Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
K150603 · Teleflexmedical, Inc. · BTR · Nov 27, 2015 · Anesthesiology
Device Facts
Record ID
K150603
Device Name
Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
Applicant
Teleflexmedical, Inc.
Product Code
BTR · Anesthesiology
Decision Date
Nov 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5730
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Rusch tracheal tubes are indicated for airway management by oral or nasal intubation of the trachea.
Device Story
Sterile, single-use tracheal tube; PVC resin construction (non-DEHP); includes cuff, inflation line, pilot balloon, one-way valve, and radiopaque line. Supplied with 15mm connector. Available in Murphy eye and Magill styles. Used in clinical settings where endotracheal intubation is performed; operated by healthcare professionals. Device maintains open airway for ventilation. Benefits include pediatric-specific sizing and non-DEHP material composition.
Clinical Evidence
Bench testing only. Testing included biocompatibility per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous activity, implantation, genotoxicity, extractables/leachables) and mechanical performance testing per ISO 5361 (tube curvature, collapse, cuff diameter, cuff herniation) and internal specifications (cuff bonding, tube compression, burst pressure, sealing pressure).
Technological Characteristics
Materials: PVC resin (non-DEHP <0.1% w/w), Polyamide connector, Barium Sulfate radiopaque marker. Dimensions: 3.0 mm – 4.5 mm ID. Sterilization: Ethylene Oxide (10-6 SAL). Biocompatibility: ISO 10993-1 compliant. Design: Cuffed, Murphy/Magill eye, beveled tip, multiple cm graduations.
Indications for Use
Indicated for airway management via oral or nasal intubation of the trachea in pediatric patients.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
Rusch Safety Silk Series Oral/Nasal Tracheal Tube (132415)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, while "U.S. FOOD & DRUG" is in a slightly smaller font, and "ADMINISTRATION" is in an even smaller font below.
October 28, 2021
Teleflex Medical, Inc. Lori Pfohl Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, North Carolina 27709
Re: K150603
Trade/Device Name: Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal tube Regulatory Class: Class II Product Code: BTR
Dear Lori Pfohl:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on November 27, 2015. Specifically, FDA is updating this SE letter due to the clearance date not appearing on the original letter.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Todd Courtney, Office of Ophthalmic, Anesthesia, Respiratory, ENT and Devices, 301-796-6371, todd.courtney@fda.hhs.gov.
Sincerely,
# Todd D. Courtney -S
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized symbol. The symbol consists of three abstract human profiles facing right, with flowing lines suggesting movement or connection. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Teleflex Medical, Inc. Lori Pfohl Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K150603
Trade/Device Name: Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal tube Regulatory Class: Class II Product Code: BTR Dated: November 5, 2015 Received: November 6, 2015
Dear Ms. Pfohl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/2/Picture/8 description: The image shows the FDA logo on the left side. On the right side, the text "Richard C. Chapman - S 2015.11.27 13:42:38 -05'00'" is displayed. The text appears to be a timestamp or signature related to the FDA logo.
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
for
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
Indications for Use (Describe)
Rusch tracheal tubes are indicated for airway management by oral or nasal intubation of the trachea.
Intended Population: Pediatric
Intended Environment of Use: Locations where ET intubation may be performed
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
### Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-491-8960 919-433-4996 Fax:
#### Contact Person
Lori Pfohl Senior Regulatory Affairs Specialist
### Date Prepared
3/9/2015
### Device Name
## Trade Name: Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
Common Name: Tracheal Tube
Classification Name: Tube, Tracheal (Class II per 21 CFR 868.5730, Product Code BTR)
## Predicate Devices
Teleflex Medical's Rusch Oral/Nasal (Safety Clear Plus) Tracheal Tube Cuffed, Magill/Murphy - K993786
Teleflex Medical's Rusch Oral/Nasal Tracheal Tube Cuffed. Magill/Murphy - K961837
## Device Description
The proposed Teleflex Medical Rusch Safety Silk Pediatric Tracheal Tube Series tracheal tubes are sterile, single use devices that are made from Polyvinyl chloride (PVC) resin that is formulated without DEHP ("Non-DEHP" = < 0.1% DEHP w/w). The tracheal tubes contain a compatible cuff, inflation line, pilot balloon and one-way valve. A radiopague line is incorporated into the full length of the tracheal tube. Each tracheal tube is supplied with an appropriately sized 15mm connector.
