K150104 · Vital Images, Inc. · JAK · Apr 17, 2015 · Radiology
Device Facts
Record ID
K150104
Device Name
Vitrea CT Myocardial Perfusion
Applicant
Vital Images, Inc.
Product Code
JAK · Radiology
Decision Date
Apr 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Vitrea® CT Myocardial Perfusion is intended to assist a trained user for the visualization of hypo/hyper dense areas in patients with angina or with a previous myocardial infarction to assess the disease state and treatment. This software provides semi-automated heart and left ventricle segmentation and color polar maps of the myocardial tissue. The information provided is intended to be qualitative in nature and, when used by a qualified physician, may aid in the identification of myocardial enhancement defects and the follow up of such findings.
Device Story
Post-processing software for Vitrea platform; processes cardiac CT studies (DICOM). Inputs: ECG-gated contrast-enhanced cardiac CT volumes. Operation: semi-automated segmentation of heart and left ventricle; calculates Perfusion Index (PI) and Transmural Perfusion Ratio (TPR) based on Hounsfield attenuation (HU) values; generates color polar maps and 3D fusion overlays. Used in clinical settings by radiologists/cardiologists. Output: visual maps, quantitative ratios, and defect size measurements. Benefit: aids identification/follow-up of myocardial enhancement defects; supports assessment of cardiac disease/function.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including software verification (requirements coverage, RTM), risk management, and validation via phantom testing and simulated usability testing by experienced professionals.
Technological Characteristics
Post-processing software for cardiac CT; leverages existing Vitrea platform (MPR, MIP, volume rendering). Connectivity: DICOM standard compliant. Software lifecycle: IEC 62304:2006. Risk management: ISO 14971:2007. Features: semi-automated segmentation, color polar maps, PI/TPR ratio calculation, 3D fusion, defect size scoring.
Indications for Use
Indicated for patients with angina or previous myocardial infarction to assist trained users (radiologists/cardiologists) in visualizing hypo/hyper dense areas, assessing disease state/treatment, and identifying myocardial enhancement defects.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the bottom of the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2015
Vital Images, Inc % Parthiv Shah Sr. Regulatory Affairs Specialist 5850 Opus Parkway, Suite 300 MINNETONKA MN 55343-4414
Re: K150104
Trade/Device Name: Vitrea CT Myocardial Perfusion Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: II Product Code: JAK, LLZ Dated: January 19, 2015 Received: January 20, 2015
Dear Parthiv Shah,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150104
Device Name Vitrea® CT Myocardial Perfusion
### Indications for Use (Describe)
Vitrea® CT Myocardial Perfusion is intended to assist a trained user for the visualization of hypo/hyper dense areas in patients with angina or with a previous myocardial infarction to assess the disease state and treatment. This software provides semi-automated heart and left ventricle segmentation and color polar maps of the myocardial tissue.
The information provided is intended to be qualitative in nature and, when used by a qualified physician, may aid in the identification of myocardial enhancement defects and the follow up of such findings.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The dot on the "i" in "VITAL" is a circle.
# 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 C.F.R. Part 807.92(c)
| Basis for the<br>Submission: | Modifications to a legally marketed device which are not qualified for a<br>Special 510(k) notification |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Vital Images, Inc.<br>5850 Opus Parkway<br>Suite 300<br>Minnetonka, MN, 55343-4414 |
| Establishment<br>Registration: | 2134213 |
| Contact Person: | Parthiv Shah<br>Sr. Regulatory Affairs Specialist<br>Phone : 952—487—9574<br>Fax: 952-487-9510<br>E-mail: pshah@vitalimages.com |
| 510(k) Type: | Traditional |
| Summary Date: | March 31, 2015 |
| Device Trade Name: | Vitrea® CT Myocardial Perfusion Software |
| Device Common<br>Name / Regulatory<br>Description: | An accessory of Computed Tomography X-ray System |
| Device<br>Classification<br>Name: | System, X-ray, Tomography, Computed |
| Regulation Number: | 21 C.F.R. 892.1750 |
| Primary Product<br>Code: | JAK |
| Secondary Product<br>Code: | LLZ |
| Regulatory<br>Classification: | Class II |
| Device Panel: | Radiology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "VITAL" in black, with a red tagline underneath that reads "A Toshiba Medical Systems Group Company". The font is sans-serif and modern. The dot on the "i" in "VITAL" is a perfect circle.
## Predicate Devices:
| Predicate Device(s) | Predicate Type | Manufacturer | FDA 510(k)<br>number |
|-------------------------|----------------------------------|----------------------------------------|----------------------|
| MyoPerfusion, CSMP-001A | Primary<br>Predicate<br>Device | Toshiba Medical<br>Systems Corporation | K132523 |
| CT Myocardial Analysis | Secondary<br>Predicate<br>Device | Vital Images, Inc. | K112531 |
## Device Description:
Vitrea CT Myocardial Perfusion is a post-processing software option for the already cleared Vitrea software platform (K071331). It leverages the existing Vitrea platform functionalities such as Multi-Planar Reconstruction (MPR) images, Maximum Intensity Projections (MIP) and volume rendering.
