# PDO Absorbable Punctum Plug (K150288)

_Riverpoint Medical · LZU · Aug 28, 2015 · OP · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150288

## Device Facts

- **Applicant:** Riverpoint Medical
- **Product Code:** [LZU](/productcode/LZU.md)
- **Decision Date:** Aug 28, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** OP
- **Attributes:** Therapeutic

## Indications for Use

The Riverpoint Medical Absorbable Canalicular Plugs are intended to temporarily block tear drainage by the obstruction of the canaliculus in the lacrimal drainage system. The Absorbable Canalicular Plugs are used for the temporary treatment of dry eye syndrome, and the dry eye components of various ocular surface diseases including contact lens intolerance secondary to dry eye. The Absorbable Canalicular Plugs can be used in the treatment of dry eye following ocular surgery or to enhance the efficacy of topical medicines and/or ocular lubricants. The Riverpoint Medical Absorbable Canalicular Plugs may also be used to determine the potential effectiveness of a nonabsorbable punctal occlusion device.

## Device Story

Monofilament synthetic absorbable device made of poly(p-dioxanone) (PDO); cylindrical shape; available in 0.2mm, 0.4mm, and 0.5mm diameters. Device inserted into lacrimal canaliculus by physician using forceps to block tear drainage; provides temporary occlusion for approximately 180 days before absorbing. Used in clinical settings for dry eye treatment or to test efficacy of permanent occlusion. Benefits patient by increasing tear volume on ocular surface, alleviating dry eye symptoms, and improving comfort for contact lens wearers or post-surgical patients.

## Clinical Evidence

Bench testing only. Includes sterilization validation (ISO 11135-1), biocompatibility (ISO 10993-1), stability/packaging testing (ISO 11607-1), and simulated use usability validation (EN 62366). All acceptance criteria met.

## Technological Characteristics

Material: Poly(p-dioxanone) (PDO). Form: Monofilament cylindrical plug. Diameters: 0.2mm, 0.4mm, 0.5mm. Length: 2.0mm. Sterilization: Ethylene oxide (EtO). Biocompatibility: ISO 10993 compliant. Single-use.

## Predicate Devices

- Comfortear® Lacrisolve™ Absorbable Punctum Plug -Paragon BioTeck, Inc. ([K140711](/device/K140711.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing right, with flowing lines extending from their heads, resembling a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 28, 2015

Riverpoint Medical Mr. Edwin Anderson Director of Quality and Regulatory 825 NE 25th Avenue Portland, OR 97232

Re: K150288

Trade/Device Name: PDO Absorbable Canalicular Plug Regulation Number: 21 CFR Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: LZU Dated: July 17, 2015 Received: July 20, 2015

Dear Mr. Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Kesia Y. Alexander -A

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150288/S002

Device Name PDO Absorbable Canalicular Plug

#### Indications for Use (Describe)

The Riverpoint Medical Absorbable Canalicular Plugs are intended to temporarily block tear drainage by the obstruction of the canaliculus in the lacrimal drainage system. The Absorbable Canalicular Plugs are used for the temporary treatment of dry eye syndrome, and the dry eye components of various ocular surface diseases including contact lens intolerance secondary to dry eye. The Absorbable Canalicular Plugs can be used in the treatment of dry eye following ocular surgery or to enhance the efficacy of topical medicines and/or ocular lubricants. The Riverpoint Medical Absorbable Canalicular Plugs may also be used to determine the potential effectiveness of a nonabsorbable punctal occlusion device.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Riverpoint Medical. The logo consists of the letters "RP" inside of a rounded green square. To the right of the square is the word "RIVERPOINT" in bold, black letters, with the word "MEDICAL" in smaller letters underneath.

