FloSure Ventilation Patch

K150459 · Simex Medizintechnik GmbH · OMP · Jan 5, 2016 · General, Plastic Surgery

Device Facts

Record IDK150459
Device NameFloSure Ventilation Patch
ApplicantSimex Medizintechnik GmbH
Product CodeOMP · General, Plastic Surgery
Decision DateJan 5, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4780
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The FloSure Ventilation Patch is intended to be used in conjunction with NPWT Dressing Kits compatible with the SIMEX NPWT Systems EX200 and EX300 for the application of negative pressure wound therapy to the wound. The FloSure Ventilation Patch is to be applied to the occlusive wound dressing. When used in conjunction with the Simex NPWT EX200 and EX300 pumps and dressing kits, the FloSure Ventilation Patch is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of wound exudates, infectious material and tissue debris. The FloSure Ventilation Patch is appropriate for use on the following wounds: - Pressure Ulcers - Diabetic/Neuropathic Ulcers - Venous Insufficiency Ulcers - Traumatic Wounds - Post-Operative and Dehisced Surgical Wounds - Skin Flap and Grafts

Device Story

FloSure Ventilation Patch is an accessory for NPWT dressing kits; applied to occlusive wound dressing. Device features proprietary hydrophobic, micro-porous filter membrane; provides additional gas permeability to wound site. Used with SIMEX NPWT Systems (EX200/EX300) to balance vacuum pressure; prevents exudate stagnation/pooling; reduces blockage alarms. Operated by clinicians in clinical settings. Benefits patient by maintaining consistent negative pressure and effective exudate removal, supporting wound healing.

Clinical Evidence

No clinical data. Bench testing only: biocompatibility (cytotoxicity, sensitization, acute systemic toxicity, intracutaneous reactivity/irritation) and comparative performance testing (pressure, flow, exudate removal, alarm functionality) with predicate device.

Technological Characteristics

Accessory patch with hydrophobic, micro-porous filter membrane. Composed of polyurethane with adhesive backing. Sterile. Designed for use with SIMEX NPWT Systems (EX200/EX300).

Indications for Use

Indicated for patients requiring negative pressure wound therapy (NPWT) for pressure, diabetic/neuropathic, venous insufficiency, traumatic, post-operative/dehisced surgical wounds, and skin flaps/grafts. Used to promote healing via removal of exudates, infectious material, and tissue debris.

