CONTAFLEX 54 (hioxifilcon D) Spherical Soft Contact Lens for Daily Wear, CONTAFLEX 49 (hioxifilcon B) Spherical Soft Contact Lens for Daily Wear,
Applicant
Contamac, Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Jul 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The CONTAFLEX 54 (hioxifilcon D) Spherical Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system. The CONTAFLEX 49 (hioxifilcon B) Spherical Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system. Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Device Story
Lathe-cut, spherical soft (hydrophilic) contact lenses; fabricated from hioxifilcon D (CONTAFLEX 54) or hioxifilcon B (CONTAFLEX 49). Lenses are non-ionic co-polymers of 2-HEMA and glycerol methacrylate, cross-linked with EGDMA. In dry state, lenses are machined and polished; in hydrated state, they become soft, pliable, and conform to corneal curvature. Used by patients under prescription from eyecare practitioners for daily wear or frequent/planned replacement. Lenses require chemical disinfection. Function as refractive media to focus light on the retina. Benefit patients by correcting refractive errors (myopia/hyperopia) and improving visual acuity.
Clinical Evidence
Bench testing only. Includes in-vitro cytotoxicity (ISO 10993-5, Agarose Overlay Method), systemic toxicity (systemic injection test), and acute ocular irritation testing. Results demonstrate safety and effectiveness equivalent to predicate materials.
Technological Characteristics
Materials: hioxifilcon D (46% polymer, 54% water) and hioxifilcon B (51% polymer, 49% water). Non-ionic co-polymers of 2-HEMA and glycerol methacrylate, cross-linked with EGDMA. Lathe-cut manufacturing. Spherical back surface. Oxygen permeability: 20.96 x 10^-11 (CONTAFLEX 54) and 16.09 x 10^-11 (CONTAFLEX 49) (revised Fatt method). Chemical disinfection required.
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes; suitable for patients with refractive astigmatism ≤ 0.75 diopters not interfering with visual acuity.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, connected by a flowing line. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 10, 2015
Contamac Ltd. % Mr. Bret Andre Official Correspondent EyeReg Consulting, Inc. 474 NE 618t PL Hillsboro, Oregon 97124
Re: K150590
Trade/Device Name: Contaflex 54 (hioxifilcon D) Spherical Soft Contact Lens for Daily Wear Contaflex 49 (hioxifilcon B) Spherical Soft Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: June 4, 2015 Received: June 8, 2015
Dear Mr. Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150590
#### Device Name
CONTAFLEX 54 (hioxifilcon D) Spherical Soft (hydrophilic) Contact Lens for Daily Wear (clear and tinted, lathe-cut); CONTAFLEX 49 (hioxifilcon B) Spherical Soft (hydrophilic) Contact Lens for Daily Wear (clear and tinted, lathe-cut)
#### Indications for Use (Describe)
The CONTAFLEX 54 (hioxifilcon D) Spherical Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive ametropia (nyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.
