K150453 · Acclarent, Inc. · EGJ · May 20, 2015 · Physical Medicine
Device Facts
Record ID
K150453
Device Name
TULA Iontophoresis System with Earset
Applicant
Acclarent, Inc.
Product Code
EGJ · Physical Medicine
Decision Date
May 20, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5525
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Iontophoresis System with Earset is indicated for the administration of drug solution, salts, or ions into the ear, including the tympanic membrane, for medical purposes.
Device Story
TULA Iontophoresis System with Earset is a single-use device for delivering drug solutions, salts, or ions into the ear/tympanic membrane via electric current; consists of Iontophoresis Control Unit, Iontophoresis Earsets, and Return Electrode Patch; accessories include syringe and earset sizers. Device uses soft pressure-sensitive adhesive (PSA) to secure ear plug in external ear canal, replacing the frame/spring mechanism of the predicate. System delivers electrical current to transport substances; used by clinicians to facilitate localized drug delivery. Benefits include targeted administration to the ear canal/tympanic membrane.
Clinical Evidence
Bench testing only. Verification testing included system tests, ear canal pressure and leak tests, fill system burst tests, plug peel force tests, tack force tests, and biocompatibility testing. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
System delivers electrical current for iontophoresis. Components: Control Unit, Integrated Ear Electrode, Return Electrode Patch. Uses soft pressure-sensitive adhesive (PSA) for ear plug retention. Non-sterile. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for administration of drug solution, salts, or ions into the ear, including the tympanic membrane, for medical purposes.
Regulatory Classification
Identification
An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined. Additionally, an iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted:
(A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended;
(B) Testing of the ability of the device to maintain a safe pH level; and
(C) If used in the ear, testing of the device to demonstrate mechanical safety.
(ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following:
(A) A description and/or graphical representation of the electrical output;
(B) A description of the electrode materials and pH buffer;
(C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and
(D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning:
*Warning:* Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile.
(vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.
Predicate Devices
Iontophoresis System with Headset or IPSHS (K110636)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 20, 2015
Acclarent, Inc. Pavan Sethi, PhD Manager, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025
Re: K150453
Trade/Device Name: Tula Iontophoresis System with Earset Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis Device Regulatory Class: Class III Product Code: EGJ Dated: April 21, 2015 Received: April 22, 2015
Dear Dr. Sethi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Our substantially equivalent decision does not apply to the drugs that you will label or promote for use with your device. Therefore, you may neither label nor promote your device for use with specific drugs, nor package drugs with your device prior to FDA having approved the drugs for iontophoretic administration. For information on the requirements for marketing new drugs, you may contact:
> Director Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857
As you are aware, there are concerns relating to the fact that no drug is currently labeled for administration via an iontophoresis device. The Agency currently is evaluating this public health concern regarding the safety and effectiveness of this route of administration of drugs, and in the near future will inform manufacturers of certain additional steps the Agency believes are necessary to address this concern.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{2}------------------------------------------------
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aguel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150453
Device Name
TULA® Iontophoresis System with Earset
Indications for Use (Describe)
The Iontophoresis System with Earset is indicated for the administration of drug solution, salts, or ions into the ear, including the tympanic membrane, for medical purposes.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 1em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------|-----------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. The first part of the word, "Acclar", is in black, while the "ent" is in a purple color. Above the word "ent" is a curved line, also in purple, that extends over the letters.
TULA Iontophoresis System with Earset
## 510(k) SUMMARY
| Sponsor/Submitter: | Acclarent, Inc.<br>1525-B O'Brien Drive<br>Menlo Park, California 94025 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Pavan Sethi, Ph.D.<br>Manager, Regulatory Affairs<br>Phone: (650) 687-5413<br>Fax: (650) 687-4847 |
| Date of Submission: | February 19, 2015 |
| Device Trade Name: | TULA® Iontophoresis System |
| Common Name: | Iontophoresis System with Earset |
| Device Classification: | Class III |
| Regulation Number: | 21 CFR 890.5525 |
| Classification Name: | Device, Iontophoresis, Other Purposes |
| Product Code: | EGJ |
| Predicate Device: | Iontophoresis System with Headset or IPSHS (K110636) manufactured by<br>Acclarent |
| Device Description: | The Acclarent Iontophoresis System with Earset (IPSES) is a single-use<br>device that employs electric current to transport drug solution, salts, or ions<br>into the ear, including the tympanic membrane. The TULA lontophoresis<br>System consists of three components, namely an Iontophoresis Control Unit,<br>Iontophoresis Earsets and a Return Electrode Patch. Accessories to the IPSES<br>include a Syringe and Earset Sizers. All components of the IPSES are provided<br>non-sterile and no sterilization is required. |
| Indications for Use: | The Iontophoresis System with Earset is indicated for the administration of<br>drug solution, salts, or ions into the ear, including the tympanic membrane, for<br>medical purposes. |
| Technological<br>Characteristics: | IPSES delivers an electrical current to the ear. The electrical current transports<br>drug solution, salts, or ions into the ear, including the tympanic membrane. The<br>subject device is identical to the predicate device in method of operation and<br>intended use. The Iontophoresis process and the components controlling it<br>including Control Unit, Integrated Ear Electrode and Return Electrode Patch<br>are unchanged between the subject and predicate device. The main differences<br>in technological characteristics include, modification of the mechanism to |
| | retain the ear plug in the external ear canal and integration of ear plugs into the<br>design of the earset instead of being a separate accessory in the predicate<br>Headset device. In the predicate device, the ear plug is attached to the Headset<br>and retains the fluid (drug solution, salts, or ions) within the ear canal by<br>forming a seal. The modified device integrates this ear plug into a smaller sized<br>Earset to achieve an analogous seal. The predicate device uses a frame to<br>maintain the Headset in position and a spring to hold the ear plug in the ear<br>canal, whereas the modified subject device uses a small amount of soft<br>pressure sensitive adhesive (PSA) to secure the ear plug to the patient. These<br>differences in technological characteristics do not raise different questions of<br>safety and effectiveness than the predicate and do not render this device Not<br>Substantially Equivalent (NSE). |
| Performance Data: | Bench verification testing was conducted to verify that the modified device<br>meets the design inputs and intended performance characteristics. The testing<br>included system test, ear canal pressure and leak test, fill system burst test,<br>plug peel force test, tack force test, and biocompatibility tests. |
| | Electrical testing demonstrated that the subject device IPSES meets all<br>applicable requirements of the standards of IEC 60601-1 and IEC 60601-1-2. |
| Summary of<br>Substantial<br>Equivalence: | The subject device IPSES is substantially equivalent to the predicate device in<br>indication for use, performance, fundamental scientific technology, safety and<br>effectiveness. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "rent" portion in purple. A purple curved line extends over the word, adding a stylized element to the logo. The logo is simple and modern, reflecting the company's focus on innovation in the medical field.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.