The AlfaSight 9000 is intended for the adjunct diagnosis of: 1) abnormalities of the female breast; 2) peripheral vascular disease; 3) musculoskeletal disorders; 4) extracranial cerebral and facial vascular disease; 5) abnormalities of the thyroid gland; and 6) various neoplastic and inflammatory conditions. The AlfaSight 9000 is not intended to serve as a sole diagnostic screening procedure.
Device Story
AlfaSight 9000 is a non-invasive, handheld, battery-operated thermographic system for clinical use by licensed practitioners. It measures skin surface temperatures to assist in diagnosing various conditions. The system includes a handheld probe, an off-the-shelf computer, and disposable polyethylene sheaths. During operation, the sheathed probe contacts the skin; capacitive pins trigger temperature measurement via infrared technology. Data is transmitted wirelessly via Bluetooth to the computer, which displays graphical thermographic charts. The software guides the user through a fixed sequence of body measurement points. The output provides healthcare providers with relative temperature data to be used as an adjunct to other clinical diagnostic procedures (e.g., X-ray, CT, mammography). It does not replace standard diagnostic tests but aids clinical decision-making by quantifying thermal patterns.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included Corvalent Design Verification, temperature probe with sheath performance testing, biocompatibility (ISO 10993-1), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Handheld probe with infrared sensor; disposable polyethylene sheaths (ISO 10993-1). Bluetooth wireless connectivity to off-the-shelf PC. Polymer Lithium-Ion battery. Measurement range 25-40°C; accuracy +/-0.2°C. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for adult patients (non-pediatric) as an adjunct diagnostic tool for breast abnormalities, peripheral vascular disease, musculoskeletal disorders, extracranial cerebral/facial vascular disease, thyroid abnormalities, and various neoplastic/inflammatory conditions. Not for use as a sole diagnostic screening procedure.
Regulatory Classification
Identification
A telethermographic system for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the outside. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2015
Alfa Thermodiagnostics, Inc. % Mr. John Smith Regulatory Counsel 555 13th Street, N.W. WASHINGTON DC 20004
Re: K150457
Trade/Device Name: AlfaSight 9000 Thermographic System Regulation Number: 21 CFR 884.2980 Regulation Name: Telethermographic system Regulatory Class: I Product Code: LHO Dated: February 20, 2015 Received: February 20, 2015
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150457
Device Name AlfaSight 9000 Thermographic System
#### Indications for Use (Describe)
The AlfaSight 9000 is intended for the adjunct diagnosis of the female breast; 2) peripheral vascular disease; 3) musculoskeletal disorders; 4) extracranial cerebral and facial vascular disease; 5) abnormalities of the thyroid gland; and 6) various neoplastic and inflammatory conditions.
The AlfaSight 9000 is not intended to serve as a sole diagnostic screening procedure.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
| DATE | February 20, 2015 |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER | Alfa Thermodiagnostics, Inc. |
| | 9057 Soquel Drive, Suite B-AA |
| | Aptos, CA 95003 |
| | Tel: (831) 901-9248 |
| | Fax: (831) 685-1128 |
| | Email: drbeilin@alfathermo.com |
| CONTACT PERSON | Daniel Beilin, O.M.D., Alfa Thermodiagnostics, Inc |
| | 9057 Soquel Drive, B-AA |
| | Aptos, CA. 95003 |
| | Tel: (831) 901-9248 |
| | Fax: (831) 685-1128 |
| | Email: drbeilin@alfathermo.com |
| DEVICE NAME | AlfaSight 9000 Thermographic System |
| Classification: | Class I |
| Common Name | Telethermographic system |
| Classification | 21 C.F.R. § 884.2980 (adjunctive use) |
| Product Code | LHQ |
| Review Panel | Obstetrics/Gynecology |
| PREDICATE DEVICE | Eidam's CRT 2000 (K971956) |
| INTENDED USE | The AlfaSight 9000 is intended for the adjunct diagnosis of: 1) abnormalities of<br>the female breast; 2) peripheral vascular disease; 3) musculoskeletal disorders;<br>4) extracranial cerebral and facial vascular disease; 5) abnormalities of the<br>thyroid gland; and 6) various neoplastic and inflammatory conditions. |
| | The AlfaSight 9000 is not intended to serve as a sole diagnostic screening<br>procedure. |
#### DEVICE DESCRIPTION
The AlfaSight 9000 system is a non-invasive, handheld and battery-operated adjunctive diagnostic device that measures and records skin surface temperatures. The system is intended for use as an adjunct to other clinical diagnostic procedures to quantify relative skin surface temperatures. It is not intended as a sole diagnostic device and does not replace or substitute any clinical test or device such as X-Ray, CT scans, mammography or other similar equipment. The system is for prescription-only use by licensed medical practitioners for use in hospitals, sub-acute healthcare settings, and private clinics.
The AlfaSight 9000 system consists of: 1) a handheld probe, 2) an off-the-shelf computer 3) a standard USB charging cable and 4) biocompatible sheaths. The single use theath is placed over the handheld probe covering the entire probe, three capacitive pins sense contact of the skin through the sheath to start the
{4}------------------------------------------------
temperature measurement. Temperature samples are then sent from the handheld probe to the computer wirelessly using Bluetooth technology and are represented in a graphical display on the computer. Samples are organized into specified regions of the body as they are sampled in a fixed order as directed by the software.
# SUBSTANTIAL EQUIVALENCE
A comparison of the AlfaSight 9000 System's indications for use, technological characteristics, and principles of operation with the predicate device indicates that the AlfaSight 9000 system is substantially equivalent to the FDA-cleared CRT 2000 thermographic device (K971956).
