CONSTRUX Mini PEEK Spacer System, CONSTRUX Mini PEEK Ti Spacer System
Applicant
Orthofix, Inc.
Product Code
ODP · Orthopedic
Decision Date
Sep 11, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K150619 · Sep 11, 2015
CONSTRUX Mini PEEK Spacer System, CONSTRUX Mini PEEK Ti Spacer System
Orthofix, Inc.
Published clinical literature
The sponsor provided published clinical data for similar PEEK interbody fusion devices to demonstrate that the use of these devices for one or two contiguous levels in anterior cervical fusion procedures poses no new or additional risks to patients.
Literature review; Cervical fusion; PEEK interbody devices
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published clinical data for cervical interbody fusion devices; Literature review of clinical outcomes
Safety and clinical outcomes for one or two contiguous levels
Indications for Use
When used as a cervical intervertebral body fusion device, the CONSTRUX Mini PEEK Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain with degeneration of the disc confirmed by patient history and radiographic studies. The CONSTRUX Mini PEEK Spacer System is intended for use with autograft and /or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation (i.e. anterior cervical plate such as the Orthofix ACP or Hallmark® System). Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Spacer System in the cervical spine. When used as a Partial Vertebral Body Replacement (VBR) System, the CONSTRUX Mini PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial replaced vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The CONSTRUX Mimi PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine. The CONSTRUX Mini PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft. The CONSTRUX Mini PEEK Spacer System is intended for use with internal fixation. The supplemental internal fixation system that may be used with the CONSTRUX Mini PEEK Spacer System is the Orthofix Spinal Fixation System (SFS) or the Firebird Spinal Fixation System. The CONSTRUX Mini PEEK Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. The CONSTRUX Mini PEEK Ti Spacer System is intended for use with autograft and /or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system (i.e. anterior cervical plate such as the Orthofix ACP or Hallmark® System). Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Ti Spacer System in the cervical spine.
Device Story
Intervertebral body fusion and partial vertebral body replacement system; implants placed in cervical or thoracolumbar disc space; stabilizes adjacent vertebrae; maintains disc height; prevents vertebral collapse; restores biomechanical integrity of spinal column. Used with autograft/allograft and supplemental internal fixation (e.g., anterior cervical plates, spinal fixation systems). Operated by surgeons in clinical/surgical settings. Output is physical stabilization of the spine to facilitate fusion. Benefits include pain relief and structural support for patients with DDD, tumors, or fractures.
Clinical Evidence
No new clinical testing performed. Evidence consists of published clinical literature regarding PEEK interbody devices for one or two contiguous level anterior cervical fusion procedures, demonstrating no new or additional patient risks.
Technological Characteristics
Implants manufactured from PEEK with titanium markers or PEEK core with integrated porous titanium end plates. Features ridged superior/inferior surfaces for stability. Available in various footprints and heights. Mechanical/structural device; no software or energy source.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) requiring spinal fusion at one or two contiguous levels (C2-T1) or partial vertebral body replacement in the thoracolumbar spine (T1-L5) for tumors, decompression, or fractures. Requires 6 weeks of prior non-operative treatment for cervical indications.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
CONSTRUX Mini PEEK Spacer System and CONSTRUX Mini PEEK Ti Spacer System (K142152)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three faces overlapping to represent the department's focus on people and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 11, 2015
Orthofix, Incorporated Natalia Volosen Senior Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056
Re: K150619
Trade/Device Name: CONSTRUX Mini PEEK Spacer System, CONSTRUX Mini PEEK Ti Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, MQP Dated: August 12, 2015 Received: August 14, 2015
Dear Ms. Volosen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
K150619
page 1 of 2
510(k) Number (if known) K150619
Device Name
CONSTRUX Mini PEEK Spacer System CONSTRUX Mini PEEK Ti Spacer System
### Indications for Use (Describe)
CONSTRUX Mini PEEK Spacer System
When used as a cervical intervertebral body fusion device, the CONSTRUX Mini PEEK Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PEEK Spacer System is intended for use with autograft and /or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation (i.e. anterior cervical plate such as the Orthofix ACP or Hallmark® System).
Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Spacer System in the cervical spine.
When used as a Partial Vertebral Body Replacement (VBR) System, the CONSTRUX Mini PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial replaced vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The CONSTRUX Mimi PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The CONSTRUX Mini PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft.
The CONSTRUX Mini PEEK Spacer System is intended for use with internal fixation. The supplemental internal fixation system that may be used with the CONSTRUX Mini PEEK Spacer System is the Orthofix Spinal Fixation System (SFS) or the Firebird Spinal Fixation System.
### CONSTRUX Mini PEEK Ti Spacer System
The CONSTRUX Mini PEEK Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PEEK Ti Spacer System is intended for use with autograft and /or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system (i.e. anterior cervical plate such as the Orthofix ACP or Hallmark® System).
Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Ti Spacer System in the cervical spine.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. The "®" symbol is located to the right of the word "ORTHOFIX".
