K150266 · Remington Medical, Inc. · BSP · Jun 4, 2015 · Anesthesiology
Device Facts
Record ID
K150266
Device Name
Remington Medical Inc. Tuohy Epidural Needles
Applicant
Remington Medical, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Jun 4, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Remington Medical, Inc. Tuohy Epidural Needles are to be used to inject local anesthetics into a patient to provide regional anesthesia or to facilitate the placement of an epidural catheter.
Device Story
Tuohy Epidural Needle; sterile, single-use, disposable device. Components: stainless steel cannula, stainless steel stylet, plastic cannula hub with stainless steel bushing, plastic stylet hub; medical grade adhesive used for assembly; polypropylene sheath for protection. Available in various gauges/lengths; optional fixed or removable wing. Used by clinicians in clinical settings to inject local anesthetics or facilitate epidural catheter placement. Device provides access for regional anesthesia administration; benefits patient by enabling pain management or surgical anesthesia.
Clinical Evidence
No clinical data. Bench testing only. Verification/validation included bond strength, catheter placement, and needle flow comparisons. Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, irritation, acute systemic toxicity, and pyrogenicity. Compliance with ISO 7864 and ISO 6009 confirmed.
Technological Characteristics
Materials: stainless steel cannula/stylet, polycarbonate hubs, stainless steel bushings, UV-cured medical grade adhesive. Sterilization: ETO. Standards: ISO 7864 (bond strength), ISO 6009 (color coding), ISO 10993-1 (biocompatibility). Form factor: needle assembly with sheath, various gauges/lengths.
Indications for Use
Indicated for patients requiring regional anesthesia via local anesthetic injection or epidural catheter placement.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
Predicate Devices
Myco Medical Supplies, Inc. Dr. Japan's Phoenix Epidural & Spinal Needles (K990519)
B. Braun Medical, Inc. Perican™ Ultra Needle (K133632)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, but instead of snakes, it features three abstract human profiles facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 4, 2015
Remington Medical, Inc. Caitlin Senter, M.S., RAC Regulatory Affairs Manager 6830 Meadowridge Court Alpharetta, GA 30005
Re: K150266
Trade/Device Name: Remington Medical Inc. Tuohy Epidural Needles Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: May 1, 2015 Received: May 4, 2015
Dear Ms. Senter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150266
Device Name
Remington Medical, Inc. Tuohy Epidural Needle
Indications for Use (Describe)
Remington Medical, Inc. Tuohy Epidural Needles are to be used to inject local anesthetics into a patient to provide regional anesthesia or to facilitate the placement of an epidural catheter.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Remington Medical. The logo consists of a blue geometric shape on the left and the words "Remington MEDICAL" on the right. The word "Remington" is in a larger, bolder font than the word "MEDICAL", which is stacked below it.
#### Traditional 510(k) Notification Remington Medical Inc. Tuohy Epidural Needle
| Preparation Date | June 2, 2015 |
|------------------------------|-------------------------------------------------------------------------------------------------------------|
| Applicant | Remington Medical, Inc.<br>6830 Meadowridge Court,<br>Alpharetta, GA, USA 30005 |
| | Registration Number: 1056553<br>Owner/Operator Number: 9006473 |
| Contact Person | Caitlin Senter, MS, RAC<br>Regulatory Affairs Manager<br>770-888-8520, extension 207<br>caitlins@remmed.com |
| Trade Proprietary<br>Name(s) | Remington Medical, Inc. Tuohy Epidural Needle |
| Common Name (s) | Needle, Conduction, Anesthetic (W/Wo Introducer) |
| Classification Name | 21 CFR 868.5150 (Anesthesia conduction needle);<br>Product Code: BSP |
| Device Class: | Class II |
## Section 5 - 510(k) Summary
### Legally Marketed Devices to Which Substantial Equivalence is Claimed:
Myco Medical Supplies, Inc. Dr. Japan's Phoenix Epidural & Spinal Needles, (K990519) B. Braun Medical, Inc. Perican™ Ultra Needle, (K133632)
### Description of the Device:
The Tuohy Epidural Needle is a device used for the injection of anesthetic agents into a patient for regional anesthesia administration or to facilitate the placement of an epidural catheter for continuous infusion of local anesthetics.
Tuohy Epidural Needles consist of a plastic cannula hub, containing a stainless steel bushing, affixed to a stainless steel cannula with medical grade adhesive and a stainless steel stylet affixed to a plastic stylet hub with medical grade adhesive. The cannula hub has the option of a permanent, fixed wing or a removable wing depending on user preference. The needle assembly is protected with a polypropylene sheath.
Tuohy Epidural Needles are provided as a sterile, single use, disposable devices. The Tuohy Epidural Needles will be available in a variety of lengths and gauges. They may be packaged individually or included in regional anesthesia trays (kits).
