K150224 · Zibo Sanying Trade Co., Ltd. · LYZ · May 12, 2015 · General Hospital
Device Facts
Record ID
K150224
Device Name
BLUE VINYL GLOVES POWDER FREE
Applicant
Zibo Sanying Trade Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
May 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The BLUE VINYL EXAMINATION GLOVES POWDER FREE is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
Device Story
Disposable vinyl examination gloves; worn on examiner's hands or fingers; barrier protection to prevent cross-contamination between patient and examiner. Device is non-sterile; available in multiple sizes (S, M, L, XL). Used in clinical settings by healthcare professionals. Blue colorant added; biocompatibility verified via ISO 10993-10 testing to ensure no irritation or sensitization. Bench testing confirms compliance with ASTM standards for physical properties, freedom from holes, and residual powder limits.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included physical property evaluation (tensile strength, elongation), freedom from holes (ASTM D5151), residual powder content (ASTM D6124), and biocompatibility (ISO 10993-10 irritation and sensitization).
Technological Characteristics
Material: Vinyl (polyvinyl chloride). Standards: ASTM D5250-06, ASTM D5151-06, ASTM D6124-06, ISO 2859-1, ISO 10993-10. Form factor: Disposable, ambidextrous, powder-free, blue-colored gloves. Dimensions: Length 240mm; width 85-115mm; thickness 0.06-0.10mm.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines above them that resemble hair or a wave-like design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 12, 2015
ZIBO SANYING TRADE CO., LTD C/O Mr. Ray Wang Official Correspondent Beijing Believe Tech. Service Co., Ltd. 1-202, Build 3, Beijing New World, No. 5 Chaoyang Rd., Chaoyang District, Beijing, 100024 CHINA
Re: K150224 Trade/Device Name: Blue Vinyl Examination Gloves Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Vinyl Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: March 29, 2015 Received: April 1, 2015
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Section II Indications for Use
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K150224
Device Name
BLUE VINYL EXAMINATION GLOVES POWDER FREE
Indications for Use (Describe)
The BLUE VINYL EXAMINATION GLOVES POWDER FREE is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block; margin-right: 5px;">▢</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;">☑</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
EF
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
#### Section III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K150224
- 1. Date of Preparation: 2015/1/30
- 2. Sponsor
ZIBO SANYING TRADE CO., LTD XILV VILLAGE FANGZHEN , ZHANGDIAN DISTRICT, ZIBO , SHANDONG , 255000 , CHINA
Contact Person: Ms. Jing Li Tel: +86- 0533-3819144 Fax: +86-0533-3818127 Email: hclijing85@126.com
- Submission Correspondent 3. Mr. Ray Wang Beijing Believe Tech. Service Co., Ltd. Tel: +86-21-50313932 Fax: +86-21-68093116 Email: Ray.Wang@believe-med.com
- 4. Proposed Device Identification
Trade Name: BLUE VINYL EXAMINATION GLOVES POWDER FREE Device Name: Vinyl Patient Examination Gloves (Powder Free) Common Name: Patient Examination Gloves
Classification: I Product Code: LYZ Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital
{5}------------------------------------------------
Intended Use Statement:
The BLUE VINYL EXAMINATION GLOVES POWDER FREE is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
- 5. Predicate Device Identification
510(k) Number: k142508 Product Name: Powder Free Yellow Vinyl Examination Gloves Manufacturer: Zibo HUAQI TRADING Co., Ltd.
## 6. Device Description
The proposed device, BLUE VINYL EXAMINATION GLOVES POWDER FREE is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
The proposed is BLUE VINYL EXAMINATION GLOVES POWDER FREE, and includes variations of different size.
- 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ASTM D5250-06, Standard Specification for Poly(vinyl chloride) Gloves for Medical Application. ASTM D 5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.
ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
- Substantially Equivalent Comparison Conclusion 8.
| ITEM | Proposed Device | Predicate Device (k142508) | Remark |
|-------------------------------|-----------------|----------------------------|--------|
| BLUE VINYL EXAMINATION GLOVES | | Powder Free Yellow Vinyl | |
#### Table 1 General Comparison
{6}------------------------------------------------
| | POWDER FREE | Patient Examination Gloves | |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|
| Product Code | LYZ | LYZ | SE |
| Regulation No. | 21 CFR 880.6250 | 21 CFR 880.6250 | SE |
| Class | I | I | SE |
| Intended Use | The BLUE VINYL EXAMINATION<br>GLOVES POWDER FREE is a disposable<br>device intended for medical purposes that is<br>worn on the examiner's hands or finger to<br>prevent contamination between patient and<br>examiner. | The Powder Free Yellow Vinyl<br>Patient Examination Gloves is a disposable<br>device intended for medical purposes that is<br>worn on the examiner's hands or finger to<br>prevent contamination between patient and<br>examiner. | SE |
| Powdered or<br>Powered free | Powdered free | Powdered free | SE |
| Table 2 Device Dimensions Comparison | | |
|--------------------------------------|--|--|
|--------------------------------------|--|--|
| | Size | | | | | Tolerance | |
|--------------------------------------------------------------------------------------|-----------------------------------------------------------------|------------|-----|-----|-----|-----------|-----|
| | Designation | S | M | L | XL | | |
| | Proposed Device<br>BLUE VINYL EXAMINATION<br>GLOVES POWDER FREE | Length, mm | 240 | 240 | 240 | | 240 |
| Width, mm | | 85 | 95 | 105 | 115 | ±5 | |
| Thickness, mm: | | | | | | | |
| | Finger | 0.10 | | | | min | |
| | Palm | 0.08 | | | | min | |
| | Cuff | 0.06 | | | | min | |
| Predicate Device (k142508)<br>Powder Free Yellow Vinyl<br>Patient Examination Gloves | Designation | S | M | L | XL | | |
| | Length, mm | 240 | 240 | 240 | 240 | min | |
| | Width, mm | 85 | 95 | 105 | 115 | ±5 | |
| | Thickness, mm: | | | | | | |
| | Finger | 0.10 | | | | min | |
| | Palm | 0.08 | | | | min | |
| | Cuff | 0.06 | | | | min | |
| Remark | SE | | | | | | |
Table 3 Performance Comparison
| ITEM | Proposed Device | Predicate Device (k142508) | Remark | | |
|------------------------|------------------------|----------------------------|-------------|-------------|----|
| | BLUE VINYL EXAMINATION | Powder Free Yellow Vinyl | | | |
| | GLOVES POWDER FREE | Patient Examination Gloves | | | |
| Colorant | Blue | Yellow | Analysis | | |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 13 MPa, min | 13 MPa, min | SE |
| | | Ultimate<br>Elongation | 400 % min | 400 % min | SE |
| | After | Tensile | 13 MPa, min | 13 MPa, min | SE |
{7}------------------------------------------------
| | Aging | Strength | | | |
|--------------------|-------|------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------|----|
| | | Ultimate<br>Elongation | 400 % min | 400 % min | SE |
| | | | Comply with ASTM D5250 | Comply with ASTM D5250 | SE |
| Freedom from Holes | | | Be free from holes when tested in<br>accordance with ASTM D5151 | Be free from holes when tested<br>in accordance with ASTM<br>D5151 | SE |
| Powder Content | | | 0.50 mg per glove | Meet the requirements of ASTM<br>5250 | SE |
### Table 4 Safety Comparison
| ITEM | Proposed Device | Predicate Device (k142508) | Remark | |
|--------------------|----------------------------------------------|--------------------------------------------------------|--------------------------|----|
| | BLUE VINYL EXAMINATION<br>GLOVES POWDER FREE | Powder Free Yellow Vinyl<br>Patient Examination Gloves | | |
| Material | Vinyl | Vinyl | SE | |
| Biocompatibility | Irritation | Under the conditions of the study, not an irritant | Comply with ISO 10993-10 | SE |
| | Sensitization | Under conditions of the study, not a sensitizer | | |
| Label and Labeling | | Meet FDA's Requirements | Meet FDA's Requirements | SE |
The color of the proposed device is different from the color of the predicate device. This difference may potentially cause biocompatibility issues. In order to mitigate this risk, we conducted biocompatibility tests according to ISO 10993-10 and the test results demonstrated that the proposed device with blue colorant did not induce skin irritation and showed no significant evidence of causing skin sensitization.
The proposed device, BLUE VINYL EXAMINATION GLOVES POWDER FREE, is determined to be Substantially Equivalent (SE) to the predicate device in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.