The CONTAFLEX 49 (hioxifilcon B) Spherical Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K150590 | |
|-----------------------------------------|---------------------------------------------------------------------------------------|--|
| Applicant information: | | |
| Date Prepared: | March 3rd, 2015 | |
| Name:<br>Address | CONTAMAC Ltd.<br>Bearwalden Business Park<br>Saffron Walden<br>Essex England CB11 4JX | |
| Contact Person:<br>Phone number: | Robert McGregor<br>44-1799 542 000 | |
| US Agent:<br>Phone number<br>Fax number | EyeReg Consulting, Inc.<br>Bret Andre<br>(503) 372-5226<br>(503) 419-4475 | |
### Device Information:
| Device Classification: | Class II |
|------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Classification Number: | LPL |
| Classification Name: | Soft (hydrophilic) Contact Lens (21 CFR 886.5925) |
| Trade Name: | CONTAFLEX 54 (hioxifilcon D) Spherical Soft<br>(hydrophilic) Contact Lens for Daily Wear (clear<br>and tinted, lathe-cut) |
| | CONTAFLEX 49 (hioxifilcon B) Spherical Soft<br>(hydrophilic) Contact Lens for Daily Wear (clear<br>and tinted, lathe-cut) |
{4}------------------------------------------------
#### Equivalent Devices:
The CONTAFLEX 54 (hioxifilcon D) & CONTAFLEX 49 (hioxifilcon B) Spherical Soft (hydrophilic)Contact Lenses are substantially equivalent to the following predicate device(s):
| Predicate device: | "BENZ-G 4X hioxifilcon D"<br>Manufactured/distributed by Benz Research and Development.<br>510(k) number; K062854 |
|-------------------|-------------------------------------------------------------------------------------------------------------------|
| | "BENZ-G 3X hioxifilcon B"<br>Manufactured/distributed by Benz Research and Development.<br>510(k) number; K964528 |
#### CONTAFLEX 54 (hioxifilcon D) Device Description:
The CONTAFLEX 54 Spherical Soft (hydrophilic) Contact Lenses are fabricated from hioxifilcon D, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The non-ionic lens material, (hioxifilcon D) is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and 2.3- Dihydroxypropyl Methacrylate (Glycerol Methacrylate), cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 46% hioxifilcon D and 54% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility-handling tint, [phthalocyaninato (2-)] copper. The hioxifilcon D name has been adopted by the United States Adopted Names Council (USAN).
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (hioxifilcon D) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 54% water by weight. The physical properties of the lens are:
Refractive Index Light Transmission (clear) Light Transmission (tinted) Water Content Specific Gravity (wet) Oxygen Permeability
```
1.5193 (hydrated)
greater than 96%
greater than 96%
54 % ± 2%
1.120
20.96 X 10-11 (cm /sec) (ml O./ml x mm Hg @ 35℃),
(revised Fatt method).
```
{5}------------------------------------------------
#### CONTAFLEX 49 (hioxifilcon B) Device Description:
The CONTAFLEX 49 Spherical Soft (hydrophilic)Contact Lenses are fabricated from hioxifilcon B, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The non-ionic lens material, (hioxifilcon B) is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and 2,3- Dihydroxypropyl Methacrylate (Glycerol Methacrylate), cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 51% hioxifilcon B and 49% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility-handling tint. [phthalocvaninato (2-)] copper. The hioxifilcon B name has been adopted by the United States Adopted Names Council (USAN).
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (hioxifilcon B) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 49% water by weight. The physical properties of the lens are:
Refractive Index Light Transmission (clear) Light Transmission (tinted) Water Content Specific Gravity (wet) Oxygen Permeability
1.5164 (hydrated) greater than 96% greater than 96% 49% ± 2% 1.140 16.09 X 10-11 (cm /sec) (ml O./ml x mm Hg @ 35℃), (revised Fatt method).
{6}------------------------------------------------
### Intended Use:
The CONTAFLEX 54 (hioxifilcon D) Spherical Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.
The CONTAFLEX 49 (hioxifilcon B) Spherical Soft (hydrophilic) Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
#### Description of Safety:
A series of preclinical testing were performed to demonstrate the safety and effectiveness of the CONTAFLEX 54 & CONTAFLEX 49 Soft (hydrophilic) Contact Lens materials. The results of all testing demonstrated that the safety and effectiveness of the CONTAFLEX 54 & CONTAFLEX 49 Soft (hydrophilic) Contact Lenses are equivalent to the currently marketed Benz hioxifilcon D (K062854) & hioxifilcon B (K964528) contact lens materials. A summary of these results from the preclinical studies is presented below.
#### Toxicology:
In-Vitro Cytotoxicity: ISO 10993-5 was conducted in accordance with standards on test article. The test article meets the requirements of the Agarose Overlay Method.
Systemic Toxicity: The lens material meets the requirements of the systemic injection test and is considered non-toxic.
Acute Ocular Irritation: Acute ocular irritation test was performed and produced no ocular irritation.
#### Shelf Life
The data presented supports substantial equivalence of this CONTAFLEX 54 & CONTAFLEX 49 Soft (hydrophilic) Contact Lens material to the already marketed Benz hioxifilcon D (K062854) & hioxifilcon B (K964528).
