K150530 · Medtronic, Inc. · DTW · May 8, 2015 · Cardiovascular
Device Facts
Record ID
K150530
Device Name
Level Sensor, Level Sensor Tape
Applicant
Medtronic, Inc.
Product Code
DTW · Cardiovascular
Decision Date
May 8, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4340
Device Class
Class 2
Indications for Use
The Medtronic Bio-Console 560 System with Level Sensing System (Level Sensor [LS100] and Level Sensing Tape [LST]) is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours). The Medtronic Performer CPB System with Level Sensing System (Level Sensor [LS100] and Level Sensing Tape [LST]) is indicated for use up to 6 hours in the extracorporeal circulation of fluids for cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Device Story
Level Sensing System (LS100 sensor and LST tape) functions as an optional safety accessory for Medtronic Bio-Console 560 and Performer CPB systems. Designed for use with Medtronic hardshell reservoirs; monitors fluid levels during cardiopulmonary bypass. System provides real-time level sensing to assist perfusionists in maintaining circuit safety. Changes from predicate include modified sensor footprint, cable configuration, and strain relief to improve usability. Operated by qualified medical professionals in clinical settings; output informs perfusionist of reservoir fluid levels to ensure continuous, safe extracorporeal support.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including visual/dimensional inspection, threshold measurements, free-fall, tensile strength, connector compatibility, splash, mounting, chemical resistance, bovine blood circuit functionality, electrical safety, and EMC emissions/immunity testing.
Technological Characteristics
System includes LS100 sensor and LST adhesive tape for hardshell reservoirs. Features modified sensor footprint, cable configuration, and strain relief. Operates as an integrated safety monitoring component for CPB pump systems. Electrical safety and EMC compliant.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass procedures requiring extracorporeal circulation of blood/fluids for up to 6 hours. Intended for use by qualified medical professionals experienced in CPB equipment operation.
Regulatory Classification
Identification
A cardiopulmonary bypass level sensing monitor and/or control is a device used to monitor and/or control the level of blood in the blood reservoir and to sound an alarm when the level falls below a predetermined value.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 8, 2015
Medtronic, Inc. Rahul Shah Regulatory Operations Specialist 7611 Northland Drive Minneapolis, MN 55428
Re: K150530
Trade/Device Name: Medtronic Bio Console 560 System with Level Sensing System (Level Sensor [LS100] and Level Sensing Tape [LST]) Medtronic Performer CPB System with Level Sensing System (Level Sensor [LS100] and Level Sensing Tape [LST]) Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary Bypass Pump Speed Control Regulatory Class: Class II Product Code: DWA, DTQ, DTW Dated: April 9, 2015 Received: April 13, 2015
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150530
#### Device Name
Medtronic Bio Console 560 System with Level Sensor [LS100] and Level Sensor [LS100] and Level Sensing Tape [LST])
#### Indications for Use (Describe)
"The Medtronic Bio-Console 560 System with Level Sensing System (LS100) and Level Sensing Tape [LST]) is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours)".
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150530
#### Device Name
Medtronic Performer CPB System with Level Sensor [LS100] and Level Sensor [LS100] and Level Sensing Tape [LST])
#### Indications for Use (Describe)
"The Medtronic Performer CPB System with Level Sensing System (LS100) and Level Sensing Tape [LST]) is indicated for use up to 6 hours in the extracorporeal circulation of fluids for cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment."
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 6. Summary of Safety and Effectiveness
| Date Prepared: | May 8, 2015 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and Address: | Medtronic Perfusion Systems<br>Medtronic, Inc.<br>7611 Northland Drive<br>Minneapolis, MN 55428 |
| Contact Person: | Rahul Shah<br>Regulatory Operations Specialist<br>Medtronic, Inc.<br>Coronary and Structural Heart Disease<br>Management<br>8200 Coral Sea Street NE, MVS 83<br>Mounds View, MN 55112<br>Phone: (763) 514-9846<br>Fax: (763) 367-8147<br>Email: rahul.m.shah@medtronic.com |
| Proprietary Name of Device: | Medtronic Bio Console 560 System with Level<br>Sensing System (level Sensor [LS100] and Level<br>Sensing Tape [LST]) |
| | Medtronic Performer CPB System with Level<br>Sensing System (Level Sensor [LS100] and Level<br>Sensing Tape [LST]) |
| Common/Usual Name: | Cardiopulmonary bypass speed control |
| Classification Name: | Control, Pump Speed, Cardiopulmonary Bypass |
| Classification: | Class II |
| Product Code: | DWA, DTQ, DTW |
| Regulation Number: | 21 CFR 870.4380 |
| Predicate Device: | Medtronic Bio Console 560 System<br>Medtronic Performer CPB System |
| Predicate 510(k): | K070286, K070213 |
#### Device Description
Level Sensing System includes the Level Sensor (LS100) and the Level Sensor Tape (LST). The Level Sensor System is one of the optional safety systems that can be used with the Medtronic Bio Console 560 Extracorporeal Blood Pumping System and the Medtronic Performer Cardiopulmonary Bypass System. It is designed to be used only with a Medtronic hardshell reservoir.
#### Indications for Use
"The Medtronic Bio-Console 560 System with Level Sensing System (Level Sensor [LS100] and Level Sensing Tape [LST]) is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass procedures (up to 6 hours)."
{5}------------------------------------------------
"The Medtronic Performer CPB System with Level Sensing System (Level Sensor [LS100] and Level Sensing Tape [LST]) is indicated for use up to 6 hours in the extracorporeal circulation of fluids for cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment."
#### Comparison to Predicate Devices:
The Level Sensing System is substantially equivalent to the previous version (predicate) of this device. The footprint of the level sensor, cable configuration and strain relief components have changed to improve the usability of the device.
The modified sensor has the following similarities to the predicate device which received 510(k) clearance:
- . Same intended use/indications
- Same operating principle ●
- Same fundamental technological characteristics ●
- Same performance requirements
- Same packaging materials and design .
- Same sterilization requirements ●
#### Summary of Performance Data:
Bench testing was used to verify the performance characteristics of these devices. The following bench studies were performed:
- Visual and Dimension Inspection ●
- Threshold Measurements
- Free-Fall Test .
- Basic Functionality with Bio-Console Test ●
- o Tensile Strength Test
- o Connector Compatibility/Wiring, Output Stage Type and Status Indicator Tests
- o Splash Test
- Mounting Test (using LST adhesive tape) .
- Chemical Resistance ●
- Basic Functionality in a circuit with bovine blood ●
- Electrical Safety Test ●
- Emissions/Immunity Test ●
Clinical testing was not required to establish substantial equivalence.
Verification and validation testing has demonstrated that the Level Sensing System is substantially equivalent to the predicate device.
#### Conclusion:
The described labeling and engineering changes do not change the indications for use, technology and performance specifications of this device. Therefore the Level Sensing System is substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.