The reprocessed EP diagnostic catheters are intended for temporary use during electrophysiology studies for intracardiac sensing, recording, and pacing for the electrophysiological mapping and evaluation of cardiac structures and arrhythmias. The catheter is designed for use in the coronary sinus.
Device Story
Steerable, multi-electrode diagnostic catheter; used for intracardiac sensing, recording, and pacing during electrophysiology studies. Features high-torque shaft; braided uni-directional deflectable tip; ten platinum electrodes (including 2 mm tip dome). Operated by physician via proximal hand piece with thumb-operated sliding piston to control tip deflection and rotation. Facilitates accurate positioning within coronary sinus for cardiac mapping. Reprocessed device; Sterilmed performs cleaning, sterilization, and functional testing. Maximum four reprocessing cycles; tracked via marking system. Output is electrical signal data from heart tissue, interpreted by external electrophysiology recording system to guide clinical decision-making regarding cardiac arrhythmias.
Clinical Evidence
Bench testing only. Includes cleaning validation, sterilization validation (ISO 11135, USP <71>), biocompatibility (ISO 10993-1), EO residual testing (ISO 10993-7), packaging validation (ASTM D 4169, F 88, F 2096), and shelf life validation (ASTM F 1980). Functional performance validated via simulated use, visual inspection, and fatigue testing.
Technological Characteristics
6 French, 115 cm length; high-torque shaft; braided uni-directional deflectable tip; 10 platinum electrodes. Sterilization: Ethylene Oxide. Reprocessed device; max 4 cycles. No changes from predicate design or materials.
Indications for Use
Indicated for patients undergoing electrophysiology studies requiring intracardiac sensing, recording, and pacing for mapping and evaluation of cardiac structures and arrhythmias; designed for use in the coronary sinus.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Predicate Devices
Webster® Coronary Sinus Uni-Directional Catheters with Auto ID Technology (K090898)
Webster® Coronary Sinus Uni-Directional Catheters with Auto ID Technology (K101345)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 27, 2015
Sterilmed. Inc. Patricia F. Kaufman Regulatory Affairs Specialist 5010 Cheshire Parkway Suite 2 Plymouth, Minnesota 55446
Re: K150357
Trade/Device Name: Reprocessed Electrophysiology Diagnostic Catheters (See Attached List of Models) Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: July 17, 2015 Received: July 20, 2015
Dear Patricia F. Kaufman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stei
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Page 3 - Patricia F. Kaufman
| Original<br>Manufacturer | Model # | Description |
|--------------------------|---------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Biosense Webster | D135303 | Webster® Coronary Sinus Uni-Directional Catheters with<br>Auto ID Technology, 6F, CS-D Curve, 115 cm useable<br>length, 2 mm tip electrode |
| Biosense Webster | D135304 | Webster® Coronary Sinus Uni-Directional Catheters with<br>Auto ID Technology, 6F, CS-F Curve, 115 cm useable<br>length, 2 mm tip electrode |
## K150357 List of Reprocessed Models Found SE:
{3}------------------------------------------------
## Indications for Use
510(k) Number: K150357
Device Name: Reprocessed Electrophysiology Diagnostic Catheters
Indications for Use: The reprocessed EP diagnostic catheters are intended for temporary use during electrophysiology studies for intracardiac sensing, recording, and pacing for the electrophysiological mapping and evaluation of cardiac structures and arrhythmias. The catheter is designed for use in the coronary sinus.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Page 1 of
{4}------------------------------------------------
## Section II. SUMMARY AND CERTIFICATION
| A. | 510(k) Summary | K150357 |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| | Submitter: | Sterilmed, Inc. |
| | Contact Person: | Patricia Kaufman<br>5010 Cheshire Pkwy N, Suite 2<br>Plymouth, MN 55446<br>Phone: 763.488.3211<br>Fax: 763.488.3350 |
| | Date Prepared: | 11 February 2015 |
| | Trade Name: | Reprocessed Electrophysiology Diagnostic Catheters |
| | Classification Name: | Electrode Recording Catheter or Electrode Recording Probe |
| | Classification Number: | Class II, 21 CFR 870.1220 |
| | Product Code: | NLH |
| Predicate<br>Devices: | The reprocessed EP diagnostic catheters are substantially equivalent to the Biosense Webster<br>Webster® Coronary Sinus Uni-Directional Catheters with Auto ID Technology (K090898 and<br>K101345). | |
| Device<br>Description: | The EP diagnostic catheters are steerable, multi-electrode catheters with a deflectable tip designed to<br>facilitate electrophysiological mapping of the heart. The catheters have a high-torque shaft with a<br>braided uni-directional deflectable tip section containing an array of ten platinum electrodes that<br>includes a 2 mm tip dome, which can be used for stimulation and recording. The catheters are 6<br>French with a usable length of 115 cm. The braided tip is controlled by a proximal hand piece that<br>features a thumb operated sliding piston, and is offered in various curve types. The high-torque shaft<br>allows the plane of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at<br>the desired site. | |
| Indications for<br>Use: | The reprocessed EP diagnostic catheters are intended for temporary use during electrophysiology<br>studies for intracardiac sensing, and pacing for the electrophysiological mapping and<br>evaluation of cardiac structures and arrhythmias. The catheter is designed for use in the coronary<br>sinus. | |
| Technological<br>Characteristics: | The reprocessed EP diagnostic catheters are identical to the predicate devices in design, materials of<br>construction, and intended use. There are no changes to the clinical applications, patient population,<br>performance specifications, or method of operation. | |
| Functional and<br>Safety Testing: | Representative samples of reprocessed EP diagnostic catheters were tested to demonstrate appropriate<br>functional characteristics. Process validation testing was performed to validate the cleaning and<br>sterilization procedures as well as device packaging. In addition, the manufacturing process includes<br>visual and validated functional testing of 100% of products reprocessed. The reprocessed EP<br>diagnostic catheters are reprocessed no more than four (4) times. Each device is marked and tracked<br>through each reprocessing cycle. After the device has reached the maximum number of reprocessing<br>cycles, the device is rejected from further reprocessing is performed only by the<br>manufacturer Sterilmed. Sterilmed restricts its reprocessings to exclude devices previously<br>reprocessed by other reprocessors. | |
| Summary of<br>Non-clinical<br>Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization validation (ISO<br>11135, USP <71>), biocompatibility testing (ISO 10993-1), ethylene oxide residual testing (ISO<br>10993-7), packaging validation (ASTM D 4169, ASTM F 88, ASTM F 2096), and shelf life<br>validation (ASTM F 1980). In addition, validation of functional performance (bench testing) was<br>performed through simulated use, visual inspection, fatigue testing, and function testing. Performance<br>testing shows the reprocessed EP diagnostic catheters to perform as originally intended. | |
| Conclusion: | Sterilmed concludes that the reprocessed EP diagnostic catheters are safe, effective, and substantially<br>equivalent to the predicate devices, Webster® Coronary Sinus Uni-Directional Catheters with Auto<br>ID Technology (K090898 and K101345) , as described in this premarket notification submission. | |
·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.