The Capnostream™35 is a portable capnograph/pulse oximeter, intended to provide professionally trained health care providers with continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath, respiration rate, arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients. The pulse oximeter is intended for use during both no motion conditions and for patients who are well or poorly perfused. The Capnostream™35 also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for age 22 and up. The device is intended for use in hospital-type facilities, during intra-hospital transport, and out-of-hospital Emergency Medical Service applications that include ground and air transport.
Device Story
Capnostream™35 is a portable, 4-inch screen respiratory monitor for hospital, transport, and EMS use. It integrates a microMediCO2 capnography module and a pulse oximetry module. Inputs include expired/inspired CO2, respiration rate, SpO2, and pulse rate. The device calculates and displays Integrated Pulmonary Index (IPI), Apnea per Hour (A/hr), and Oxygen Desaturation Index (ODI). Data is displayed as numeric values, CO2 waveforms, and SpO2 pleth waveforms. Clinicians use these real-time parameters and integrated indices to monitor respiratory status and detect potential adverse events. The device features IP54 ingress protection, shock resistance, and altitude compensation for EMS environments. It provides audible/visual alarms when thresholds are crossed, assisting in clinical decision-making and patient management.
Clinical Evidence
Bench testing only. Performance verified via testing to ISO 80601-2-55 (capnography) and ISO 80601-2-61 (pulse oximetry). Electrical safety, electromagnetic compatibility, and environmental testing performed per IEC 60601 series. Software validation and wireless communication testing conducted per FDA guidance. Hazard analysis performed per ISO 14971:2012.
Technological Characteristics
Portable monitor with 4-inch color display. Sensing: NDIR for CO2, spectrophotometry/plethysmography for SpO2. Features: IP54 ingress protection, 1.25M shockproof, hot-swap battery, altitude compensation up to 15,000 ft. Standards: ISO 80601-2-55, ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-49, IEC 60601-1-12, IEC 60601-1-6.
Indications for Use
Indicated for continuous non-invasive monitoring of EtCO2, respiration rate, SpO2, and pulse rate in neonatal, pediatric, and adult patients. IPI indicated for pediatric and adult patients. A/hr and ODI indicated for patients age 22 and up. Suitable for well or poorly perfused patients and no-motion conditions.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Bedside Respiratory Patient Monitoring System With Respiration Rate Software (K130320)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three faces in profile, stacked on top of each other. The faces are black and are oriented to the right. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 23, 2015
Oridion Medical 1987 Ltd Ms. Dalia Givony Regulatory Affairs Manager 7 Hamarpe Street Jerusalem, Israel 9777407
Re: K150272
Trade/Device Name: Capnostream™ 35 Portable Respiratory Monitor Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK, DQA, MNR Dated: June 11, 2015 Received: June 16, 2015
Dear Ms. Givony:
This letter corrects our substantially equivalent letter of July 17, 2105.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Givony
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Traditional 510(k) Submission for Capnostream®35 Portable Respiratory Monitor
001 Statement Indications for Use
Capnostream™35 Portable Respiratory Monitor (This document is not confidential)
February 1, 2015
510(k) Number (if known)
# Device Name: Capnostream™35 Portable Respiratory Monitor Indications for Use:
Intended Use
The Capnostream™35 is a portable capnograph/pulse oximeter, intended to provide professionally trained health care providers with continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath, respiration rate, arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients. The pulse oximeter is intended for use during both no motion conditions and for patients who are well or poorly perfused.
The Capnostream™35 also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for age 22 and up.
The device is intended for use in hospital-type facilities, during intra-hospital transport, and out-of-hospital Emergency Medical Service applications that include ground and air transport.