#### Intended Use/Indications for Use
Remington Medical, Inc. Tuohy Epidural Needles are to be used to inject local anesthetics into a patient to provide regional anesthesia or to facilitate the placement of an epidural catheter.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Remington Medical. The logo consists of a blue geometric shape on the left and the words "Remington MEDICAL" in blue on the right. The word "Remington" is in a larger font than the word "MEDICAL", which is directly below it.
#### Traditional 510(k) Notification Remington Medical Inc. Tuohy Epidural Needle
### Technological Characteristics:
The technological characteristics (design, specifications, and performance) of the subject device and the predicate devices are substantially equivalent. Remington Medical Inc. Tuohy Epidural Needles meet performance standards where applicable for:
Hub to Needle Bond Strength: ISO 7864
Color: ISO 6009
| | Subject Device:<br>Remington Medical Inc. Tuohy<br>Epidural Needles | Predicate Device:<br>Myco Medical Supplies, Inc.<br>Dr. Japan's Phoenix Epidural &<br>Spinal Needles (K990519) | Predicate Device:<br>B. Braun Medical, Inc.<br>Perican™ Ultra Needle,<br>(K133632) |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class | Class II | Class II | Class II |
| FDA Product<br>Code | BSP | BSP | BSP |
| Regulation | 21 CFR 868.5150 | 21 CFR 868.5150 | 21 CFR 868.5150 |
| Indications for<br>Use Statement | Remington Medical, Inc. Tuohy<br>Epidural Needles are to be used to<br>inject local anesthetics into a<br>patient to provide regional<br>anesthesia or to facilitate the<br>placement of an epidural catheter. | Spinal and Epidural needles are to<br>be used to inject local anesthetics<br>into a patient to provide regional<br>anesthesia. | The B. Braun Perican Ultra<br>needle is intended for use in<br>injecting local anesthetics<br>and/or analgesics into a patient<br>to provide regional anesthesia<br>and/or to facilitate the<br>placement of a catheter. |
| Cannula<br>Material | Metal, Stainless Steel | Metal, Stainless Steel | Metal, Stainless Steel |
| Cannula Depth<br>Marking | 10 mm, Wide Band | Wide Band | Wide Band |
| Cannula Hub<br>Material | Plastic (Polycarbonate), Stainless<br>Steel Bushing and UV Cured<br>Medical Grade Adhesive | Plastic, Metal Bushing, and<br>Adhesive | Plastic, Spacer (Bushing) and<br>Adhesive |
| Stylet Material | Metal, Stainless Steel | Metal, Stainless Steel | Metal, Stainless Steel |
| Stylet Hub<br>Material | Color-coded Plastic<br>(Polycarbonate) and UV Cured<br>Medical Grade Adhesive | Color-coded Plastic and Adhesive | |
| Gauge/Length<br>(inches)/Color<br>Depiction | 17 GA 3.5", 5.0" Violet | 16 GA 3.5" White | |
| | 18 GA 3.5", 5.0" Pink | 17 GA 3.5", 6.0" Violet | |
| | 19 GA 2.0" Cream | 18 GA 2.5", 3.5", 6.0" Pink | |
| | 20 GA 3.5", 5.0", 6.0" Yellow | 20 GA 3.5", 4.5", 6.0" Yellow | |
| | 22 GA 3.5" Black | 22 GA 2.5", 3.5" Black | |
| Final Needle<br>Assembly<br>Protection | Sheath | Sheath | Sheath |
| Sterilization<br>Method | ETO | ETO | ETO |
{5}------------------------------------------------
### Traditional 510(k) Notification Remington Medical Inc. Tuohy Epidural Needle
### Performance Data:
Tests were performed on the Remington Medical Inc. Tuohy Epidural Needles including verification/validation testing to internal functional specifications (e.g. catheter placement, and bond strength between cannula, hub, and stylet) including comparison data (e.g., needle flow) to the K990519 predicate device. These tests demonstrated that the Remington Medical Inc. Tuohy Epidural Needles are substantially equivalent to the predicate device. Visual inspection of the test catheter under magnification confirmed no negative impact to the catheter post-placement. Testing confirmed that the Remington Medical Inc. Tuohy Epidural Needles comply with relevant standards, specifically ISO standards 7864 and 6009. In addition, evaluations and validations have been performed to demonstrate compliance to the applicable standards for biocompatibility (ISO 10993-1) and sterilization. Biocompatibility testing performed includes cytotoxicity, sensitization, irritation or intracutaneous reactivity, and acute systemic toxicity. Pyrogenicity was also tested.
## Clinical testing:
Clinical testing was not required.
## Conclusion:
The results of the non-clinical testing demonstrated that the subject device, Remington Medical Inc. Tuohy Epidural Needles, is substantially equivalent to the predicate devices, Myco Medical Supplies, Inc. Dr. Japan's Phoenix Epidural & Spinal Needles and B. Braun Medical, Inc. Perican™ Ultra Needle, (K133632) with respect to intended use, materials, design, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.