The oral version of the device (Super Safety Silk) will be sold in both Murphy eye and Magill styles.
The nasal version of the device (Nasal Safety Silk) will be offered in the Murphy eye style only.
# Teleflex Medical
{5}------------------------------------------------
# Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
# Indications for Use
- Rusch tracheal tubes are indicated for airway management by oral or nasal intubation of the . trachea.
Intended Population
- Pediatric patients ●
Intended Environment of Use
- Locations where ET intubation may be performed. .
Contraindications
None
# Substantial Equivalence
The proposed device is substantially equivalent to the predicate devices:
| Features | Teleflex Medical<br>Safety Silk Series<br>Oral / Nasal<br>Tracheal Tube<br>(Proposed) | Teleflex Medical<br>Oral / Nasal Tracheal<br>Tube<br>(Safety Clear Plus)<br>(K993786) | Teleflex<br>Medical<br>Oral / Nasal<br>Tracheal Tube<br>(K961837) | Teleflex Medical<br>Safety Silk Series<br>Oral / Nasal Tracheal<br>Tube<br>(132415) |
|------------------------|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Classification<br>Name | Tube, Tracheal<br>(w/wo connector) | Same | Same | Same |
| Product Code | 73 BTR | Same | Same | Same |
| Regulation<br>Number | 868.5730 | Same | Same | Same |
| Indications for<br>Use | Rusch tracheal tubes<br>are indicated<br>for airway<br>management by oral<br>or nasal intubation<br>of the trachea | Rusch Oral / Nasal<br>Trachea Tube,Cuffed,<br>Magill/Murphy is a<br>device inserted into a<br>patient's trachea via<br>the nose or mouth<br>and used to maintain<br>an open airway. | Oral or nasal<br>intubation and<br>airway<br>management | Same |
| Environment of<br>Use | Locations where ET<br>intubation may be<br>performed | Not stated | Not stated | Same |
| Patient<br>Population | Pediatric | Not stated | Not stated | Adult |
| Contraindications | None | Same | Same | Same |
| Single Use | Yes | Same | Same | Yes |
{6}------------------------------------------------
| Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube |
|-------------------------------------------------------------|
|-------------------------------------------------------------|
| Features | Teleflex Medical<br>Safety Silk Series<br>Oral / Nasal<br>Tracheal Tube<br>(proposed) | Teleflex Medical<br>Oral / Nasal Tracheal<br>Tube<br>(Safety Clear Plus)<br>(K993786) | Teleflex<br>Medical<br>Oral / Nasal<br>Tracheal Tube<br>(K961837) | Teleflex Medical<br>Safety Silk Series<br>Oral / Nasal Tracheal<br>Tube<br>(Reference Predicate) |
|-------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Size Range | 3.0 mm – 4.5 mm | 2.0 mm - 4.0 mm | 4.5 mm - 11.0<br>mm | 5.0 mm – 10.0 mm |
| Cuffed | Yes | Same | Same | Same |
| Radiopaque | Yes | Same | Same | Same |
| Connection to<br>ventilation source | 15 mm connector | Same | Same | Same |
| Method of<br>Sterilization | Ethylene Oxide 10-6<br>SAL | Same | Same | Same |
| Biocompatibility | Materials have been<br>tested per ISO<br>10993 | Same | Same | Same |
| Made without<br>DEHP | Yes | No | No | Yes |
| Packaging<br>Configuration | Ten (10) individual<br>units in a shelf box.<br>Ten shelf boxes in a<br>shipper | Not stated | Ten (10)<br>individual units<br>in a shelf box. | Same |
| Sterile | Yes | Same | Same | Same |
| Eye | Murphy and Magill | Same | Same | Same |
| Tip | Beveled | Same | Same | Same |
| Graduations | Multiple cm<br>markings | Same | Same | Same |
| Shaft | PVC | Same | Same | Same |
| Cuff | PVC | Same | Same | Same |
| Pilot Balloon | PVC | Same | Same | Same |
| Inflation Valve | PVC | Same | Same | Same |
| Inflation Tube | PVC | Same | Same | Same |
| X-Ray Marker,<br>Stripe | PVC with Barium<br>Sulfate | Same | Same | Same |
| Connector | Polyamide | Polypropylene | Polypropylene | Same |
# Comparison to Predicate Device:
The proposed Safety Silk series tracheal tubes are substantially equivalent to the predicate devices with respect to indications for use, technology and construction. The differences between the predicate and the proposed devices are minor and any risks have been mitigated through testing. The proposed device is designed with essentially the similar materials as the predicate, and identical materials to the reference predicate
{7}------------------------------------------------
# Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
#### Indications for Use -●
The indications for use are substantially equivalent for the proposed device when compared to the predicates. Each device is indicated for Oral or Nasal use in airway management.
