K150540 · Summit Medical, Inc. · KCT · Jun 22, 2015 · General Hospital
Device Facts
Record ID
K150540
Device Name
Instru-Safe Instrument Protection System
Applicant
Summit Medical, Inc.
Product Code
KCT · General Hospital
Decision Date
Jun 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.
Device Story
Instrument protection system consisting of perforated cassettes/trays with silicone inserts; used to organize and protect surgical instruments during sterilization, storage, and transport. Cassettes feature rectangular base with latchable cover; perforations allow sterilant penetration. Used in healthcare facilities by sterile processing staff. Device does not maintain sterility independently; must be used with legally marketed wrap or Aesculap rigid container. Validated for specific Amsco V-PRO Low Temperature Sterilization cycles (Standard, Lumen, Non-Lumen). Benefits include organized instrument handling and protection during sterilization cycles.
Clinical Evidence
Bench testing only. Sterilization validation performed to demonstrate compatibility with Amsco V-PRO Low Temperature Sterilization systems. Biocompatibility testing (MEM Elution Cytotoxicity per ISO 10993-5) confirmed samples are non-cytotoxic.
Technological Characteristics
Materials: Anodized aluminum, stainless steel, silicone. Form factor: Perforated rectangular base and cover with latching mechanism and internal silicone inserts (hold-it/hold-down). Sterilization: Compatible with Amsco V-PRO Low Temperature Sterilization cycles. No software or electronic components.
Indications for Use
Indicated for use by healthcare providers to organize and protect medical devices during sterilization in Amsco V-PRO Low Temperature Sterilization Cycles. Used in conjunction with legally marketed sterilization wraps or Aesculap rigid containers. Not intended to maintain sterility independently.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Instru-Safe Instrument Protection System (K133015)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2015
Summit Medical Inc. Ms. Nicole Dove Quality Assurance/Regulatory Affairs Manager 815 Northwest Pkwy, Suite 100 St. Paul, MN 55121
Re: K150540
Trade/Device Name: Instru-Safe® Instrument Protection System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: May 20, 2015 Received: May 22, 2015
Dear Ms. Dove:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Dove
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150540
#### Device Name
Instru-Safe® Instrument Protection System
#### Indications for Use (Describe)
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.
| | | | Amsco V-PRO Low Temperature Sterilization Cycles: | | |
|--|--|--|---------------------------------------------------|--|--|
|--|--|--|---------------------------------------------------|--|--|
| Sterilizer | Standard<br>Cycle | Lumen<br>Cycle | Non Lumen<br>Cycle |
|--------------|-------------------|----------------|--------------------|
| V-PRO 1 | X | N/A | N/A |
| V-PRO 1 Plus | N/A | X | X |
| V-PRO maX | N/A | X | X |
| Summit Cassette Model | Aesculap Container Model |
|-----------------------|--------------------------|
| IN-8823 | *JM444 |
| IN-6105 | *JM742 |
* Validated by Summit Medical for use in Amso V-PRO Low Temperature Sterilization systems ONLY. When using the Aesculap container as a sterile barrier, the load (Summit tray and contents), should not exceed the load claims for the container in weight or load type.
| Lumen size of instrumentation validated includes: | | | |
|---------------------------------------------------|-------------------------|----------------|------------------|
| Summit Cassette Model | Minimum Inside Diameter | Maximum Length | Number of Lumens |
| IN-8823 | 3 mm | 400 mm | 2 |
| IN-6105 | 3 mm | 200 mm | 1 |
| IN-2681 | 1 mm | 64 mm | 1 |
The worst case validated load by vent-to-volume calculation is the IN-2681 tray.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller, sans-serif font underneath. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo has a clean and modern design.
