← Product Code [KNT](/productcode/KNT) · K150350

# Flexi-Seal Signal Fecal Management System (K150350)

_Convatec, Ltd. · KNT · Jul 21, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150350

## Device Facts

- **Applicant:** Convatec, Ltd.
- **Product Code:** [KNT](/productcode/KNT.md)
- **Decision Date:** Jul 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.

## Device Story

Indwelling fecal management system; manages fecal incontinence in non-pediatric patients. Components: soft silicone catheter tube, retention balloon, collection bag, syringe, cinch clamp. Clinician inserts catheter into rectum; inflates low-pressure retention balloon with water/saline via dedicated port; balloon provides visual/tactile signal of optimal inflation. Catheter features odor-adsorbing material to reduce fecal odor. Sampling port allows stool collection; irrigation port allows medication administration. Device collects liquid/semi-liquid stool into external bag. Used in clinical settings by healthcare providers. Benefits: contains waste, reduces odor, protects skin, facilitates medication delivery.

## Clinical Evidence

Bench testing only. No clinical data presented. Functional and structural parameters (material strength, catheter size, balloon size) were compared to the predicate device. Biocompatibility evaluated per ISO-10993.

## Technological Characteristics

Collapsible silicone rubber catheter with odor-adsorbing additive. Features: low-pressure silicone retention balloon, Luer-lock inflation/irrigation ports, sampling port, 1L collection bag with gas filter. Non-sterile, single-use. Dimensions: 1,550mm-1,670mm length. No animal products, natural rubber latex, or DEHP.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Flexi-Seal® SIGNAL™ Fecal Management System ([K112342](/device/K112342.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 21, 2015

ConvaTec, Ltd. Clare Williamson Regulatory Affairs Documentation Compliance Manager Unit 20, First Avenue, Deeside Industrial Park Deeside, Flintshire CH5 2NU UK

Re: K150350 Trade/Device Name: Flexi-Seal® Signal™ Fecal Management System Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: June 18, 2015 Received: June 24, 2015

Dear Clare Williamson,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Form Approved: OMB No. 0910-0120  |  |
|-----------------------------------|--|
| Expiration Date: January 31, 2017 |  |
| See PRA Statement below.          |  |

DEPARTMENT OF HEALTH AND HUMAN SERVICES
  

Food and Drug Administration

**Indications for Use**

| 510(k) Number (if known)       | K150350                                                                                                                                     |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | Flexi-Seal® Signal™ Fecal Management System                                                                                                 |
| Indications for Use (Describe) | For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. |

| Type of Use (Select one or both, as applicable)                                                     |                                                                                     |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.
  

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
  

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  

Food and Drug Administration
  

Office of Chief Information Officer
  

Paperwork Reduction Act (PRA) Staff
  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

| FORM FDA 3881 (1/14) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for ConvaTec. The logo consists of the word "ConvaTec" in blue, followed by a blue circle with a stylized image of a building with three columns inside. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.

Unit 20, First Avenue Deeside Industrial Park Deeside Flintshire CH5 2NU Telephone: 01244 584000 Facsimile: 01244 584011

## 510(k) Summary

Subject Device: Flexi-Seal® SIGNAL™ Fecal Management System

Date Prepared: February 06, 2015

Applicant: ConvaTec Ltd.

First Avenue Deeside Industrial Park Deeside Flintshire CH5 2NU UK

Contact: Clare Williamson

> Regulatory Affairs Documentation Compliance Manager Tel: +44 (0)1244 584155 Fax: +44 (0)1244 584311 Email: clare.williamson@convatec.com

Device Trade Name: Flexi-Seal® SIGNAL™ Fecal Management System

Common Name: Fecal Management System

Classification Name: Gastrointestinal Tube and Accessories (21 CFR §876.5980, Product Code: KNT)

Device Class: Class II

#### Predicate Device:

Device Trade Name: Flexi-Seal® SIGNAL™ Fecal Management System -ConvaTec Inc.

