K150232 · Ashitaka Factory of Terumo Corp. · DQO · Oct 19, 2015 · Cardiovascular
Device Facts
Record ID
K150232
Device Name
Radifocus Optitorque Angiographic Catheter
Applicant
Ashitaka Factory of Terumo Corp.
Product Code
DQO · Cardiovascular
Decision Date
Oct 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Radifocus Optitorque Angiographic Catheter is intended for use in cardiac and vascular procedures. It is designed to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the vascular system. The different shapes are designed to selectively engage arteries from access sites such as the femoral, radial, and brachial artery.
Device Story
Radifocus Optitorque Angiographic Catheter is a disposable, single-use diagnostic intravascular catheter. It functions as a conduit to deliver radiopaque contrast media, guide wires, and therapeutic agents to specific vascular sites. The device is operated manually by a physician during cardiac and vascular procedures. It features a multi-layer shaft construction (polyurethane/polyamide elastomers with a stainless steel mesh braid) and a soft distal tip/tube designed to minimize vessel wall trauma. The device is inserted via femoral, radial, or brachial access sites. Visibility under fluoroscopy is provided by barium sulfate incorporated into the shaft layers. The physician uses the catheter to selectively engage arteries; the output is the successful delivery of diagnostic or therapeutic agents to the target site, facilitating clinical imaging or intervention. The device benefits patients by enabling precise, minimally invasive access to the vascular system for diagnostic and therapeutic purposes.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ISO 10555-1 (radio-detectability, tensile force, leakage, power injection, flow rate, burst pressure) and internal standards (torque transmission, bending strength, cleanliness, etc.). Biocompatibility testing (ISO 10993) performed on non-aged and 3-year accelerated-aged devices.
Technological Characteristics
Two-layer shaft construction: inner/outer layers of polyurethane and polyamide elastomers with a stainless steel mesh braid. Barium sulfate for radiopacity. Soft tip/tube: polyurethane elastomer. Dimensions: 4-6 Fr, 65-120 cm lengths. Manual operation. Sterilization: Ethylene oxide (ISO 11135). Complies with ISO 10555-1 and ISO 10993.
Indications for Use
Indicated for use in cardiac and vascular procedures to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the vascular system via femoral, radial, or brachial access. Contraindicated in patients with a contraindication to anti-platelet and/or anti-coagulation therapy.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 19, 2015
Ashitaka Factory of Terumo Corporation Monika McDole-Russell Senior Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Ave., Suite 320 Somerset, New Jersey 08873
Re: K150232
Trade/Device Name: Radifocus® Optitorque™ Angiographic Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO Dated: September 17, 2015 Received: September 18, 2015
Dear Monika McDole-Russell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
M. A. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K150232
#### Device Name
Radifocus Optitorque Angiographic Catheter
#### Indications for Use (Describe)
The Radifocus Optitorque Angiographic Catheter is indicated for use in carcular procedures. It is designed to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the different shapes are designed to selectively engage arteries from access sites such as the femoral, radial, and brachial artery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K150232 – 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
## Table of Contents
| A. SUBMITTER INFORMATION (807.92(A)(1)). | 2 |
|-------------------------------------------------------|----|
| B. DEVICE NAME (807.92(A)(2)). | 3 |
| C. PREDICATE DEVICE (807.92(A)(3)). | 3 |
| D. REASON FOR 510(K) SUBMISSION. | 3 |
| E. DEVICE DESCRIPTION (807.92(A)(4)). | 4 |
| F. INTENDED USE (807.92(A)(5)). | 4 |
| G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(A)(6)). | 5 |
| H. NON CLINICAL TESTS (807.92(B)(1)) | 7 |
| I. CLINICAL TESTS (807.92(B)(2)) | 11 |
| J. CONCLUSION (807.92(B)(3)). | 11 |
{5}------------------------------------------------
#### 510(k) SUMMARY
## A. SUBMITTER INFORMATION (807.92(a)(1))
| Prepared by: | Monika McDole-Russell, MSRA, RAC |
|--------------|--------------------------------------|
| | Senior Regulatory Affairs Specialist |
| | Terumo Medical Corporation |
| | 265 Davidson Avenue |
| | Suite 320 |
| | Somerset, NJ 08873 |
| | Tel. (732) 302-4900 ext. 4273 |
| | Fax (410) 398-6079 |
#### Prepared for: Owner/Operator
Terumo Corporation 44-1, 2-Chome, Hatagaya Shibuya-Ku, Tokyo Japan 151-0072 Registration Number: 801 002 6
#### Manufacturer and Sterilization Facility
Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya Shizuoka 418-0015 Japan Registration No: 968 183 4
- Contact Person: Monika McDole-Russell, MSRA, RAC Senior Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue Suite 320 Somerset, NJ 08873 Tel. (732) 302-4900 ext. 4273 Fax (410) 398-6079
{6}------------------------------------------------
E-mail: monika.mcdolerussell(@terumomedical.com
Date prepared: October 19, 2015
#### B. DEVICE NAME (807.92(a)(2))
| Proprietary Name: | Radifocus Optitorque Angiographic Catheter |
|-----------------------|--------------------------------------------|
| Common Name: | Angiographic, Catheter |
| Classification Name: | Diagnostic Intravascular Catheter |
| Classification Panel: | Cardiovascular |
| Regulation: | 21 CFR 870.1200 |
| Product Code: | DQO |
| Classification: | Class II |
#### C. PREDICATE DEVICE (807.92(a)(3))
The legally marketed device(s) to which substantial equivalence is claimed is K082736 Radifocus Optitorque Angiographic Catheter, manufactured by Ashitaka Factory of Terumo Corporation in Japan.
