K150089 · AngioDynamics, Inc. · OAB · Jun 18, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150089
Device Name
NanoKnife System
Applicant
AngioDynamics, Inc.
Product Code
OAB · General, Plastic Surgery
Decision Date
Jun 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NanoKnife System with six outputs is indicated for the surgical ablation of soft tissue.
Device Story
Software-controlled low-energy direct-current (LEDC) generator; delivers electrical pulses between pairs of probes placed in soft tissue; creates electroporation for ablation. Clinician-operated; parameters (volts/cm, pulse length, pulse count, timing mode) set by user; activated via foot-pedal. Supports up to six probes in an array. Used in surgical settings; requires visual confirmation of ablation via standard imaging (CT or ultrasound). Benefits patient through targeted soft tissue ablation.
Clinical Evidence
Bench testing only. Performance testing included ECG synchronization, power supply regulation, hardware safety measures (over/under current pulse intervention), pulse accuracy (amplitude, duration, frequency, interval, count), and maximum energy delivery. Reliability, vibration/handling, EMC (EN 60601-1-2), and electrical safety (IEC 60601-1) testing performed. Software validation completed. Additionally, Pennes bioheat model simulations using finite element analysis confirmed thermal ablation volumes remain within electroporation boundaries.
Technological Characteristics
LEDC generator; six-output architecture. Probes: single electrode, 15cm/25cm lengths. Energy: pulsed DC. Connectivity: foot-pedal activation. Standards: IEC 60601-1 3rd ed., EN 60601-1-2. Software: firmware-controlled, includes FPGA. Sterilization: identical to predicate. Features: ECG synchronization, 90PPM/240PPM timing modes.
Indications for Use
Indicated for the surgical ablation of soft tissue in patients requiring such procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a way that creates a sense of depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2015
Angiodynamics Incorporated Ms. Suzanne Goodman Senior Regulatory Affairs Manager 26 Forest Street Marlborough, Massachusetts 01752
Re: K150089
Trade/Device Name: NanoKnife System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: OAB Dated: May 18, 2015 Received: May 19, 2015
Dear Ms. Goodman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150089
Device Name NanoKnife System
#### Indications for Use (Describe)
The NanoKnife System with six outputs is indicated for the surgical ablation of soft tissue.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D) D Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K150089 Page 1 of 4
# 510(k) Summary - NanoKnife System
Date Prepared: 16 January, 2015
#### A. Sponsor
AngioDynamics, Inc. 603 Queensbury Avenue Queensbury Avenue Queensbury, NY 12804
#### B. Contact
Suzanne Goodman Senior Regulatory Affairs Manager AngioDynamics, Inc. 26 Forest Street Marlborough, MA 01752 508-658-7942
#### C. Device Name
Trade Name: NanoKnife System Common/Usual Name: Low energy direct current ablation device Classification Name: 21 CFR 878.4400, Electrosurgical Cutting and Coagulation Device and Accessories OAB Product Code: : D. Predicate Device(s) Trade Name: NanoKnife System (K102329)
Common/Usual Name: Classification Name:
Classification Panel:
Low energy direct current ablation device 21 CFR 878.4400, Electrosurgical Cutting and Coagulation Device and Accessories General and Plastic Surgery
#### E. Device Description
The NanoKnife System is a software-controlled low-energy direct-current (LEDC) generator which surgically ablates soft tissue. Included for use with the system are the NanoKnife Single Electrode Probes and optional Probe Spacer. With the NanoKnife System, electrical current is delivered between pairs of probes in a series of pulses. The waveform of the current is adjustable as determined by clinician-chosen parameters. These parameters include volts/cm, pulse length, number of pulses to be delivered between electrode pairs, distance between probes, and the timing mode (90PPM, 240PPM or ECG synchronization). Up to six probes may be placed in an array within the tissue. The array are matched as pairs by the system. When probes are activated via a foot-pedal, the scheduled current is delivered to tissue between subsequent pairs of probes. Soft tissue between the probes is ablated.
{4}------------------------------------------------
#### F. Intended Use and Indications for Use
The intended use of the NanoKnife System is identical to the intended use of the predicate device, the NanoKnife System subject of K102329:
"The NanoKnife System with six outputs is indicated for the surgical ablation of soft tissue. "
The NanoKnife System is used in the same manner as the predicate device, the NanoKnife System subject of K102329.
