The Optilite Freelite Kappa Free Kit is intended for the quantitative in vitro measurement of kappa free light chains in serum using the Binding Site Optilite turbidimetric analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. The Optilite Freelite Lambda Free Kit is intended for the quantitative in vitro measurement of kappa free light chains in serum using the Binding Site Optilite turbidimetric analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings.
Device Story
Optilite Freelite Kappa and Lambda Free Kits are in vitro diagnostic reagents for use on the Binding Site Optilite turbidimetric analyser. The system measures free light chains in serum samples via turbidimetric immunoassay. The analyser automates sample processing and optical measurement; results are provided to clinicians to assist in diagnosing and monitoring plasma cell dyscrasias and related connective tissue disorders. The device is intended for professional use in clinical laboratory settings. Output data supports clinical decision-making regarding disease status and treatment efficacy.
Clinical Evidence
No clinical data provided; substantial equivalence is supported by analytical performance and bench testing on the Optilite turbidimetric analyser.
Technological Characteristics
Turbidimetric immunoassay reagents for use on the Optilite turbidimetric analyser. Measures kappa and lambda free light chains in serum. System utilizes automated optical turbidimetry. Connectivity and software are integrated into the Optilite analyser platform.
Indications for Use
Indicated for quantitative measurement of kappa and lambda free light chains in serum to aid in diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinaemia, AL amyloidosis, light chain deposition disease, and connective tissue diseases (e.g., SLE) in adult patients.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Predicate Devices
Freelite® Human Kappa Free kit for use on Roche Cobas Integra® 400/400plus (K070900)
Freelite® Human Lambda Free kit for use on Roche Cobas Integra® 400/400plus (K070900)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K150658
B. Purpose for Submission:
New device
C. Measurand:
Kappa (κ) free light chains
Lambda (λ) free light chains
D. Type of Test:
Turbidimetry, quantitative
E. Applicant:
The Binding Site Group, Ltd.
F. Proprietary and Established Names:
Optilite® Freelite® Kappa Free Kit
Optilite® Freelite® Lambda Free Kit
G. Regulatory Information:
1. Regulation section:
21 CFR§866.5550 – Immunoglobulin (light chain specific) immunological test system
2. Classification:
Class II
3. Product code:
DFH, Kappa, antigen, antiserum, control
{1}
DEH, Lambda, antigen, antiserum, control
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
The Optilite Freelite Kappa Free Kit is intended for the quantitative in vitro measurement of kappa free light chains in serum using the Binding Site Optilite turbidimetric analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings.
The Optilite Freelite Lambda Free Kit is intended for the quantitative in vitro measurement of kappa free light chains in serum using the Binding Site Optilite turbidimetric analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
For prescription use only.
Warning: The kappa free light chain results for a given specimen determined with assays from different manufacturers or on different systems can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the kappa free light chain assay used. Values obtained with different assays or systems cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay or system used for determining kappa free light chain levels is changed, additional sequential testing should be carried out. Prior to changing assay or system, the laboratory MUST confirm baseline values for patients being serially monitored.
Warning: The lambda free light chain results for a given specimen determined with assays from different manufacturers or on different systems can vary due to differences in assay methods and reagent specificity. The results reported by the
{2}
laboratory to the physician must include the identity of the lambda free light chain assay used. Values obtained with different assays or systems cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay or system used for determining lambda free light chain levels is changed, additional sequential testing should be carried out. Prior to changing assay or system, the laboratory MUST confirm baseline values for patients being serially monitored.
4. Special instrument requirements:
Optilite® Analyzer (K141100)
I. Device Description:
The Optilite® Freelite® Kappa Free Kit and Optilite® Freelite® Lambda Free Kit are comprised of the following reagents:
- Latex Reagent: Consisting of polyclonal monospecific antibody coated onto polystyrene latex. Supplied in stabilized liquid form. Preservatives: 0.1% E-amino-n-caproic acid (EACA), 0.01% benzamidine, 0.05% ProClin.
