elance Vital Signs Monitor, Central Monitor Software
Applicant
Spacelabs Healthcare
Product Code
MHX · Cardiovascular
Decision Date
Nov 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Spacelabs élance Vital Signs Monitor and élance Central Station is indicated for use in patient populations for: - Adult - Pediatric (1 year old and above) The Spacelabs élance Vital Signs Monitor and élance Central Station facilitates the monitoring of: - ECG with arrhythmia detection - Respiration - Non-invasive blood pressures - Invasive blood pressures - Body temperature - Functional arterial oxygen saturation, and - End tidal CO2. The Spacelabs élance Vital Signs Monitor and élance Central Station is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Device Story
Portable patient monitor family (10.2" or 12.1" displays) and central station software; monitors physiological signals including ECG, respiration, blood pressure (invasive/non-invasive), temperature, SpO2, and EtCO2. Used by clinicians in healthcare facilities. Central station software runs on customer-acquired hardware to aggregate data from monitors. Provides real-time monitoring, arrhythmia detection, and ST-segment analysis. Output allows clinicians to assess patient status, adjust care, and respond to alarms. Benefits include continuous physiological surveillance and timely clinical intervention.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified against standards including ANSI/AAMI EC57 (arrhythmia/ST-segment), ANSI/AAMI SP10 (NIBP), EN 12470-4 (thermometers), ISO 80601-2-55 (gas monitors), and ISO 80601-2-61 (pulse oximeters). Electrical safety and EMC testing performed per IEC 60601 series.
Technological Characteristics
Portable patient monitor with LCD display (10.2"/12.1"). Parameters: ECG, respiration, NIBP, IBP, temperature, SpO2, EtCO2. Connectivity: Central station software for networked monitoring. Software class: Major level of concern. Standards: IEC 60601-1, IEC 60601-1-2, ANSI/AAMI EC57, ANSI/AAMI SP10, EN 12470-4, EN 60601-1-8, EN 62366, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 80601-2-55, ISO 80601-2-61.
Indications for Use
Indicated for adult and pediatric patients (1 year and older) in healthcare facilities for monitoring ECG (with arrhythmia detection), respiration, non-invasive/invasive blood pressure, body temperature, functional arterial oxygen saturation, and end-tidal CO2.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Spacelabs Healthcare élance Vital Signs Monitor and élance Central Station (K093501)
Nellcor Puritan Bennett Incorporated OxiMAX Pulse Oximetry System with N-595 Pulse Oximeter and OxiMAX Sensors and Cables (K012891)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 5, 2015
Spacelabs Healthcare % Thomas Kroenke, Principal Consultant Speed To Market, Inc. PO Box 3018 Nederland, Colorado 80466
Re: K150329
Trade/Device Name: élance Vital Signs Monitor and élance Central Station Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX Dated: October 1, 2015 Received: October 6, 2015
Dear Thomas Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
K150329 510(k) Number (if known):
Device Name: élance Vital Signs Monitor and élance Central Station
Indications For Use: The Spacelabs élance Vital Signs Monitor and élance Central Station is indicated for use in patient populations for:
- Adult ।
- -Pediatric (1 year old and above)
The Spacelabs élance Vital Signs Monitor and élance Central Station facilitates the monitoring of:
- ECG with arrhythmia detection -
- -Respiration
- Non-invasive blood pressures -
- -Invasive blood pressures
- -Body temperature
- Functional arterial oxygen saturation, and -
- -End tidal CO2.
The Spacelabs élance Vital Signs Monitor and élance Central Station is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Page 1 of 1
{3}------------------------------------------------
| Submission Date: | 01 October 2015 | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare<br>35301 SE Center St<br>Snoqualmie, WA 98065 | | |
| Submitter Contact: | Mr. Al Van Houdt<br>Spacelabs Healthcare<br>Phone: +1 (425) 363- 5970<br>Fax: +1 (425) 363-5762<br>Email: al.vanhoudt@spacelabs.com | | |
| Application<br>Correspondent: | Thomas Kroenke<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 | | |
| Manufacturing Site: | Spacelabs Healthcare<br>35301 SE Center St<br>Snoqualmie, WA 98065 | | |
| Trade Name: | élance Vital Signs Monitor and élance Central Station | | |
| Common Name: | Patient Physiological Monitor (with arrhythmia detection or alarms) | | |
| Classification Name: | Patient Physiological Monitor (with arrhythmia detection or alarms) | | |
| Primary<br>Classification<br>Regulation: | 21 CFR §870.1025 | | |
| Primary Product<br>Code: | MHX | | |
| Substantially<br>Equivalent Devices: | New Spacelabs Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
| | Spacelabs Healthcare élance<br>Vital Signs Monitor and<br>élance Central Station | K093501 | Spacelabs Healthcare élance<br>Vital Signs Monitor and élance<br>Central Station |
| | | K033296 | Masimo Corporation Masimo<br>SET® Rad-5 Pulse Oximeter |
| | | K012891 | Nellcor Puritan Bennett<br>Incorporated OxiMAX Pulse<br>Oximetry System with N-595 Pulse<br>Oximeter and OxiMAX Sensors<br>and Cables |
{4}------------------------------------------------
| Device Description: | The Spacelabs Healthcare (Spacelabs) élance Vital Signs Monitor is a<br>family of portable patient monitors intended to be used by clinicians<br>and medical qualified personnel for monitoring ECG with arrhythmia<br>detection, respiration, NIBP, temperature, SPO2, invasive blood<br>pressure and EtCO2. Models within the Spacelabs élance family come<br>in two different sized viewing areas (10.2" and 12.1"), two different<br>housing colors (white and black) and offer selected monitoring<br>features. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Spacelabs élance Central Station software package is available for<br>use with a customer acquired computer based on specifications<br>provided by Spacelabs. This package allows monitoring of the élance<br>Vital Signs Monitor at a central workstation. |
