← Product Code [HWC](/productcode/HWC) · K150309

# Extremity Medical Calcaneal Osteotomy Device (K150309)

_Extremity Medical, LLC · HWC · Mar 4, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150309

## Device Facts

- **Applicant:** Extremity Medical, LLC
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Mar 4, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Extremity Medical Calcaneal Osteotomy Device is intended for fixation of osteotomies of the calcaneus.

## Device Story

Bone fixation device for calcaneal osteotomies; consists of bracket and lag screws; includes surgical instruments for site preparation and delivery. Bracket available in 6mm to 13mm heights; modified design accommodates 3 lag screws. Used by orthopedic surgeons in clinical/surgical settings to stabilize bone segments post-osteotomy. Provides mechanical fixation to facilitate bone healing.

## Clinical Evidence

No clinical testing was performed. Equivalence established via non-clinical bench testing, including pull-out, torque, and bending strength assessments of worst-case constructs.

## Technological Characteristics

Metallic bone fixation fastener (bracket and lag screws). Materials and mechanical properties are equivalent to the predicate. Dimensions: bracket heights 6mm to 13mm. Design allows for 3-screw fixation.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Calcaneal Osteotomy Device ([K143402](/device/K143402.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.

March 4, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Extremity Medical, LLC Mr. Brian Smekal Vice President, Regulatory Affairs and Quality Assurance 300 Interpace Parkway, Suite 410 Parsippany, New Jersey 07054

Re: K150309

Trade/Device Name: Extremity Medical Calcaneal Osteotomy Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, HRS Dated: February 6, 2015 Received: February 9, 2015

Dear Mr. Brian Smekal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K150309

Device Name

Extremity Medical Calcaneal Osteotomy Device

Indications for Use (Describe)

The Extremity Medical Calcaneal Osteotomy Device is intended for fixation of osteotomies of the calcaneus.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Special 510(k) Summary of Safety and Effectiveness:

## Calcaneal Osteotomy Device

| Submitter:                                  | Extremity Medical, LLC.<br>300 Interpace Parkway<br>Suite 410<br>Parsippany, NJ 07054                                                                                                                                                                                                                                                                                                                              |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                              | Brian Smekal, MS, RAC<br>Director, Regulatory Affairs<br>Phone: (973) 588-8988<br>Email: bsmekal@extremitymedical.com                                                                                                                                                                                                                                                                                              |
| Date Prepared                               | February 6, 2015                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade Name                                  | Calcaneal Osteotomy Device                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Name<br>and Number           | Smooth or threaded metallic bone fixation fastener<br>21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                              |
| Product Code                                | HWC (screw, fixation, bone), HRS (plate, fixation, bone)                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Devices                           | K143402 - Calcaneal Osteotomy Device (Extremity Medical)                                                                                                                                                                                                                                                                                                                                                           |
| Device Description                          | The Calcaneal Osteotomy Device is a bone fixation device consisting of a<br>bracket and lag screws and a set of instruments used for implant site preparation<br>and delivery. The bracket is offered in height sizes of 6mm to 13mm. The<br>modified device allows for placement of the implant with 3 lag screws. The<br>Calcaneal Osteotomy Device is intended for fixation of osteotomies of the<br>calcaneus. |
| Indications for use                         | The Extremity Medical Calcaneal Osteotomy Device is intended for fixation of<br>osteotomies of the calcaneus.                                                                                                                                                                                                                                                                                                      |
| Statement of<br>Technological<br>Comparison | The Calcaneal Osteotomy Device and predicate devices are equivalent in terms<br>of indications for use, design, and material mechanical properties.                                                                                                                                                                                                                                                                |
| Non-clinical Testing                        | Pull-out, torque and bending strength was assessed compared to the predicate<br>device and worse case constructs have been tested to demonstrate equivalence.                                                                                                                                                                                                                                                      |
| Clinical Testing                            | No clinical testing was performed.                                                                                                                                                                                                                                                                                                                                                                                 |
| Conclusion                                  | The Calcaneal Osteotomy Device is substantially equivalent to its predicate<br>device. This conclusion is based upon indications for use, materials, design, test<br>data and principles of operation.                                                                                                                                                                                                             |

---

**Source:** [https://fda-staging.innolitics.com/device/K150309](https://fda-staging.innolitics.com/device/K150309)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).