- Calibrator and Controls: Pooled human serum, supplied in stabilized liquid form. Containing 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives.
- Reaction Buffer: Containing 0.099% sodium azide as a preservative.
J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s):
Freelite® Human Kappa Free kit for use on Roche Cobas Integra® 400/400plus (K070900)
Freelite® Human Lambda Free kit for use on Roche Cobas Integra® 400/400plus (K070900)
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device
Optilite® Freelite® Kappa Free Kit and
Optilite® Freelite® Lambda Free Kit | Predicate
Freelite® Human Kappa Free Kit and
Freelite® Human Lambda Free Kit |
| Intended use | Quantitative in vitro measurement of kappa free light chains (FLC) or lambda FLC in serum | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device
Optilite® Freelite®
Kappa Free Kit and
Optilite® Freelite®
Lambda Free Kit | Predicate
Freelite® Human
Kappa Free Kit and
Freelite® Human
Lambda Free Kit |
| Indication for use | Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. | Same |
| Analyte | Kappa: Kappa FLC
Lambda: Lambda FLC | Same |
| Measurement | Quantitative | Same |
| Detection Method | Turbidimetric | Same |
| Sample type | Serum | Same |
| Detection Antibody | Kappa: Polyclonal sheep anti-human kappa antibody (F(ab)2 fragment) coated onto latex particles
Lambda: Polyclonal sheep anti-human lambda antibody (F(ab)2 fragment) coated onto latex particles | Kappa: Same
Lambda: Same |
| Open Vial Stability | 3 months | Same |
| Reference Interval | Kappa: 3.30–19.40 mg/L
Lambda: 5.71–26.30 mg/L
Ratio: 0.26–1.65 mg/L | Same |
| Lambda Sample dilutions | Lambda: 1+1, 1+7, 1+79, 1+799, 1+7999 | Same |
| Lambda Measuring Range (mg/L) | Lambda: At 1+ 7 standard dilution: 52–1390 mg/L
Extended Measuring range: 1.3–34.7 mg/L (1+1) | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device
Optilite® Freelite®
Kappa Free Kit and
Optilite® Freelite®
Lambda Free Kit | Predicate
Freelite® Human
Kappa Free Kit and
Freelite® Human
Lambda Free Kit |
| | 52–1390 mg/L (1+79)
520–13900 mg/L (1+799)
5200–139000 mg/L (1+7999) | |
| Differences | | |
| --- | --- | --- |
| Item | Device
Optilite® Freelite®
Kappa Free Kit and
Optilite® Freelite®
Labmda Free Kit | Predicate
Freelite® Human
Kappa Free Kit and
Freelite® Human
Lambda Free Kit |
| Instrument | Binding Site Optilite | Roche Cobas Integra®
400/400plus |
| Kappa
Sample dilutions | Kappa:
1+1, 1+9, 1+99, 1+999 | Kappa:
1+1, 1+9, 1+99 |
| Kappa
Measuring Range
(mg/L) | Kappa:
At 1+9 Standard Dilution”
2.9–127 mg/L
Extended Measuring range:
0.6–25.3 mg/L (1+1)
29–1270 mg/L (1+99)
290–12700 mg/L (1+999)
1450–63500 mg/L (1+4999) | Kappa:
At 1+9 Standard
Dilution”
2.9–127 mg/L
Extended Measuring
range:
0.6–25.3 mg/L (1+1)
29–1270 mg/L (1+99) |
| On-board
Stability | 30 days | 3 months |
| Open Vial
Stability | 12 weeks | 3 months |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2 “Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition”.
CLSI EP06-A “Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline”.
CLSI EP07-A2 “Interference Testing in Clinical Chemistry, Approved Guideline - Second Edition”.
{5}
CLSI EP17-A “Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline”.
CLSI C28-A3 “Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory”.