The AlfaSight 9000, like the predicate device CRT-2000, is intended as an adjunct to other clinical diagnostic procedures. The indications for use are identical to the indications for the predicate CRT-2000 device.
The AlfaSight 9000 system differs from its predicate, the CRT-2000, in that it measures the skin temperature (through a biocompatible sheath) using infrared technology rather than the thermocouple measurements employed by the CRT-2000 device. The use of infrared measurements does not affect the safety or effectiveness of the AlfaSight 9000 system for its intended use, as demonstrated by the completed bench testing, including Corvalent Design Verification Testing and Temperature Probe with Sheath Performance Testing.
The points measured for the AlfaSight 9000 systems are identical to the points measured by the CRT 2000 predicate device.
| Parameter | AlfaSight 9000 | CRT 2000 (K971956) |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use / Indications for use | Adjunct diagnosis of:<br>1) abnormalities of the female breast;<br>2) peripheral vascular disease;<br>3) musculoskeletal disorders;<br>4) extracranial cerebral and facial vascular disease;<br>5) abnormalities of the thyroid gland; and<br>6) various neoplastic and inflammatory conditions.<br><br>The AlfaSight 9000 is not intended to serve as a sole diagnostic screening procedure. | Adjunct diagnosis of:<br>1) abnormalities of the female breast;<br>2) peripheral vascular disease;<br>3) musculoskeletal disorders;<br>4) extracranial cerebral and facial vascular disease;<br>5) abnormalities of the thyroid gland; and<br>6) various neoplastic and inflammatory conditions.<br><br>The CRT 2000 is not intended to serve as a sole diagnostic screening procedure. |
| Parameter | AlfaSight 9000 | CRT 2000 (K971956) |
| Functionality | Captures skin surface<br>temperatures, provides that<br>data in a thermographic chart<br>for use by the healthcare<br>practitioner as an adjunct to<br>other clinical procedures. | Captures skin surface<br>temperatures, provides that<br>data in a thermographic chart<br>for use by the healthcare<br>practitioner as an adjunct to<br>other clinical procedures. |
| Where used | Doctor's Office, Clinic or<br>Hospital | Doctor's Office, Clinic or<br>Hospital |
| Anatomical Sites | Same | Same |
| Target Population/ demographic | Adult use, non pediatric | Adult use, non pediatric |
| Device Regulatory Classification | 21 CFR 884.2980 | 21 CFR 884.2980 |
| Product Code | LHQ | LHQ |
| Device Class | Class I | Class I |
| Biocompatibility | (Sheath) ISO 10993-1 | Data Not Available |
| EMC and Safety | IEC 60601-1; IEC 60601-1-2 | Data Not Available |
| 510(k) number | To be determined | K971956 |
| PC Supply voltage | AC 85 -264V | AC 110 - 240 V |
| PC Power Input | 60 Watts | 180 Watt |
| PC Frequency | Hz 47-63 Hz | Hz 50/60 |
| PC Storage Temperature Range | -40°C to +85°C | 0°C - 50°C |
| PC Screen Display | Graphic LCD Display | Graphic LCD Display |
| Probe Battery | Polymer Lithium-Ion | None |
| Probe Rechargeable | Yes | N/A |
| Probe Status Indicator | Yes | Yes |
| Probe Error Margin | +/-0.2 degrees C | +/-0.2 degrees C |
| Probe Temp. Collection method | Infrared thermocouple | bi-metal thermocouple |
| Probe Skin Contact | Sheathed probe tips contact<br>the skin to start the<br>temperature measurement.<br>The IR sensor does not touch<br>the skin. | Thermocouple sensor and edge<br>surface of probe tip contact<br>the skin to take the<br>measurement. The sensor<br>touches the skin. |
| Disposable Sheath Material | Polyethylene | N/A |
| Probe Optimum Range | 25-40°C | 27-35°C |
| Probe Communication | Blue Tooth Wireless<br>Communication of<br>measurement data from Probe<br>to PC | Wired Communications |
| Parameter | AlfaSight 9000 | CRT 2000 (K971956) |
| Software Main functions | Receives temperature signal<br>and generates multiple point<br>temperature measurement<br>graphics for healthcare<br>practitioners on measurement<br>locations and temperature<br>measurement readings. | Receives temperature signal<br>and generates multiple point<br>temperature measurement<br>graphics for healthcare<br>practitioners on measurement<br>locations and temperature<br>measurement readings. |
| Correct Orientation and location | Measurement points are<br>highlighted on the figure on<br>the display screen for<br>measurement location and<br>sequence of measurements. | Measurement points are<br>highlighted on the figure on<br>the display screen for<br>measurement location and<br>sequence of measurements. |
The following table provides more detailed information regarding the basis for the determination of substantial equivalence:
{5}------------------------------------------------
{6}------------------------------------------------
## PERFORMANCE TESTING
The AlfaSight 9000 System has been determined through engineering to be substantially equivalent to the predicate device. Testing demonstrates that the AlfaSight 9000 conforms to ISO 10993-1, IEC 60601-1, and IEC 60601-1-2.
The following non-clinical bench performance testing has been completed for the subject device:
- Biocompatibility
- Software Risk Analysis
- Software Requirements Specification ●
- Software Validation Protocol
- Software Verification Summary and Test Case Report
- Software Test Results Report and Revision History in
- Software Design Specification
- Software Development Environment Description Document
- Electromagnetic Compatibility and Electrical Safety
## CONCLUSION
In sum, the results of non-clinical testing demonstrate that the AlfaSight 9000 System is as safe, as effective and performs as well as the predicate device. In sum, the AlfaSight 9000 is substantially equivalent to the CRT 2000 predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.