PowerDot(R) PD-01 Muscle Stimulator (with PowerDot(R) Mobile Application)
Applicant
Smartmissimo Technologies Pte, Ltd.
Product Code
NGX · Physical Medicine
Decision Date
Sep 30, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerDot PD-01 device, used with PowerDot Mobile Application, is intended for the stimulation of healthy muscles in order to improve or facilitate muscle performance. The PowerDot PD-01 device and PowerDot Mobile Application are not intended to be used in conjunction with therapy or treatment diseases of medical or medical conditions of any kind.
Device Story
PowerDot PD-01 is a battery-powered neuromuscular stimulator; stimulates healthy muscles to improve performance. Device uses Bluetooth Low Energy (4.0) to communicate with a mobile application on smartphones/tablets. User operates device via mobile app interface; app controls stimulation parameters (frequency, pulse width, duty cycle). Device delivers electrical pulses through hydrogel-based self-adhesive electrode pads applied to skin. Used in home or athletic training facilities by consumers. Provides regulated voltage output; includes safety features like automatic overload/no-load trips and shut-off. Healthcare provider/user monitors status via app. Benefits include muscle conditioning/training without treating medical conditions.
Clinical Evidence
No clinical data. Effectiveness based on literature review. Bench testing included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and output waveform analysis under various loads (500 Ω, 2 kΩ, 10 kΩ).
Indicated for stimulation of healthy muscles to improve or facilitate muscle performance. Contraindicated for individuals with cardiac pacemakers, implanted defibrillators, or other implanted metallic or electronic devices.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name arranged in a circular fashion around a central emblem. The emblem consists of a stylized caduceus, a symbol often associated with healthcare, formed by three intertwined figures in profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 30, 2015
Smartmissimo Technologies Pte Ltd Alexey Pisarev Chief Executive Officer #28-01, Sgx Centre II, 4 Shenton Way Singapore, 068807 SG
Re: K150078
Trade/Device Name: PowerDot Pd-01 Muscle Stimulator (with PowerDot Mobile Application) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: August 18, 2015 Received: August 31, 2015
Dear Alexey Pisarev:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150078
Device Name
PowerDot® PD-01 Muscle Stimulator (with PowerDot® Mobile Application)
Indications for Use (Describe)
The PowerDot PD-01 device, used with PowerDot Mobile Application, is intended for the stimulation of healthy muscles in order to improve or facilitate muscle performance.
The PowerDot PD-01 device and PowerDot Mobile Application are not intended to be used in conjunction with therapy or treatment diseases of medical or medical conditions of any kind.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: underline;">☐</span> Registration Use (Part 21 CFR 201.Subpart D) |
|-------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Over-The-Counter Use (21 CFR 201.Subpart C) |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary
This summary is being submitted in accordance with the reguirements of 21 CFR 807.92.
#### 5.1 Submitter
Smartmissimo Technologies Pte Ltd. 4 Shenton Way, #28-01 SGX Centre II, Singapore, 068807
Contact: Alexey Pisarev, Chief Executive Officer Tel. +65 83216256 Email: alexey@getpowerdot.com
#### 5.2 Date Prepared: Sept 29, 2015
#### 5.3 Device Name and Classification Information:
Trade/Proprietary Name: PowerDot® PD-01 Muscle Stimulator (with PowerDot® Mobile Application) Common Name: Smart Wearable Sports Muscle Stimulator Classification Name: Powered Muscle Stimulator for Muscle Conditioning Classification: 21 CFR 890.5850, Class II Product Code: NGX Panel: 89, Physical Medicine
#### 5.4 Predicate Device: Compex® Sport Plus (K083140)
#### 5.5 Device Description:
PowerDot PD-01 Muscle Stimulator is a battery-powered neuromuscular stimulator intended to stimulate healthy muscles in order to improve or facilitate muscle performance and, with that regard, may be considered a technique or method for muscle training.
PowerDot PD-01 device is designed to be used together with PowerDot Mobile Application.
PowerDot PD-01 device uses Bluetooth™ Low Energy (Bluetooth 4.0, Class II) wireless radio frequency protocol for communication with supported range of mobile devices (such as smartphones and/or tablets) via PowerDot Mobile Application.
