K150409 · Sung Hwan E&B Co.,Ltd · OUH · Jan 15, 2016 · General, Plastic Surgery
Device Facts
Record ID
K150409
Device Name
VIVACE Electrosurgical System
Applicant
Sung Hwan E&B Co.,Ltd
Product Code
OUH · General, Plastic Surgery
Decision Date
Jan 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. This Vivace Electrosurgical System is intended for use with Skin Type I to Skin Type V.
Device Story
System uses bipolar radiofrequency (RF) energy to perform electrocoagulation and hemostasis in dermatologic/general surgery and treat facial wrinkles. Device consists of main console with LCD touch screen, RF generator, and power supply; handpiece uses disposable sterile 36-needle electrode cartridges (6x6 array). Physician operates handpiece, placing electrode tip in light contact with epidermis. RF energy delivered between adjacent electrode pairs generates electro-thermal reaction in target tissue. Output parameters (voltage, frequency) controlled via console. Clinical benefit includes reduction in wrinkle severity and solar elastosis. Adverse events (erythema, edema, sensitivity) typically resolve within 24 hours. Device intended for clinical use by physicians.
Clinical Evidence
Prospective clinical study of 31 participants (28 completed). Primary endpoint: change in Fitzpatrick Wrinkle and Elastosis scale at 30, 60, and 90 days. Results: 60.7% (17/28) of subjects showed ≥1 point improvement at 90 days. Adverse events (erythema, edema, sensitivity) occurred but decreased significantly across three treatment cycles. Pain levels also decreased over treatment sessions.
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis, and for the percutaneous treatment of facial wrinkles in patients with Fitzpatrick Skin Types I through V.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized image of three human profiles. The profiles are stacked on top of each other and are connected by a flowing line that resembles a ribbon or a wave.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
Sung Hwan E&B Company, Ltd % Mr. Kachi Enyinna 510K Technology Group, LLC 263 Huntington Avenue, #332 Boston, Massachusetts 02115
Re: K150409
Trade/Device Name: VIVACE Electrosurgical System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: OUH Dated: November 30, 2015 Received: December 3, 2015
Dear Mr. Enyinna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K150409
Device Name
VIVACE Electrosurgical System
Indications for Use (Describe)
The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. This Vivace Electrosurgical System is intended for use with Skin Type I to Skin Type V.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right:5px">☒</span>Prescription Use (Part 21 CFR 801 Subpart D) </span> <span style="padding-left:20px"> <span style="padding-right:5px">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Sung Hwan E&B Co., Ltd.
Vivace Electrosurgical System 510(k) Notification
### 510(k) Summary
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.
#### I. Sponsor's Information
| Name: | Sung Hwan E&B Co., Ltd. | | |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Address: | Suite #502, SK Techno Bldg<br>16-4 Sung Su-Dong 1 Ga<br>Sung Dong Ku, Seoul Korea | | |
| Phone: | 82-2-466-0010 | | |
| Fax: | 82-2-466-5473 | | |
| Contact Person: | Sun Young Kang<br>Tel: 82-2-466-0010<br>Fax: 82-2-466-5473<br>Email: ksy66820@naver.com | | |
| Official Correspondent: | Kachi Enyinna, MS<br>Principal Consultant<br>510K Technology Group, LLC<br>263 Huntington Avenue, Suite 332<br>Boston, MA 02115<br>(240) 994-4242 | | |
| Date Prepared: | January 6, 2016 | | |
| II. Device Name | | | |
| Trade Name: | VIVACE | | |
| Common Name: | Electrosurgical System and Accessories | | |
| Classification Name: | Electrosurgical Cutting and Coagulation device and<br>Accessories | | |
| Classification Number: | 21 CFR 878.4400 | | |
| Product Code: | OUH | | |
| Classification Panel: | General and Plastic Surgery | | |
#### III. Predicate Device
The VIVACE Electrosurgical System is substantially equivalent to the INFINI Radiofrequency System (K121481).
{4}------------------------------------------------
#### IV. Intended Use
The VIVACE Electrosurgical System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. This Vivace Electrosurgical System is intended for use with Skin Type I to Skin Type V.
#### V. Device Description
The VIVACE Electrosurgical System is comprised of the following components: the system main body of the device which consists of the LCD touch screen control panel, the high frequency generating output section (main P.C.B board or RF Generator), and the power supply component, the Switching Power Supply (SMPS). The accessories to the device include the handpiece with disposable micro-need le cartridge (electrode) insertion, and foot switch. Radio frequency current (RF energy) is delivered from the RF Generator, through the handpiece and electrode tip into the target tissue. The hand piece being held at right angles, the tip is placed in light contact with the epidermis. As RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating (causing minor dermal damage) the tissue. The bi-polar RF energy is delivered between independent adjacent electrode pairs (total 36 needle electrode, 6 x 6 array insertion). The RF generator, hand piece are not disposable. Each disposable micro-needle cartridge (electrode) is supplied sterile and is for single patient use only and cannot be resterilized.