## D. REASON FOR 510(K) SUBMISSION
This premarket notification is being submitted due to a raw material change in the catheter shaft (outer layer) of the device. The current polyamide elastomer material is no longer available from the vendor and has been replaced with an equivalent polyamide elastomer. Biocompatibility testing has been completed to demonstrate that the raw material change has not affected the safety and performance of the device. The IFU is also being revised to update the flow rate table and add a contraindication; the addition of the contraindication is to bring the IFU into alignment with wording recommended in FDA draft guidance Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (issued April 23, 2013). Additionally, a clarification is being made to the wording of the Indications for Use; however, no changes are being made to the actual indications previously cleared under K082736.
{7}------------------------------------------------
## E. DEVICE DESCRIPTION (807.92(a)(4))
The Radifocus Optitorque Angiographic Catheter is comprised of a two-layer construction featuring stainless steel mesh sandwiched between layers of polyurethane and polyamide elastomers. The shaft inner layer and outer layer contain barium sulfate for visibility and contrast under fluoroscopy. A soft tip is attached to the distal portion of some 4 Fr and all 5 Fr and 6 Fr catheters except those with a pigtail design; a soft tube is attached to the distal portion of 5 Fr and 6 Fr catheters with a pigtail design; the 4 Fr products are available with or without a soft tip. Constructed of flexible, supple polyurethane that is permanently welded to the catheter shaft, the soft tip and soft tube are designed to minimize trauma to the vessel wall. There is no change to the design of the device as a result of this submission.
The device is offered in lengths of 65-120 cm. French sizes and shaft inner diameters are as follows:
| French Size | Shaft Inner Diameter |
|-------------|----------------------|
| 4 | 1.05mm |
| 5 | 1.22mm |
| 6 | 1.32mm |
It is a disposable device intended for single use only. This device is individually packaged and sterilized by ethylene oxide gas.
## F. INTENDED USE (807.92(a)(5))
The Radifocus Optitorque Angiographic Catheter is intended for use in cardiac and vascular procedures. It is designed to deliver radiopaque media, guide wires, catheters, and therapeutic agents to selected sites in the vascular system. The different shapes are designed to selectively engage arteries from access sites such as the femoral, radial, and brachial artery.
Note: This intended use is identical to the predicate device, Radifocus Optitorque Angiographic Catheter (K082736).
{8}------------------------------------------------
#### G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))
The Radifocus Optitorque Angiographic Catheter with the new material, the subject of this Traditional 510(k), is substantially equivalent in its intended use/indications for use, technology/principal of operation, materials, and performance to the currently marketed Radifocus Optitorque Angiographic Catheter, manufactured by Ashitaka Factory of Terumo Corporation and cleared under K082736.
A comparison of the intended use/indications for use and technological characteristics are summarized in the table below. The raw material change and addition of a contraindication to the IFU do not negatively impact the safety and effectiveness of the device.
