SpiroThor spirometer is a freestanding laboratory instrument for performing basic lung function tests in adults and children over the age of four years. It is intended to be used by physicians or professional medical personnel for testing in physicians' offices, industrial medical, and hospital settings.
Device Story
Hand-held portable diagnostic spirometer; measures patient breath flow and volume via ultrasonic flowmeter sensor. Device consists of main unit and pulmonary function filter. Ultrasonic technology (WaveFront™) measures bidirectional flow; algorithms calculate lung function parameters (e.g., FVC, FEV1, PEF). Data displayed on LCD; supports USB/Bluetooth connectivity. Used by physicians/medical personnel in clinics/hospitals. Output provides diagnostic lung function metrics to assist clinical decision-making. Benefits include portable, objective assessment of respiratory function.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by bench testing, including ATS 24 standard waveforms, ISO 26782:2009 compliance, flow resistance testing, and electrical safety/EMC testing (EN 60601-1, EN 60601-1-2, EN 60601-1-6).
Technological Characteristics
Ultrasonic flow measurement (WaveFront™ technology); semiconductor temperature sensor (-25°C to +85°C). Dimensions: 85x120x35 mm; weight: 300g. Power: 3.7V Li-Ion battery. Connectivity: USB, Bluetooth. Materials: Identical to predicate, compliant with ISO 10993-1:2009 for limited duration (<24h) mucosal/gas pathway contact. Software: Embedded algorithms for flow/volume calculation.
Indications for Use
Indicated for adults and children over age four requiring basic lung function testing in clinical, industrial, or hospital settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stylized to resemble a bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 27, 2015
Cohero Health LLC c/o Mr. Paul Dryden Consultant 335 Madison Avenue, 3rd Floor New York, New York 10017
Re: K150137
Trade/Device Name: SpiroThor Regulation Number: 21 CFR 868.1840 Regulation Name: Spirometer, diagnostic Regulatory Class: II Product Code: BZG Dated: April 21, 2015 Received: April 22, 2015
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
## K150137
Device Name
## SpiroThor
Indications for Use (Describe)
SpiroThor spirometer is a freestanding laboratory instrument for performing basic lung function tests in adults and children over the age of four years.
It is intended to be used by physicians or professional medical personnel for testing in physicians' offices, industrial medical, and hospital settings.
Type of Use (Select one or both, as applicable)
| XX Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Page 1 of 1
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 5
| Date Prepared: | 26-May-2015 |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Cohero Health LLC<br>335 Madison Ave., 3rd Floor<br>New York, NY 10017 | Phone - 917.363.5408<br>Fax - 212.423.2998 |
| Official Contact: | Melissa Manice, Ph.D., CEO |
| Proprietary or Trade Name: | SpiroThor |
| Common/Usual Name: | Spirometer |
| Classification Name: | 21CFR 868.1840<br>73 BZG – Spirometer, Diagnostic<br>Class II |
| Predicate Devices: | K092813 – AstraSonic Spirometer<br>K051712 – Air Safety Model 2800 PFT Filter |
## Device Description:
The Cohero Health SpiroThor Spirometer is a hand-held portable diagnostic spirometer for the measurement of patient breath flow and volume. The proposed device consists of a compact main unit and an ultrasonic flowmeter sensor used with a pulmonary function filter. The device uses an ultrasonic sensor that measures flow. Algorithms are used to determine values based on this flow measurement. Tabular and graphical data are displayed on the spirometer LCD display.
We note that the proposed SpiroThor is identical to the AstraSonic Spirometer cleared under K092813 with these differences, name change and that another cleared bacterial filter may be used with the device.
#### Indications for Use:
SpiroThor spirometer is a freestanding laboratory instrument for performing basic lung function tests in adults and children over the age of four years.
It is intended to be used by physicians or professional medical personnel for testing in Physicians' offices, industrial medical, and hospital settings.
## Substantial Equivalence Discussion:
The proposed device is substantially equivalent to the predicate device, manufactured by Thor Medical for SDI Diagnostics and cleared under the SDI name K092813 as the AstraSonic spirometer except as noted above, i.e. name change and another bacterial filter may be used.
Tables 1 and 2 compare the key features of the proposed SpiroThor with the identified predicate and demonstrate that the proposed device is substantially equivalent.