## L. Test Principle:
The concentration of the soluble antigen is assessed by turbidimetry. The test sample is added to a solution containing the appropriate antibody in a reaction cuvette. A beam of light is passed through the cuvette and is increasingly scattered by the formation of insoluble immune complexes. Light scatter is monitored by measuring the decrease in intensity of the incident beam of light. The antibody in the cuvette is in excess so the amount of immune complex formed is proportional to the antigen concentration. A series of calibrators of known antigen concentration are assayed to produce a calibration curve of measured light scatter versus antigen concentration. Samples of unknown antigen concentration can then be assayed and the results read from the calibration curve.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision studies were based on CLSI EP5-A2, where eight serum samples were evaluated with both assays. The samples chosen covered the assays' measuring ranges: two samples were close to the low medical decision points (Kappa: 2.48 mg/L and 4.13 mg/L, Lambda: 4.28 mg/L and 7.14 mg/L), one sample was a negative (healthy) sample with a concentration within the reference range (Kappa: 8 mg/L, Lambda: 13 mg/L), two samples were close to the high medical decision points (Kappa: 14.55 mg/L and 24.25 mg/L, Lambda: 19.73 mg/L and 32.88 mg/L), and three samples were highly elevated clinical samples (both Kappa and Lambda: 75 mg/L, 110 mg/L and 400 mg/L)) were tested in two runs per day (in duplicate) over 21 days using three analyzers. Samples were run at the standard dilution (1+9 for Kappa and 1+7 for Lambda) unless otherwise indicated. Results met the sponsor's predetermined acceptance criteria for total precision (%CV<8.5%), within-run precision (%CV<5%), between-run precision (%CV<8.5%), between-day precision (%CV<8.5%) and between-instrument precision (%CV<10%).
Optilite® Freelite® Kappa Free Kit
| Kappa FLC | N | Mean g/L | Within-Run | | Between-Run | | Between-Day | | Between-Instrument | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| 1* | 84 | 2.5 | 0.1 | 2.6 | 0.00 | 0.8 | 0.1 | 3.5 | 0.1 | 4.1 | 0.1 | 4.4 |
| 2 | 84 | 4.9 | 0.1 | 2.6 | 0.1 | 2.9 | 0.3 | 6.6 | 0.1 | 2.4 | 0.4 | 7.7 |
{6}
7
| Kappa FLC | N | Mean g/L | Within-Run | | Between-Run | | Between-Day | | Between-Instrument | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| 3 | 84 | 8.2 | 0.2 | 2.1 | 0.1 | 1.7 | 0.3 | 4.1 | 0.2 | 2.7 | 0.4 | 5.0 |
| 4 | 84 | 13.7 | 0.2 | 1.7 | 0.1 | 1.0 | 0.6 | 4.0 | 0.4 | 3.2 | 0.6 | 4.5 |
| 5 | 84 | 23.2 | 0.3 | 1.5 | 0.4 | 1.8 | 0.6 | 2.7 | 0.5 | 2.2 | 0.8 | 3.5 |
| 6 | 84 | 71.8 | 2.4 | 3.3 | 1.3 | 1.8 | 2.1 | 3.0 | 1.6 | 2.3 | 3.4 | 4.8 |
| 7 | 84 | 105.1 | 5.5 | 5.2 | 0.0 | 0.0 | 4.5 | 4.2 | 4.1 | 3.9 | 7.1 | 6.7 |
| 8** | 84 | 329.8 | 13.3 | 4.0 | 5.7 | 1.7 | 13.2 | 4.0 | 5.4 | 1.6 | 19.6 | 5.9 |
*performed at the 1+1 sample dilution
**performed at the 1+99 sample dilution
Optilite® Freelite® Lambda Free Kit
