← Product Code [JWH](/productcode/JWH) · K150544

# Columbus Total Knee System, Columbus Revision Knee System, EnduRo Knee System (K150544)

_Aesculap Implant Systems, LLC · JWH · May 29, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150544

## Device Facts

- **Applicant:** Aesculap Implant Systems, LLC
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** May 29, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Columbus Total Knee System, Columbus Revision Knee System are indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture. Posterior Stabilized (PS) components are also for absent or non-functioning posterior cruciate ligament and severe anteroposterior instability of the knee joint. Hinge Knee Systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws. The Columbus Total Knee System, Columbus Revision Knee System are designed for use with bone cement.

## Device Story

The Columbus Total Knee, Columbus Revision Knee, and EnduRo Knee systems are orthopedic implants for knee joint reconstruction. This submission introduces the VEGA patella—a previously cleared component (K101281)—as a compatible option for these systems. The VEGA patella is manufactured from UHMWPE with a wrought stainless steel radiographic marker. It is intended for single-use, sterile, and cemented application. Surgeons use these components in primary or revision knee arthroplasty to restore joint function, alleviate pain, and correct deformities. The device functions as a mechanical prosthesis; it does not involve software or algorithms. Clinical benefit is derived from the restoration of joint stability and mobility in patients with degenerative or post-traumatic knee conditions.

## Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence was supported by geometrical analysis and lateral stability evaluation comparing the VEGA patella with the subject knee systems.

## Technological Characteristics

Implant system consisting of metal/polymer components. VEGA patella material: Ultra High Molecular Weight Polyethylene (UHMWPE) with wrought stainless steel radiographic marker. Fixation: Bone cement. Design: Semi-constrained cemented prosthesis. Sterilization: Sterile, single-use.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- VEGA patella ([K101281](/device/K101281.md))
- Columbus Total Knee System ([K120955](/device/K120955.md))
- Columbus Revision Knee System ([K122985](/device/K122985.md))
- EnduRo Knee System ([K120955](/device/K120955.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

May 29, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Aesculap Implant Systems, LLC Ms. Lisa M. Boyle Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K150544

Trade/Device Name: Columbus Total Knee System, Columbus Revision Knee System, EnduRo Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: May 4, 2015 Received: May 5, 2015

Dear Ms. Boyle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K150544

Device Name

Columbus Total Knee System, Columbus Revision Knee System, EnduRo Knee System

#### Indications for Use (Describe)

Columbus Total Knee System, Columbus Revision Knee System are indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture.

Posterior Stabilized (PS) components are also for absent or non-functioning posterior cruciate ligament and severe anteroposterior instability of the knee joint.

Hinge Knee Systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.

The Columbus Total Knee System, Columbus Revision Knee System are designed for use with bone cement.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED,

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Page 1 of 2

#### B. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

Columbus Total Knee System Columbus Revision Knee System EnduRo Knee System

February 24, 2015

| COMPANY:      | Aesculap® Implant Systems, LLC<br>3773 Corporate Parkway<br>Center Valley, PA 18034 |
|---------------|-------------------------------------------------------------------------------------|
| ESTABLISHMENT |                                                                                     |

REGISTRATION NUMBER: 3005673311

CONTACT:

Lisa M. Boyle 610-984-9274 (phone) 610-791-6882 (fax) lisa.boyle@aesculap.com

### DEVICE

| TRADE NAMES:         | Columbus Total Knee System<br>Columbus Revision Knee System<br>EnduRo Knee System            |
|----------------------|----------------------------------------------------------------------------------------------|
| COMMON NAME:         | Total Knee System                                                                            |
| DEVICE CLASS:        | CLASS II                                                                                     |
| PRODUCT CODE:        | JWH                                                                                          |
| REGULATION NUMBER:   | 888.3560                                                                                     |
| CLASSIFICATION NAME: | Knee Joint Patellofemorotibial Polymer/Metal/Polymer<br>Semi constrained Cemented Prosthesis |

## SUBSTANTIAL EQUIVALENCE

Aesculap Implant Systems, LLC believes allowing the use of the previously cleared VEGA patella (K101281) with Aesculap's Columbus Total Knee System, Columbus Revision Knee System and EnduRo Knee System is a line extension that remains substantially equivalent to the currently marketed subject knee systems. There have been no changes to any of the subject components.

{4}------------------------------------------------

## DEVICE DESCRIPTION

The VEGA patella is manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE) that incorporates a wrought stainless steel radiographic marker. The VEGA patella is intended to articulate with the femoral and tibial components of the Columbus Total Knee System, Columbus Revision Knee Svstem or EnduRo Knee System.

The VEGA patella is sterile and intended for single use only.

## INDICATIONS FOR USE

Columbus Total Knee System, Columbus Revision Knee and EnduRo Knee System are indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture.

Posterior Stabilized (PS) components are also for absent or non-functioning posterior cruciate ligament and severe anteroposterior instability of the knee joint.

Hinge Knee Systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.

The Columbus Total Knee System, Columbus Revision Knee System and EnduRo Knee System are designed for use with bone cement.

## TECHNOLOGICAL CHARACTERISTICS (Compared to the Predicate)

The VEGA patella is a legally marketed component of Aesculap VEGA Knee System that was originally cleared via 510(k) K101281. The introduction of the VEGA patella to the subject knee systems does not impact the fundamental scientific technology of Columbus Total Knee System (K120955), Columbus Revision Knee System (K122985), and EnduRo Knee System (K120955).

## PERFORMANCE DATA

As a result of the risk analysis, a geometrical comparison and an evaluation of lateral stability were used to determine cross compatibility of Aesculap VEGA patella with Aesculap's Columbus Revision and EnduRo Knee Systems. Results of the geometrical analysis and lateral stability demonstrated substantial equivalence and showed that there are no new risks associated with the alternate use of the VEGA patella in the subject Columbus Total Knee System (K120955), Columbus Revision Knee System (K122985), and EnduRo Knee System (K120955).

---

**Source:** [https://fda-staging.innolitics.com/device/K150544](https://fda-staging.innolitics.com/device/K150544)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com