{4}------------------------------------------------
# 510(k) Summary Page 2 of 5
26-May-2015
## Table 1 – Comparison to Predicate
| Specification | K092813 - SDI AstraSonic<br>Diagnostic Spirometer | Proposed Device<br>SpiroThor |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Indicationsfor Use | The SDI AstraSonic Spirometer<br>is a freestanding laboratory<br>instrument for performing basic<br>lung function tests. | The SpiroThor Spirometer is a<br>freestanding laboratory<br>instrument for performing basic<br>lung function tests. |
| Environment of Use | Physicians' offices, industrial<br>medical and hospital settings | Physicians' offices, industrial<br>medical and hospital settings |
| Patient Population | Adults and children over the age<br>of four years | Adults and children over the age<br>of four years |
| Temperature Sensor | Semiconductor<br>(-25 °C - +85 °C) | Semiconductor<br>(-25 °C - +85 °C) |
| Display | Graphic Quarter VGA (320x240<br>pixels), 262K colors | Graphic Quarter VGA (320x240<br>pixels), 262K colors |
| Keyboard | 20-key keyboard | 20-key keyboard |
| Connection Type | USB, BlueTooth | USB, BlueTooth |
| Flow Tube<br>Dimensions | Ø30 mm x 150 mm | Ø30 mm x 150 mm |
| Power Supply | Internal 3.7 V Li-Ion battery<br>(Rechargeable via 5V 500mA<br>mini USB charger) | Internal 3.7 V Li-Ion battery<br>(Rechargeable via 5V 500mA<br>mini USB charger) |
| Dimensions | 85x120x35 mm<br>(Flow tube: 26 mm(ID)x150 mm) | 85x120x35 mm<br>(Flow tube: 26 mm(ID)x150 mm) |
| Weight | 300g | 300g |
| Flow/Volume | Bidirectional, ultrasound | Bidirectional, ultrasound |
| Measurement System | Ultrasonic technology (referred to<br>as "WaveFront™ technology" | Ultrasonic technology (referred<br>to as "WaveFront™ technology" |
| Volume Range | ± 20 L (BTPS) | ± 20 L (BTPS) |
| Flow Range | ± 18 L/s (BTPS) | ± 18 L/s (BTPS) |
| Volume Accuracy | ± 3% or 50 mL | ± 3% or 50 mL |
| Flow Accuracy | ± 3% or 20 mL/s | ± 3% or 20 mL/s |
| Dynamic<br>Resistance at 14 L/s | < 1.1 cmH2O/L/s | < 1.1 cmH2O/L/s |
| Disposable in-line<br>filter | AstraGuard - K062913 | Air Safety - K051712 |
| Electrical<br>Protection | Internal battery power | Internal battery power |
| Printer | External printer via RS232 or<br>BlueTooth | External printer via RS232 or<br>BlueTooth |
{5}------------------------------------------------
## 510(k) Summary Page 3 of 5 26-May-2015
#### Table 2 - Comparison to Predicate
| | Predicate<br>K092813 | SpiroThor |
|---------------------------------------------------------|----------------------|-----------|
| FVC (Forced Vital Capacity) | Yes | Yes |
| FEV.5 (Volume Expired in the first half second of test) | No | No |
| FEV 1 (Volume Expired in the first second of test) | Yes | Yes |
| FEV3 (Volume Expired in the first three second of test) | Yes | Yes |
| FEV6 (Volume Expired in the first six second of test) | Yes | Yes |
| FEV1/FEV6 - Ratio | No | No |
| FEV.5/FVC - Ratio | No | No |
| FEV1/FVC - Ratio | Yes | Yes |
| FEV1/VC - Ratio | No | No |
| PEF (Peak Expiratory Flow) | Yes | Yes |
| FEF25 (Maximum flow at 25% of FVC) | Yes | Yes |
| FEF50 (Maximum flow at 50% of FVC) | Yes | Yes |
| FEF75 (Maximum flow at 75% of FVC) | Yes | Yes |
| FEF25-75 (Average flow between 25% and 75% of FVC) | Yes | Yes |
| FEF75-85 (Average flow between 75% and 85% of FVC) | No | No |
| FET25-75 (Time between 25% and 75% of FVC) | No | No |
| FET100 (Forced Expiratory Time) | Yes | Yes |
| FIVC (Forced Inspiratory Vital Capacity) | Yes | Yes |
| *FVC (Best FVC) | No | No |
| IVC (Inspiratory Slow Vital Capacity) | Yes | Yes |
| MVV (Maximum Voluntary Ventilation) | Yes | Yes |
| PIF (Peak Inspiratory Flow) | Yes | Yes |
| Te (Expiratory Time) | No | No |
| Ti (Inspiratory Time) | No | No |
| Ti/Tt (Ratio of Inspiratory Time to Total Time) | No | No |
| TV (Tidal Volume) | Yes | Yes |
| TV/Ti Ratio | No | No |
| VE (Minute Ventilation at Rest) | No | No |