| Lambda FLC | N | Mean g/L | Within-Run | | Between-Run | | Between-Day | | Between-Instrument | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| 1* | 84 | 4.7 | 0.1 | 2.1 | 0.1 | 2.7 | 0.2 | 3.2 | 0.1 | 3.0 | 0.2 | 4.7 |
| 2 | 84 | 7.5 | 0.2 | 3.0 | 0.3 | 3.7 | 0.5 | 7.2 | 0.67 | 8.6 | 0.7 | 8.6 |
| 3 | 84 | 14.4 | 0.2 | 1.6 | 0.2 | 1.5 | 0.5 | 3.2 | 0.5 | 3.6 | 0.6 | 3.9 |
| 4 | 84 | 19.8 | 0.4 | 1.8 | 0.4 | 20. | 0.4 | 2.1 | 0.5 | 2.9 | 0.7 | 3.4 |
| 5 | 84 | 30.0 | 0.6 | 2.0 | 0.5 | 1.8 | 0.8 | 2.6 | 0.9 | 2.9 | 1.1 | 3.7 |
| 6 | 84 | 71.1 | 1.3 | 1.8 | 1.5 | 2.0 | 1.7 | 2.4 | 1.3 | 1.8 | 2.6 | 3.6 |
| 7 | 84 | 115.6 | 4.0 | 3.4 | 4.9 | 4.2 | 5.3 | 4.6 | 3.2 | 2.7 | 8.2 | 7.1 |
| 8** | 84 | 335.4 | 4.6 | 1.4 | 10.7 | 3.2 | 15.4 | 4.6 | 18.3 | 5.4 | 19.2 | 5.8 |
*performed at the 1+1 sample dilution
**performed at the 1+79 sample dilution
Reagents used in the Optilite® Freelite® Kappa and Lambda Free Kits are identical to those of the predicate device. Lot-to-lot studies were previously conducted in K070900.
b. Linearity/assay reportable range:
A linearity study was performed following CLSI EP06-A. The linearity of this assay has been confirmed using a full range pool spanning the claimed measuring range and a low range pool covering the normal range. These samples were serially diluted to cover the range at the standard dilution (1+9 for Kappa and 1+7 for Lambda).
Samples were run in triplicate. The linearity of the Optilite® Freelite® Kappa Kit was confirmed using serially diluted serum samples spanning the range of 2.9–127 mg/L with deviation from linearity <10%. The linearity of the Optilite® Freelite® Lambda Free Kit was confirmed using a serially diluted serum samples spanning the range of 5.2–139 mg/L with deviation from linearity <10%.
{7}
| Kit | Assay reportable range
Standard Dilution | Assay reportable range
Expanded dilutions |
| --- | --- | --- |
| Optilite® Freelite® Kappa Free Kit | 2.9–127 mg/L | 0.6–63500 mg/L |
| Optilite® Freelite® Lambda Free Kit | 5.2–139 mg/L | 1.3–39000 mg/L |
For each sample pool, the mean observed value was also plotted against the calculated concentrations, and the relationship between the two was determined by both ordinary and weighted linear regression. Regression equations for the Optilite® Freelite® Kappa Free Kit and Optilite® Freelite® Lambda Free Kit high/full linear range and for the low linear range are as follows:
| Analyte | Pool series | Sample Range (g/L) | Ordinary Linear Regression equation | Weighted Linear Regression equation | R |
| --- | --- | --- | --- | --- | --- |
| Kappa FLC | Full Range Pool | 2.55–140.34 | y = 1.00x + 0.00 | y = 1.01x - 0.09 | 0.99 |
| | Low Range Pool | 2.60–40.7 | y = 1.00x + 0.00 | y = 0.99x + 0.12 | 0.99 |
| Lambda FLC | Full Range Pool | 4.33–155.45 | y = 1.00x - 0.00 | y = 1.01x - 0.17 | 0.99 |
| | Low Range Pool | 4.09–46.94 | y = 1.00x - 0.00 | y = 1.00x + 0.02 | 0.99 |
The approximate measuring range of the Optilite® Freelite® Kappa Free Kit at the standard dilution (1+9) is 2.9–127 mg/L.
The approximate measuring range of the Optilite® Freelite® Lambda Free Kit at the standard dilution (1+7) is 5.2–139 mg/L.