Common Name: Fecal Management System

Classification Name: Gastrointestinal Tube and Accessories (21 CFR §876.5980, Product Code: KNT)

Device Class: Class II

510(k) Substantial Equivalence: K112342 - determined substantially equivalent on April 26, 2012

Device Description:

{4}------------------------------------------------

The Flexi-Seal® SIGNAL™ Fecal Management System is comprised of a soft catheter tube assembly, a Luer-lock syringe, a collection bag with cap and a cinch clamp to pinch off flow in the catheter when required for medication retention or to stop waste flow. The components are contained in a rigid thermoformed plastic clamshell.

The catheter main drain tube is fabricated from collapsible silicone rubber with the addition of an odor adsorber, which is added in order to help contain the fecal odor which often permeates through the wall of the catheter.

The drain tube has a low-pressure silicone retention balloon at the distal end and a connector for attaching a collection bag (provided with the device and also available separately) at the proximal end. There is a recess (pocket) under the balloon for the clinician's finger, which allows the device to be positioned digitally.

Two ports are attached to the side of the catheter. One port is used to inflate the retention balloon with water or saline after the device has been inserted into the patient's rectum. This port also provides a visual and tactile signal of when the low pressure retention balloon is filled to its optimal volume. The other port is used to flush the device if needed and administer medication if prescribed.

An additional sampling port is located on the side of the catheter, which allows access to clinicians for stool sample collection.

A syringe (provided with the device) is used to fill and evacuate the retention balloon for insertion and removal.

The device, collection bag and syringe are intended for single us, are provided non-sterile and contain no components made with animal products, natural rubber latex or DEHP.

This Traditional 510(k) concerns modifications which do not alter the indications for use of the predicate device, the Flexi-Seal® SIGNAL™ Fecal Management System (ref. K112342).

### Intended use:

The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.

## Summary of Technological Characteristics:

The Flexi-Seal® SIGNAL™ Fecal Management System has the same intended use and indications for use as the predicate device, however the tubing now incorporates odor adsorbing properties. The adsorber reduces the odor which often permeates through the wall of the catheter. These unpleasant odors during catheter use can cause major discomfort to the patients, caregivers and visitors. Therefore, the reduction of odor is advantageous to overall patient wellbeing.

The device specifications are essentially identical to those of the predicate device, although minor adjustments have been made to the catheter tubing to

{5}------------------------------------------------

ensure that the material softness remains. Catheter tube diameter, catheter tube hardness, catheter tube tensile strength, and balloon-catheter bonding were evaluated via bench testing and are substantially equivalent to the predicate device. There are no differences in operation between the Flexi-Seal® SIGNAL ™ Fecal Management System and the predicate device.

## Summary of Performance (Non-Clinical Testing) Data:

Non-clinical testing of the subject device for functional and structural parameters has been performed. In this testing, the device's performance has been found to be substantially equivalent to the aforementioned predicate device both functionally and structurally (material strength, catheter size, balloon size, etc.). The device has also been evaluated for biocompatibility in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', issued May 1, 1995, and has been found safe in such respect for its intended use.

In conclusion, the subject device has been demonstrated as safe and effective and substantially equivalent to the predicate device.

{6}------------------------------------------------

### Summary of Technological Characteristics

The following table summarises the technological characteristics of the subject and predicate devices (ref. K112342, determined substantially equivalent on April 26, 2012) and outlines the product characteristics and specifications which form the basis of the substantial equivalence discussion.