| Intended Use: | The Spacelabs élance Vital Signs Monitor and élance Central Station is<br>indicated for use in patient populations for:<br>- Adult<br>- Pediatric (1 year old and above)<br>The Spacelabs élance Vital Signs Monitor and élance Central Station<br>facilitates the monitoring of:<br>- ECG with arrhythmia detection<br>- Respiration<br>- Non-invasive blood pressures<br>- Invasive blood pressures<br>- Body temperature<br>- Functional arterial oxygen saturation, and<br>- End tidal CO2.<br>The Spacelabs élance Vital Signs Monitor and élance Central Station is<br>a prescription device intended to be used by healthcare professionals in<br>all areas of a healthcare facility. |
{5}------------------------------------------------
### Technology Comparison:
The Spacelabs Spacelabs élance Vital Signs Monitor and élance Central Station employs the same technological characteristics as the predicate device.
| Characteristic | Predicate Device | Proposed Device |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Parameters | ECG with arrhythmia<br>detection<br>Respiration<br>Non-invasive blood pressure<br>Invasive blood pressure<br>Body temperature<br>Functional arterial oxygen<br>saturation, and<br>End tidal CO2. | Same |
| Display Type | LCD | Same |
| Display Sizes | 10.2" and 12.1" | Same |
| VSM Variants | 5, 5i, 5c, 5 elite, 5i elite, 5c<br>elite<br>7, 7i, 7c, 7 elite, 7i elite, 7c<br>elite | Same |
| Additional Options | - | 93300-M, Masimo SpO2<br>93300-N, Nellcor SpO2<br>93300-A: Arrhythmia<br>93300-S: ST Segment Analysis |
#### Summary of Performance Testing:
Software
The élance VSM and élance CS contain MAJOR level of concern software. Software was designed and developed according to a robust software development process, and was rigorously verified and validated. Software information is provided in accordance with
- · FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05;
- · FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99:
- FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; and
- · FDA guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, 14 Jun 13
Test results indicated that the élance VSM and élance CS software comply with predetermined specification.
{6}------------------------------------------------
| Electrical Safety | The élance VSM was tested for patient safety in accordance with:<br>IEC 60601-1:1988, Am1: 1991, Am2: 1995, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. Test results indicated that the élance VSM complies with the applicable standards. |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electromagnetic Compatibility | The élance VSM was tested for EMC in accordance with:<br>IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests. Test results indicated that the élance VSM complies with the applicable standards. |
| Performance Testing – Bench | The élance VSM was tested for parameter performance in accordance with:<br>ANSI/AAMI EC57: 1998/(R) 2003, Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms; ANSI/AAMI SP10: 2002, Am1: 2003, Manual, electronic, or automated sphygmomanometers; EN 12470-4: 2009, Clinical thermometers – Part 4: Performance of electrical thermometers for continuous measurement; EN 60601-1-6: 2010, Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability; EN 60601-1-8 2007, Am1: 2013, AC: 2014, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance – Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems; EN 62366: 2008, Medical devices - Application of usability engineering to medical devices; IEC 60601-2-27: 2005, Medical electrical equipment – Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment; |
{7}------------------------------------------------
- Performance Testing – Bench (continued)
- · IEC 60601-2-30: 1999, Medical electrical equipment Part 2-30: Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment;
- · IEC 60601-2-34: 2000, Medical electrical equipment Part 2-34: Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment;
- · IEC 60601-2-49: 2011, Medical electrical equipment Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment;
- ISO 80601-2-55: 2011, Medical electrical equipment Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors; and
- ISO 80601-2-61: 2011, Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
The élance CS was tested for parameter performance in accordance with:
- EN 60601-1-6: 2010, Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability;
- EN 60601-1-8 2007, Am1: 2013, AC: 2014, Medical electrical equipment – Part 1-8: General requirements for basic safety and essential performance – Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems;
- · EN 62366: 2008, Medical devices Application of usability engineering to medical devices;
- · Test results indicated that the élance VSM and élance CS comply with these Standards. Additionally, the élance VSM and élance CS were tested in accordance with internal requirements and procedures, and test results indicated that the devices comply with the predetermined requirements.
#### Verification and validation activities were conducted to establish the Conclusion performance and safety characteristics of the élance VSM and élance CS. The results of these activities demonstrate that the élance VSM and élance CS are as safe, as effective, and perform as well as or better than the predicate devices. Therefore, the élance VSM and élance CS are considered substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.