## Antigen Excess:
Prozone/Hook effect parameters are in effect to protect the Optilite® analyzer from antigen excess effects. A sample with atypical reaction kinetics will have a 'Prozone detected' flag (prozone warning) in the results entry and the sample will be automatically re-measured at a higher analyzer dilution.
To test this warning system, reaction kinetics of normal and high level clinical samples were tested. No samples within the normal rage produced false prozone detection warnings. All samples exceeding the measuring range gave either a 'greater
{8}
than' or 'prozone detected' flag and the sample was automatically rerun at the next appropriate dilution.
Because the possibility of antigen excess can never be completely excluded and to address the potential for monoclonal samples to exhibit non-linearity, the following warning statement is included in the reagent package inserts:
**Important Note:** No automated check will identify all cases of antigen excess and a very small percentage of samples in antigen excess have normal reaction kinetics so will not prompt the "Prozone Detected" flag. It is recommended that the following statement accompany all free light chain results.
"Undetected antigen excess is a rare event but cannot be excluded. If these free light chain results do not agree with other clinical or laboratory findings, or if the sample is from a patient that has previously demonstrated antigen excess, the result must be checked by retesting at a higher sample dilution. Results should always be interpreted in conjunction with other laboratory tests and clinical evidence; any anomalies should be discussed with the testing laboratory."
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Traceability:**
In the absence of an international reference standard, the calibration of the assay is traceable to an internally assigned master calibrator.
**Kit Stability:**
**Real-time stability:** Studies to establish shelf-life stability (from the date of manufacture when stored at recommended temperature 2–8°C) of the predicate devices were carried out between 2008 and 2010. The reagents in the Optilite® Freelite® Kappa and Lambda Free kits are identical to those in the predicate devices; the predicate stability claim was able to be transferred. Data supports an 18-month stability claim.
**Open-vial stability:** Open vial stability claims from the predicate were revalidated on the Optilite® analyzer. The Optilite® Freelite® Kappa and Lambda Free Kit reagents can be stored opened at 2–8°C for up to 12 weeks.
**On-board stability:** On-board stability claims from the predicate were revalidated on the Optilite® analyzer. The Optilite® Freelite® Kappa and Lambda Free Kit reagents can be stored on-board the Optilite® Analyzer for 30 days.
d. Detection limit:
The analytical sensitivity of both kits was determined in accordance with CLSI EP17-
{9}
A. The Limit of Blank (LoB) was based on 60 determinations of a sample containing analyte depleted serum and sample diluent and was estimated as the $95\%$ percentile of the distribution. The Limit of Detection (LoD) was calculated according to the equation $\mathrm{LoD} = \mathrm{LoB} + 1.645 \times \mathrm{SDs}$ where SDs, the standard deviation was based on 12 determinations each of five LoQ samples with analyte levels near the lower limit of the reportable range. Total error $(0.29 \mathrm{mg} / \mathrm{L}$ for Kappa and $0.17 \mathrm{mg} / \mathrm{L}$ for Lambda) at LoQ was within the sponsor's pre-determined criteria for maximum allowable total error.
| Optilite® Freelite Assay | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| Kappa Free Kit | 0.100 mg/L | 0.158 mg/L | 0.580 mg/L |
| Lambda Free Kit | 0.185 mg/L | 0.274 mg/L | 1.300 mg/L |
# e. Analytical specificity:
Interferences were assessed according to CLSI EP7-A2 by testing six serum samples for each assay with Kappa FLC and Lambda FLC concentrations falling below (deficient), within (healthy), above (elevated) or close to the limits of the reference interval. Each sample was spiked with interfering substances and tested in a minimum of three replicates. For non-interference to be claimed, the mean results from the spiked samples must be within $10\%$ of the mean of the control samples.
Kappa results: The data demonstrated that the assay was not affected by the following substances at the concentrations given below.