#### ● Technology and construction -
The design, fabrication, shape, size, etc. are substantially equivalent to the predicates. This design includes the disposable tracheal tube, cuff side arm assembly and 15mm connector.
#### ● Environment of use -
While not specifically stated, the environments of use are equivalent to predicates
#### ● Patient Population -
While not specifically stated, the patient population is equivalent to the predicate
#### . Materials -
All patient contacting materials are in compliance with ISO 10993-1. Testing included Cytotoxicity, sensitization, intracutaneous activity, implantation and genotoxicity testing, and Extractables/Leachables Testing.
| Test | Reference to<br>Standard (if<br>applicable) | Principle of Test | Acceptance Criteria |
|---------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cuff Bonding Leak<br>Evaluation | N/A | To perform a visual check at the welding<br>point of the cuff | No leak around the welding area |
| Tube Curvature<br>Test | ISO 5361 | To measure the curvature of the tube | Must meet the product requirement<br>of 140mm ± 20mm |
| Tube Collapse | ISO 5361 | The patency of the ET tube airway lumen<br>is tested by passing a steel ball through<br>the tracheal tube lumen with the cuff<br>inflated within a transparent tube | The steel ball (OD = 75% of the<br>stated ID) must pass through the<br>lumen freely. |
| Cuff Resting<br>Diameter | ISO 5361 | The resting diameter of the cuff is<br>measured when the cuff is inflated to a<br>reference pressure which is intended to<br>remove creases but minimize stretching<br>of its walls | The cuff resting diameter shall be<br>within ± 15% of specification for<br>each individual size |
| Cuff Herniation | ISO 5361 | The tendency of the cuff to herniate<br>beyond the plane perpendicular to the<br>long axis of the tube at the nearest edge<br>of the bevel is tested by applying an axial<br>force with the cuff inflated within a<br>transparent tube. A cuff which protrudes<br>excessively at its patient end may<br>partially or completely occlude the orifice<br>at the patient end | No abnormality or defect on the cuff<br>(any part of the inflated cuff reaches<br>beyond the nearest edge of the<br>bevel will be considered as defect).<br>No abnormality on the configuration<br>of the cuff during deflating the cuff<br>over a period of not less than 10s<br>(any abnormality will be considered<br>as defect). |
# Performance Testing
{8}------------------------------------------------
| Test | Reference to<br>Standard (if<br>applicable) | Principle of Test | Acceptance Criteria |
|----------------------------------------|---------------------------------------------|--------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Tube<br>Compression<br>Evaluation | N/A | To measure the rigidity of the tube | Must be within the range of current<br>data |
| Cuff Unrestrained<br>Burst Evaluation | N/A | To determine minimum cuff burst<br>pressure under unrestrained conditions | Must be within the range of current<br>data |
| Cuff Restrained<br>Burst Evaluation | N/A | To determine cuff burst volume under<br>restrained conditions to simulate<br>conditions in the trachea | Must be within the range of current<br>data |
| Cuff Sealing<br>Pressure<br>Evaluation | N/A | To determine the minimum cuff pressure<br>required to seal the corrugated tube | Must be within the range of current<br>data |
# Rusch Safety Silk Pediatric Series Oral/Nasal Tracheal Tube
The Rusch Safety Silk Pediatric Tracheal Tube Series has the same indications for use,
technological characteristics and construction as its predicates. Performance test results demonstrate that the proposed device meets its intended use. It is for these reasons that the proposed device can be found substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.