The STM9000 is intended for stimulation of peripheral nerves for diagnostic purposes.
Device Story
STM9000 is a magnetic stimulator for non-invasive stimulation of neural and cortical tissue. Device consists of power electronics generating magnetic pulses delivered via external magnetic coils (circular or butterfly configurations). Operator controls stimulus sequences, frequency (up to 100 pps), and waveforms (biphasic/monophasic) via a graphic LCD, keyboard, and multifunction knob interface. Used in clinical settings for diagnostic peripheral nerve stimulation. Output is a magnetic pulse that induces electrical currents in tissue; healthcare providers use the resulting neural response to assess nerve function. Benefits include non-invasive diagnostic capability for neurological assessment.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993), electrical safety (IEC 60601-1, IEC 60601-1-4), electromagnetic compatibility (IEC 60601-1-2), and software verification and validation.
Technological Characteristics
Magnetic stimulator; power electronics drive external magnetic coils. Waveforms: monophasic and biphasic. Repetition rate: up to 100 pps. Interface: graphic LCD, keyboard, multifunction knob. Standards: IEC 60601-1, IEC 60601-1-4, IEC 60601-1-2, ISO 10993. Coils: circular and butterfly types. Software level of concern: Moderate.
Indications for Use
Indicated for stimulation of peripheral nerves for diagnostic purposes in patients requiring non-invasive neural tissue stimulation.
Regulatory Classification
Identification
An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol of a stylized caduceus, which is a staff entwined by two snakes. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the seal. The caduceus symbol is composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 19, 2015
EB Neuro, S.p.A. c/o Allison Scott Navigant Consulting, Inc. 9001 Wesleyan Road, Suite 200 Indianapolis IN 46268
Re: K150220
Trade/Device Name: STM9000 Basic, STM9000 Standard, STM9000 Fast, STM9000 Ultra-Fast Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF Dated: July 16, 2015 Received: July 20, 2015
Dear Ms. Scott,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and 8091); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Hoffmann -S
- for Carlos L. Peña, PhD, MS Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150220
Device Name
STM9000 Basic, STM9000 Standard, STM9000 Fast, STM9000 Ultra-Fast
Indications for Use (Describe)
The STM9000 is intended for stimulation of peripheral nerves for diagnostic purposes.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter Information l.
EB Neuro, S.p.A. Via Pietro Fanfani 97/a Florence, Italy 50127
| Contact Person: | Allison Scott |
|-----------------|----------------------------|
| | 317.228.8719 |
| | Allison.Scott@Navigant.com |
Date Prepared: July 17, 2015
807.92(a)(2)
#### II. Devices
| Trade Name: | STM9000 Basic, STM9000 Standard, STM9000 Fast, STM9000 Ultra-Fast |
|-------------------------|-------------------------------------------------------------------|
| Common Name: | Evoked Response Electrical Stimulator |
| Classification Name(s): | Evoked Response Electrical Stimulator (21 CFR 882.1870) |
| Product Code: | GWF |
| Regulatory Class: | Class II |
#### III. Predicate & Reference Device(s)
| Device | Owner | 510(k) | |
|-------------------------------------------------------------------------|-----------------------|---------|-----------|
| MagPro, Model R30 | Tonica Elektronik A/S | K061645 | Predicate |
| MagPro R30 incl. MagOption, MagPro<br>X100, MagPro X100 incl. MagOption | Tonica Elektronik A/S | K091940 | Reference |
These devices have not been subject to a design-related recall.
#### IV. Device Description
The STM 9000 are magnetic stimulators used for magnetic stimulation. Magnetic stimulation is a noninvasive technique for stimulating neural and cortical tissue. Application areas of magnetic stimulation are a subset of the application areas for current stimulation.
The STM 9000 is connected to a magnetic coil which transfers the magnetic stimulation to the tissue. The following coils can be used with the STM 9000 family of magnetic stimulators:
| Code | Description |
|-------------|----------------------|
| B9621086002 | Circular Coil 90 mm |
| B9621086001 | Circular Coil 70 mm |
| B9621086003 | Butterfly Coil 70 mm |
{4}------------------------------------------------
| B9621086004 | Butterfly Coil 70 mm air cooled |
|-------------|---------------------------------|
| B9621086005 | Butterfly Coil 80 mm, 120° |
| B9621086006 | Circular Coil 90 mm air cooled |
The STM 9000 consists of power electronics to generate the magnetic field in the magnetic coil. The STM 9000 is controlled via a simple user interface, enabling the operator to overview all functions, stimulus sequences, controls, status and measured data. The magnetic pulse is capable of biphasic and monophasic waveforms and the stimulate with a frequency up to 100 pulses per second (pps).
#### V. Intended Use
The STM9000 is intended for stimulation of peripheral nerves for diagnostic purposes.
# VI. Comparison of Technological Characteristics
The technological characteristics of STM9000 are similar to those of MagPro X100. The devices are magnetic stimulators used for magnetic stimulation as a non-invasive technique for stimulating neural tissue. All devices are capable of single stimulation mode and repetitive stimulation mode.
The subject and predicate devices are connected to magnetic coils which transfer the magnetic stimulation to the tissue. The operator controls the functions and status of the devices through an interface consisting of a graphic LCD, a keyboard and a multifunction knob, integrated in the stimulator. The mechanical, electrical, and magnetic parameters of the coils in this submission are very near to the parameters of the predicate coils.
The main differences between the subject device and predicate are:
- -In some configurations, the STM9000 has a lower maximum repetition rate and/or allowable energy level than the predicate configurations
{5}------------------------------------------------
# VII. Performance Data
# Biocompatibility Testing
The biocompatibility evaluation for the STM9000 was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- . Sensitization
- . Irritation
The STM9000 coils are considered a surface device contacting intact skin for a duration of less than 24 hours.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the STM9000. The system complies with the IEC 60601-1, IEC 60601-1-4 standards for safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator.
# VIII. Conclusion
The biocompatibility, electrical safety and EMC, and software verification and validation demonstrate that the STM9000 should perform as well as the predicate device in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.