Molnar Disc<br>- Temporary flange<br>Delivery System<br>- Pusher<br>- Dilator<br>- Sheath with Captor valve<br>- Transfer tube |
| Plug<br>dimensions | Length: 180 mm<br>Diameter: 4 mm | Length: 180 mm<br>Diameter: 4 and 7 mm |
| Implant<br>materials | - SIS lyophilized, acellular, collagenous, porcine small intestinal mucosa (plug)<br>- Nitinol (flange)<br>- 316 stainless steel (pin and radio-opaque marker)<br>- Polyurethane (flange)<br>- Polydioxanone (PDO) (tether)<br>- Crosslinked SIS (gasket) | - SIS lyophilized, acellular, collagenous, porcine small intestinal mucosa (plug)<br>- Nitinol (flange)<br>- 316 stainless steel (radio-opaque marker)<br>- Polyurethane (flange)<br>- Polydioxanone (PDO) (tether)<br>- Poly (lactic-co-glycolic acid) (PLGA) (pin) |
| Supplied sterile | Yes | Yes |
| Sterilization<br>method | Ethylene Oxide | Ethylene Oxide |
| Intended for | Yes | Yes |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.