Vitrea CT Myocardial Perfusion enables the visualization and analysis of perfusion defects in the myocardium. The software is intended for use with cardiac CT (Computed Tomography) studies to analyze cardiovascular anatomy and pathology and to assess the presence of hypo/hyper dense areas of myocardial tissue.
The software visualization tools provide semi-automated heart and Left Ventricle (LV) segmentation, and color overlay and polar maps of the myocardial tissue based on the Hounsfield attenuation (HU) values. The software displays the values associated with the generation of the Perfusion Index (PI) and Transmural Perfusion Ratio (TPR) maps. Perfusion Index (PI) is the ratio of the Mean Myocardial CT value to the LV blood pool CT value. Transmural Perfusion Ratio (TPR) is provided as a ratio per sector of the Endocardial CT value to the mean Epicardial CT value. The defect size scoring tool allows a user to delineate one or more contiguous 3D regions within the myocardium for independent size measurements. The user may observe the interpolated result as they are constructing the defect.
The software analysis tools include measurements and comparison ratios. The CT Myocardial Perfusion application allows users to load one or two volumes. In dual-volume cases, the volumes are displayed based on time. It also includes reporting tools for formatting findings and user selected areas of interest.
### Intended Use / Indications for Use:
Vitrea® CT Myocardial Perfusion is intended to assist a trained user for the visualization of hypo/hyper dense areas in patients with angina or with a previous myocardial infarction to assess the disease state and treatment. This software provides semi-automated heart and left ventricle segmentation and color polar maps of the myocardial tissue.
The information provided is intended to be qualitative in nature and, when used by a qualified physician, may aid in the identification of myocardial enhancement defects and the follow up of such findings.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The word "VITAL" is the main focus of the image.
# Intended for Disease / Condition / Patient Population:
The software provides Radiologists and Cardiologists with a robust dedicated suite of software tools to aid in the creation of evidence to support these physicians with their assessment of cardiac disease and functions for patients with angina or those who have had a previous myocardial infarction.
## Substantial Equivalence Comparison:
#### . Regulatory Comparison
| Characteristic | Vitrea CT<br>Myocardial<br>Perfusion<br>software<br>(Submission<br>Subject) | MyoPerfusion,<br>CSMP-001A<br>(K132523)<br>(Predicate<br>Device-1)<br><br>Primary<br>Predicate Device | CT Myocardial<br>Analysis<br>(K112531)<br>(Predicate<br>Device-2)<br><br>Secondary<br>Predicate Device | Noted Differences |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Classification<br>Name | Computed<br>tomography<br>x-ray System | Computed<br>tomography<br>x-ray System | Picture Archiving<br>and<br>Communications<br>System | None<br>Note: The subject<br>device has both JAK<br>and LLZ product<br>codes. |
| Regulatory<br>Number | 892.1750 | 892.1750 | 892.2050 | None<br>Note: The subject<br>device has both JAK<br>and LLZ product<br>codes. |
| Product Code | Primary: JAK<br>Secondary:<br>LLZ | JAK | LLZ | None |
| Classification | Class II | Class II | Class II | None |
| Review Panel | Radiology | Radiology | Radiology | None |
| Decision Date | Under Review | December 12,<br>2013 | November 18,<br>2011 | Both predicates are<br>cleared. |
| Characteristic | Vitrea CT Myocardial<br>Perfusion software<br>(Submission Subject) | MyoPerfusion,<br>CSMP-001A (K132523)<br>(Predicate Device-1) | Noted Differences | |
| Intended Use | Vitrea® CT Myocardial<br>Perfusion is intended<br>to assist a trained user<br>for the visualization of<br>hypo/hyper dense<br>areas in patients with<br>angina or with a<br>previous myocardial<br>infarction to assess<br>the disease state and<br>treatment. This<br>software provides<br>semi-automated heart<br>and left ventricle<br>segmentation and<br>color polar maps of<br>the myocardial tissue.<br>The information<br>provided is intended to<br>be qualitative in nature<br>and, when used by a<br>qualified physician,<br>may aid in the<br>identification of<br>myocardial<br>enhancement defects<br>and the follow up of<br>such findings. | This software is<br>intended to be used for<br>the visualization of non-<br>reversible perfusion<br>defects (hypo/hyper<br>dense areas) in<br>patients with angina or<br>with a previous<br>myocardial infarct.<br>Included software tools<br>may aid a trained user<br>in monitoring the<br>disease state and<br>treatment over time.<br>This software provides<br>maps and the values<br>used to generate the<br>maps.<br>The information<br>provided is intended to<br>be qualitative in nature<br>and when used by a<br>qualified physician may<br>aid in the identification<br>of myocardial<br>enhancement defects<br>and the follow up of<br>such findings. | None | |