{7}------------------------------------------------
#### Substantial Equivalence:
The CONTAFLEX 54 & CONTAFLEX 49 Soft (hydrophilic) Contact Lenses are substantially equivalent to the predicate device as depicted in the following table, and do not raise different questions of safety and effectiveness than the predicate device identified previously.
The following table depicts the pre-clinical characteristics of the CONTAFLEX 54 & CONTAFLEX 49 materials, as well as the predicate devices.
{8}------------------------------------------------
## Substantial Equivalence Matrix
| | Contamac ltd.<br>CONTAFLEX 54<br>(Subject Device) | Contamac ltd.<br>CONTAFLEX 49<br>(Subject Device) | Benz Research and<br>Development<br>BENZ-G 4X<br>(Predicate Device) | Benz Research and<br>Development<br>BENZ-G 3X<br>(Predicate Device) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Indicated for daily wear<br>for the correction of<br>visual acuity in aphakic<br>and not aphakic persons<br>with non-diseased eyes<br>with myopia or<br>hyperopia. | Indicated for daily<br>wear for the correction<br>of visual acuity in<br>aphakic and not<br>aphakic persons with<br>non-diseased eyes with<br>myopia or hyperopia. | Indicated for daily<br>wear for the<br>correction of visual<br>acuity in aphakic<br>and not aphakic<br>persons with non-<br>diseased eyes with<br>myopia or<br>hyperopia. | Indicated for daily<br>wear for the<br>correction of visual<br>acuity in aphakic<br>and not aphakic<br>persons with non-<br>diseased eyes with<br>myopia or<br>hyperopia. |
| Functionality | After machining from<br>the optical blank, the<br>contact lenses act as a<br>refractive medium that<br>focus light rays from<br>near and distant objects<br>on the retina. | After machining from<br>the optical blank, the<br>contact lenses act as a<br>refractive medium that<br>focus light rays from<br>near and distant<br>objects on the retina. | After machining<br>from the optical<br>blank, the contact<br>lenses act as a<br>refractive medium<br>that focus light rays<br>from near and<br>distant objects on<br>the retina. | After machining<br>from the optical<br>blank, the contact<br>lenses act as a<br>refractive medium<br>that focus light rays<br>from near and<br>distant objects on<br>the retina. |
| Indications | Soft (hydrophilic)<br>Contact Lens (21 CFR<br>886.5925) | Soft (hydrophilic)<br>Contact Lens (21 CFR<br>886.5925) | Soft (hydrophilic)<br>Contact Lens (21 CFR<br>886.5925) | Soft (hydrophilic)<br>Contact Lens (21 CFR<br>886.5925) |
| Production Method | Lathe cut | Lathe cut | Lathe cut | Lathe cut |
| USAN name | hioxifilcon D | hioxifilcon B | hioxifilcon D | hioxifilcon B |
| Water Content (%) | 54±2% | 49±2% | 54±2% | 49±2% |
| Oxygen<br>Permeability | $20.96 x 10^{-11} (cm^2/sec)(mlO_2)/(ml x mmHg @ 35°C)) (revised Fatt method)$ | $16.09 x 10^{-11} (cm^2/sec)(mlO_2)/(ml x mmHg @ 35°C)) (revised Fatt method)$ | $20.09 x 10^{-11} (cm^2/sec)(mlO_2)/(ml x mmHg @ 35°C)) (revised Fatt method)$ | $15.71 x 10^{-11} (cm^2/sec)(mlO_2)/(ml x mmHg @ 35°C)) (revised Fatt method)$ |
| FDA Group | FDA Group 2<br>(>50% H2O, non-ionic<br>polymer) | FDA Group 1<br>(<50% H2O, non-ionic<br>polymer) | FDA Group 2<br>(>50% H2O, non-<br>ionic polymer) | FDA Group 1<br>(<50% H2O, non-<br>ionic polymer) |
| Specific Gravity | 1.12 (hydrated) | 1.14 (hydrated) | 1.10 (hydrated) | 1.15 (hydrated) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.