A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
Device Story
GammaBeam 500 is a cobalt-60 teletherapy unit for total body irradiation (TBI). It delivers gamma radiation via a large fixed treatment field at extended source-to-skin distance. The device comprises a source head, main frame, base, and controls. Operated by healthcare professionals in radiation therapy clinics, it features redundant safety mechanisms including auxiliary source drawer locks, secondary/manual source return, emergency stops, and operational interlocks. The system uses a main controller and graphical user interface to manage treatment delivery. By providing a dedicated TBI platform with fixed field sizes, it reduces collision hazards and simplifies clinical workflow compared to rotational units. The device benefits patients by providing a stable, safe environment for large-field radiation therapy.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including mechanical validation, physics validation, simulated use testing, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and software verification/validation.
Technological Characteristics
Cobalt-60 radionuclide radiation therapy system. Components: source head, main frame, base, patient couch. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), EN 60601-2-11 (gamma beam therapy), EN 61217 (coordinates/movements). Control system: separate main controller and GUI with battery backup. Fixed-field treatment mode. Vertical-only patient table motion.
Indications for Use
Indicated for the treatment of cancer in patients requiring radiation therapy, administered by healthcare professionals in a clinical setting.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three figures facing to the right. The figures are connected and appear to be in motion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2015
Best Theratronics Limited % Mr. Mike de van der Schueren Quality & Regulatory Manager 413 March Road Ottawa, Ontario K2K 0E4 CANADA
Re: K150191
Trade/Device Name: GammaBeam 500 Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: IWB Dated: February 18, 2015 Received: February 19, 2015
Dear Mr. de van der Schueren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
for
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150191
Device Name GammaBeam 500
Indications for Use (Describe)
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| X Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Date Summary Prepared | January 23, 2015 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Submitted by | Best Theratronics<br>413 March Road<br>Ottawa, Ontario K2K 0E4<br>Canada<br>Tel. (613) 591-2100<br>Fax. (613) 592-5680 |
| Contact Person | Mr. Mike de van der Schueren<br>Quality & Regulatory Manager |
| Trade Name | GammaBeam 500 |
| Common Name | Cobalt Teletherapy Device |
| Classification Name | Radionuclide Radiation Therapy System |
| Legally Marketed Predicate Device | Theratron Phoenix (K863180) |
| Device Classification | Class II, 21 CFR 892.5750 |
| Product Code | IWB |
## DESCRIPTION OF DEVICE
The GammaBeam 500 is a radiotherapy treatment unit with a cobalt-60 radiation delivery system that is designed to be a dedicated, standalone device for performing total body irradiation (TBI). The device consists of a source head, main frame, base, and controls. The design of this device is substantially equivalent to the predicate device Theratron Phoenix.
The GammaBeam 500 is aimed to deliver total body irradiation to patients by generating a large fixed radiation treatment field at an extended Source - to - Skin Distance (SSD). This is a common radiotherapy technique performed by many clinics worldwide. A dedicated device improves the safety, effectiveness and clinical workflow of these treatments.
The unit will be operated in a similar manner as the predicate and has the same redundant mechanisms for safety, auxiliary source drawer lock, secondary and manual source return, emergency stop switches on the unit and control console, source position indicator, and operational and inhibit state interlocks.
### INTENDED USE OF DEVICE
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
{4}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The GammaBeam 500 is substantially equivalent to the predicate device (K863180).
This GammaBeam 500 includes several changes to the design of Phoenix teletherapy unit.
| Device Characteristic | Predicate Device | Proposed Device |
|-----------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Control System | Single computer in control room | Separate main controller and graphic user<br>interface that performs same functionality<br>with added safety features, improved<br>useability and battery backup |
| Treatment Mode | Rotational and/or fixed treatments with<br>a variable field size | Fixed treatments only and fixed field size<br>reduces complexity and eliminates collision<br>hazards and hazards related to treatment<br>setup |
| Patient Table | Vertical, longitudinal, lateral and<br>isocentric positioning of the patient | Optional accessory - Vertical motion only for<br>ease of lowering patient to the treatment<br>height from a comfortable seated position |
Usability is being improved by adding functions relevant to contemporary cancer treatment techniques and removing those that are not.
The major components of the Theratron Phoenix, including the head, source mechanism, main frame and base have had minor modifications to accommodate the above changes.
There are no changes to the mechanical structure or radiological shielding of the head.
The irradiation source and radioactivity of the cobalt-60 source remains unchanged as does the source drawer mechanism.
The control system has been designed to meet the same intended use as the current model.
### PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the GammaBeam 500, consisting of the main unit and patient couch. The system complies with the following standards for safety:
- 0 IEC 60601-1:2013 Edition 3.1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2007 Edition 3.0 Medical electrical equipment Part 1-2: General requirements 0 for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
- 0 EN 60601-2-11:2011 Edition 3.0 Medical electrical equipment -- Part 2: Particular requirements for the safety of gamma beam therapy equipment
- EN 61217 (2014) Radiotherapy equipment Coordinates, movements and scales 0
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in a serious injury or death to the patient or operator.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the words "Best" and "Theratronics" stacked on top of each other. The word "Best" is in a cursive font, while the word "Theratronics" is in a sans-serif font. The text is gray in color and appears to be a logo or brand name.
### Mechanical Testing
The performance of the device was tested against a set of functional specifications in an environment that simulated, as much as possible, the actual operating environment. This included:
- · Mechanical validation of the main unit and patient couch
- · Physics validation
- · Simulated use testing
#### CONCLUSIONS
The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the GammaBeam 500 performs as intended in the specified use conditions. The performance data demonstrates that the GammaBeam 500 is as safe and effective as the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.