## Indications for Use Statement
| Table 1 – Device Models | | |
|-------------------------|--------------------------|----------------------------------------------|
| Part Number | Maximum # of Instruments | Estimated Weight - Tray w/ instruments (lbs) |
| IN-1315 | 30 | 3.5 |
| IN-2006 | 8 | 2 |
| IN-2681 | 13 | 0.64 |
| IN-2682 | 13 | 0.89 |
| IN-2683 | 13 | 1.39 |
| IN-2840 | 36 | 8.75 |
| IN-2842 | 24 | 6.5 |
| IN-2843 | 36 | 8.75 |
| IN-2880 | 56 | 12.1 |
| IN-2900 | 22 | 4.18 |
| IN-2950 | 12 | 4.1 |
| IN-3030 | 34 | 9.5 |
| IN-4000 | 20 | 3.2 |
| IN-4003 | 30 | 3.25 |
| IN-4010 | 10 | 1.98 |
| IN-5009 | 8 | 5 |
| IN-5401 | 6 | 2 |
| IN-5401-02 | 2 | 1 |
| IN-5401-03 | 2 | 1 |
| IN-5401-08 | 8 | 3.2 |
| IN-5401-12 | 12 | 3.25 |
| IN-6103 | 2 | 2.15 |
| IN-6105 | 2 | 2.15 |
| IN-6110 | 3 | 2.15 |
| IN-6203 | 2 | 2.75 |
| IN-6205 | 2 | 2.75 |
| IN-6210 | 2 | 2.75 |
| IN-6240 | 2 | 2.75 |
| IN-6303 | 2 | 3.28 |
| IN-6305 | 2 | 3.28 |
| IN-6310 | 2 | 3.28 |
| IN-6403 | 2 | 3.28 |
| IN-6405 | 2 | 3.28 |
| IN-6410 | 2 | 3.28 |
| IN-6500 | 30 | 12.5 |
| IN-7010 | 2 | 2 |
| IN-7012 | 1 | 1.07 |
| IN-7020 | 30 | 8.5 |
| IN-7030 | 30 | 10 |
| IN-7032 | 2 | 1.1 |
| IN-7073 | 10 | 5 |
| IN-7120 | 45 | 11.25 |
| IN-7123 | 45 | 12 |
| IN-7130 | 45 | 13.5 |
| IN-7140 | 45 | 14.5 |
| IN-7150 | 8 | 1.9 |
| IN-7153 | 6 | 1.7 |
| IN-7220 | 45 | 14.5 |
| IN-7223 | 10 | 9.2 |
| IN-7230 | 45 | 14.5 |
| IN-7234 | 45 | 14.5 |
| IN-7240 | 45 | 14.5 |
| IN-7250 | 45 | 14.5 |
| IN-7252 | 25 | 8 |
| IN-7260 | 45 | 14.5 |
| IN-7273 | 10 | 6 |
| IN-7274 | 30 | 8 |
| IN-7322 | 45 | 14.5 |
| IN-7323 | 45 | 14.5 |
| IN-7343 | 45 | 14.5 |
| IN-7344 | 1 | 4 |
| IN-7360 | 45 | 14.5 |
| IN-7423 | 45 | 14.5 |
| IN-7452 | 10 | 8 |
| IN-7453 | 10 | 8 |
| IN-7540 | 45 | 14.5 |
| IN-7560 | 45 | 14.5 |
| IN-7644 | 45 | 14.5 |
| IN-7723 | 15 | 7.18 |
| IN-7724 | 15 | 7.2 |
| IN-7725 | 10 | 9.5 |
| IN-7781 | 45 | 14.5 |
| IN-7823 | 45 | 14.5 |
| IN-7830 | 45 | 14.5 |
| IN-7840 | 45 | 13.5 |
| IN-7940 | 20 | 13.25 |
| IN-8240 | 20 | 13.5 |
| IN-8610 | 2 | 6.65 |
| IN-8612 | 2 | 6.8 |
| IN-8613 | 2 | 6.1 |
| IN-8615 | 2 | 5.8 |
| IN-8616 | 2 | 5.8 |
| IN-8620 | 3 | 7.2 |
| IN-8621 | 4 | 7.18 |
| IN-8622 | 4 | 7.18 |
| IN-8630 | 3 | 6.5 |
| IN-8632 | 3 | 6.45 |
| IN-8633 | 3 | 6.8 |
| IN-8640 | 4 | 5.35 |
| IN-8642 | 4 | 5.35 |
| IN-8643 | 5 | 5.35 |
| IN-8645 | 4 | 5.35 |
| IN-8650 | 4 | 5.85 |
| IN-8660 | 4 | 5.35 |
| IN-8662 | 4 | 5.35 |
| IN-8663 | 3 | 5.5 |
| IN-8700 | 40 | 14 |
| IN-8810 | 20 | 13.5 |
| IN-8820 | 15 | 8.75 |
| IN-8823 | 45 | 14 |
| IN-8830 | 15 | 8.75 |
| IN-8833 | 45 | 14 |
| IN-8840 | 20 | 13.75 |
| IN-8850 | 15 | 8.75 |
| IN-8853 | 45 | 14 |
| IN-8860 | 15 | 8.75 |
| IN-8862 | 30 | 10.5 |
| IN-8863 | 45 | 14 |
| IN-8880 | 2 | 3.28 |
| IN_8882 | 16 | 12.1 |
| IN-8883 | 2 | 3.28 |
| IN-8884 | 4 | 5.35 |