Note: A statement of substantial equivalence to another product is required by 21CFR807.87, and relates only to whether the present product can be marketed without prior reclassification or clinical approval. The present submission is, therefore, not related to the coverage of any patent, and is not to be interpreted as an admission or used as evidence in a patent infringement lawsuit. As the Commissioner of the FDA has stated, "…a determination of substantial equivalence under the federal Food, Drug, and Cosmetic Act relates to the fact that the product can lawfully be marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whatever on the resolution of patent infringement suits" 42 Fed. Reg. 42,520, et seq. (1977)
{9}------------------------------------------------
| Device<br>Characteristic | | Predicate Device: Radifocus Optitorque<br>Angiographic Catheter (K082736) | Subject Device: Radifocus Optitorque<br>Angiographic Catheter | | | | | | | | | | | | | | | |
|----------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|------|
| Manufacturer | | Ashitaka Factory of Terumo Corporation | Same | | | | | | | | | | | | | | | |
| Intended Use/Indications for Use | Indications<br>for Use | The Radifocus Optitorque Angiographic<br>Catheter is intended for cardiac and<br>vascular procedures. It is designed to<br>deliver radiopaque media, guide wires,<br>catheters, and therapeutic agents to<br>selected sites in the vascular system. The<br>different shapes are designed to<br>selectively engage arteries from access<br>sites such as the femoral, radial, and<br>brachial artery. | The Radifocus Optitorque Angiographic<br>Catheter is indicated for use in cardiac<br>and vascular procedures. It is designed to<br>deliver radiopaque media, guide wires,<br>catheters, and therapeutic agents to<br>selected sites in the vascular system. The<br>different shapes are designed to<br>selectively engage arteries from access<br>sites such as the femoral, radial, and<br>brachial artery. | | | | | | | | | | | | | | | |
| | Contraindic<br>-ations | Contraindications<br>cleared<br>under<br>K082736 | Contraindications<br>cleared<br>under<br>K082736, plus the following statement:<br>"Patients with a contraindication to anti-<br>platelet and/or anti-coagulation therapy" | | | | | | | | | | | | | | | |
| | Warnings<br>and<br>Precautions | Warnings and precautions cleared under<br>K082736 | Warnings and precautions cleared under<br>K082736, plus the maximum flow rate of<br>saline was added in accordance with 6.3.<br>(i) of ISO 10555-1:2013. | | | | | | | | | | | | | | | |
| Operation<br>Principle | | Manual | Same | | | | | | | | | | | | | | | |
| Design /<br>Construction | | Catheter assembly (shaft, hub, soft<br>tip/soft tube, strain relief/inserter, hub<br>support tube (4 Fr only))<br>Fr Distal Part Proximal Part 4 Soft tip Strain Relief Inserter 5, 6 Soft tip Strain Relief Soft tube Inserter | | | | | | | | | | | | | | | | Same |
| Materials | | Shaft<br>●<br>o Inner Layer - Polyamide<br>elastomer, Polyurethane<br>elastomer<br>o Braid - Stainless steel<br>o Outer Layer - Polyamide<br>elastomer, Polyurethane<br>elastomer | Shaft<br>●<br>o Inner Layer - Same<br>o Braid - Same<br>o Outer Layer - Polyamide<br>elastomer, Polyurethane<br>elastomer | | | | | | | | | | | | | | | |
## Table 5.1: Device Comparison Table
{10}------------------------------------------------
| Device<br>Characteristic | | Predicate Device: Radifocus Optitorque<br>Angiographic Catheter (K082736) | Subject Device: Radifocus Optitorque<br>Angiographic Catheter |
|--------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Package | | Soft tip – Polyurethane elastomer<br>●<br>Soft tube (5 and 6 Fr only) -<br>●<br>Polyurethane elastomer<br>Strain relief – Polyamide elastomer<br>●<br>Hub – Polyamide<br>●<br>Hub support tube (4 Fr only) -<br>●<br>Polyurethane elastomer<br>Inserter – Polyethylene<br>●<br>Adhesive - Cyanoacrylate<br>· Paperboard mount<br>• individual package<br>• Unit box<br>· Shipping carton<br>• Pouch | Soft tip - Same<br>●<br>Soft tube (5 and 6 Fr only) – Same<br>●<br>Strain relief - Same<br>●<br>Hub - Same<br>●<br>Hub support tube (4 Fr only) – Same<br>●<br>Inserter - Same<br>●<br>Adhesive - Same<br>●<br>Same |
| pecifications | Sizes/Shaft<br>Inner<br>Diameter | 4 Fr / 1.05mm<br>5 Fr / 1.22mm<br>6 Fr / 1.32mm | Same |
| | Lengths | 65-120 cm | Same |
| | Maximum<br>Labeled<br>Injection<br>Pressure | 4 Fr: 750 psi<br>5 Fr and 6 Fr: 1000 psi | Same |
| Sterilization | | Ethylene oxide | Same |
# H. NON CLINICAL TESTS (807.92(b)(1))
## Performance
Performance testing was conducted to ensure the safety and effectiveness of the Radifocus Optitorque Angiographic Catheter with the new raw material throughout the device's shelf life, verify conformity to the applicable parts of ISO standards, and demonstrate substantial equivalence to the predicate device.
No new issues of safety and effectiveness were raised with the testing performed.
