CLIP PLUS Disposable Clip Applier,ML;CLIP PLUS Disposable Clip Applier,L

K150259 · Taiwan Surgical Corporation · FZP · Nov 6, 2015 · General, Plastic Surgery

Device Facts

Record IDK150259
Device NameCLIP PLUS Disposable Clip Applier,ML;CLIP PLUS Disposable Clip Applier,L
ApplicantTaiwan Surgical Corporation
Product CodeFZP · General, Plastic Surgery
Decision DateNov 6, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The use of the product is indicated in endoscopic procedures, including laparoscopic surgical procedures where instruments are introduced into the body through a cannula for the purpose of ligating vessels and tubular structures.

Device Story

Device consists of handle, trigger, 360-degree rotational knob, cartridge housing shaft, and jaws; used with sterile, single-use disposable titanium clip cartridges. Operated by surgeons in laparoscopic/endoscopic procedures. Surgeon inserts cartridge into shaft via rear slot; squeezes trigger to deploy and close titanium clips onto vessels or tubular structures. Design allows multiple clip applications without withdrawing instrument from cannula. Output is secured titanium clip placement for vessel ligation. Benefits include reduced procedure time and minimized tissue trauma by avoiding repeated instrument insertion/withdrawal.

Clinical Evidence

Bench testing only. Evaluated clip formation, clip gap measurement, perpendicular/parallel clip pull strength, and airtight capability (15-30 PSI) compared to predicate. Biocompatibility testing performed per ISO 10993. Sterilization validated per ISO 11137-2:2012.

Technological Characteristics

Materials: Stainless steel (handle/shaft), titanium (clips). Dimensions: 10mm shaft diameter; 300mm/350mm length. Principle: Mechanical clip applier with rotational jaw control. Sterilization: Gamma irradiation (ISO 11137-2:2012).