Accessories include lead cables of 2 different lengths, USB charging cable, 2 types of hydrogel-based self-adhesive electrode pads and carrying case.
#### 5.6 Indications for Use:
The PowerDot PD-01 device, used with PowerDot Mobile Application, is intended for the stimulation of healthy muscles in order to improve or facilitate muscle performance. The PowerDot PD-01 device and PowerDot Mobile Application are not intended to be
{4}------------------------------------------------
used in conjunction with therapy or treatment diseases of medical conditions of any kind.
#### 5.7 Contraindications
PowerDot PD-01 device and PowerDot Mobile Application should not be used by anyone with a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death.
| Parameter | PowerDot PD-01 (with<br>PowerDot Mobile<br>Application) | Compex Sport Plus<br>(K083140) | Substantial<br>Equivalence |
|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use and Indications<br>For Use | The PowerDot PD-01 device<br>with PowerDot Mobile<br>Application is intended for<br>the stimulation of healthy<br>muscles in order to improve<br>or facilitate muscle<br>performance. The PowerDot<br>PD-01 device and PowerDot<br>Mobile Application are not<br>intended to be used in<br>conjunction with therapy or<br>treatment diseases of medical<br>or medical conditions of any<br>kind. | Intended for the<br>stimulation of healthy<br>muscles in order to<br>improve or facilitate<br>muscle performance.<br>Compex Sport is not<br>intended to be used in<br>conjunction with<br>therapy or treatment of<br>medical diseases or<br>medical conditions of<br>any kind. Usage of any<br>of Compex Sport<br>programs on injured or<br>ailing muscles is<br>contraindicated. | Same |
| Use environment | Over-the-counter (non-<br>prescriptive) use in athletic<br>training facilities or at home | Over-the-counter (non-<br>prescriptive) use in<br>athletic training<br>facilities or at home | Same |
| Anatomical Sites | Electrodes can be applied to<br>multiple anatomical sites. | Electrodes can be<br>applied to multiple<br>anatomical sites. | Same |
| Stimulation Parameters | | | |
| Power Source | Battery, Li-Po, rechargeable.<br>Not replaceable by user. | Battery, Ni-MH,<br>rechargeable | Substantially<br>equivalent. Both<br>devices are<br>powered with<br>rechargeable<br>batteries. |
| Method of Line Current<br>Isolation | N/A | N/A | |
| Patient Leakage Current | | | Unable to<br>compare.<br>Leakage current<br>of PowerDot<br>PD-01 is<br>extremely low<br>and meets IEC<br>60601-1 safety<br>criteria. |
| - Normal Condition | < 25 μΑ | Unknown | |
| - Single Fault<br>Condition | < 25 μΑ | Unknown | |
| Average DC current through<br>electrodes when device is on<br>but no pulses are being<br>applied | 0 μΑ | 0 μΑ | Same |
| Number of output modes | One output mode, but with | One output mode, but | Substantially |
| | varying stimulation<br>frequency, pulse width and<br>duty cycle ranges | with varying<br>stimulation frequency<br>and duty cycle ranges | equivalent |
| Number of output channels | 2 channels (or 4 channels<br>when two PowerDot PD-01<br>devices are used with one<br>PowerDot Mobile<br>Application application) | 4 channels | Substantially<br>equivalent. |
| Synchronous or alternating | Synchronous (with<br>asynchronous support) | Synchronous | |
| Method of channel isolation | Galvanic | Galvanic | Substantially<br>equivalent |
| Regulated current or<br>regulated voltage | Regulated voltage | Regulated current | Substantially<br>equivalent.<br>Difference is<br>regulation<br>technology is<br>not substantial<br>as both<br>technologies are<br>efficient for<br>power output<br>control. |
| Microprocessor controlled | Yes | Yes | Same |
| Automatic overload trip | Yes | Yes | Same |
| Automatic no-load trip | Yes | Yes | Same |
| Automatic shut-off | Yes | Yes | Same |
| User over-ride control | Yes | Yes | Same |
| Indicator display:<br>- On/Off status | Yes | Yes | Same |
| - Low battery | Yes (available from<br>PowerDot Mobile<br>Application) | Yes | Substantially<br>equivalent.<br>PowerDot and<br>Compex<br>indicators serve<br>the same |
| - Voltage/current<br>level | Yes (available from<br>PowerDot Mobile<br>Application) | Yes | purpose. |