#### G. Summary of Similarities and Differences in Technological Characteristics and Performance between the Predicate and Modified Device
The NanoKnife Generator, subject of this 510(k) Premarket Notification, is a modification to the NanoKnife Generator submitted in K102329. The available ablation parameters, pulse amplitude, pulse length, maximum energy per pulse, and energy storage between recharges of the NanoKnife Generator are all identical to the predicate device. The design modifications to the NanoKnife Generator consist of hardware, firmware, software and labeling changes. Table 1 below provides a summary of these changes.
| Hardware and Specifications |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component changes due to availability of components and to improve<br>the electromagnetic (EMC)/electromagnetic (EMI) profile of the device<br>to comply with IEC 60601-1 3rd ed. standards. |
| Increased shelf life specification of Single Electrode Probes to 3 years<br>(no design modification or change to functionality) |
| Firmware |
| Programming and code changes to: |
| • eliminate system shutdowns upon detection of over-current<br>conditions |
| • debug programming (continuous device life-cycle improvement) |
| • enhance software diagnostics |
| Software |
| Programming and code changes to: |
| • remove non-English Languages |
| • disable radio frequency identification (RFID) functionality |
| • enable functions in administrator mode (not available to users) |
| • add clearer warning messages to the user when using 240 PPM and<br>90 PPM pulse modes |
| • debug programming (continuous device life-cycle improvement) |
| Labeling |
| System Manual Updates - reorganized and revised safety section,<br>including expanded discussion of High Current and Over-Current<br>conditions |
| User Manual - Removed discussion of RFID functions |
| User Manual - Add information per IEC 60601-1 3rd ed. |
| Pouch/Box Label - Administrative changes |
#### Table 1. Summary of Design Modifications
{5}------------------------------------------------
The NanoKnife Single Electrode Probes and optional Probe Spacer are identical to the NanoKnife Single Electrode Probes and Probe Spacer utilized by the predicate NanoKnife System. Two Single Electrode Probes (15cm and 25cm lengths) are offered with the modified device. The manual actuation, material, tip configuration, and sterilization method of the NanoKnife Single Electrode Probes are all identical to the predicate device.
Furthermore, there have been no changes made to the way ablation procedures will be performed using the NanoKnife System as compared to the predicate device. Users trained in the use of the NanoKnife System predicate device are instructed to confirm that the ablation was successful using standard imaging techniques. Like its predicate device, the NanoKnife System is a tool for surgically ablating soft tissue. Both the predicate device and the NanoKnife System require visual confirmation of ablation using standard imaging techniques (e.g., CT or ultrasound) following the procedure.
#### H. Performance Data
The modified NanoKnife System Generator was tested per the following to support a determination of substantial equivalence:
- Performance testing, including: ●
- ECG synchronization operation O
- Power supply regulation O
- AC/DC board voltages O
- Charge bank verification O
- Hardware safety measures (i.e., over/under current pulse intervention, overlength pulse, O Field Programmable Gate Array firmware functionality, etc.)
- Voltage and current accuracy O
- Pulse duration O
- Minimum capacitance O
- Pulse amplitude O
- Pulse frequency O
- Pulse intervals O
- O Pulse count
- Maximum energy delivered O
- Emergency Stop Button functionality o
- Ship testing environmental and vibration/handling ●
- Reliability testing
- EMC testing (EN 60601-1-2:2001/A1:2006) ●
- . Electrical safety testing (IEC 60601-1-2:2001/A1:2004)
- Software validation testing .
The test results verify that the NanoKnife System meets all the specified performance specifications and thus, is substantially equivalent to the predicate device.
Additionally, AngioDynamics performed a series of ablation and thermal zone calculations with physical parameters input to the Pennes bioheat model utilizing an industry standard differential equation and finite element analysis tool for physics and biological systems. The results of worst-case modeling indicate that potential thermal ablation volumes lie within the boundaries of electroporation volumes where clinically-relevant input parameters are utilized.
#### I. Conclusion
The modified NanoKnife System is substantially equivalent to the predicate NanoKnife System (K102329). The design modifications included in this 510(k) Premarket Notification do not affect the cleared intended use or indications for use. Nor do the design modifications alter the fundamental
{6}------------------------------------------------
technological characteristics of the predicate or raise new questions of safety and effectiveness. Where changes in technology occur, the results of performance testing confirm the modified NanoKnife System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.