← Product Code [JWH](/productcode/JWH) · K150496

# Total Knee Arthroplasty System (K150496)

_Responsive Orthopedics, LLC · JWH · Jun 12, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150496

## Device Facts

- **Applicant:** Responsive Orthopedics, LLC
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Jun 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Responsive Orthopedics Total Knee Arthroplasty System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: - noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; - inflammatory degenerative joint disease including rheumatoid arthritis; - correction of functional deformity: - revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques. The Responsive Orthopedics Total Knee Arthroplasty System is for cemented use only.

## Device Story

Total knee arthroplasty system designed to restore alignment, stability, and range of motion while alleviating pain by replacing knee joint articulating surfaces. System includes femoral, tibial, and patellar components in cruciate retaining, ultra-conforming cruciate sacrificing, medial-pivot cruciate sacrificing, and posterior stabilizing versions. Used by orthopedic surgeons in clinical settings to treat degenerative joint disease or trauma. Components implanted via cement fixation. Output is a reconstructed knee joint; clinical benefit includes improved joint function and pain reduction.

## Clinical Evidence

Bench testing only. Testing included tibial baseplate fatigue (ASTM F1800), UHMWPE bearing thickness justification, posterior-stabilized bearing shear fatigue, range of motion/constraint (ASTM F1223), tibiofemoral contact area/stress (ASTM F2083), and component interlock strength (ASTM F1814).

## Technological Characteristics

Femoral components and tibial trays: cast Co-Cr-Mo alloy (ASTM F75). Poly inserts and patella components: noncrosslinked GUR 1050 Ram Extruded UHMWPE (ASTM F648). Semi-constrained cemented prosthesis. System includes multiple versions: cruciate retaining, ultra-conforming cruciate sacrificing, medial-pivot cruciate sacrificing, and posterior stabilizing.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- DePuy Sigma® Knee System ([K943462](/device/K943462.md), [K961685](/device/K961685.md), [K032151](/device/K032151.md), [K950010](/device/K950010.md))
- Wright Medical Technology EVOLUTION™ Medial Pivot ([K093552](/device/K093552.md))
- Consensus Knee System ([K932837](/device/K932837.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized image of three human profiles facing to the right, stacked one behind the other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2015

Responsive Orthopedics LLC % Jude Paganelli Cor Medical Ventures LLC 101 North Acacia Avenue, Suite 106 Solana Beach, California 92075

Re: K150496 Trade/Device Name: Responsive Orthopedics Total Knee Arthroplasty System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 13, 2015 Received: March 16, 2015

Dear Jude Paganelli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 -Jude Paganelli

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K150496

Device Name

Responsive Orthopedics Total Knee Arthroplasty System

#### Indications for Use (Describe)

The Responsive Orthopedics Total Knee Arthroplasty System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- · noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- · inflammatory degenerative joint disease including rheumatoid arthritis;
- · correction of functional deformity:
- · revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

The Responsive Orthopedics Total Knee Arthroplasty System is for cemented use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# SECTION 5

# 510(K) SUMMARY

## SUBMITTER:

Submitted By:

| Company Name: | Responsive Orthopedics LLC                 |
|---------------|--------------------------------------------|
| Address:      | 1755 Concordia Street<br>Wayzata, MN 55391 |
| Telephone:    | 858-774-7891                               |

CONTACT PERSON: Jude Paganelli

DATE PREPARED: February 23, 2015

TRADE NAME: Responsive Orthopedics Total Knee Arthroplasty System

COMMON NAME: RO Knee System

CLASSIFICATION NAME: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis. (21 CFR 888.3560)

PRODUCT CODE: JWH

### SUBSTANTIAL EQUIVALENCE:

The Responsive Orthopedics Total Knee Arthroplasty System is substantially equivalent to the predicate devices in all facets including: function, design, performance, material, and intended use. The predicate devices for the Responsive Orthopedics Total Knee Arthroplasty System are:

DePuy Sigma® Knee System (K943462, K961685, K032151, K950010) Wright Medical Technology EVOLUTION™ Medial Pivot (K093552) Consensus Knee System (K932837)

#### DEVICE DESCRIPTION:

The Responsive Orthopedics Total Knee Arthroplasty System is a total knee system for the restoration of alignment, stability and range of motion, and alleviation of pain, by replacing the articulating surfaces of the knee joint. The system includes femoral, tibial and patellar components available in cruciate retaining, ultra-conforming cruciate sacrificing, medial-pivot cruciate sacrificing and posterior stabilizing versions.

{4}------------------------------------------------

#### MATERIALS:

The RO TKA femoral components and tibial trays are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75.

The RO TKA poly inserts and patella components are manufactured from noncrosslinked GUR 1050 Ram Extruded UHMWPE conforming to ASTM F648.

#### INDICATIONS FOR USE:

The Responsive Orthopedics Total Knee Arthroplasty System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- noninflammatory degenerative joint disease including osteoarthritis, . traumatic arthritis, or avascular necrosis;
- inflammatory degenerative joint disease including rheumatoid arthritis:
- . correction of functional deformity;
- revision procedures where other treatments or devices have failed; and . treatment of fractures that are unmanageable using other techniques.

The Responsive Orthopedics Total Knee Arthroplasty System is for cemented use only.

#### PERFORMANCE TESTING:

The following bench testing was performed on the Responsive Orthopedics Total Knee Arthroplasty System:

Tibial Baseplate Fatique Testing per ASTM F1800 UHMWPE Tibial Bearing Components (minimum thickness justification) Posterior-Stabilized Tibial Bearing Component Shear Fatigue Testing Range of Motion and Constraint Testing per ASTM F1223 Tibiofemoral Contact Area/Stress per ASTM F2083 Component Interlock Strength per ASTM F1814

#### CONCLUSIONS:

The Responsive Orthopedics Total Knee Arthroplasty System has shown to be substantially equivalent to legally marketed predicates based on indications for use, technological characteristics, performance testing and comparison to predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K150496](https://fda-staging.innolitics.com/device/K150496)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).