The intended use, technological characteristics and principles of operation of the Flexi-Seal® SIGNAL™ Fecal Management System remains the same as those of the predicate device.

|                                            | Subject Device                                                                                                                           | Predicate                                                                                                                                | Comparison                |             |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------|
| Parameter                                  | Flexi-Seal® SIGNAL™ Fecal Management<br>System                                                                                           | Flexi-Seal® SIGNAL™ Fecal Management<br>System<br>(K112342)                                                                              | Similarities              | Differences |
| FDA Product<br>Code                        | KNT                                                                                                                                      | KNT                                                                                                                                      | Both devices are the same | None        |
| FDA<br>Classification<br>Regulation        | 21 CFR 876.5980                                                                                                                          | 21 CFR 876.5980                                                                                                                          | Both devices are the same | None        |
| Regulatory<br>Class                        | Class II                                                                                                                                 | Class II                                                                                                                                 | Both devices are the same | None        |
| Intended Use                               | To manage fecal incontinence through the<br>collection of liquid to semi-liquid stool and to<br>provide access to administer medications | To manage fecal incontinence through the<br>collection of liquid to semi-liquid stool and to<br>provide access to administer medications | Both devices are the same | None        |
| Functional<br>configuration                | Collapsible catheter with distal end secured in<br>the rectum and proximal end connected to a<br>collection bag                          | Collapsible catheter with distal end secured in<br>the rectum and proximal end connected to a<br>collection bag                          | Both devices are the same | None        |
| Retention<br>feature                       | Soft annular balloon                                                                                                                     | Soft annular balloon                                                                                                                     | Both devices are the same | None        |
| Balloon material                           | Silicone rubber                                                                                                                          | Silicone rubber                                                                                                                          | Both devices are the same | None        |
| Inflation<br>management                    | Sealed water filled balloon                                                                                                              | Sealed water filled balloon                                                                                                              | Both devices are the same | None        |
| Retention<br>balloon inflation<br>material | Water/saline (45ml)                                                                                                                      | Water/saline (45ml)                                                                                                                      | Both devices are the same | None        |
| Catheter tube<br>length                    | 1,550mm - 1,670mm                                                                                                                        | 1,550mm - 1,670mm                                                                                                                        | Both devices are the same | None        |
| Length of<br>sphincter section             | >10cm                                                                                                                                    | >10cm                                                                                                                                    | Both devices are the same | None        |

{7}------------------------------------------------

| Parameter                                       | Subject Device                                 | Predicate                                                   | Comparison                |             |
|-------------------------------------------------|------------------------------------------------|-------------------------------------------------------------|---------------------------|-------------|
|                                                 | Flexi-Seal® SIGNAL™ Fecal Management<br>System | Flexi-Seal® SIGNAL™ Fecal Management<br>System<br>(K112342) | Similarities              | Differences |
| Drain channel                                   | Large main drain tube                          | Large main drain tube                                       | Both devices are the same | None        |
| Inflation and<br>irrigation port<br>connections | Luer                                           | Luer                                                        | Both devices are the same | None        |
| Inflation lumen<br>diameter                     | Two lumens at 1.45 mm                          | Two lumens at 1.45 mm                                       | Both devices are the same | None        |
| Inflation lines                                 | 2                                              | 2                                                           | Both devices are the same | None        |
| Irrigation lines                                | 1                                              | 1                                                           | Both devices are the same | None        |
| Sampling port                                   | Closable snap seal for catheter syringe access | Closable snap seal for catheter syringe access              | Both devices are the same | None        |
| Collection bag<br>configuration                 | Disposable with gas filter                     | Disposable with gas filter                                  | Both devices are the same | None        |
| Collection bag<br>size                          | 1 litre disposable                             | 1 litre disposable                                          | Both devices are the same | None        |
| Method of bed<br>connection                     | Hanging strap on bag connector                 | Hanging strap on bag connector                              | Both devices are the same | None        |
| Accessory<br>components                         | Syringe and cinch clamp                        | Syringe and cinch clamp                                     | Both devices are the same | None        |
| Flow stop<br>mechanism                          | External cinch clamp                           | External cinch clamp                                        | Both devices are the same | None        |
| Sterility status<br>(how provided)              | Non-sterile                                    | Non-sterile                                                 | Both devices are the same | None        |
| Packaging                                       | Thermoformed plastic clamshell                 | Thermoformed plastic clamshell                              | Both devices are the same | None        |

---

**Source:** [https://fda-staging.innolitics.com/device/K150350](https://fda-staging.innolitics.com/device/K150350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).