K150554 · Wellquest International, Inc. · ILY · Oct 6, 2015 · Physical Medicine
Device Facts
Record ID
K150554
Device Name
NuBrilliance Light Pain Therapy
Applicant
Wellquest International, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Oct 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nu Brilliance Light Pain Therapy system is an Over-The-Counter handheld device intended for the relaxation of muscles and relief of muscle spasms; temporary relief of minor muscle and joint aches, pains, and stiffness; temporary relief of minor pain and stiffness associated with arthritis; and to temporarily increase local blood circulation.
Device Story
Handheld LED device; emits infrared/red light for therapeutic heating; 4 red diodes (625 nm) and 56 infrared diodes (940 nm). Operator-controlled; no user settings/adjustments; includes internal cooling fan and automatic shut-off safety feature. Used by patients in home setting for temporary pain relief and increased local blood circulation. Device elevates skin temperature via light energy absorption. Clinical decision-making not applicable; device provides symptomatic relief.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including biocompatibility and electrical safety, and historical market presence of predicate devices.
Technological Characteristics
Handheld LED infrared lamp; 60 total diodes (4 red @ 625nm, 56 infrared @ 940nm); internal cooling fan; automatic shut-off; electrical safety and biocompatibility standards met.
Indications for Use
Indicated for OTC use by adults for muscle relaxation, relief of muscle spasms, temporary relief of minor muscle/joint aches, pains, and stiffness, temporary relief of arthritis-associated pain/stiffness, and temporary increase of local blood circulation.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
Predicate Devices
LED Technologies, LLC DPL™ Therapy System (K081570)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a central emblem. The emblem consists of a stylized image of an eagle or bird with three human profiles nested within its wing or body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2015
Wellquest International Incorporated % Ms. Carrie Hetrick Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, Texas 78701
Re: K150554
Trade/Device Name: NuBrilliance Light Pain Therapy Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Dated: August 27, 2015 Received: August 28, 2015
Dear Ms. Hetrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS)
{1}------------------------------------------------
regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Joshua
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150554
Device Name NuBrilliance Light Pain Therapy
#### Indications for Use (Describe)
The Nu Brilliance Light Pain Therapy system is an Over-The-Counter handheld device intended for the relaxation of muscles and relief of muscle spasms; temporary relief of minor muscle and joint aches, pains, and stiffness; temporary relief of minor pain and stiffness associated with arthritis; and to temporarily increase local blood circulation.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for NuBrilliance Light Pain Therapy K150554
#### 1. Submission Sponsor
Wellquest International, Inc. 230 Fifth Ave. Suite 808 New York, NY 10001 USA Phone: (212) 689.9094 Fax: (212) 213.1599 Contact: Ms. Yveth Kan
#### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (720) 838.4113 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant, RA Email: project.management@emergogroup.com
#### 3. Date Prepared
October 5, 2015
#### 4. Device Identification
| Trade/Proprietary Name: | NuBrilliance Light Pain Therapy |
|----------------------------|---------------------------------|
| Common/Usual Name: | Infrared lamp |
| Classification Name: | Infrared lamp |
| Classification Regulation: | 21 CFR § 890.5500 |
| Product Code: | ILY |
| Device Class: | Class II |
| Classification Panel: | Physical Medicine |
#### 5. Legally Marketed Predicate Device(s)
LED Technologies, LLC DPL™ Therapy System (K081570)
#### 6. Device Description
The NuBrilliance Light Pain Therapy device is a lightweight, light emitting diode (LED) device that emits light energy. The NuBrilliance Light Pain Therapy device is an Over-The-Counter
{4}------------------------------------------------
(OTC), hand held, ergonomically designed pain relief device that uses the infrared lamp for therapeutic heating. The head consists of four (#4) visible diodes (red lights) in the 625 nm (±5%) spectrum, and (#56) fifty-six infrared diodes (purple lights) in the 940 nm (±5%) spectrum that are used in combination for therapeutic heating. It delivers infrared light to the skin resulting in a safe elevation of the skin temperature for therapeutic effects. The handle of the device contains the electronics of the device including a fan for cooling and an automatic shut-off safety feature. Treatment is controlled by the operator. There are no user settings or adjustments.
## 7. Indication for Use Statement
The NuBrilliance Light Pain Therapy system is an Over-The-Counter handheld device intended for the relaxation of muscles and relief of muscle spasms; temporary relief of minor muscle and joint aches, pains, and stiffness; temporary relief of minor pain and stiffness associated with arthritis; and to temporarily increase local blood circulation.
## 8. Substantial Equivalence Discussion
The Wellquest NuBrilliance Light Pain Therapy device utilizes the same technological characteristics and has the same intended use as the K081570 predicate device. The K081570 device emits energy from light-emitting diodes (LED)s with wavelengths of 660 nm and 880 nm. The NuBrilliance Light Pain Therapy device emit energy from light-emitting diodes (LED)s with wavelengths of 625 nm and 940 nm. Both devices provide similar irradiance output to the treatment area. The desired therapeutic effects from both devices are generated by their emissions creating a local increased temperature in the area being treated. Differences between the NuBrilliance Light Pain Therapy device and the predicate do not raise new types of questions regarding the safety and effectiveness when used for the above indications for use.
#### 9. Non-Clinical Performance Data
The device has been tested for applicable safety requirements. The NuBrilliance Light Pain Therapy complies with the applicable voluntary standards for biocompatibility. As part of demonstrating safety and effectiveness of NuBrilliance Light Pain Therapy and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Wellquest International, Inc. completed a number of tests. The NuBrilliance Light Pain Therapy meets all the requirements for the overall design, biocompatibility, and electrical safety confirm that the output meets the design inputs and specifications. The NuBrilliance Light Pain Therapy passed all testing stated above as shown by the acceptable results obtained.
{5}------------------------------------------------
## 10. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Statement of Substantial Equivalence
It has been shown in this 510(k) submission that the differences, between the NuBrilliance Light Pain Therapy device and the predicate device listed above, do not raise new types of questions regarding safety and effectiveness. Further, the NuBrilliance Light Pain Therapy device utilizes the same type of technology as the predicate device. The NuBrilliance Light Pain Therapy, as designed and manufactured, is determined to be substantially equivalent to the above-listed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.