← Product Code [GEX](/productcode/GEX) · K150516

# DEKA SYNCHRO REPLA:Y FAMILY OF LASER SYSTEMS (K150516)

_El.En Electronic Engineering Spa · GEX · Apr 13, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150516

## Device Facts

- **Applicant:** El.En Electronic Engineering Spa
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Apr 13, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

755mn laser: Temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick 1 - VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. Photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). 1064nm laser: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques. The laser is indicated for pigmented lesion size. for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles. Pulsed light FT handpiece: Permanent hair reduction, photocoagulation of vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions. Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types, as indicated in the following table: [Table omitted]

## Device Story

Platform system utilizing 755nm Alexandrite laser, 1064nm Nd:YAG laser, or pulsed light (FT) handpiece; intended for dermatology/plastic surgery. Laser heads deliver energy via optical fiber to interchangeable handpieces; FT handpiece uses sapphire lightguide with integrated cooling. Operated by clinician via touch panel, footswitch, or fingerswitch. Modifications include larger spot size handpieces (22mm, 24mm) and adjusted pulse repetition rates for Nd:YAG. Output enables photothermolysis/coagulation for hair removal, vascular/pigmented lesion treatment, and soft tissue management. Clinical benefit derived from selective photothermolysis of target chromophores (melanin, hemoglobin).

## Clinical Evidence

No clinical data provided. Substantial equivalence based on technological characteristics and performance specifications compared to predicate.

## Technological Characteristics

Platform system with 755nm Alexandrite and 1064nm Nd:YAG laser sources and pulsed light (FT) handpiece. Features interchangeable quick-release handpieces with electronic spot recognition. FT handpiece includes sapphire lightguide and integrated skin cooling. Connectivity via touch panel interface. Software manages handpiece recognition and pulse repetition parameters.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- DEKA Synchro Repla:Y family of laser systems ([K131095](/device/K131095.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 13, 2015

El. En. Electronic Engineering S.p.A. Mr. Paolo Peruzzi Regulatory Affairs Manager 17 Via Baldanzese Calenzano 50041 Italy

Re: K150516 Trade/Device Name: DEKA SYNCHRO REPLA:Y family of Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 13, 2015 Received: March 18, 2015

Dear Mr. Paolo Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K150516

### Device Name

Deka Synchro Repla: Y family of laser systems

Indications for Use (Describe)

#### 755mn laser:

Temporary hair reduction.

Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

On all skin types (Fitzpatrick 1 - VI) including tanned skin.

Treatment of benign pigmented lesions.

Treatment of wrinkles.

Photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

### 1064nm laser:

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin.

Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins.

Coagulation and hemostasis of soft tissue.

Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques.

The laser is indicated for pigmented lesion size. for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

Treatment of wrinkles.

{3}------------------------------------------------

### Pulsed light FT handpiece:

Permanent hair reduction, photocoagulation of vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.

Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types, as indicated in the following table:

| Pulsed light     | Hair reduction   | Vascular lesions | Blood vessels    | Pigmented lesions |
|------------------|------------------|------------------|------------------|-------------------|
| Wavelength range |                  |                  |                  |                   |
| 500-1200nm       | -                | Skin Types I, II | Skin Types I, II | -                 |
| 520-1200nm       | -                | Skin Type III    | Skin Type III    | Skin Types I, II  |
| 550-1200nm       | Skin Types I, II | -                | -                | Skin Type III     |
| 600-1200nm       | Skin Type III    | -                | -                | -                 |
| 650-1200nm       | Skin Type IV     | -                | -                | Skin Type IV      |

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{4}------------------------------------------------

# 510(K) Summary

# K150516 Deka Synchro Repla: Y family of Laser Systems - Special 510(k)

## Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

## Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

## Date Summary Prepared:

April 9, 2015

Device Trade Name: Deka Synchro Repla: Y family of Laser Systems

Common Name: Medical Laser and pulsed Light system

## Classification Name:

Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology (GEX)

## Classification Number:

21 CFR 878.4810

## Equivalent Devices:

DEKA Synchro Repla: Y family of laser systems (K131095).

## Device Description:

The Deka Synchro Repla: Y family of lasers is a platform which can be equipped with two separate solid state laser heads (755nm and 1064nm) and a pulsed light handpiece (FT handpiece).

The two laser heads can only be activated separately and they deliver the laser output through the same single laser delivery system, consisting of a lens coupled user replaceable optical fiber with a wide range of interchangeable, quick release laser handpieces with electronic spot recognition.

The FT handpiece delivers the pulsed light to the treatment area by means of a sapphire lightguide; it is provided with has an integrated skin cooling system and is equipped with five different emission spectra interchangeable filters.

The system allows to switch from one of the three available sources to the other by pressing a key on touch panel.