## 510(k) SUMMARY

## Riverpoint Medical 's PDO Absorbable Canalicular Plug

#### Submitter Information

| Submitter's Name:    | Riverpoint Medical                     |
|----------------------|----------------------------------------|
| Address:             | 825 NE 25th Ave.<br>Portland, OR 97232 |
| Phone Number:        | (503) 517-8001 or 866 445-4923         |
| Fax Number:          | (503) 517-8002                         |
| Registration Number: | 3006981798                             |
| Contact Person:      | Edwin Anderson<br>(503) 517-8001       |
| Date of Preparation: | February 3, 2015                       |

#### Device Name

| Trade Name:            | PDO Absorbable Canalicular Plug                               |
|------------------------|---------------------------------------------------------------|
| Common or Usual Names: | Punctum/Punctal Plug, Intracanalicular Plug, Canalicular Plug |
| Classification Name:   | Plug, Punctal                                                 |

#### Device Classification

| FDA Class:              | Unclassified                                            |
|-------------------------|---------------------------------------------------------|
| Product Classification: | Unclassified                                            |
| Classification Code:    | LZU                                                     |
| Review Panel            | Ophthalmic                                              |
| Premarket Review        | Office of Device Evaluation                             |
|                         | Division of Ophthalmic and Ear, Nose and Throat Devices |
|                         | Intraocular and Corneal Implants Devices Branch         |

#### Predicate Devices

K140711 - Comfortear® Lacrisolve™ Absorbable Punctum Plug -Paragon BioTeck, Inc.

## Device Description:

The Riverpoint Medical PDO Absorbable Canalicular Plug is a monofilament synthetic absorbable device prepared from poly(p-dioxanone). PDO Absorbable Canalicular Plugs are available undyed or violet. When dyed, only FDA-approved color additives such as D&C Violet No. 2 are used. PDO Absorbable Canalicular Plugs are used to temporarily block tear drainage by the occlusion of the lacrimal canaliculi for approximately 180 days.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Riverpoint Medical. The logo consists of the letters "RP" inside of a green rounded square. To the right of the square is the text "RIVERPOINT" in bold black letters, with the word "MEDICAL" in smaller letters underneath.

## Intended Use / Indications for Use

The Riverpoint Medical Absorbable Canalicular Plugs are intended to temporarily block tear drainage by the obstruction of the canaliculus in the lacrimal drainage system. The Absorbable Canalicular Plugs are used for the temporary treatment of dry eye syndrome, and the dry eye components of various ocular surface diseases including contact lens intolerance secondary to dry eve. The Absorbable Canalicular Plugs can be used in the treatment of dry eve following ocular surgery or to enhance the efficacy of topical medicines and/or ocular lubricants. The Riverpoint Medical Absorbable Canalicular Plugs may also be used to determine the potential effectiveness of a non-absorbable punctal occlusion device.

## Technological Characteristics

The Riverpoint Medical Absorbable Canalicular Plugs are provided sterilized via ethylene oxide, for single use with two plugs per package. Plugs are available in 0.2mm, 0.4mm and 0.5mm diameters.

## Substantial Equivalence

The Riverpoint Medical Absorbable Canalicular Plug is as safe and effective as the Comfortear® Lacrisolve™ Absorbable Punctum Plug. The Riverpoint Medical Absorbable Canalicular Plugs have the same intended use and similar indications for use, principles of operation, and technical characteristics, including implant duration, material, shape, color additives, diameter availability, sterilization method, and insertion or removal method as the predicate device, see the following table. The Riverpoint Medical Absorbable Canalicular Pluqs minor technological differences as compared to the predicate device do not raise new issues of safety or effectiveness. Performance, biocompatibility and sterility testing demonstrated that the Riverpoint Medical product is as safe and effective as the predicate device and additional testing is performed on each lot of product to verify that requirements have been met prior to release. The Riverpoint Medical Absorbable Canalicular Plugs are substantially equivalent to the predicate device.

