Healgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card

K150356 · Healgen Scientific,, LLC · JXM · May 13, 2015 · Clinical Toxicology

Device Facts

Record IDK150356
Device NameHealgen Multi-Drug Urine Test Cup, Healgen Multi-Drug Urine Test Dip Card
ApplicantHealgen Scientific,, LLC
Product CodeJXM · Clinical Toxicology
Decision DateMay 13, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3170
Device ClassClass 2

Indications for Use

The Healgen multi-test devices are intended for the qualitative detection of drugs of abuse in human urine. The devices are intended for professional use only.

Device Story

Multi-test device for qualitative detection of up to 7 drugs of abuse in human urine; combines previously cleared single-drug test strips, cassettes, dip cards, and cups into a single platform; intended for professional use; provides visual qualitative results; assists clinicians in identifying drug presence; utilizes lateral flow immunoassay technology; no change to fundamental scientific technology from predicate devices.

Clinical Evidence

Bench testing only. Verification studies included interference studies and a lay-user study to support the multi-drug format modification. No clinical trial data provided.

Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding format. Multi-drug cup or dip card configuration containing multiple test strips. Qualitative visual readout. No electronic components, software, or energy source required.

Indications for Use

Indicated for qualitative, simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, and Oxycodone in human urine. Intended for both prescription and over-the-counter (OTC) use. Provides preliminary results; requires confirmatory testing (GC/MS or LC/MS).

Regulatory Classification

Identification

A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Decision Summary To: THE FILE RE: DOCUMENT NUMBER K150356 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) K142280, Healgen Oxazepam Test (Strip, Cassette, Dip Card, Cup), Healgen Morphine Test (Strip, Cassette, Dip Card, Cup); K143187, Healgen Amphetamine Test (Strip, Cassette, Dip Card, Cup), Healgen Oxycodone Test (Strip, Cassette, Dip Card, Cup); K141647, Healgen COC One Step Cocaine Test (Strip, Cassette, Dip Card, Cup); K140546, Healgen THC One Step Marijuana Test, Healgen mAMP One Step Methamphetamine Test 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: The combination of the previously cleared drugs of abuse test devices (cleared in k142280, k143187, k141647, and k140546) into the candidate multi-test devices that can detect a combination of up to 7 drugs of abuse at the same time. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and cutoffs. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...