{5}------------------------------------------------

Handpiece activation is either by footswitch or fingerswitch.

The modifications to the device are two laser handpieces, allowing for a larger treatment area, 22mm and 24mm, and a change of maximum pulse repetition rate for some intermediate settings for the Nd: YAG laser source.

The device hardware did not require any modification to support the new features. Minor changes to the software have been made in order to manage the new handpieces and the modified intermediate settings for the Nd:YAG laser source.

The intended use of the modified device, as described in the labeling, has not changed as a result of the modifications.

## Intended Use:

## 755mn laser:

Temporary hair reduction.

Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6,9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I - VI) including tanned skin.

Treatment of benign pigmented lesions.

Treatment of wrinkles.

Photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

### 1064nm laser:

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin.

Photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins.

Coagulation and hemostasis of soft tissue.

Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

Treatment of wrinkles.

Pulsed light FT handpiece:

Permanent hair reduction, photocoagulation of vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.

Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types, as indicated in the following table:

| Pulsed light<br>Wavelength range | Hair reduction   | Vascular lesions | Blood vessels    | Pigmented lesions |
|----------------------------------|------------------|------------------|------------------|-------------------|
| 500-1200nm                       | -                | Skin Types I, II | Skin Types I, II | -                 |
| 520-1200nm                       | -                | Skin Type III    | Skin Type III    | Skin Types I, II  |
| 550-1200nm                       | Skin Types I, II | -                | -                | Skin Type III     |
| 600-1200nm                       | Skin Type III    | -                | -                | -                 |
| 650-1200nm                       | Skin Type IV     | -                | -                | Skin Type IV      |

{6}------------------------------------------------

## Substantial equivalence discussion:

The DEKA Synchro Repla: Y family of laser systems is substantially equivalent to the DEKA Synchro Repla: Y family of laser systems (K131095).

| Device Trade<br>Name   | DEKA Synchro<br>Repla:Y family of<br>laser systems | Predicate Device<br>K131095<br>DEKA Synchro<br>Repla:Y family of<br>laser systems |
|------------------------|----------------------------------------------------|-----------------------------------------------------------------------------------|
| Laser Type             | Alexandrite, Nd:YAG                                | Alexandrite, Nd:YAG                                                               |
| Wavelength             | 755 nm, 1064 nm                                    | 755 nm, 1064 nm                                                                   |
| Max Fluence            | 600 J/cm²                                          | 600 J/cm²                                                                         |
| Spot Sizes             | 2.5 to 24mm                                        | 2.5 to 20mm                                                                       |
| Pulse Duration         | 0.25 to 300 ms                                     | 0.25 to 300 ms                                                                    |
| Pulse Rep Rate         | up to 10 Hz                                        | up to 10 Hz                                                                       |
| Skin Cooling<br>System | Yes, optional                                      | Yes, optional                                                                     |

| Device Trade<br>Name      | DEKA Synchro<br>Repla:Y family of<br>laser systems                             | Predicate Device<br>K131095<br>DEKA Synchro<br>Repla:Y family of<br>laser systems |
|---------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Device Type               | Pulsed Flashlamp                                                               | Pulsed Flashlamp                                                                  |
| Wavelengths<br>(nm)       | 500 -  1200nm<br>520 - 1200nm<br>550 - 1200nm<br>600 -  1200nm<br>650 - 1200nm | 500 - 1200nm<br>520 - 1200nm<br>550 -  1200nm<br>600 -  1200nm<br>650 -  1200nm   |
| Pulse Width               | 3 - 124 ms                                                                     | 3 - 124 ms                                                                        |
| Fluence                   | 3 - 25 J/cm²                                                                   | 3 - 25 J/cm²                                                                      |
| Repetition Rate           | 0.5 Hz max.                                                                    | 0.5 Hz max.                                                                       |
| Spot Sizes                | 48x13 mm                                                                       | 48x13 mm                                                                          |
| Method of Skin<br>Cooling | Integrated - provided<br>via the handpiece<br>lightguide                       | Integrated - provided<br>via the handpiece<br>lightguide                          |
| Method of<br>Output       | Direct Delivery<br>Handpiece                                                   | Direct Delivery<br>Handpiece                                                      |

{7}------------------------------------------------

The DEKA Synchro Repla: Y family of laser systems has the same indications for use as the abovementioned predicate device, with same principle of operation and essentially the same performances.

## Clinical Performance Data:

None

## Non-Clinical Performance Data:

None

## Conclusion:

The Deka Synchro Repla: Y family of laser systems is a family of safe and effective devices for the applications mentioned above.

## Additional Information:

None

---

**Source:** [https://fda-staging.innolitics.com/device/K150516](https://fda-staging.innolitics.com/device/K150516)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).