Prescription Use _ X AND/OR AND/OR AND/OR (Per 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{3}------------------------------------------------
### 510K Summary
This summary is being submitted in accordance with the requirements of 21 CFR 807.92
#### (This section is not confidential)
#### DATE THIS SUMMARY WAS PREPARED
July 15, 2015
#### SUBMITTER'S NAME AND ESTABLISHMENT ADDRESS:
Oridion Medical 1987 Ltd
7 Hamarpe St. P.O. Box 45025
Jerusalem, Israel 9777407
### ESTABLISHMENT REGISTRATION NUMBER
8044004
# CONTACT PERSON:
Dalia Givony, Regulatory Affairs Manager
Oridion Medical 1987 Ltd
7 Hamarpe St. P.O. Box 45025
Jerusalem, Israel 9777407
Telephone: +972-2-589 9235
FAX: +972-2-586-6680
{4}------------------------------------------------
# DEVICE INFORMATION
Trade Name: Capnostream™35 Portable Respiratory Monitor; PM35MN
Common Name: Portable Respiratory Monitor
Device Classification Name: analyzer, gas, carbon-dioxide, gaseous-phase
Regulation Number:
868.1400, Carbon Dioxide Analyzer (Classification CCK) Subsequent product codes: 870.2700 Pulse Oximeter (Classification DQA) 868.2375 Ventilatory Effort Recorder (Classification MNR)
## PREDICATE DEVICE
Capnostream™35 is substantially equivalent to the following commercially available devices:
| Manufacturer | Device | 510(k) No. | Clearance Date |
|----------------------------------------------------|------------------------------------------------------------------------------------------|------------|------------------|
| Primary predicate:<br>Oridion Medical 1987<br>Ltd. | Capnostream20p<br>With Smart A/hr &<br>ODITM | K112368 | July 19, 2012 |
| Secondary predicate:<br>Covidien, LLC | Bedside Respiratory<br>Patient Monitoring<br>System With<br>Respiration Rate<br>Software | K130320 | February 4, 2014 |
{5}------------------------------------------------
# DEVICE DESCRIPTION
The Capnostream™35 is a 4-inch color screen portable two-parameter monitor consisting of a microMediCO2 capnography module and a pulse oximetry module implemented in a host device. The host device displays parameters received from the respective modules and generates alarms when preset alarm thresholds are crossed.
The microMediCO2 module provides the following inputs to the host monitor:
EtCO2 numeric, Respiratory Rate, IPI (integrated Pulmonary Index), Continuous CO2 waveform, Apnea per Hour (A/hr) and Oxygen Desaturation Index (ODI).
The SpO2 module integrated in the Capnostream™35 monitor presented in this submission provides SpO2 and Pulse Rate parameters to the host for display.
The SpO2 measurements are also provided to the microMediCO2 module, enabling the calculation of IPI and ODI.
The host monitor will display this data to the user on a screen as numeric values, and will also display the CO2 waveform and SpO2 (pleth) waveform or pulse bar graph.
The device is intended for use in hospital-type facilities, and during intra-hospital transport as well as during out-of-hospital Emergency Medical Service applications. The device features IP54 Liquids & Solids ingress protection, 1.25M Shockproof status, sunlight readable display, automatic display brightness, hot swap battery capability and altitude use up to 15000 feet (4572m) for use in out-of-hospital Emergency Medical Service applications.
#### INTENDED USE
The Capnostream™35 is a portable capnograph/pulse oximeter, intended to provide professionally trained health care providers with continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath, respiration rate, arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neonatal patients. The pulse oximeter is intended for use during both no motion conditions and for patients who are well or poorly perfused.
{6}------------------------------------------------
The Capnostream™35 also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for patients age 22 and up. The device is intended for use in hospital-type facilities, during intra-hospital transport, and out-of-hospital Emergency Medical Service applications that include ground and air transport.