{11}------------------------------------------------
The following tests were performed on non-aged and accelerated aged (3 yr.) samples according to currently accepted ISO standards:
| Test | Standard |
|-----------------------------|------------------------------------------------------|
| Radio-detectability | ISO 10555-1:2013 Section 4.2<br>ASTM F640-12 |
| Surface | ISO 10555-1:2013 Section 4.4 |
| Peak tensile force | ISO 10555-1:2013 Section 4.6 |
| Freedom from leakage | ISO 10555-1:2013 Section 4.7 |
| Power injection | ISO 10555-1:2013 Section 4.10<br>Annex F and Annex G |
| Distal tip | ISO 10555-1:2013 Section 4.12 |
| Flow rate<br>Burst pressure | ISO 10555-1:2013 Section 6.3 |
Table 5.2: Performance Testing per ISO Standards
All samples tested met the standard applicable to each test.
Additionally, performance testing other than to the above ISO Standards was performed on the device in accordance with Terumo's internal standards. The device complies with the acceptance criteria established for each test based on the predicate:
Table 5.3: Performance Testing per Internal Standards
| Performance Test | Results |
|---------------------------------------------|---------------------------|
| Shaft stiffness (Stiffness of braided-area) | Meets acceptance criteria |
| Torque transmission | Meets acceptance criteria |
| Torque strength | Meets acceptance criteria |
| Tensile strength (Shaft) | Meets acceptance criteria |
{12}------------------------------------------------
| Performance Test | Results |
|----------------------------------------------------------------|---------------------------|
| Bending strength | Meets acceptance criteria |
| Cleanliness | Meets acceptance criteria |
| Strength of hub junction | Meets acceptance criteria |
| Tensile Strength of soft tip | Meets acceptance criteria |
| Product dimensions<br>(I.D.; O.D.; Tip I.D.; Tip O.D.; Length) | Meets acceptance criteria |
| Embolic cube testing | Meets acceptance criteria |
Performance testing demonstrates that the Radifocus Optitorque Angiographic Catheter with the new raw material conforms to the recognized consensus ISO standards, is substantially equivalent to the predicate device, and is acceptable for clinical use throughout its shelf life.
## Biocompatibility
The Radifocus Optitorque Angiographic Catheter with the new raw material is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). This is the same classification as the predicate Radifocus Optitorque Angiographic Catheter (K082736).
The finished device's blood/body contacting parts were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and Draft Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." Tests were also performed on accelerated aged devices to ensure that biocompatibility is maintained throughout the shelf life of the product: results of the testing demonstrate that the device remains biocompatible throughout its shelf life.
{13}------------------------------------------------
Table 5.4: Biocompatibility Testing
| Testing performed on non-aged, sterile, whole devices |
|----------------------------------------------------------------------------|
| Cytotoxicity |
| Sensitization |
| Intracutaneous Reactivity |
| Systemic Toxicity (acute) |
| Pyrogenicity |
| Hemolysis |
| Thrombogenicity |
| Complement Activation (Immunology) |
| Chemical Extraction |
| Testing performed on accelerated-aged (3 years),<br>sterile, whole devices |
| Cytotoxicity |
| Hemolysis |
| Chemical Extraction |
## Pyrogen Testing
The Radifocus Optitorque Angiographic Catheter with the new raw material is certified to be non-pyrogenic in the unopened and undamaged package. Pyrogen testing was performed in accordance with applicable ISO standards 10993-11 (2006) and ISO 10993-12 (2012) as well as United States Pharmacopeia 35, National Formulary 30, 2012 <151 > Pyrogen Test (USP Rabbit Test).
#### Sterilization
The sterilization conditions have been validated according to ISO 11135: 2014, Sterilization of Health Care Products - Ethylene Oxide - Requirements for development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10th. No changes have been made to the sterilization process for this device as a result of this submission.
Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) will meet requirements for limited exposure devices (contact up to 24 hours) prior to use based on ISO 10993-7, Biological Evaluation of medical devices- Part 7: Ethylene Oxide Sterilization residuals. Residual EO will not exceed 4 mg per device and residual ECH will not exceed 9 mg per device.
{14}------------------------------------------------
#### Risk Analysis
A Product Risk Analysis was conducted in accordance with ISO 14971, taking into account the modifications to the previous device, and it was determined that any new risks were adequately captured and mitigated.
#### I. CLINICAL TESTS (807.92(b)(2))
This 510(k) does not include data from clinical tests.
## J. CONCLUSION (807.92(b)(3))
The Radifocus Optitorque Angiographic catheter with the new materials is substantially equivalent in intended use, principles of operation, design features, materials, performance and fundamental scientific technology when compared to the predicate Radifocus Optitorque (K082736). Verification testing was conducted and demonstrated that the modified device meets the design inputs and meets the same or equivalent requirements as the predicate Radifocus Optitorque for both ISO 10555-1 and internal standards. The differences between the predicate and proposed devices do not raise any new issues regarding safety and effectiveness. Therefore, the Radifocus Optitorque Angiographic catheter with the new materials is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.