← Product Code [LPH](/productcode/LPH) · K150503

# Echo Bi-Metric Microplasty Line Extension (K150503)

_Biomet, Inc. · LPH · Mar 25, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150503

## Device Facts

- **Applicant:** Biomet, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Mar 25, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2) Rheumatoid arthritis. 3) Correction of functional deformity. 4) Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques. 5) Revision of previously failed total hip arthroplasty. Porous coated components are intended for uncemented, biological fixation.

## Device Story

Echo Bi-Metric Microplasty Femoral Stem is a monolithic, collarless, straight hip prosthesis; designed to reduce hip pain and restore joint biomechanics/stability. Implanted during primary or revision total hip arthroplasty; used by orthopedic surgeons in clinical/OR settings. Device provides structural replacement for femoral head/neck; porous plasma spray coating facilitates uncemented biological fixation to patient bone. Surgeon selects appropriate stem size to fit patient anatomy; device functions as mechanical support for hip joint. Benefits include restoration of joint function and pain relief.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance validated via mechanical testing: Proximal Fatigue Testing (ISO 7206-6), Distal Fatigue Testing (ISO 7206-4), and Range of Motion Analysis (ISO 21535).

## Technological Characteristics

Monolithic, collarless, straight femoral stem. Material: Ti-6Al-4V (ASTM F136/ASTM F620). Surface: Proximal circumferential porous plasma spray (PPS) coating (ASTM F1580). Instruments: Stainless steel with AlTiN or TiN PVD coating. Sterilization: Sterile, single-use.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Echo Bi-Metric Microplasty Hip System ([K143009](/device/K143009.md))

## Reference Devices

- Balance Microplasty Hip System ([K050251](/device/K050251.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

Biomet, Incorporated Ms. Amy Walriven Manager, Regulatory Affairs 56 East Bell Drive Warsaw, Indiana 46581

Re: K150503 Trade/Device Name: Echo Bi-Metric Microplasty Line Extension Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, KWL, LZO, LWJ, KWZ, KWY, JDI, OQG, OQH, OQI, PBI Dated: February 25, 2015 Received: February 26, 2015

Dear Ms. Amy Walriven:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K150503

Device Name Echo Bi-Metric Microplasty Line Extension

Indications for Use (Describe)

1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.

2) Rheumatoid arthritis.

3) Correction of functional deformity.

4) Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques.

5) Revision of previously failed total hip arthroplasty.

Porous coated components are intended for uncemented, biological fixation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The color of the letters is a dark gray, contrasting with the white background. The overall impression is clean and professional.

# 510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Echo Bi-Metric Microplasty Line Extension 510(k) premarket notification.

| Sponsor:        | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:        | Amy L Walriven<br>Manager, Regulatory Affairs<br>Phone: (574) 372-6660<br>Fax: (574) 372-1683                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Date:           | March 13, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Subject Device: | Trade Name: Echo Bi-Metric Microplasty Hip System<br>Common Name: Hip Prosthesis<br>Product Code(s): LPH, LZO, KWZ, JDI, KWL, LWJ, KWY, OQG, OQH, OQI, PBI<br>Regulation/Description:<br>• 21 CFR 888.3358 - Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis<br>• 21 CFR 888.3353 - Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis<br>• 21 CFR 888.3310 - Hip joint metal/polymer constrained<br>cemented or uncemented prosthesis<br>• 21 CFR 888.3350 - Hip joint metal/polymer semi-constrained<br>cemented prosthesis<br>• 21 CFR 888.3360 – Hip joint femoral (hemi-hip) metallic cemented<br>or uncemented prosthesis<br>• 21 CFR 888.3390 – Hip joint femoral (hemi-hip) metal/polymer<br>cemented or uncemented prosthesis |

# Legally marketed devices to which substantial equivalence is claimed:

- Echo Bi-Metric Microplasty Hip System (K143009) ●
- Balance Microplasty Hip System (K050251) - Reference

#### Device Description

The Echo Bi-Metric Microplasty Femoral Stem is a monolithic, collarless, straight stem designed to reduce hip pain for patients and restore joint biomechanics and stability. The femoral stem is designed to fit patient femoral anatomies for primary or revision hip arthroplasties. The stems

{4}------------------------------------------------

are manufactured from Ti-6Al-4V per ASTM F136/ASTM F620 and feature a proximal circumferential porous plasma spray (PPS) coating for biological fixation per ASTM F1580.

Patient contacting, reusable implant specific instruments are manufactured from stainless steel and include an aluminum titanium nitride (AlTiN) PVD or titanium nitride (TiN) PVD coating.

# Intended Use and Indications for Use

- 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis. 2)
- Correction of functional deformity. 3)
- 4) Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed total hip arthroplasty. 5)

Porous coated components are intended for uncemented, biological fixation.

# Summary of Technological Characteristics

The rationale for substantial equivalence is based on consideration of the following characteristics:

- . Intended Use: The proposed and predicate Echo Bi-Metric Microplasty Femoral Stem devices have the identical intended use.
- Indications for Use: The proposed and predicate Echo Bi-Metric Microplasty Femoral Stem devices have identical indications for use.
- . Materials: The proposed and predicate Echo Bi-Metric Microplasty Femoral Stem devices are manufactured from Ti-6Al-4V per ASTM F136 and ASTM F620 and feature a porous plasma spray (PPS) coating per ASTM F1580.
- Design Features: The proposed and predicate Echo Bi-Metric Microplasty Femoral Stem devices incorporate the same design features.
- Sterilization: The proposed and predicate Echo Bi-Metric Microplasty Femoral Stem devices are provided sterile via the same sterilization methods for single-use.

# Summary of Performance Data

Results from mechanical tests demonstrate that the proposed Echo Bi-Metric Microplasty Femoral Stems are substantially equivalent to the predicate femoral stems. A description of the tests performed on the proposed device is as follows:

- . Proximal Fatigue Testing - ISO 7206-6
- . Distal Fatigue Testing - ISO 7206-4
- Range of Motion Analysis — ISO 21535

# Substantial Equivalence Conclusion

The proposed Echo Bi-Metric Microplasty Femoral Stems have the same intended use and indications for use as the predicate devices. Performance test data demonstrates the device is as safe and effective and is substantially equivalent to the legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K150503](https://fda-staging.innolitics.com/device/K150503)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).