{7}------------------------------------------------
# COMPARISON TO PREDICATE DEVICE
The Capnostream™35 is substantially equivalent to the following predicate devices:
| Attribute | Primary Predicate<br>Device: | Secondary<br>Predicate Device: | Subject Device : |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Capnostream20p<br>With Smart A/hr &<br>ODI (K112368) | Bedside Respiratory<br>Patient Monitoring<br>System With<br>Respiration Rate<br>Software (K130320) | Capnostream™35<br>Portable Respiratory<br>Monitor |
| Classification | II | II | II |
| Product Code | CCK<br>DQA<br>MNR | DQA<br>Subsequent code:<br>BZQ | CCK<br>Subsequent code:<br>DQA<br>MNR |
| Purpose and<br>Function | Continuous non-<br>invasive monitoring of<br>expired and inspired<br>CO2, EtCO2,<br>respiration rate<br>(provided from the<br>Capnograph module),<br>SpO2 and pulse rate | Continues non-<br>invasive monitoring<br>of SpO2, pulse rate<br>and respiratory rate | Same as K112368<br>(Respiration Rate is<br>derived by<br>Capnography) |
| Target<br>population | Neonatal, pediatric,<br>and adult patients.<br>A/hr and ODI<br>indication for use is<br>for adult patients age<br>22 and up. IPI is<br>intended for pediatric<br>and adult patients<br>only. | SpO2, Pulse Rate:<br>Adult, pediatric, and<br>neonatal patients<br>during both no<br>motion and motion<br>conditions, and for<br>patients who are well<br>or poorly perfused.<br>Respiration Rate:<br>adult patients who<br>are well perfused<br>during no motion | Same as K112368.<br>SpO2, Pulse Rate:<br>same as K130320 |
| | | conditions | |
| Where used | In hospitals, hospital-<br>type facilities, intra-<br>hospital transport and<br>home environments | SpO2, Pulse Rate: In<br>hospitals, hospital-<br>type facilities, and<br>during intra hospital<br>transport.<br>Respiration Rate:<br>hospitals and<br>hospital-type<br>facilities. | In hospitals, hospital-<br>type facilities, intra-<br>hospital transport,<br>out-of-hospital<br>Emergency Medical<br>Service applications<br>that include ground<br>and air transport. |
| Fundamental<br>Technology | NDIR (CO2)<br>Spectrophotometry<br>and Plethysmography | Spectrophotometry<br>and<br>Plethysmography | Same as K112368 |
| Performance<br>Standards | ISO 21647<br>ISO 9919 | EN 80601-2-61 | ISO 80601-2-55<br>ISO 80601-2-61 |
| Main Safety<br>Standards | IEC/EN 60601-1<br>IEC/EN 60601-1-2<br>IEC 60601-1-8<br>IEC 60601-2-49 | EN 60601-1<br>IEC 60601-1-2<br>EN 60601-1-4<br>EN 60601-1-8<br>EN 60601-1-6 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-8<br>IEC 60601-2-49<br>IEC 60601-1-12<br>IEC 60601-1-6 |
{8}------------------------------------------------
{9}------------------------------------------------
### TECHNOLOGICAL AND PERFORMANCE CHARACTERISTICS
The Capnostream™35 features the same technology (NDIR, Spectrophotometry and Plethysmography) as its primary predicate K112368 and the same performance characteristics as its predicates K112368 and K130320. The capnograph module (microMediCO2) integrated in the Capnostream™35 performs as in the predicate K112368 and the SpO2 module integrated in the Capnostream™35 performs as in the predicate K130320. The device was designed to support its use in out-of-hospital Emergency Medical Service applications for which it features IP54 Liquids & Solids ingress protection, 1.25M Shockproof status, sunlight readable display, automatic display brightness, hot swap battery capability and altitude use up to 15000 feet (4572m).
## PERFORMANCE TESTING
Non-clinical tests were performed to support the determination of substantial equivalence. Appropriate safety, environmental and performance tests were conducted to ensure that the specifications of the Capnostream™35 were met. The device has successfully undergone performance testing according to ISO 80601-2-55 and ISO 80601-2-61 as well as electrical safety, electromagnetic and environmental testing according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-49, IEC 60601-1-12 and IEC 60601-1-6. Software testing was performed to validate the performance of the new monitor and its substantial equivalence to the predicate devices.
Wireless communication testing was conducted according to FDA guidance document from Aug 13,2013- Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and Food and Drug Administration Staff.
A hazard analysis was carried out on the Capnostream™35 monitor in compliance with ISO 14971:2012. This hazard analysis concluded that any residual risks were judged as acceptable when weighed against the intended benefits of use of the device.
{10}------------------------------------------------
### Substantial Equivalence
The Capnostream™35 has the same intended use, technological characteristics and performance as its predicate devices. Compared to its predicate devices, the Capnostream™35 is intended also for out-of-hospital Emergency Medical Service applications. Its functionality in its intended environments was verified.
The Capnostream™35 has successfully undergone performance, electrical safety, electromagnetic, environmental and wireless communication testing to ensure it complies with its performance testing requirements. Testing did not raise any concerns when compared to its predicate devices therefore the subject device is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.