| Timer range in minutes | 60 minutes maximum (for<br>the longest running<br>'Endurance' program<br>possible) | Maximum program: 60<br>minutes | Substantially<br>equivalent.<br><br>Both devices can<br>be used for<br>extended periods<br>of time. |
| Compliance with voluntary<br>standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10 | Substantially<br>equivalent |
| Compliance with 21 CFR 898 | Yes | Yes | Same |
| | | | equivalent.<br>Both devices are<br>small and<br>lightweight. |
| Dimensions | 2.4"x1.7"x0.5" | 5.6"x1.5"x3.9" | Substantially equivalent |
| Housing material and<br>construction | Plastic injection molding,<br>TPU | Plastic injection<br>molding | Substantially equivalent<br>PowerDot<br>housing<br>materials are<br>evaluated for<br>biocompatibility. |
#### 5.8 Brief Technical Comparison with the Predicate Device and Discussion of Differences
{5}------------------------------------------------
{6}------------------------------------------------
### 5.9 Non-clinical Tests
The following non-clinical testing was provided in this 510(k):
#### Biocompatibility Testing
Skin contacting hydrogel, which is used in PowerDot electrode pads, has been tested to ISO 10993-1:2009 standard.
TPU. ABS and PVC Compound materials, used in PowerDot PD-01 device housing and lead cables, were evaluated for biocompatibility based on Material Safety Data Sheets provided by the material vendors, and were concluded to be safe and biocompatible.
### Software Verification and Validation
PowerDot Firmware and PowerDot Mobile Application software documentation, consistent with a moderate level of concern, is provided with this 510(k).
System Validation Testing proves that all software requirement specifications were met and all software hazards were mitigated to Accepted risk level.
#### Electrical Safety and Electromagnetic Compatibility Testing
PowerDot PD-01 device together with its controlling PowerDot Mobile Application has been designed to comply and tested for compatibility with the applicable clauses of the following FDA-recognized standards:
- . IEC/EN 60601-1:2005 "Medical Electrical Equipment - Part 1: General Requirements for Safety"
- IEC 60601-2-10:2012 "Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. Edition 2.0"
{7}------------------------------------------------
- IEC/EN 60601-1-2:2014 "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements and tests. Edition 4.0"
- . IEC/EN 60601-1-11:2010 "Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. Edition 1.0"
#### Battery Testing
Lithium-Polymer battery, used in PowerDot PD-01 was tested by battery manufacturer for compliance with FDA-recognized UL 1642 Standard for Lithium Batteries (Cells).
#### Engineering Bench Testing
In addition to the full system validation testing, the 510(k) also included testing in accordance with the recommendations of FDA's "Draft Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Powered Muscle Stimulator for Muscle Conditioning", Attachment 11, Section 1 - Output Waveforms. Oscilloscope tracings were obtained of the device output waveforms under maximum supported voltage and pulse widths under loads of 500 Ω, 2 kΩ and 10 kΩ.
Also, a number of system validation testing scenarios covering mitigation of wireless risks in accordance with FDA's "Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and Food and Drug Administration Staff" were added to our full system testing protocol.
#### 5.10 Brief Description of clinical performance data
No applicable. This device does not diagnose, cure, mitigate, treat or prevent disease or affect the function of the human body.
Clinical effectiveness of the device was based on a literature review.
#### 5.11 Conclusion
Test results demonstrate the PowerDot PD-01 device with PowerDot Mobile Application is safe and effective for its intended use and the results support determination of substantial equivalent.
The usage of PowerDot Mobile Application as a wireless user interface for PowerDot PD-01 device as well as any other engineering differences do not: (1) affect the intended use (2) alter the fundamental scientific technology of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.