The device is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate); Transesophagea1; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
Device Story
Vivid E80/E90/E95 is a Track 3 diagnostic ultrasound system for cardiac, vascular, and general radiology imaging. It utilizes electronic array transducers (linear, curved, sector/phased, matrix, pencil) to acquire ultrasound signals. The system processes these inputs to generate real-time 2D, 3D, and 4D images, Doppler flow analysis, and various functional measurements. Operated by physicians or qualified personnel in clinical settings, the system features a mobile console with a keyboard, touch panel, and high-resolution display (LCD/OLED). Output is visualized on the display for diagnostic assessment, supporting clinical decision-making through advanced features like Automated Functional Imaging (AFI), tissue tracking, and speckle reduction. Networking and DICOM capabilities allow for image storage and integration into hospital workflows. The device benefits patients by providing non-invasive, real-time diagnostic visualization of internal structures and blood flow.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output safety, biocompatibility, and electrical/mechanical safety verification.
Technological Characteristics
Mobile ultrasound console with keyboard, touch panel, and LCD/OLED display. Supports linear, curved, sector/phased, matrix, and pencil array transducers. Modes: B, M, PW/CW Doppler, Color Doppler, Color M, Power Doppler, Harmonic Imaging, Coded Pulse, and Real-time 3D/4D. Connectivity: DICOM. Standards: ANSI/AAMI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2/UD 3, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for ultrasound imaging and analysis of fetal/obstetrics, abdominal (renal, GYN/pelvic), pediatric, small organ (breast, testes, thyroid), neonatal/adult cephalic, cardiac (adult/pediatric), peripheral vascular, musculo-skeletal, urology (prostate), transesophageal, transrectal, transvaginal, and intraoperative (abdominal, thoracic, vascular) applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Vivid S60/S70 Diagnostic Ultrasound System (K142323)
L8-18i-D Diagnostic Ultrasound Transducer on Logiq E9 (K142160)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2015
GE Vingmed Ultrasound AS % Mr. Bryan Behn Regulatory Affairs Manager GE Healthcare 9900 W Innovation Drive, RP-2138 WAUWATOSA WI 53226
Re: K150087
Trade/Device Name: Vivid E90, Vivid E90 and Vivid E95 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 13, 2015 Received: January 15, 2015
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
## K150087
Device Name Vivid E80 and Vivid E90 and Vivid E95
Indications for Use (Describe)
The device is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate); Transesophagea1; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The color of the logo is blue, and the background is white. The logo is simple and recognizable.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | N | N | N | N | N | N | N | N | N | N | N | |
| Abdominal[1] | N | N | N | N | N | N | N | N | N | N | N | |
| Pediatric | N | N | N | N | N | N | N | N | N | N | N | |
| Small Organ[2] | N | N | N | N | N | N | N | N | N | N | N | |
| Neonatal Cephalic | N | N | N | N | N | N | N | N | N | N | N | |
| Adult Cephalic | N | N | N | N | N | N | N | N | N | N | N | |
| Cardiac[3] | N | N | N | N | N | N | N | N | N | N | N | |
| Peripheral Vascular | N | N | N | N | N | N | N | N | N | N | N | |
| Musculo-skeletal Conventional | N | N | N | N | N | N | N | N | N | N | N | |
| Musculo-skeletal Superficial | N | N | N | N | N | N | N | N | N | N | N | |
| Other[4] | N | N | N | N | N | N | N | N | N | N | N | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | N | N | N | N | N | N | N | N | N | N | N | |
| Transrectal | N | N | N | N | N | N | N | N | N | N | N | |
| Transvaginal | N | N | N | N | N | N | N | N | N | N | N | |
| Transuretheral | N | N | N | N | N | N | N | N | N | N | N | |
| Intraoperative[5] | N | N | N | N | N | N | N | N | N | N | N | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
System provides real-time 3D and 4D acquisition when used with special 4D probes.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 2 of 18
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a decorative border that resembles a stylized wave or ribbon pattern.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with 12S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation | | | | | | | | | | RT3D<br>Mode * | |
|----------------------------------------------------|-------------------|---|---|---|---|---|---|---|---|---|----------------|--|
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | | |
| Small Organ | | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[1] | P | P | P | P | P | P | P | P | P | P | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [1] Cardiac is Adult and Pediatric.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ♦] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 3 of 18
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and the circle has a decorative border. The logo is simple and recognizable, and it is associated with a well-known company.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with 4V-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac [2] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[3] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Cardiac is Adult and Pediatric.