Lambda results: The data demonstrated that the assay was not affected by the following substances at the concentrations given below.
| Category | Interferent | Optilite® Freelite® Kappa Free Kit | Optilite® Freelite® Lambda Free Kit |
| --- | --- | --- | --- |
| | | Concentration tested | Concentration tested |
| Endogenous Substances | Bilirubin | 200 mg/L | 200 mg/L |
| | Hemoglobin | 0.625 g/L* | 5 g/L |
| | Intralipid | 62.5 mg/dL* | 1000 mg/dL |
| | Triglyceride | 1000 mg/dL | 2000 mg/dL |
| OTC Drugs | Acetaminophen | 1324 μmol/L | 1324 μmol/L |
| | Acetylsalicylic acid | 3.63 μmol/L | 3.63 μmol/L |
| | Ascorbic acid | 342 μmol/L | 342 μmol/L |
| | Caffeine | 308 μmol/L | 308 μmol/L |
| | Ibuprofen | 1212.5 μmol/L | 2425 μmol/L |
| Common Drugs | Cimetidine | 79.2 μmol/L | 79.2 μmol/L |
| | Digoxin | 7.8 nmol/L | 7.8 nmol/L |
{10}
| Category | Interferent | Optilite® Freelite® Kappa Free Kit | Optilite® Freelite® Lambda Free Kit |
| --- | --- | --- | --- |
| | | Concentration tested | Concentration tested |
| | Penicillin | 75 mg/L | 75 mg/L |
| | Phenotoin | 198 μmol/L | 198 μmol/L |
| | Theophylline | 222 μmol/L | 222 μmol/L |
| Specific Drugs | Bortezomib | 6 mg/mL | 3 mg/mL |
| | Cyclophosphamide Monohydrate | 60 μg/mL | 60 μg/mL |
| | Pomalidomide | 100 μg/mL | 100 μg/mL |
| | Prednisolone | 100 μg/mL | 100 μg/mL |
*Hemoglobin and Intralipid passed at very low interferent concentrations. The Optilite® Freelite® Kappa Free Kit is therefore not suitable for use with hemolyzed or lipemic samples. This is clearly stated in the product insert.
f. Assay cut-off:
See expected values/reference range.
2. Comparison studies:
a. Method comparison with predicate device:
The Binding Site Optilite FREELITE™ assays for Kappa and Lambda were evaluated and compared to the results obtained with the kits on the predicate analyzer. A total of 292 serum samples spanning the dynamic range of one or both assays were used in this study; 238 of these samples were run on the Kappa assays, 192 were run on the Lambda assays, 143 of these samples were run on both Kappa and Lambda assays. Samples were tested in singlicate. Samples included nine normal donors and 74 clinical samples with relevant admission diagnosis (including multiple myeloma, Waldenström's Macroglobulinemia, lymphocytic neoplasms and systemic lupus erythematosus). The remainder of samples had unknown clinical status.
| Analyte | N | Sample Range (g/L) | Passing Bablok Regression equation | Slope (95% CI) | Y-Intercept (95% CI) | R² |
| --- | --- | --- | --- | --- | --- | --- |
| Kappa FLC | 238 | 3.42 – 45578.13 | y = 0.95x - 0.74 | 0.95 (0.92 to 0.97) | -0.74 (-1.29 to -0.18) | 0.979 |
| Lambda FLC | 192 | 1.91 – 5089.37 | y = 1.01x + 0.39 | 1.01 (0.98 to 1.03) | 0.39 (0.03 to 0.84) | 0.981 |
| κ/λ FCL Ratio | 143 | 0.002 – 603.42 | y = 0.88x + 0.02 | 0.88 (0.84 to 0.93) | 0.02 (-0.03 to 0.05) | 0.991 |
{11}
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable. See assay cut-off.
5. Expected values/Reference range:
The reference ranges were transferred from the predicate devices and were verified by testing 50 adult donor samples.
Kappa FLC reference range: 3.30–19.40 mg/L
Lambda FLC reference range: 5.71–26.30 mg/L
Kappa/Lambda Ratio reference Range: 0.26–1.65 mg/L
In the product insert, the sponsor recommends that each user of the kit should verify the transferability of the expected values to its own population and if necessary determine its own reference interval.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.