Regulatory Classification

Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 5, 2016 SIMEX Medizintechnik, GmbH Mr. Hamid Khosrowshahi FloSure Technologies LLC PO Box 123 Tarrytown, New York 10591 Re: K150459 Trade/Device Name: FloSure Ventilation Patch Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: December 3, 2015 Received: December 7, 2015 Dear Mr. Khosrowshahi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K150459 Device Name FloSure Ventilation Patch Indications for Use (Describe) The FloSure Ventilation Patch is intended to be used in conjunction with NPWT Dressing Kits compatible with the SIMEX NPWT Systems EX200 and EX300 for the application of negative pressure wound therapy to the wound. The FloSure Ventilation Patch is to be applied to the occlusive wound dressing. When used in conjunction with the Simex NPWT EX200 and EX300 pumps and dressing kits, the FloSure Ventilation Patch is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of wound exudates, infectious material and tissue debris. The FloSure Ventilation Patch is appropriate for use on the following wounds: - · Pressure Ulcers - · Diabetic/Neuropathic Ulcers - · Venous Insufficiency Ulcers - · Traumatic Wounds - · Post-Operative and Dehisced Surgical Wounds - · Skin Flap and Grafts Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) SUMMARY | Date Prepared: | January 2, 2016 | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor and<br>Manufacturer: | SIMEX Medizintechnik, GmbH<br>Post Box 1207<br>D-78649 Deisslingen, Germany<br>FDA Registration Number 3005813597 | | 510(k) Contact: | Mr. Hamid Khosrowshahi<br>FloSure Technologies LLC<br>PO Box 123<br>Tarrytown, NY 10591<br>Telephone: 914-772-7326<br>e-mail: hkhosrow@optonline.net | | Trade Name: | FloSure Ventilation Patch | | Common Name: | Occlusive Wound Dressing | | Classification: | Powered Suction Pump<br>21 CFR 878.4780<br>Class II | | Device Product Code: | OMP - Pump, Portable, Aspiration (manual or powered) | | Predicate Device: | UNI NPWT Foam Wound Dressing Kit (K133333) | Indications for Use: The FloSure Ventilation Patch is intended to be used in conjunction with NPWT Dressing Kits compatible with the SIMEX NPWT Systems EX200 and EX300 for the application of negative pressure wound therapy to the wound. The FloSure Ventilation Patch is to be applied to the occlusive wound dressing. When used in conjunction with the Simex NPWT EX200 and EX300 pumps and dressing kits, the FloSure Ventilation Patch is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of wound exudates, infectious material and tissue debris. The FloSure Ventilation Patch is appropriate for use on the following wounds: - Pressure Ulcers - . Diabetic/Neuropathic Ulcers - Venous Insufficiency Ulcers - Traumatic Wounds - Post-Operative and Dehisced Surgical Wounds - Skin Flap and Grafts Device Description: The FloSure Ventilation Patch is used with the NPWT wound dressings to improve air-flow through the dressing when desired. The FloSure Ventilation Patch augments the transparent dressing by providing additional gas permeability to the wound dressing site through its {4}------------------------------------------------ proprietary hydrophobic and micro-porous filter membrane. The access to additional air-flow can aid in keeping the vacuum in balance resulting in a continuous flow of exudate. Stagnation in the flow of exudate can cause pooling at the wound site and result in blockage alarms. | Feature | FloSure Ventilation Patch | UNI NPWT Foam Wound Dressing Kit<br>K133333 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name | FloSure Ventilation Patch | UNI NPWT Foam Wound Dressing Kits | | Product Code | OMP | OMP | | Description | Accessory for use with occlusive<br>dressings to allow filtered air to flow<br>through the occlusive dressing when<br>desired. Used with SIMEX Negative<br>Pressure Wound Therapy Systems (EX200<br>and EX300) | Foam based dressing including an<br>occlusive drape to create a sealed wound<br>environment<br>Used with SIMEX Negative Pressure<br>Wound Therapy Systems (EX200 and<br>EX300) | | Sterility | Provided Sterile | Provided Sterile | | Mode of operation | The FloSure Ventilation Patch is applied<br>to the occlusive (transparent) drape of the<br>NPWT wound dressing. It maintains the<br>moist environment of the NPWT dressing<br>and allows the exchange of gases through<br>a hydrophobic micro-porous filter<br>membrane. | The occlusive drape of the wound dressing<br>kit provides a moist wound environment,<br>and allows the exchange of gases through<br>the device. | | Materials | Consists of a piece of synthetic polymeric<br>material with an adhesive backing. | Occlusive drape consists of a piece of<br>synthetic polymeric material with an<br>adhesive backing. | | Indications for Use | The FloSure Ventilation Patch is intended<br>to be used in conjunction with NPWT<br>Dressing Kits compatible with the SIMEX<br>NPWT Systems EX200 and EX300 for the<br>application of negative pressure wound<br>therapy to the wound. The FloSure<br>Ventilation Patch is to be applied to the<br>occlusive wound dressing. When used in<br>conjunction with the Simex NPWT EX200<br>and EX300 pumps and dressing kits, the<br>FloSure Ventilation Patch is indicated for<br>patients who would benefit from a suction<br>device, particularly as the device may<br>promote wound healing by removal of<br>wound exudates, infectious material and<br>tissue debris.<br>The FloSure Ventilation Patch is<br>appropriate for use on the following<br>wounds:<br>• Pressure Ulcers<br>• Diabetic/Neuropathic Ulcers<br>• Venous Insufficiency Ulcers<br>• Traumatic Wounds<br>• Post-Operative and Dehisced Surgical<br>Wounds | The UNI NPWT Foam Dressing Kit is<br>intended to be used in conjunction with the<br>SIMEX Negative Pressure Wound<br>Therapy Pumps (K113291) for the<br>application of negative pressure wound<br>therapy to the wound. When used in<br>conjunction with the SIMEX Negative<br>Pressure Wound Therapy Pumps, the UNI<br>NPWT Foam Dressing Kit is indicated for<br>patients who would benefit form a suction<br>device, particularly as the device<br>may promote wound healing by removal of<br>excess exudates, infectious material and<br>tissue debris.<br>The UNI NPWT Foam Dressing Kit is<br>appropriate for use on the following<br>wounds:<br>• Pressure Ulcers<br>• Diabetic/Neuropathic Ulcers<br>• Venous Insufficiency Ulcers<br>• Traumatic Wounds<br>• Post-Operative and Dehisced Surgical<br>Wounds<br>• Skin Flap and Grafts | Feature Comparison Chart {5}------------------------------------------------ ## Non-clinical Testing: The following non-clinical testing was performed: - Cytotoxicity - Sensitization - Acute Systemic Toxicity - Intracutaneous Reactivity or Irritation Comparative performance bench testing was performed with the predicate and with the FloSure Ventilation Patch using the SIMEX EX200 and EX300 NPWT systems. Testing demonstrated equivalence in ability to maintain consistent pressure, flow and removal of exudate and demonstrated that the FloSure Ventilation patch did not interfere with the proper functioning of the NPWT pump alarms. Clinical Testing: No clinical study was conducted. ## Substantial Equivalence: The FloSure Ventilation Patch is substantially equivalent in function and intended use to the UNI NPWT Foam Dressing Kit (K13333). Like the occlusive drape of the predicate UNI NPWT Foam Dressing Kit, the FloSure Ventilation Patch is semipermeable, maintains a moist environment for the wound by being hydrophobic, allows the exchange of gases, and maintains a protective barrier to the wound site. The FloSure Ventilation Patch performs the same basic function as the occlusive drape of NPWT Wound Dressing Kit. The support layer of the FloSure Ventilation Patch, which adheres to the NPWT occlusive drape, is composed of polvurethane, the same material that the occlusive drape is composed of. Performance testing with the predicate device and with the FloSure Ventilation Patch demonstrated that they were substantially equivalent in the ability to maintain consistent pressure, flow and removal of exudate when tested with the SIMEX EX200 and EX300 NPWT Systems. Conclusion: The FloSure Ventilation Patch is substantially equivalent to the currently marketed UNI NPWT Foam Wound Dressing Kits (K13333), in indications for use, basic technological characteristics, and does not raise new issues of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...