| IN-8885 | 1 | 2.25 |
| IN-8886 | 6 | 12.1 |
| IN-8889 | 6 | 12.1 |
| IN-8891-S | 1 | 2 |
| IN-8891-SI-12-S | 1 | 2 |
| IN-8891-SI-85-S | 1 | 2 |
| IN-8892 | 12 | 12.1 |
| IN-8893 | 9 | 7.5 |
| IN-8894 | 5 | 16.1 |
| IN-8897 | 8 | 6 |
| IN-8898 | 10 | 10.25 |
| IN-8899 | 7 | 6.5 |
| IN-8901 | 1 | 2.25 |
| IN-8902-G2 | 22 | 17 |
| IN-8903 | 15 | 13.25 |
| IN-8904 | 22 | 17 |
| IN-8907 | 7 | 12.5 |
| IN-8931 | 1 | 2.4 |
| IN-8932 | 9 | 9.5 |
| IN-8933 | 3 | 3.75 |
| IN-8936 | 6 | 11.5 |
| IN-8937 | 16 | 14.5 |
| IN-8938 | 8 | 12.5 |
| IN-8939 | 10 | 11.6 |
| IN-8940 | 5 | 5.18 |
| IN-8942 | 11 | 10 |
| IN-8943 | 1 | 2.7 |
| IN-8944 | 6 | 4.7 |
| IN-8945 | 2 | 5.18 |
| IN-8946 | 9 | 6.1 |
| IN-8980 | 20 | 9.5 |
| IN-8982-01 | 17 | 9.5 |
| IN-8983-01 | 16 | 9.5 |
| IN-8984-01 | 15 | 9.5 |
| IN-8986-S | 2 | 6.5 |
| IN-8987-S | 2 | 6.5 |
| IN-8988-S | 2 | 6 |
| IN-8989-S | 2 | 6 |
| IN-9999-160 | 6 | 12.1 |
| IN-9999-162-S | 2 | 5.8 |
| IN-9999-168-S | 2 | 5.8 |
| IN-9999-172-S | 2 | 5.8 |
| IN-9999-178-S | 2 | 5.8 |
Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font underneath. Above the word "SUMMIT" are three blue diamonds arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the blue color gives it a professional look.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds, arranged horizontally. The logo is simple and modern, and the use of blue suggests trust and reliability.
# 510(k) Summary
Following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92
| Submitter: | Summit Medical Inc.<br>815 Northwest Parkway, Suite 100<br>St. Paul, MN 55121<br>Tel: (651) 789-3939 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ER Number: | 3008719017 |
| Contact<br>Person: | Nicole Dove<br>QA/RA Manager Tel: (651)<br>789-3921<br>ndove@summitmedicalusa.com |
| Date Prepared: | June 22, 2015 |
| Subject<br>Device: | Trade Name(s):<br>Instru-Safe® Instrument Protection System<br>Classification Name:<br>Sterilization wrap containers, trays, cassettes & other accessory (21 CFR 880.6850<br>Common Name:<br>Instrument Tray, Sterilization Tray, Sterilization Cassettes, Instrument Delivery<br>System<br>Device Class:<br>Class II<br>Device Code:<br>KCT<br>Panel:<br>General Hospital |
| Predicate<br>Device: | Tradename: Instru-Safe Instrument Protection System<br>510(k) Holder: Summit Medical Inc.<br>510(k) #: K133015 |
| Device<br>Description: | Summit Medical Inc. Instru-Safe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays by themselves do not maintain sterility. |
Page 1 of 11
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of three blue diamonds arranged horizontally above the word "SUMMIT" in a stylized font. Below "SUMMIT" is the word "medical" in a smaller, sans-serif font. The overall design is clean and professional, suggesting a healthcare or medical-related organization.
