← Product Code [GZT](/productcode/GZT) · K150378

# NICO BrainPath & Accessories, NICO BrainPath, NICO Shepherd's Hook (K150378)

_Nico Corporation · GZT · Jun 12, 2015 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150378

## Device Facts

- **Applicant:** Nico Corporation
- **Product Code:** [GZT](/productcode/GZT.md)
- **Decision Date:** Jun 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4800
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following: - Primary/Secondary Brain Tumors - Vascular Abnormalities/Malformations - Intraventricular Tumors/Cysts

## Device Story

NICO BrainPath is a self-retaining neurosurgical retractor system providing minimally invasive access to neurological tissues. Device consists of reusable aluminum obturators and single-use cyclic olefin copolymer (COC) sheaths. In the OR, surgeon assembles obturator and sheath; inserts assembly to create atraumatic surgical corridor; removes obturator; leaves sheath in place to provide 13.5 mm corridor for visualization and instrumentation. Accessories include manipulation tool for sheath positioning and shepherd's hooks for attachment to third-party retractors. System supports trajectory planning software, navigation, optics, and resection tools. Device facilitates subcortical access to tumors, vascular malformations, and cysts. Benefits include reduced tissue trauma and improved surgical access. Operated by neurosurgeons.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included cytotoxicity, sensitization, irritation, simulated use with third-party instruments, packaging/shelf-life, cleaning validation, and sterility validation (autoclave, hydrogen peroxide gas plasma, gamma).

## Technological Characteristics

Materials: Aluminum (obturator), COC (sheath). Principle: Mechanical self-retaining retractor. Dimensions: 13.5 mm inner diameter, 15.8 mm outer diameter; lengths 50-75 mm. Connectivity: None (standalone). Sterilization: Gamma (disposable), Autoclave/Hydrogen Peroxide Gas Plasma (reusable).

## Regulatory Identification

A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.

## Predicate Devices

- NICO Brain Port ([K120691](/device/K120691.md))

## Reference Devices

- Synaptive Medical Inc. BrightMatter Planning Software ([K140337](/device/K140337.md))
- Synaptive BrightMatter Navigation System ([K142024](/device/K142024.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol composed of three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2015

NICO Corporation Mr. Sean Spence, RAC Regulatory Affairs Manager 250 E. 96th Street, Suite 125 Indianapolis, Indiana 46240

Re: K150378

Trade/Device Name: NICO® BrainPath® and Accessories Regulation Number: 21 CFR 882.4800 Regulation Name: Self-Retaining Retractor for Neurosurgery Regulatory Class: Class II Product Code: GZT Dated: May 13, 2015 Received: May 14, 2015

Dear Mr. Spence:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S 同心

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150378

Device Name NICO BrainPath and Accessories

#### Indications for Use (Describe)

To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following:

- Primary/Secondary Brain Tumors
- Vascular Abnormalities/Malformations
- Intraventricular Tumors/Cysts

| Type of Use (Select one or both, as applicable)                                                                                                      |  |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:inline-block;"><span style="font-size: 16px; vertical-align:middle;">☑</span></div> Research Use (Part 21 CFR 201.26(a) and (b)) |  |
| <div style="display:inline-block;"><span style="font-size: 16px; vertical-align:middle;">☐</span></div> Over-The-Counter Use (21 CFR 201.66)         |  |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NICO Corporation. The letters "NIC" are in black, and the "O" is in gold. Underneath the letters is the word "CORPORATION" in a smaller font.

# NICO® BrainPath®

21 CFR §807.92 Date Prepared: 03 June 2015

510(k) Number: K150378

| Submitter/Manufacturer: | NICO Corporation<br>250 E. 96th Street, Suite 125<br>Indianapolis, IN 46240 |
|-------------------------|-----------------------------------------------------------------------------|
| Contact Person:         | Sean Spence, RAC<br>Regulatory Affairs Manager<br>Office: 317.660.7118      |
| Trade Name:             | NICO® BrainPath® and Accessories                                            |
| Common/Usual Name:      | Self-retaining retractor for neurosurgery                                   |
| Classification:         | 21 CFR §882.4800                                                            |
| Product Codes:          | GZT                                                                         |
| Predicate Device:       | K120691, NICO Brain Port (previous name)                                    |

## Device Description

The NICO BrainPath and Accessories are designed to provide minimally invasive access to neurological tissues. The design specifically supports the creation of an atraumatic surgical corridor to access most areas of the brain. The BrainPath also facilitates the expanding neurosurgical armamentarium of trajectory planning software (e.g., Synaptive® Medical Inc. BrightMatter® Planning Software, K140337), navigation (e.g., Synaptive BrightMatter Navigation System, K142024), optics, corridor resection (e.g., NICO Myriad), and biopsy. To date, the BrainPath technology has been used to successfully access primary and secondary brain tumors, vascular abnormalities or malformations, and intraventricular tumors and cysts.