Indications for Use

Indicated for patients undergoing endoscopic or laparoscopic surgical procedures requiring ligation of vessels and tubular structures.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 6, 2015 Taiwan Surgical Corporation Ms. Hsiu-Ping Huang Regulatory Specialist 3F., No. 12, Sec. 2. Sheng Yi Rd. Zhubei City, Hsinchu County 30261 Taiwan Re: K150259 Trade/Device Name: CLIP PLUS Disposable Clip Applier, ML CLIP PLUS Disposable Clip Applier, L Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP. GDO Dated: January 30, 2015 Received: October 5, 2015 Dear Ms. Huang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # David Krause -S - for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150259 #### Device Name CLIP PLUS Disposable Clip Applier, ML CLIP PLUS Disposable Clip Applier, L #### Indications for Use (Describe) The use of the product is indicated in endoscopic procedures, including laparoscopic surgical procedures where instruments are introduced into the body through a cannula for the purpose of ligating vessels and tubular structures . Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The Assigned 510(K) Number: K150259 Date Prepared: 11/06/2015 #### I. SUBMITTER: ## Submitter: TAIWAN SURGICAL CORPORATION Address: 3F., No.12, Sec.2, ShengYi Rd., Zhubei City, Hsinchu County 302, Taiwan Phone Number: +886-3-6588129 Fax Number: +886-3-6588355 Contact Person: Ging-Wen Lu Email: lulu@taiwansurgical.com.tw Phone Number: +886-3-6588129 ext. 260 Fax Number: +886-3-6588355 ### II. DEVICE Trade Name: - CLIP PLUS™ Disposable Clip Applier, ML . - CLIP PLUS™ Disposable Clip Applier, L . Common Name and Classification: Table 1. Common Name and Classification | No. | Common Name | Product Code | Classification | Regulation<br>Section | Panel | |-----|-------------------------------------------------|--------------|-----------------|-----------------------|---------------------------------| | 1 | Applier, surgical, clip<br>and implantable clip | GDO<br>FZP | I, exempt<br>II | 878.4800<br>878.4300 | General &<br>Plastic<br>Surgery | {4}------------------------------------------------ ### III. PREDICATE DEVICE Predicate Device 1: MICROLINE PENTAX, INC., K013695 | Subject Device | Predicate Device | | | |-------------------------------------------|-------------------------------------------------------------------------|---------------------------|---------------| | | Predicate Device | Manufacturer | 510(k) Number | | CLIP PLUSTM<br>Disposable Clip<br>Applier | Reusable laparoscopic clip<br>applier with implantable<br>titanium clip | MICROLINE<br>PENTAX, INC. | K013695 | Table 2. Predicate Device Identification ### IV. DEVICE DESCRIPTION The Clip Applier consists of a handle piece and implantable titanium clips. The Clip Applier handle piece consists of a molded handle and trigger, a 360 degrees rotational knob, a cartridge housing shaft and a pair of jaws which provide secured placement of the clip to the desired vessel. The design enables the surgeon to apply several clips during a laparoscopic procedure without the need for withdrawing and reinserting the Clip Applier each time a clip is put in place and closed. The Clip Applier is designed to be used with a black loaded, sterile cartridge supplied in a sterile pouch. The single use disposable clip cartridge sits inside the shaft of the Clip Applier handle piece. While fully squeezing the hand piece trigger, the cartridge is inserted into the shaft through a slot located in the instrument rear. #### v. INDICATIONS FOR USE The use of the product is indicated in endoscopic procedures, including laparoscopic surgical procedures where instruments are introduced into the body through a cannula for the purpose of ligating vessels and tubular structures. ## VI. PRODUCT SPECIFICATION The shaft of the Clip Applier handle piece is sized to fit through 10mm cannula ports and the overall length of the shaft is 300mm for TDC-10-ML while the overall length of the shaft is 350mm for TDC-10-L. For TDC-10-ML, the clip length is 9mm long when closed and the clip {5}------------------------------------------------ depth is 7.5mm; for TDC-10-L, the clip length is 10.5mm long when closed and the clip depth is 9.5mm. # VII. COMPARISION OF TECHNOLOGY CHARACTERISTICS WITH THE PRIDICATE DEVICE | Features | | Candidate Device | Predicate Device | | |----------|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------| | 1 | Hand<br>Piece | Long tube<br>diameter | 10mm | 10mm | | 2 | Long tube length | | 290mm/340mm | 300mm/350mm | | 3 | Staple quantity | | 20 | 20 | | 4 | Counter | | N | Y | | 5 | Instructions | | Y | Y | | 6 | Staple fallen out<br>prevention<br>mechanism | | N | Y | | 7 | Surgical field | | 15° | 15° | | 8 | Titanium<br>Clip | Staple wire | 0.9x0.6mm | 0.9x0.6mm | | 9 | Staple wide | | 4.3mm | 4.3mm | | 10 | Staple depth | | 6.9mm/9.5mm | 6.9mm/9.5mm | | 11 | Pinch flat length | | 9mm/10.5mm | 9mm/10.5mm | | 12 | Function | Clip Formation | Pass | Pass | | | Clip Gap<br>measurement | | $0.215\pm0.024$ mm | $0.223\pm0.005$ mm | | | Perpendicular<br>Clip Pull | | $315\pm11$ g | $305\pm11$ g | | | Parallel Clip Pull | | $627\pm41$ g | $622\pm52$ g | | | Airtight<br>Capability | | Pass | Pass | | 13 | Regulatory Classification | | class II | class I/class II | | | | | | | | | | 878.4800 | 878.4800 | | | | | 878.4300 | 878.4300 | | | | Intended use | Use of these products is indicated in endoscopic, including laparoscopic surgical procedures where instruments are introduced into the body through a cannula for the purpose of ligating vessels with Titanium clips. | The Microline reusable laparoscopic clip applier with implantable titanium surgical clip is intended for use to occlude vessels, ducts, tracts and other tubular structures during laparoscopic and general surgical procedures. | | | 15 | Package | Blister with Tyvek lid | Blister with Tyvek lid | | Table 3. Specification Comparison Table {6}------------------------------------------------ #### PERFORMANCE DATA VIII. - Biocompatibility : Because the stainless steel and titanium are standard materials for current clip applier on the market, they have been thoroughly tested in the past for biocompatibility. The product has passed the biocompatibility tests by following ISO 10993. The contact time for the device is permanent (≥ 30 days) with the body. - Performance Testing : The Bench report of Disposable Clip Applier indicates the following results. In clip formation test, the clips were all properly formed by the clip applier. In clip gap measurement, there is no significant difference between Taiwan Surgical Corporation (TWSC) and MICROLINE applier. The perpendicular clip pull test shows that the TWSC result is similar with the MICROLINE result suggested that the TWSC clips is not easily dislodged from the blood vessel. The parallel clip pull test shows that the TWSC result is similar with the MICROLINE result suggested that the TWSC clips is not easily sliding over the blood vessel. The airtight capability test result reveals that no tested clips applied to silicone tube (3.0 and 2.0 mm O.D.) allowed for any air {7}------------------------------------------------ leakage of both clip Appliers through the air pressure from 15 to 30 PSI. - Sterilization Verification Testing: The sterilization validation of gamma irradiation for disposable clip applier was successful and had met the requirements of ISO 11137-2:2012 VD max25 method on substantiation of 25kGy as a sterilization dose. This study therefore supports the multiple batch products to be irradiated at the sterilization dose kGy for a SAL of 10 °. # IX. CONCULSIONS The CLIP PLUS™ Disposable Clip Applier with implantable titanium clip has the same intended use and same basic technology as the predicate device, thus is able to achieve same effectiveness and safety as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...