Hematocrit and Metabolite QUALICHECK

K150226 · Radiometer Medical Aps · JJS · May 13, 2015 · Clinical Chemistry

Device Facts

Record IDK150226
Device NameHematocrit and Metabolite QUALICHECK
ApplicantRadiometer Medical Aps
Product CodeJJS · Clinical Chemistry
Decision DateMay 13, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

This Hematocrit and Metabolite QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: cGlucose, cLactate, Hct. For prescription use only.

Device Story

Hematocrit and Metabolite QUALICHECK is a two-level aqueous quality control system; used to verify accuracy and precision of blood gas/metabolite analyzers (ABL77, ABL555, ABL 605/615/625, ABL 80 FLEX, ABL 80 Basic, EML 105). System consists of 2 mL ampoules containing organic buffer, acid salts, metabolites, and preservative. Clinicians use the control solution to validate analyzer performance; results are compared against provided control ranges. Single-use only; contents must be used immediately upon opening.

Clinical Evidence

No clinical data. Bench testing only. Stability testing confirmed 24-month shelf life at 2–25 °C. Value assignment performed via comparative measurement against reference batches across analyzer platforms.

Technological Characteristics

Aqueous solution containing organic buffer, acid, salts, metabolites, and ProClin 950 preservative. Two-level system provided in 2 mL ampoules. Storage 2 °C to 25 °C. Compatible with specific Radiometer ABL and EML analyzer models. Value assignment uses validated algorithms and statistical analysis of measurement data.

Indications for Use

Indicated for use as an assayed quality control system to evaluate accuracy and precision of cGlucose, cLactate, and Hct measurements on specified Radiometer analyzers.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k150226 B. Purpose for Submission: New Device C. Measurand: Quality control material for blood gases D. Type of Test: Not Applicable E. Applicant: Radiometer Medical ApS F. Proprietary and Established Names: Hematocrit and Metabolite QUALICHECK G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1660, Quality Control Material (Assayed and Unassayed) 2. Classification: Class I, reserved 3. Product code: JJS 4. Panel: Clinical Chemistry (75) {1} H. Intended Use: 1. Intended use(s): See Indications for Use 2. Indication(s) for use: This Hematocrit and Metabolite QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. Analytes are: cGlucose, cLactate, Hct 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Not applicable I. Device Description: The Hematocrit and Metabolite QUALICHECK is a two-level quality control system, which can be used for quality control of the ABL77, ABL555, ABL 605/615/625, ABL 80 FLEX, ABL 80 Basic, and EML 105. The system consists of 30 ampoules per box. Each ampoule contains 2 mL of solution. Two levels of different concentrations of analytes for glucose, lactate, and hematocrit are provided. The quality control is an aqueous solution containing an organic buffer, acid salts, metabolites, and a preservative. J. Substantial Equivalence Information: 1. Predicate device name(s): Radiometer QUALICHECK5+ 2. Predicate 510(k) number(s): k980135 3. Comparison with predicate: {2} | Similarities | | | | --- | --- | --- | | Item | Candidate Device | Predicate Device | | | Radiometer Hematocrit and Metabolite QUALICHECK (Candidate Device) | Radiometer QUALICHECK5+ k980135 (Predicate Device) | | Intended Use | The Hematocrit and Metabolite QUALICHECK solution is an assayed quality control system for evaluating the accuracy and precision of all parameters listed on the insert specifying the control ranges. | Same | | Form | Liquid | Same | | Matrix | Aqueous solution | Same | | Differences | | | | --- | --- | --- | | Item | Candidate Device | Predicate Device | | | Radiometer Hematocrit and Metabolite QUALICHECK (Candidate Device) | Radiometer QUALICHECK5+ k980135 (Predicate Device) | | Analytes | cGlucose, cLactate, Hct | pH, pCO2, pO2, cNa+, cK+, cCa2+, cCl- cGlucose- cLatate, ctHb- sO2- FO2Hb- FCOHb, FMetHb, FHbF, ctBil | | Storage | 2 °C to 25 °C for 2 years | 2 °C to 25 °C until expiration date including up to 15 days at up to 32 °C | | Levels | Two levels | Four levels | # K. Standard/Guidance Document Referenced (if applicable): None was referenced # L. Test Principle: Not applicable # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable {3} b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: The materials are traceable to the following standard materials: | Analyte | Unit | Traceable to | | --- | --- | --- | | cGlucose | mmol/L | NIST Standard Reference Material (SRM) 917b | | cLactate | mmol/L | L-Lactic Acid Lithium Salt (commercial source) | | Hematocrit (Hct) | g/dL | Conductivity | Stability: The shelf-life stability (real time testing) was conducted with three lots of Hematocrit and Metabolite QUALICHECK Level 1 and 2 at 32 °C for 3 months followed by 25 °C for a total of 25 months. Multiple time points were tested, with six replicates at each test point. The real-time stability study protocol and acceptance criteria have been reviewed and found to be adequate. The data indicated that the control solutions are stable for 24 months at 2 – 25 °C. There is no open-vial (in use) stability claim since the controls can only be used once. The contents should be used immediately after opening. Expected Values/Value assignment Assigned values and control ranges are established based on the true values and predetermined specifications according to the internal procedure. The sponsor provided assigned values and control ranges for each analyte on each of the 3 different analyzers. To determine the assigned values and control ranges for the Hematocrit and Metabolite QUALICHECK, six trays of 1000 ampoules are sampled randomly from the Hematocrit and Metabolite QUALICHECK batch. Five ampoules are sampled from each of the 6 trays and 30 ampoules are sampled from the reference batch. The samples are conditioned and shaken at 25 °C in a water bath for 6 hours. Measurements of each parameter is performed alternately on the reference ampoules and the sample ampoules and repeated 5 times on one analyzer platform. {4} d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected values for the representative analyzers are provided in the labeling for each specific lot. The labeling indicates each laboratory should establish its own acceptable ranges and use those provided only as a guide. 5 {5} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...