[3] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ◆] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 4 of 18
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The color of the logo is a light blue.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with iC5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|--------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------------|---------|----------------|--------------|
| Clinical<br>Application<br>Anatomy/ Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined Harmonic<br>Modes | Imaging | Coded<br>Pulse | RT3D<br>Mode |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | |
| Abdominal | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | P | P | P | | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P | P | P | | P | P | P | P | | P | |
| Transvaginal | P | P | P | | P | P | P | P | | P | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [4] Other use includes Urology/Prostate;
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[*] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 5 of 18
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The script is white, creating a contrast against the blue background.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with 8C Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | P | P | P | P | P | | P | P |
| Pediatric | P | P | P | P | P | P | P | P | | P | P |
| Small Organ (specify) | P | P | P | P | P | P | P | P | | P | P |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | | P | P |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | | P | P |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [1] Abdominal includes GYN/Pelvic, Renal and Aorta-iliac artery;
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ◆] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 6 of 18
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a decorative border with a swirling pattern.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with C2-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | | |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | | |
| Small Organ (specify) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other[4] | P | P | P | | P | P | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [1] Abdominal includes GYN/Pelvic, Renal and Aorta-iliac artery;
[4] Other use includes Urology/Prostate;
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 7 of 18
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are white and have a flowing, interconnected design. The circle is surrounded by a decorative border, also in blue, that resembles a series of stylized waves or flourishes.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with 9L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|-------|--------------------|-------|------|--|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | | Combined Harmonic | Coded | RT3D | |
| Anatomy/Region of Interest | | | Doppler | Doppler | Doppler | Doppler | Doppler | Modes | Imaging | Pulse | Mode | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | | |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [2] Small organ includes breast, testes, thyroid.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ◆] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 8 of 18
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are white and have a swirling design around them. The logo is simple and recognizable, and it is often used to represent the company's brand.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with 11L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | | |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative[5] | P | P | P | | P | P | P | P | P | P | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [2] Small organ includes breast, testes, thyroid.
[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ♦] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 9 of 18
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a decorative border that resembles a stylized wave or ribbon pattern.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with M5Sc-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|---------------|
| Clinical<br>Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | P |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[4] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ◆ ] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please do not WRITE BELOW THIS LINE - Continue on another Page if needed)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 10 of 18
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are white and have a flowing, interconnected design. The circle is surrounded by a decorative border, also in blue, that resembles a series of stylized waves or flourishes.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with 6S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------------|---|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P | |
| Abdominal | P | P | P | P | P | P | P | P | P | P | P | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | P | P |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | P | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
P = previously cleared by FDA on Vivid E9 (K131514)
Notes: [2] Small organ includes breast, testes, thyroid.
[3] Cardiac is Adult and Pediatric.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ♦] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 11 of 18
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by a decorative, swirling pattern, also in blue. The logo is simple, recognizable, and associated with a well-known multinational corporation.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with C1-6-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|---------------|
| Clinical<br>Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | P |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | P |
| Pediatric | P | P | P | P | P | P | P | P | P | P | P |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | P |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | P | P | P | P | P | P | P | P | P | P | P |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
P = previously cleared by FDA on Vivid S60/S70 (K142323)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes, thyroid.
[4] Other use includes Urology/Prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD.
[ ◆] RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(Please do not WRITE BELOW THIS LINE - Continue on another Page if needed)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
Page 12 of 18
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by a decorative, swirling pattern, also in blue. The logo is simple, recognizable, and associated with a well-known multinational corporation.
#### Diagnostic Ultrasound Indications for Use Form GE Vivid E80 / Vivid E90 / Vivid E95 with L8-18i-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|-------|-----…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.