The BrainPath consists of multiple-sized reusable and re-sterilizable obturators with coordinating single patient use disposable sheaths. The obturator and sheath are assembled in the operating room immediately prior to use. After placement, the obturator is removed leaving behind the sheath which provides a 13.5 mm surgical corridor to the lesion or abnormality. Table 1 outlines the four currently available BrainPath configurations.

| General Name                | Sheath<br>Length (mm) | Obturator Tip<br>Length (mm) |
|-----------------------------|-----------------------|------------------------------|
| 50 mm Shallow-Tip (ST-Gold) | 50                    | 8                            |
| 50 mm BrainPath             | 50                    | 15                           |
| 60 mm BrainPath             | 60                    | 15                           |
| 75 mm BrainPath             | 75                    | 15                           |

## Table 1: BrainPath Configurations

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for NICO Corporation. The letters "NICO" are in a bold, sans-serif font, with the "O" being a golden color. Below the letters is the word "CORPORATION" in a smaller, sans-serif font.

The BrainPath Accessories include a "manipulation tool," which is similar to a dental probe and is used for manipulating the position of the sheath after it has been placed. The accessories also include "shepherd's hooks" for attaching to various commercially available retractors, and a sterilization tray for the reusable components (i.e., obturators and manipulation tools).

| BrainPath Component          | Sterile or Non-Sterile | Single-Use or Reusable |
|------------------------------|------------------------|------------------------|
| Sheath (all sizes)           | Sterile                | Single-Use             |
| Obturators (all sizes)       | Non-Sterile            | Reusable               |
| Manipulation Tool            | Non-Sterile            | Reusable               |
| Sterilization Tray           | Non-Sterile            | Reusable               |
| Shepherd's Hooks (all types) | Sterile                | Single-Use             |

#### Table 2: NICO BrainPath Components Supplied

## Intended Use

To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following:

- Primary/Secondary brain tumors -
- -Vascular abnormalities/malformations
- -Intraventricular tumors/cysts

#### Comparison to Predicate

The subject BrainPath iteration is nearly identical to the predicate NICO Brain Port (K120691). Following the previous clearance the Brain Port name was changed to BrainPath®. Modifications to the design include creation of the ST-Gold shallow tip 50 mm obturator that provides access to shallow abnormalities by having a 7.5 mm obturator tip instead of 15 mm. The proximal end of the sheath was modified to include slots which enable the management of surgical patties, which are commonly used during neurosurgical procedures. Additionally, tabs were added to the proximal end of the sheath to facilitate retention of Shepherd's hooks, which may be used to help position the sheath during surgery. Finally, the surface finish on the inner diameter and proximal head of the sheath was textured to reduce optical glare during surgery. Table 3 provides additional details on how the subject and predicate devices compare.

|                                                     | Subject Device<br>NICO BrainPath                                                                                                                                                               | Predicate Device<br>NICO Brain Port                                                                                              |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                         | BrainPath                                                                                                                                                                                      | Brain Port                                                                                                                       |
| Intended Use                                        | To provide for access and allow for visualization of the surgical field during brain and spinal surgery.                                                                                       | Identical                                                                                                                        |
| Indications for Use                                 | Indications may include subcortical access to diseases such as the following:<br>- Primary/Secondary brain tumors<br>- Vascular abnormalities/malformations<br>- Intraventricular tumors/cysts | None specifically identified                                                                                                     |
| Principle of Operation                              | Consists of an “obturator-like” component and a “sheath-like” component which are assembled, inserted, and disassembled to provide corridor access                                             | Identical                                                                                                                        |
|                                                     | Subject Device                                                                                                                                                                                 | Predicate Device                                                                                                                 |
|                                                     | NICO BrainPath                                                                                                                                                                                 | NICO Brain Port                                                                                                                  |
| Operation and<br>placement                          | Handheld and can be assisted by third-party navigation<br>(if desired)                                                                                                                         | Identical                                                                                                                        |
| Shipping<br>Configuration                           | Obturator and sheath packaged and shipped separately<br>and paired during surgical case.                                                                                                       | Identical                                                                                                                        |
| Reusable or Single<br>Patient Use                   | Single Patient Use and Reusable                                                                                                                                                                | Identical                                                                                                                        |
| Method of<br>Sterilization                          | Gamma for disposable component (sheath)<br><br>Autoclave/hydrogen peroxide gas plasma for reusable<br>components                                                                               | Identical                                                                                                                        |
| Biocompatibility                                    | Demonstrated based on externally communicating<br>device in direct contact with tissue/bone/dentin for a<br>limited duration                                                                   | Identical                                                                                                                        |
| Materials of<br>Manufacture                         | Obturator: Aluminum<br>Sheath: COC                                                                                                                                                             | Identical                                                                                                                        |
| Cross Sectional<br>analysis of<br>Obturator/ Sheath | Obturator/Sheath combination has a circular cross<br>section.                                                                                                                                  | Identical                                                                                                                        |
| Depth markings                                      | Incremental depth markings on both sheath and<br>obturator.                                                                                                                                    | Identical                                                                                                                        |
| Sheath diameter<br>dimensions                       | Available sheath diameter is 13.5 mm (inner diameter)<br>and 15.8 mm (outer diameter).                                                                                                         | Identical                                                                                                                        |
| Sheath lengths                                      | Available sheath lengths are 50 mm, 60 mm, and 75<br>mm.                                                                                                                                       | Identical                                                                                                                        |
| General Shape of<br>Obturator Tip                   | Distal end of obturator has a conical shape with a<br>rounded tip and no opening.                                                                                                              | Identical                                                                                                                        |
| Obturator Tip                                       | For the blue, standard tip obturators, the tip extends 15<br>mm beyond the sheath. For the Gold-ST, shallow-tip<br>obturator, the tip extends 7.5 mm beyond the sheath.                        | The three blue standard tip<br>obturators are identical. The<br>Gold-ST was not part of the<br>original Brain Port submission.   |
| 3rd party<br>instrumentation                        | Obturator component interfaces with third party<br>instruments.                                                                                                                                | Identical                                                                                                                        |
| Surface of Sheath                                   | Used Ink Y. Inner diameter and horizontal proximal<br>surface of the knurled ring are textured.                                                                                                | Used Ink X. Sheath was "clear"<br>and smooth. Surface texturing<br>was not part of the previous<br>submission.                   |
| Proximal End of<br>Sheath                           | Knurled ring and distal portion of the sheath has holes,<br>slots, and ears.                                                                                                                   | Knurled ring on sheath includes<br>holes for securing. Slots and ears<br>were not part of the original<br>Brain Port submission. |
| Manipulation Tool                                   | "Manipulation Tool" is textured to prevent glare/shine<br>from surgical lighting.                                                                                                              | "Manipulation Tool" was<br>offered but was not textured on<br>the distal portion.                                                |
| Shepherd's Hooks                                    | Three Shepherd's Hooks available to facilitate<br>interfacing with common third party retractors.                                                                                              | Shepherd's Hooks were not<br>mentioned in the previous<br>submission                                                             |