| Intended<br>Users | Radiologists and<br>Cardiologists | Radiologists and<br>Cardiologists | None | |
| Patient<br>Population | Patients with angina or<br>those who have had a<br>previous myocardial<br>infarction | Patients with angina or<br>those who have had a<br>previous myocardial<br>infarction | None | |
| Modality<br>Support | CT | CT | None | |
| Characteristic | Vitrea CT Myocardial<br>Perfusion software<br>(Submission Subject) | CT Myocardial Analysis<br>(K112531)<br>(Predicate Device-2) | Noted Differences | |
| Intended Use | Vitrea® CT Myocardial<br>Perfusion is intended<br>to assist a trained user<br>for the visualization of<br>hypo/hyper dense<br>areas in patients with<br>angina or with a<br>previous myocardial<br>infarction to assess<br>the disease state and<br>treatment. This<br>software provides<br>semi-automated heart<br>and left ventricle<br>segmentation and<br>color polar maps of<br>the myocardial tissue.<br>The information<br>provided is intended to<br>be qualitative in nature<br>and, when used by a<br>qualified physician,<br>may aid in the<br>identification of<br>myocardial<br>enhancement defects<br>and the follow up of<br>such findings. | Vitrea® CT Myocardial<br>Analysis is an image<br>analysis software<br>application for cardiac<br>Computer Tomography<br>(CT) studies to<br>visualize cardiovascular<br>anatomy and pathology<br>and to highlight and<br>color code the<br>presence of hypo/hyper<br>dense areas of<br>myocardial tissue. | None<br>The language of Intended<br>Use has been updated to<br>be similar with the 510(k)<br>cleared predicate products<br>available in the US market. | |
| Intended<br>Users | Radiologists and<br>Cardiologists | Radiologists and<br>Cardiologists | None | |
| Patient<br>Population | Patients with angina or<br>those who have had a<br>previous myocardial<br>infarction | Patients with angina or<br>those who have had a<br>previous myocardial<br>infarction | None | |
| Modality<br>Support | CT | CT | None | |
| Characteristic | Vitrea CT Myocardial<br>Perfusion software<br>(Submission Subject) | MyoPerfusion,<br>CSMP-001A (K132523)<br>(Predicate Device-1) | Noted Differences | |
| Device<br>Description | Vitrea CT Myocardial<br>Perfusion is a post-<br>processing software<br>option for the already<br>cleared Vitrea<br>software platform<br>(K071331).<br>It leverages existing<br>Vitrea® functionality<br>such as Multi-Planar<br>Reconstruction (MPR)<br>images, Maximum<br>Intensity Projections<br>(MIP) and volume<br>rendering.<br>Vitrea CT Myocardial<br>Perfusion enables the<br>visualization and<br>analysis of perfusion<br>deficits in the<br>myocardium. The<br>software is intended<br>for use with cardiac<br>CT (Computed<br>Tomography) studies<br>to analyze<br>cardiovascular<br>anatomy and<br>pathology and to<br>assess the presence<br>of hypo/hyper dense<br>areas of myocardial<br>tissue.<br>The software<br>visualization tools<br>provide semi-<br>automated heart and<br>left ventricle<br>segmentation, and<br>color overlay and polar<br>maps of the<br>myocardial tissue | This device processes<br>ECG-gated contrast<br>enhanced cardiac scan<br>data using MPR<br>generated images<br>according to the cardiac<br>axis.<br>The software generates<br>polar maps, perfusion<br>index (PI) map and<br>Transmural Perfusion<br>Ration (TPR) maps<br>based upon the<br>measured CT values of<br>the tissue within the<br>specified region of<br>interest.<br>The software displays<br>the values associated<br>with the generation of<br>the Perfusion Index and<br>TPR Maps.<br>PI is the ratio of the<br>Mean Myocardial CT<br>value to the LV blood<br>pool CT value.<br>TPR is provided as a<br>ratio per sector of the<br>Endocardial CT value<br>to the mean Epicardial<br>CT value. | None<br>Note:<br>The added key features,<br>except Defect Size Scoring<br>tool, are similar to those<br>which are available in the<br>510(k) cleared Predicate<br>Device-1. | |
| Characteristic | Vitrea CT Myocardial<br>Perfusion software<br>(Submission Subject) | MyoPerfusion,<br>CSMP-001A (K132523)<br>(Predicate Device-1) | Noted Differences | |
| | based on the<br>Hounsfield attenuation<br>(HU) values. | | | |
| | The software displays<br>the values associated<br>with the generation of<br>the Perfusion Index<br>(PI) and Transmural<br>Perfusion Ratio (TPR)<br>maps. | | | |
| | PI is the ratio of the<br>Mean Myocardial CT<br>value to the LV blood<br>pool CT value. | | | |
| | TPR is provided as a<br>ratio per sector of the<br>Endocardial CT value<br>to the mean Epicardial<br>CT value. | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.