| Device<br>Technological<br>Characteristic | Riverpoint Medical PDO Absorbable<br>Canalicular Plug                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Paragon Bioteck, Inc.,Comfortear®<br>Lacrisolve™ Absorbable Punctum<br>Plugs (K140711)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                              | The Riverpoint Medical Absorbable<br>Canalicular Plugs are intended to temporarily<br>block tear drainage by the obstruction of the<br>canaliculus in the lacrimal drainage system.<br>The Absorbable Canalicular Plugs are used<br>for the temporary treatment of dry eye<br>syndrome, and the dry eye components of<br>various ocular surface diseases including<br>contact lens intolerance secondary to dryeye. The Absorbable Canalicular Plugs can<br>be used in the treatment of dry eye following<br>ocular surgery or to enhance the efficacy of<br>topical medicines and/or ocular<br>lubricants. The Riverpoint Medical<br>Absorbable Canalicular Plugs may also be<br>used to determine the potential effectiveness<br>of a non-absorbable punctal occlusion | The Comfortear® Lacrisolve™<br>Absorbable Punctum Plugs are intended<br>to temporarily block tear drainage by the<br>occlusion of the canaliculus in order to:<br>• Determine the potential effectiveness<br>of permanent occlusion,<br>• Temporarily treat dry eye syndrome,<br>and the dry eye components of<br>various ocular surface diseases,<br>• Temporarily enhance the efficacy of<br>topical medications or ocular<br>lubricants,<br>• Temporarily treat contact lens<br>intolerance secondary to dry eye,<br>and<br>• Temporarily treat dry eye after ocular<br>surgery |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Riverpoint Medical. The logo consists of the letters "RP" inside of a green rounded square. To the right of the square is the text "RIVERPOINT" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font below it.

| Device<br>Technological<br>Characteristic | Riverpoint Medical PDO Absorbable<br>Canalicular Plug | Paragon Bioteck, Inc., Comfortear®<br>Lacrisolve™ Absorbable Punctum<br>Plugs (K140711) |
|-------------------------------------------|-------------------------------------------------------|-----------------------------------------------------------------------------------------|
|                                           | device.                                               |                                                                                         |
| Intracanalicular<br>Punctum Plug          | Yes                                                   | Yes                                                                                     |
| Intended<br>Duration                      | Approximately 180 Days                                | Approximately 180 Days                                                                  |
| Material                                  | Polydioxanone (PDO)                                   | Polydioxanone (PDO)                                                                     |
| Color Additive                            | D&C Violet No. 2                                      | D&C Violet No. 2                                                                        |
| Shape                                     | Cylindrical                                           | Cylindrical                                                                             |
| Diameter<br>Availability                  | 0.2mm-0.5mm                                           | 0.2mm-0.5mm                                                                             |
| Length                                    | 2.0mm                                                 | 1.75mm                                                                                  |
| Packaging                                 | Foil Pouch                                            | Tyvek/Poly Pouch                                                                        |
| Biocompatibility                          | Biocompatible per ISO10993                            | Biocompatible per ISO10993                                                              |
| Sterilization                             | EtO Sterilization                                     | EtO Sterilization                                                                       |
| Insertion<br>Methods                      | Manual with Forceps                                   | Manual with Forceps                                                                     |
| Removal<br>Methods                        | Irrigation, Probe, or Dissolution                     | Irrigation, Probe, or Dissolution                                                       |

## Performance Data

Non-clinical performance testing for the Riverpoint Medical Absorbable Canalicular Plugs included sterilization validation per ISO11135-1:2007 - Sterilization of Health Care Products Ethylene Oxide Part 1: Requirements for the Development, Validation, and Routine Control of a Sterilization Process for medical devices, biocompatibility testing per ISO10993-1:2009 - Biological Evaluation of Medical Devices, stability testing on the product and packaging per ISO 11607-1:2006 - Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems and packaging systems, and a simulated use Usability Validation performed per EN62366: 2008-Medical devices - Application of usability engineering to medical devices. All acceptance criteria were met, and the Riverpoint Medical Absorbable Canalicular Plug performed as intended.

## Conclusion

Based on the information provided within this 510(k) submission, Riverpoint Medical concludes that the proposed Absorbable Canalicular Plugs do not raise significant questions of safety or effectiveness and is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K150288](https://fda-staging.innolitics.com/device/K150288)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).