K150187 · Cordis Corporation, A Johnson & Johnson Company · PDU · May 19, 2015 · Cardiovascular
Device Facts
Record ID
K150187
Device Name
ELITECROSS Support Catheter
Applicant
Cordis Corporation, A Johnson & Johnson Company
Product Code
PDU · Cardiovascular
Decision Date
May 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Cordis ELITECROSS™ Support Catheter is intended to facilitate the intraluminal placement of diagnostic/ interventional devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention, and to deliver saline or contrast medium. Additionally, ELITECROSS™ can be used as an accessory with the FRONTRUNNER® XP CTO Catheter.
Device Story
ELITECROSS Support Catheter is a single-use, torqueable, single-lumen catheter; used in peripheral vasculature to provide distal support for ancillary diagnostic/interventional devices (e.g., FRONTRUNNER XP CTO Catheter) and to deliver saline or contrast medium. Device features PTFE inner liner, stainless steel braid, polymer jacket, and distal hydrophilic coating. Radiopaque marker band allows fluoroscopic visualization. Operated by physicians in clinical settings. Facilitates guide wire placement across occlusions. Benefits include improved device delivery and navigation through stenotic lesions.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including catheter sheath introducer (CSI) compatibility, static burst testing, and flow rate performance testing, supported by risk analysis and verification/validation activities.
Indicated for patients requiring peripheral vascular intervention to facilitate intraluminal placement of diagnostic/interventional devices beyond stenotic lesions, including chronic total occlusions, and for delivery of saline or contrast medium.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them, possibly representing water or movement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2015
Cordis Corporation, a Johnson & Johnson Company Michelle Ragozzino Rodgers, Ph.D. Senior Regulatory Affairs Specialist 6500 Paseo Padre Parkway Fremont, California 94555
Re: K150187
Trade/Device Name: ELITECROSS Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: April 13, 2015 Received: April 14, 2015
Dear Dr. Rodgers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150187
Device Name ELITECROSS™ Support Catheter
### Indications for Use (Describe)
The Cordis ELITECROSS™ Support Catheter is intended to facilitate the intraluminal placement of diagnostic/ interventional devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention, and to deliver saline or contrast medium. Additionally, ELITECROSS™ can be used as an accessory with the FRONTRUNNER® XP CTO Catheter.
Type of Use (Select one or both, as applicable)
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|--------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
### I. SUBMITTER
Cordis Corporation 6500 Paseo Padre Parkway Fremont, CA 94555
Contact Person: Michelle Ragozzino Rodgers, Ph.D. Tel: (510) 248-2450 Fax: (510) 248-2533
Date Prepared: January 23, 2015
### II. DEVICE
Name of Device: ELITECROSS™ Support Catheter Common Name: Percutaneous catheter Classification Name: Catheter for Crossing Total Occlusions (21 CFR §870.1250) Regulatory Class: Class II Product Code: PDU
### III. PREDICATE DEVICE
Primary Predicate: Cordis Micro Guide Catheter ELITE, previously cleared on 11/14/2014 under K140438. Regulation Number: 21 CFR §870.1250. Product Code: PDU This predicate has not been the subject of a recall.
Reference Device: Cook CXI Support Catheter (K122796) Regulation Number: 21 CFR §870.1210 Product Code: KRA
#### DEVICE DESCRIPTION IV.
The ELITECROSS™ Support Catheter is a single-use, 5F sheath-compatible sterile catheter designed to provide additional support to the distal portion of ancillary diagnostic/interventional devices, including the FRONTRUNNER® XP CTO Catheter, as well as to deliver saline or contrast medium. After crossing the occlusion with the ancillary device, the ELITECROSS™ may be used to facilitate placing a guide wire across the occlusion.
The ELITECROSS™ Support Catheter device is a single lumen torqueable tube containing a full-length PTFE inner liner that is surrounded by a stainless steel braid, which is further encompassed by a polymer jacket, and features a final external hydrophilic coating on the distal 40 cm of the catheter body. The proximal end utilizes a molded hub with a luer fitting for flushing, with winged tabs designed to facilitate maneuvering and torqueing in the vasculature, while the distal tip contains a radiopaque marker band for visibility under fluoroscopy. The ELITECROSS™ Support Catheter will be available in various configurations and several
{4}------------------------------------------------
lengths. The ELITECROSS™ Support Catheter is provided sterile (by EO) and intended for single use only.
# V. INDICATIONS FOR USE
The Cordis ELITECROSS™ Support Catheter is intended to facilitate the intraluminal placement of diagnostic/interventional devices beyond stenotic lesions (including chronic total occlusions) in the peripheral vasculature prior to further percutaneous intervention, and to deliver saline or contrast medium. Additionally, ELITECROSS™ can be used as an accessory with the FRONTRUNNER® XP CTO Catheter.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject ELITECROSS™ Support Catheter device is identical to the predicate Micro Guide Catheter ELITE with respect to intended use, design, materials, dimensions, configurations, operating principle, control mechanism, and site of manufacture of the device. The subject and predicate devices are based on the following same technological elements:
- Catheter-Facilitates placement of and provides support to ancillary diagnostic/interventional devices, including Frontrunner® XP CTO Catheter
- Single lumen torqueable tube containing a full-length PTFE inner liner that is surrounded 0 by a stainless steel braid, which is further encompassed by a polymer jacket
- Use of hydrophilic coating for distal lubricity to advance through lesion
- Use of marker bands and radiopaque materials for fluoroscopic visualization of ● catheter tip
- Use of a standard male luer fitting for flush port
- o EO sterilized, single use device
Relative to the predicate, changes are limited to the labeling of ELITECROSS™ Support Catheter, including the trade name, labeled sheath compatibility (with no dimensional changes to the device), and edits to the Indications for Use statement (with no change to the Intended Use of the device). Additional testing supported the labeling of ELITECROSS for 5F Catheter Sheath Introducer (CSI) compatibility, with no dimensional changes to the subject device relative to the predicate. The labeling of ELITECROSS for the delivery of saline or contrast was supported by verification and validation activities, risk analyses, and similarities to the cleared reference device. The labeling modifications do not raise new questions of safety and effectiveness.
## VII. PERFORMANCE DATA
## Device Dimensional and Functional Testing
- Catheter Sheath Introducer (CSI) Compatibility ●
- Static Burst Testing ●
- o Flow Rate Performance Testing
{5}------------------------------------------------
## VIII. CONCLUSIONS
The subject ELITECROSS Support™ Catheter has the same design and intended use as the predicate, Micro Guide Catheter ELITE. The labeling modifications to the ELITECROSS™ product family do not alter the fundamental scientific technology, operating principles, mechanism of action, or intended use of device. Design verification and validation testing demonstrate that the ELITECROSS™ Support Catheter meets applicable performance requirements. The labeling modifications do not raise new questions of safety and effectiveness. The ELITECROSS™ continues to meet all previous performance specifications for Micro Guide Catheter ELITE, and none of the critical clinical performance parameters have changed. ELITECROSS Support™ Catheter can be used according to its intended use and in an equivalent manner to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.