| | Electrical voltage and frequency | AC 230V 50/60 Hz |
|--------------------------|----------------------------------|--------------------------------------------------------|
| | Power conception | 40 VA |
| | Maximum output current | 62mA ± 20% (load resistance<br>500Ω |
| | Maximum output voltage | 134V ± 20% (load resistance<br>500Ω |
| | Accuracy of the output frequency | 1 MHz |
| Microneedle<br>Cartridge | Electrode<br>Exposed<br>length | 36 each, 6 x 6 array<br>0.5 ~ 3.5mm (0.1mm increments) |
| | Outer<br>diameter | ∅0.3mm |
| Weight | | |
| Dimension | | (W)380mm x (L)340mm x<br>(H)1200mm |
#### VI. Technical Specifications
{5}------------------------------------------------
#### VII. Performance Data
Electrical safety test, Electromagnetic compatibility test, Sterility, and Biocompatibility testing were completed to demonstrate the safety and performance of the VIVACE Electrosurgical System.
Tests were performed on the Vivace Electrosrugical System and its sterile disposable single use micro-needle cartridge. The biological safety of the device has been demonstrated through biocompatibility studies of all patient contact materials in accordance with the standards outlined in IS010993-1
Results of the clinical testing demonstrated that the VIVACE Electrosurgical System is safe and effective for its intended use. The micro-need le cartridge is supplied sterile and sterility conforms to a Sterility Assurance Level (SAL) of 10° . The supplied Instructions for Use provide the user with the applicable warnings and cautions during use. There are no new safety or effectiveness issues related to this device.
#### VIII. Clinical Summary
The study had 31 participants with 28 subjects completing all three sessions and are included in the final study analysis. The data were evaluated against baseline at 30 days. 60 days and 90 days by three. blinded, independent licensed physicians using the 9-point Fitzpatrick Wrinkle and Elastosis scale scoring methodology for the peri-oral and periorbital regions. The scores showed improvement over the 90 day period after treatments. Among the 3 blinded evaluators, by 2/3 evaluator agreement (≥ 1 point change) criteria, 17 out of 28 (60.7%) subjects had one point improvement at 90 day follow up.
Adverse Events that occurred in greater than 5% of the subjects were Erythema (77.42%), Edema (19.35%), Edema (16.13%), and Sensitivity (6.45%). Adverse event cases decreased with each re-treatment phase for all AE types and were resolved within 24 hours. Overall adverse event rates were highest at the first treatment cycle, decreasing by a third after the second treatment and with little to no AEs reported after the third and final treatment (from 70.97% at treatment cycle 1 down to 6.45% at treatment cycle 3). Across treatments, each AE consistently decreased after the first, second and third (final) treatment cycles for every AE.
{6}------------------------------------------------
# Sung Hwan E&B Co., Ltd.
# Vivace Electrosurgical System 510(k) Notification
Most of the patients felt less pain overall through 3 treatment cycles. At 1st treatment cycle, for the peri-orbital wrinkle area, 9/28 subjects experienced pain at level 1-3 (32%) range, 12/28 at level 4 to 6 (43%) range, and 7/28 at level 7 to 10 (25%) range. During 3rd treatment cycle, 15/28 subjects reported experiencing pain at levels 1-3(53%), 12/28 at levels 4 to 6 (43%), and 1/28 at levels over 8 (3.5%). We hypothesize that the decrease in pain may be related to tissue coagulation and increased dermal thickness from initial exposure to RF energy treatment. During the 7-10 day interval, histological changes in the dermis may affect sensitivity during follow up treatment cycles with increase to RF energy.
The data shows gradual wrinkle class improvement at 90 days. The scores and change in wrinkle class demonstrate improvement in rhytids and in solar elastosis. The independent blinded evaluations support post treatment efficacy based on the Fitzpatrick Wrinkle Scores. All studied parameters suggest a favorable treatment profile of the VIVACE device to reduce wrinkles in the peri-orbital and peri-oral regions. The device showed moderate, but positive improvement in wrinkle severity and elastosis damage.
#### IX. Conclusion
The intended use of the VIVACE Electrosurgical System is substantially equivalent to the predicate device in intended use, design and RF technology. The clinical study demonstrates that the VIVACE functions as intended with no significant impact on the safety and effectiveness profile of the device for the treatment of facial wrinkles.
Data to support substantial equivalence to the predicate device was generated in a clinical study. The clinical study was designed to demonstrate the Vivace device is safe and effective for the treatment of facial wrinkles. Clinical data from a 31 patient US study using a validated scale, Fitzpatrick Wrinkle Severity Scale, supported substantial equivalence of the VIVACE to the predicate device for the percutaneous treatment of facial wrinkles in peri-oral and peri-orbital regions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.