RT2 Trabeculite Titanium Cervical Cage System

K150455 · Rhausler, Inc. · ODP · Nov 19, 2015 · Orthopedic

Device Facts

Record IDK150455
Device NameRT2 Trabeculite Titanium Cervical Cage System
ApplicantRhausler, Inc.
Product CodeODP · Orthopedic
Decision DateNov 19, 2015
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Rhausler RT2 Trabeculite Titanium Cervical Cage System is intended for spinal fusion procedures at one level (C3-T1) in skeletally mature patients with cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Implants are to be implanted via an open, anterior approach and packed with autogenous bone. This device is to be used in patients who have at least 6 weeks of nonoperative treatment. This device is intended to be used with supplemental spinal fixation systems.

Device Story

Cervical intervertebral body fusion device; rectangular, lordotic titanium alloy (Ti6Al4V) cage; hollow center for autogenous bone graft. Used in open, anterior cervical spine surgery; requires supplemental spinal fixation. Implanted by orthopedic or neurosurgeons to facilitate fusion in patients with discogenic neck pain. Provides structural support to vertebral space during bone healing; benefits patient by stabilizing segment and promoting fusion.

Clinical Evidence

Bench testing only. Mechanical testing per ASTM F2077-11 (static/dynamic compression and torsion) and ASTM F2267-04 (subsidence and expulsion). Biocompatibility testing per ISO 10993 and GLP regulations. Sterilization validation per ANSI/AAMI/ISO 11137-1/2.

Technological Characteristics

Material: Medical grade Titanium alloy (Ti6Al4V). Form: Rectangular cage with 4-degree lordotic profile, hollow cylindrical fenestration. Dimensions: Two cross-sectional sizes, various heights. Sterilization: ANSI/AAMI/ISO 11137-1/2 (SAL 10^-6). Mechanical standards: ASTM F2077-11, ASTM F2267-04.

Indications for Use

Indicated for skeletally mature patients with cervical disc disease (neck pain of discogenic origin with disc degeneration confirmed by history/radiography) requiring spinal fusion at one level (C3-T1). Requires at least 6 weeks of prior nonoperative treatment. Contraindications: not specified.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 19, 2015 Rhausler, Incorporated % Ms. Dionicia Reblando Regulatory Consultant 837 Industrial Road, Unit E San Carlos, California 94070 Re: K150455 Trade/Device Name: RT2 Trabeculite Titanium Cervical Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: October 21, 2015 Received: October 23, 2015 Dear Ms. Reblando: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K150455 Device Name RT2 Trabeculite Titanium Cervical Cage System #### Indications for Use (Describe) The Rhausler RT2 Trabeculite Titanium Cervical Cage System is intended for spinal fusion procedures at one level (C3-T1) in skeletally mature patients with cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Implants are to be implanted via an open, anterior approach and packed with autogenous bone. This device is to be used in patients who have at least of nonoperative treatment. This device is intended to be used with supplemental spinal fixation systems. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 5: 510(k) Summary | Assigned 510(k) number: | K150455 | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company: | Rhausler Inc.<br>837 Industrial Road, Unit E<br>San Carlos, CA 94070 USA<br>Phone: 650-631-4515<br>Fax: 650-631-4555 | | Contact: | Dionicia B. Reblando<br>Email: joann.reblando@yahoo.com | | Date Prepared: | October 17, 2015 | | Proprietary Names: | RT2 Trabeculite Titanium Cervical Cage System | | Classification Name: | Intervertebral body fusion device | | Classification: | 21 CFR 888.3080, Class II, Product Code ODP | | Predicate Device: | Primary predicate device-Nubic and Rabea Device (K082848) Interbody<br>Fusion device with bone graft, cervical.<br>Other predicate device: The Rhausler Plage Anterior Cervical<br>Fusion System (K111272) | | Device Description: | The Rhausler RT2 Trabeculite Titanium Cervical Cage System is a<br>cervical intervertebral body fusion device. The system is comprised<br>of implants and instruments. The instruments are utilized for the<br>implantation procedure. All implants are manufactured from<br>medical grade Titanium alloy (Ti6 Al 4V).<br>The implant is rectangular in shape with lordotic configuration, i.e.,<br>with 4 degree profiles. It is offered in two cross sectional sizes and<br>is available in various height options to accommodate variations in<br>patient anatomy. The center of the implant has a hollow cylindrical<br>shape fenestration for placement of autogenous bone. Trabeculite<br>Titanium Cervical Cage implant is provided sterile. | | Indications For Use: | The Rhausler RT2 Trabeculite Titanium Cervical Cage System is<br>intended for spinal fusion procedures at one level (C3-T1) in<br>skeletally mature patients with cervical disc disease (defined as<br>neck pain of discogenic origin with degeneration of the disc<br>confirmed by history and radiographic studies). Implants are to be | {4}------------------------------------------------ | | implanted via an open, anterior approach and packed with autogenous bone. This device is to be used in patients who have at least 6 weeks of nonoperative treatment. | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | This device is intended to be used with supplemental spinal fixation systems. | | Technological<br>Comparison to Predicate<br>Device: | The Rhausler RT2 Trabeculite Titanium Cervical Cage System is equivalent to the predicate devices. Both are intervertebral body fusion devices that attach to the anterior portion of the cervical spine during the development of spinal fusion. Additionally, both use the same fundamental scientific technology, principle of operation, and materials. Performance test results confirm that design differences do not pose new issues of safety or effectiveness. | | Performance Testing: | Mechanical testing has been performed on the RT2 Trabeculite Titanium Cervical Cage System in accordance to the FDA recommendation entitled "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device", which included Compression tests (Static and Dynamic) and Torsion tests (Static and Dynamic) per ASTM F2077-11, Subsidence testing per ASTM F2267-04 and Expulsion test. The results of testing and analyses conducted, demonstrate that the proposed system adequately meets the predetermined requirements established for its mechanical performance. | | | Biocompatibility tests for the RT2 Trabeculite Titanium Cervical Cage were selected in accordance with blue book memorandum G95-1 entitled Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing". Biocompatibility tests were conducted in compliance with U.S. Food and Drug Administration Good Laboratory Practice (GLP) regulations set forth in 21 CFR 58. All tests conducted passed biocompatibility requirements. | | | Sterilization validation of the RT2 Trabeculite Titanium Cervical Cage implant is to be based on the requirements of ANSI/AAMI/ISO 11137-1:2006 (R) 2010 and 11137-2:2013 with a minimum Sterility Assurance Level of 1X10-6 | | Conclusion: | Based upon the design, technology, performance, and intended use, the Rhausler RT2 Trabeculite Titanium Cervical Cage System is substantially equivalent to the predicate device/s currently marketed under the Food, Drug and Cosmetic Act. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...