#### Table 3: Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for NICO Corporation. The logo is in black and gold. The word "NICO" is in large, bold letters, with the "O" being gold. Below the word "NICO" is the word "CORPORATION" in smaller, thinner letters.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for NICO Corporation. The letters "NICO" are in a bold, sans-serif font, with the "O" being a golden color. Below the letters is the word "CORPORATION" in a smaller, sans-serif font.

## Non-Clinical Testing

The following tests were confirmed or repeated to demonstrate that the subject device modifications met applicable design and performance requirements:

#### Table 4: Non-Clinical Testing

| Testing                                                                                                                                                                                                                                             | Device(s)                                                   | Result/Conclusion |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------|
| Cytotoxicity - MEM Elution: 72 hour incubation                                                                                                                                                                                                      | BrainPath Sheath,<br>Obturator, and<br>Manipulation Tool    | Non-cytotoxic     |
| Sensitization - Maximization (2 extracts)                                                                                                                                                                                                           | BrainPath Sheath and<br>Obturator                           | Non-sensitizer    |
| Irritation - Intracutaneous Reactivity (2 extracts)                                                                                                                                                                                                 |                                                             | Non-irritant      |
| Simulated Use to demonstrate the BrainPath has the ability to<br>interface with 3rd Party Instruments and meets design input<br>requirements                                                                                                        | BrainPath Sheath,<br>Obturator, and<br>Manipulation Tool    | Pass              |
| Packaging & Shelf Life - shipping/distribution simulation,<br>environmental conditions, aging, visual packaging inspection,<br>bubble and burst packaging testing, and functional testing<br>following aging, environmental and shipping simulation | BrainPath Sheath and<br>Obturator                           | Pass              |
| Specification Review                                                                                                                                                                                                                                | BrainPath Sheath,<br>Obturator, and<br>Manipulation Tool    | Pass              |
| Cleaning Validation (Reusable Device) – Establishment of<br>cleaning validation per miles soil test using bioburden<br>endotoxin and protein testing                                                                                                | BrainPath Obturator                                         | Pass              |
| Sterility Validation (Reusable Device) - Steam autoclaving,<br>IUSS, and hydrogen peroxide gas plasma                                                                                                                                               | Obturators, Manipulation<br>Tool, and Sterilization<br>Tray | Pass              |
| Sterility Validation (Single-Use) - B&F testing, VDmax for<br>SAL 10-6, along with routine Pyrogenicity testing                                                                                                                                     | BrainPath Sheath and<br>Shepherd's Hooks                    | Pass              |
| Sterilization Tray Drop Test from 4 ft.                                                                                                                                                                                                             | Obturator and<br>Manipulation Tool                          | Pass              |
| Cytotoxicity - MEM Elution: 72 hour incubation                                                                                                                                                                                                      | Shepherd's Hooks                                            | Non-cytotoxic     |
| Specification Review                                                                                                                                                                                                                                | Shepherd's Hooks                                            | Pass              |

## Conclusion

Risk assessments, biocompatibility, non-clinical testing, design validation, and compliance with recognized standards have demonstrated that the subject NICO BrainPath does not raise different questions of safety or effectiveness when compared to the predicate. Therefore, the results of these tests provide reasonable assurance that the NICO BrainPath has a similar safety and effectiveness profile as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K150378](https://fda-staging.innolitics.com/device/K150378)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).