| The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray. | | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|------------------|---------------------|--|
| Intended Use: | Instru-Safe ® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1. | | | | |
| Amsco V-PRO Low Temperature Sterilization Systems | | | | | |
| Sterilizer | Standard Cycle | Lumen Cycle | Non Lumen Cycle | | |
| V-PRO 1 | X | N/A | N/A | | |
| V-PRO 1 Plus | N/A | X | X | | |
| V-PRO maX | N/A | X | X | | |
| Summit Cassette Model | Aesculap Container Model | | | | |
| IN-8823 | *JM444 | | | | |
| IN-6105 | *JM742 | | | | |
| *Validated by Summit Medical for use in Amsco V-PRO Low Temperature Sterilization Systems ONLY. When using the Aesculap container as a sterile barrier, the load (Summit tray and contents), should not exceed the load claims for the container in weight or load type. | | | | | |
| Lumen size of instrumentation validated includes: | | | | | |
| Summit Cassette Model | Minimum Inside Diameter | Maximum Length | Number of Lumens | | |
| IN-8823 | 3mm | 400mm | 2 | | |
| IN-6105 | 3mm | 200mm | 1 | | |
| IN-2681 | 1mm | 64mm | 1 | | |
| The worst case validated load by vent-to-volume calculation is the IN-2681 tray. The intended use of the subject device includes the Amsco V-PRO Low Temperature Sterilization Systems. Performance testing has been completed for the Amsco V-PRO Low Temperature Sterilization Systems. These new sterilization cycles do not affect safety and effectiveness of the Instru-Safe Instrument Protection System. | | | | | |
| Comparison of Characteristics to Predicate Device: | Based on a comparison of the design, technology, materials, manufacturing, performance, specifications and methods of use, the Instru-Safe Instrument Protection System is equivalent to the identified 510(k) cleared predicate device. | | | | |
| Element | New Device | | | Predicate (K133015) | |
Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com
Page 2 of 11
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in a stylized font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three blue diamonds. The logo is simple and modern, and the use of blue gives it a professional look.
| Intended Use | Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1. | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| | Instru-Safe Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycle. The Instru-Safe System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instru-Safe System cassettes are not intended on their own to maintain sterility. | | |
| | Sterilization methods and configurations | Sterilization methods and configurations | |
| | Amsco V-PRO Low Temperature Sterilization Systems | • Autoclave Sterilization Parameter:<br>Cycle: Pre-vacuum<br>Temperature: 270°F (132°C)<br>Exposure Time: 4 minutes<br>Minimum Dry Time: 30 minutes | |
| Sterilizer | Standard Cycle | Lumen Cycle | Non Lumen Cycle |
| V-PRO 1 | X | N/A | N/A |
| V-PRO 1 Plus | N/A | X | X |
| V-PRO MAX | N/A | X | X |
| Summit Cassette Model | Aesculap Container Model | | |
| IN-8823-AE | *JN444 | | |
| IN-2880 | *JK444 | | |
| IN-6105 | *JN742 | | |
| IN-8823 | *JM444 | | |
| IN-6105 | *JM742 | | |
| *Validated by Summit Medical for use in Amsco V-PRO Low Temperature Sterilization Systems ONLY. When using the Aesculap container as a sterile barrier, the load (Summit tray and contents), should not exceed the load claims for the container in weight or load type. | | *Validated by Summit Medical for use in steam prevacuum sterilizers ONLY operating at 270°F (132°C) for 4 minutes exposure time. Consult container instructions to ensure that contents do not exceed the sterilization container intended load claims. | |
Page 3 of 11
{12}------------------------------------------------
| Lumen size of instrumentation validated includes: | | | |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summit<br>Cassette<br>Model | Minimum<br>Inside<br>Diameter | Maximum<br>Length | Number of<br>Lumens |
| IN-8823 | 3mm | 400mm | 2 |
| IN-6105 | 3mm | 200mm | 1 |
| IN-2681 | 1mm | 64mm | 1 |
| The worst case validated load by vent-to- volume calculation<br>is the IN-2681 tray. | | | |
| Material<br>Composition | No changes from predicate device | | The cassette contains<br>components made<br>of anodized aluminum,<br>stainless steel, blue |
| Physical<br>Properties | Instru-Safe Instrument Protection<br>System cassettes include<br>- perforated base<br>- perforated cover<br>- silicone inserts (hold-it / hold down)<br>- Handles<br>- Latches<br>- Feet<br>- Posts (optional)<br>- Divider (optional)<br>- Shelf (optional) | | Instru-Safe Instrument<br>Protection<br>System cassettes include<br>- perforated base<br>- perforated cover<br>- silicone<br>inserts (hold-<br>it / hold<br>down)<br>- Handles<br>- Latches<br>- Feet |
| Chemical<br>Properties | Not Applicable | | Not Applicable |
| Configurations<br>/ Dimensions | Various configurations / dimensions | | See table located in<br>predicate device |
| Air permeance | Not Applicable | | Not Applicable |
Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com
Page 4 of 11
{13}------------------------------------------------
| Percent of<br>surface<br>performations | Not Applicable | Not Applicable |
|----------------------------------------|----------------------------------------|---------------------------|
| Performance | New Device | Predicate |
| Sterilant | Amsco V-PRO Low Temperature | Pre-Vacuum Steam |
| Penetration | Sterilization Systems. | Cycle: Pre- |
| | | vacuum |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.