Conditioning of silica-based ceramic surfaces for the adhesive bond with luting composites
Device Story
Monobond Etch & Prime is a dental conditioning agent used extra-orally by dental technicians to prepare silica-based ceramic surfaces for bonding with luting composites. It functions as a single-step solution that combines etching and silanization, replacing the traditional separate step of etching with 5% hydrofluoric acid. The device contains silane, a ceramic etching agent, and adhesive monomers. By eliminating the need for separate hydrofluoric acid etching, the device simplifies the preparation process for silica glass ceramics. The healthcare provider applies the agent to the ceramic surface to enable subsequent bonding with luting composites (e.g., Variolink or Multilink lines). The device benefits the patient by ensuring a reliable adhesive bond for ceramic restorations.
Clinical Evidence
Bench testing only. Tensile bond strength testing was performed to compare the subject device to the predicate. Results demonstrated that the performance of Monobond Etch & Prime conforms to specifications and is substantially equivalent to the predicate. Biocompatibility was assessed according to EN ISO 10993-1:2009 and EN ISO 7405:2008.
Technological Characteristics
Liquid solution containing silane, ceramic etching agent, and adhesive monomers. Applied extra-orally. Tested per ISO 1641:2009. Biocompatibility tested per EN ISO 10993-1:2009 and EN ISO 7405:2008. No software or electronic components.
Indications for Use
Indicated for the conditioning of silica-based ceramic surfaces to facilitate adhesive bonding with luting composites. Limited to extra-oral application on silica-based ceramics; not indicated for oxide ceramics, metal, composite, or fiber-reinforced composites.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol consisting of three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 3, 2015
Ivoclar Vivadent, Inc. Ms. Donna Marie Hartnett, Esq. Director, QA/Regulatory Affairs 175 Pineview Drive Amherst, NY 14228
Re: K150164
Trade/Device Name: Monobond® Etch & Prime Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Codes: KLE Dated: January 21, 2015 Received: January 26, 2015
Dear Ms. Hartnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
### Page 2 - Ms. Donna Marie Hartnett
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tina
Kiang-S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150164
Device Name MONOBOND ETCH & PRIME
Indications for Use (Describe) Conditioning of silica-based ceramic surfaces for the adhesive bond with luting composites
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo consists of the company name in blue, with the word "Ivoclar" stacked on top of "Vivadent". Below the name is the slogan "passion vision innovation" in red. Above the company name is a curved line of green and blue squares and circles.
| Contact: | Donna Marie Hartnett |
|----------------------|----------------------------------------------------------------------------------------------|
| Company: | Ivoclar Vivadent, AG<br>Bendererstrasse 2, Schaan<br>FL-9494, Liechtenstein<br>+423-235-3535 |
| Date Prepared: | February 10, 2015 |
| Proprietary Name: | Monobond® Etch & Prime |
| Classification Name: | Agent, Tooth bonding, Resin (872.3200)<br>(Classification Code KLE) |
| Predicate Device: | Monobond Plus (K090826) |
Device Description: Monobond Etch & Prime is used to condition silica-based ceramic surfaces for the adhesive bond with luting composites from the Variolink or Multilink product lines, for example, Monobond Etch & Prime etches and silanizes silicate ceramic suffaces in one working step. The new product combines the role of bonding agent with that of etchant. Typically the dental technician would etch the ceramic with 5% hydrofluoric acid but this step is achieved using Monobond Etch & Prime. The etching step was carried out outside the mouth and the new product is also limited to extra-oral applications of the new product are limited to silica ceramics, Oxide ceramics, metal, composite and fibre-reinforced composites are not indicated. This is because only silica glass ceramics required etching with hydrofluoric acid and therefore the benefit of using Monobond Etch & Prime is primarily relevant for this group of materials.
### Intended Use:
Conditioning of silica-based ceramic surfaces for the adhesive bond with luting composites
Comparison to Predicate: The predicate device to which Monobond Etch & Prime has been compared is Monobond Plus (K090826). For this application, Monobond Etch & Prime has been compared to its predicate with regard to chemical composition, physical properties, and indications for use. The comparison shows that Monobond Etch & Prime is substantially equivalent to the predicate device.
{4}------------------------------------------------
# 510(K) SUMMARY (rev 2.10.15) Monobond® Etch & Prime
Image /page/4/Picture/1 description: The image is a logo for Ivoclar Vivadent. The logo features the company name in blue, with the word "Ivoclar" stacked on top of "Vivadent". Above the company name is an arc of green and blue dots. Below the company name is the phrase "passion vision innovation" in red.
| Predicate - Monobond Plus (K090826) | Subject Device |
|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bonding agent between luting composites and glass/oxide ceramic, metal, composite and fiber-reinforced composite restorations. | Conditioning of silica-based ceramic surfaces for the adhesive bond with luting composites. |
| Monobond Plus is a dilute solution of adhesive monomers and silane which enable it to be used with various substrates. | Monobond Etch & Prime is a dilute solution of silane, a ceramic etching agent and an adhesive monomer, and can be used only with silica ceramics |
| Tensile bond strength to glass ceramics was tested for comparison purposes to the subject device. | Tensile bond strength testing showed that the performance of the subject device is conforming to the specification and substantially equivalent to the predicate device. |
Technological Characteristics and Testing Summary: As no product specific standard exists for the subject device, testing was conducted under ISO 1641:2009 Dentistry - Medical Devices for dentistry - Materials.
The device was tested for Tensile bond strength. The predicate referred to shear bond strength. The methods and specifications cannot be directly compared. However, tests for Monobond Etch & Prime and Monobond Plus tensile bond strength showed that the performance of the new product is conforming to the specification and substantially equivalent to the predicate. Only the bond strength to glass ceramics is considered as use with oxide ceramics is not indicated for Monobond Etch & Prime. The key function of Monobond Etch & Prime is to silanise silica glass ceramics. This is achieved using exactly the same silane agent as used in Monobond Plus. Although the compositions are different, the subject device has been assessed for biocompatibility through testing according EN ISO 10993-1:2009 and EN ISO 7405:2008.
The device design, i.e. delivery form, and the predicate device are the same. They are both applied extra-orally. The Intended use of the subject and the predicate device are the same except that the subject device is limited for use with silica-based ceramic surfaces. This be because only silica glass ceramics require etching with hydrofluoric acid and therefore the benefit of using the subject device is relevant for this group of materials.
CONCLUSION: The above data and analysis demonstrates that Monobond Etch & Prime is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.