# 510(k) SUMMARY
## PEEK Cervical Interbody Devices
| 510(k) Owner Information<br>Name:<br>Address: | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056 |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number:<br>Fax Number:<br>Email: | 214.937.2145<br>214-937-3322<br>nataliavolosen@orthofix.com |
| Registration Number: | 3008524126 |
| Contact Person: | Natalia Volosen<br>Senior Regulatory Affairs Specialist |
| Date Prepared: | September 3, 2015 |
| Name of Device<br>Trade Name / Proprietary<br>Name: | CONSTRUX Mini PEEK Spacer System<br>CONSTRUX Mini PEEK Ti Spacer System |
| Common Name: | Intervertebral body fusion device<br>Spinal intervertebral body fixation orthosis |
| Product Code: | ODP, MQP |
| Regulatory Classification: | Class II per 21 CFR § 888.3080<br>Class II per 21 CFR § 888.3060 |
| Review Panel: | Orthopedic Device Panel |
| Predicate Devices: | K142264 - Valeo Spacer System, SE 12/8/2014, Amedica<br>Corporation (primary predicate device)<br>K142152 – CONSTRUX Mini PEEK Spacer System and<br>CONSTRUX Mini PEEK Ti Spacer System (additional<br>predicate)<br>No reference devices were used in this submission |
| Reason for 510(k) Submission:<br>Expanded Indications for Use | |
### Device Description
CONSTRUX Mini PEEK Spacer System:
The CONSTRUX Mini PEEK Spacer System is comprised of a variety of implants manufactured from PEEK and with titanium markers. The implants are available in two footprint sizes, a small and a large. The implants are available in various heights, in onemillimeter increments. The superior and inferior surfaces of the implant have a pattern of
{5}------------------------------------------------
ridges to provide increased stability and help prevent anterior/posterior movement of the device.
### CONSTRUX Mini PEEK Ti Spacer System:
The CONSTRUX Mini PEEK Ti Spacer System is comprised of a variety of implants that has a PEEK core with integrated porous Titanium end plates. The superior and inferior Titanium plate surfaces of the implant provide increased stability for the implant. The implants are available in three footprint sizes and various heights. The CONSTRUX Mini PEEK Ti spacer is implanted in the cervical intervertebral disc space and is intended to facilitate vertebral fusion by stabilizing adjacent vertebrae, maintaining disc height, and preventing the collapsing of one vertebrate onto another.
### Intended Use / Indications for Use
### CONSTRUX Mini PEEK Spacer System
When used as a cervical intervertebral body fusion device, the CONSTRUX Mini PEEK Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PEEK Spacer System is intended for use with autograft and /or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation (i.e. anterior cervical plate such as the Orthofix ACP or Hallmark® System).
Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Spacer System in the cervical spine.
When used as a Partial Vertebral Body Replacement (VBR) System, the CONSTRUX Mini PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The CONSTRUX Mini PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The CONSTRUX Mini PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft.
The CONSTRUX Mini PEEK Spacer System is intended for use with internal fixation. The supplemental internal fixation system that may be used with the CONSTRUX Mini PEEK Spacer System is the Orthofix Spinal Fixation System (SFS) or the Firebird Spinal Fixation System.
### CONSTRUX Mini PEEK Ti Spacer System
The CONSTRUX Mini PEEK Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in bold, black letters. The "R" in "ORTHOFIX" has a registered trademark symbol next to it.
discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PEEK Ti Spacer System is intended for use with autograft and /or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system (i.e. anterior cervical plate such as the Orthofix ACP or Hallmark® System).
Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Ti Spacer System in the cervical spine.
### Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices
The CONSTRUX Mini PEEK Spacer system and CONSTRUX Mini PEEK Ti Spacer system has the same intended use, similar indications for use, similar technological characteristics and design, similar materials and the same principles of operation as the Amedica Valeo Spacer System (K142264).
### PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
The purpose of this 510(k) submission is to seek clearance to allow one or two contiguous levels use for the CONSTRUX Mini PEEK Spacer System and CONSTRUX Mini PEEK Ti Spacer System implants in the cervical spine. No changes have been made to the actual implants and therefore no mechanical testing was performed to help determine substantial equivalence.
### PERFORMANCE DATA - Summary of Clinical Test Conducted for Determination of Substantial Equivalence
Published clinical data for the cervical interbody fusion devices similar to the CONSTRUX Mini PEEK Spacer System and CONSTRUX Mini PEEK Ti Spacer System, implants was provided in support of this application. The published clinical outcomes demonstrated that the use of PEEK interbody devices for one or two contiguous levels in anterior cervical fusion procedures, as defined above, poses no new or additional risks to patients. No changes were made to the existing devices, nor were any new components added to the system. Therefore, no additional clinical testing was required or performed.
### Basis of Substantial Equivalence
The CONSTRUX Mini PEEK Spacer system and CONSTRUX Mini PEEK Ti Spacer system has the same intended use, similar indications for use, similar technological characteristics and design, similar materials and the same principles of operation as the Amedica Valeo Spacer System (K142264). The minor differences between CONSTRUX Mini PEEK Spacer system and CONSTRUX Mini PEEK Ti Spacer system implant design and the Amedica Valeo Spacer System (K142264) predicate does not introduce any additional patient risk.
No design changes have been made to CONSTRUX Mini PEEK Spacer System and CONSTRUX Mini PEEK Ti Spacer System since their last 510(k) clearance K142152, SE 1/8/2015. The purpose of this 510(k) submission is to seek clearance to allow one or
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. There is a registered trademark symbol after the word "ORTHOFIX".
two contiguous levels use for the CONSTRUX Mini PEEK Spacer System and CONSTRUX Mini PEEK Ti Spacer System implants in the cervical spine.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.