| Vext (Extrapolated Volume) | Yes | Yes |
| VC (Slow Vital Capacity) | Yes | Yes |
Differences:
The SpiroThor and predicate, K092813 – SDI AstraSonic Diagnostic Spirometer, are identical devices with minor differences. Namely, both products are produced by Thormed ("THOR") and the only differences are the name of the device and that another bacterial filter may be used with the proposed device. THOR produced the SDI AstraSonic that was cleared in 2009, under the SDI AstraSonic name K092813.
{6}------------------------------------------------
## 510(k) Summary Page 4 of 5 26-May-2015
There are no changes in technology, materials, design, performance, or any other difference that would render the device not substantially equivalent except as noted above.
In summary, one can conclude that substantial equivalence is met based upon the following:
## Indications for Use -
The indications for use are identical for the proposed device when compared to the predicate -K092813 - SDI AstraSonic Diagnostic Spirometer.
Discussion – Each device is indicated for use as a freestanding laboratory instrument for performing basic lung function tests in adults and children over the age of four years.
## Technology and construction -
The design, components, shape, size, etc. are identical to the predicate, K092813 - SDI AstraSonic Diagnostic Spirometer.
Discussion - The technology and construction are identical to the predicate, K092813 - SDI AstraSonic Diagnostic Spirometer. It is noted that the proposed SpiroThor uses a different bacterial filter than the predicate, but comparative testing demonstrated that there is no differences in the performance of the SpiroThor with either filter.
## Environment of Use -
The environments of use are identical to the predicate, K092813 - SDI AstraSonic Diagnostic Spirometer.
Discussion - physicians' offices, industrial medical and hospital settings.
## Patient Population -
The patient population of adults and children over 4 years old.
Discussion - The patient populations are equivalent to the predicate, K092813 - SDI AstraSonic Diagnostic Spirometer.
## Non-Clinical Testing Summary -
For K092813, THOR performed a number of bench tests.
These included:
- । American Thoracic Society (ATS) 24 standard waveforms
- -ISO 26782:2009
- Accuracy of waveform generator -
- -Flow resistance test
- EN 60601-1:2006/AC:2010 -
- EN 60601-1-2:2007/AC:2010 -
- -EN 60601-1-6:2010
- Performance to specifications with each bacterial filter -
The testing demonstrated that the SpiroThor performed as intended and met its performance specifications and was found to be substantially equivalent to the predicate K092813 - SDI AstraSonic Diagnostic Spirometer.
{7}------------------------------------------------
## 510(k) Summary Page 5 of 5 26-May-2015
#### Materials:
The materials for each component, patient contact and predicate for the material per G95-1 and ISO 10993-1:2009 would be considered as both:
- External Communicating (Indirect gas pathway),
- . Tissue contacting
- Duration of Use limited (< 24 hours) .
And
- Surface Contact
- Mucosal contacting
- Duration of Use limited (< 24 hours) ●
The following tests would be required if a material certification cannot be provided.
- Cytotoxicity ●
- Sensitization
- Intracutaneous / Irritation
The materials are identical to the predicates and we have provided a material certification.
## Clinical Testing -
No clinical testing was required or performed.
## Discussion of Differences -
There are no significant differences between the proposed device and the predicate except that the proposed device may utilize the Air Safety PFT filter and a name change, but we have demonstrated that there are no functional performance differences between the AstraGuard and Air Safety filters.
## Substantial Equivalence Conclusion -
All testing demonstrated that the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.