Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L
K150076 · Dong IL Technology , Ltd. · DZI · Jun 2, 2015 · Dental
Device Facts
Record ID
K150076
Device Name
Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L
Applicant
Dong IL Technology , Ltd.
Product Code
DZI · Dental
Decision Date
Jun 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied in dental use.
Device Story
Piezoelectric ultrasonic surgical units; transform electrical energy into 26±3kHz kinetic energy at surgical tip. Used in dental clinics by clinicians for bone cutting, grinding, drilling, and prophylaxis. Features include automatic frequency tuning, irrigation pump, and LED illumination. Normal mode for scaling; Boost mode for surgical procedures. Output controlled via touch LCD and foot switch. Provides mechanical ultrasound treatment to bone/tissue; facilitates precise osteotomy/osteoplasty. Benefits include controlled bone removal and surgical versatility. Tips are interchangeable, cleanable, and autoclavable.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (General safety) and IEC 60601-1-2 (EMC) standards. Performance testing verified output power, irrigation flow rates, and LED brightness levels against predicate specifications.
Technological Characteristics
Piezoelectric transducer; 26±3kHz operating frequency; automatic frequency tuning. Materials: PC/ABS enclosure, silicone hoses, PPSU handpiece, stainless steel tips with ZrN/Diamond coating. Class I, Type B electrical safety. Connectivity: Foot switch, touch LCD. Sterilization: Autoclavable tips. Power: 100-240V AC, 50/60Hz.
Indications for Use
Indicated for dental surgical procedures including osteotomy, osteoplasty, periodontal surgery, implantation, and tooth scaling applications.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2015
Dong Il Technology Ltd. c/o Mr. Peter Chung Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K150076
Trade/Device Name: Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: II Product Code: DZI Dated: April 29, 2015 Received: April 30, 2015
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K150076
#### Device Name
Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L
Indications for Use (Describe)
Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied in dental use.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K150076
#### 510(k) Summary[as required by 807.92(c)]
### A. Submission Information
- 1) Submission: 510(k) Submission
- 2) Submission Date: June 1, 2015
- 3) Submission Type: Traditional
- 4) 510(k) number: K150076
### B. Applicant:
- 1) Company name: DONG IL TECHNOLOGY LTD.
- 2) Address: 28, Namyang-ro, 930(gubaeksamsip)beon-gil, Bugyang-dong, Hwaseong-si, Gyeonggi-do, Republic of Korea
- 3) Tel: +82(31)229-5500
- 4) Fax: +82(31)357-2610
- 5) Homepage: http://www.dit-med.co.kr
- 6) Contact person: Peter Chung 412-687-3976
- 7) Contact person address: 300, Atwood ST, Pittsburgh, PA, 15213
#### C. Proprietary and Established Names:
- 1) Trade Name: Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L
- 2) Common Name: Bone cutting instrument and accessories
- 3) Regulation Name: Sonic surgical instrument and accessories/attachments
- 4) Regulation Number: 872.4120
- 5) Classification: Class 2
- 6) Panel: Dental
- 7) Product Code: DZI
- 8) Identify contact classification:
Category: External communicating device Contact: Tissue/bone/dentin Contact duration: Limited exposure (A)
#### D. legally marketed predicate devices:
Primary: Sonic Surgeon 300 (K110881) Reference: UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)
#### E. Device Description:
Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L are ultrasonic surgical units intended for mechanical ultrasound treatment in prophylaxis, periodontics. The normal mode is tooth scaling applications, and the Boost Mode of Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L can be used for surgical procedures, including osteotomy, osteoplasty, periodontal surgery and implantation.
The tips can easily be changed during the treatment and must also be cleaned and autoclaved. Materials for Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are as follows:
| Component | Material |
|--------------------------------------------|-------------------------------------------------|
| Enclosure (Generator Unit, Foot<br>Switch) | PC(Polycarbonate) |
| Hose, and seals | ABS(Acrylonitrile-Butadiene-Styrene Terpolymer) |
| Enclosure (Handpiece) | Silicone |
| Tip | PES(Polyethersulfone) |
| | PPSU(Polyphenylsulfone) |
{4}------------------------------------------------
### F. Intended use
Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied to use.
### G. Technological Characteristics:
Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L have the following features:
- . Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L transform generated ultrasonic (26±3kHz) energy to the kinetic energy and transmits it to the tip.
- . The generator of the Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L performs automatic tuning of the operating frequency, the efficiency of the piezoelectric transducer in the handpiece. This feature makes it possible to perform bone cutting, grinding and drilling action.
Our device is very similar to predicate devices, Sonic Surgeon 300 (K110881), UBS ultrasonic bone surgery and UDD ultrasonic debridement device(K080220), because Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L have the following equivalent characteristics: intended use, sterilization method, material used, electrical input power, frequency, and power output as predicate devices.
| Manufacturer | Model | Insulation<br>Class | Input Voltage | Input Power<br>Current or<br>Power | Frequency | Ultrasonic<br>Operation<br>Frequency |
|-------------------------------|-------------------------------------------------------------------------------|---------------------|-----------------------------|------------------------------------|------------|------------------------------------------|
| K150076 | | | | | | |
| DONG IL<br>TECHNOLOG<br>Y LTD | Sonic Surgeon<br>310L, 600L, 800L | Class I,<br>Type B | AC 100-240 V | AC Max 2A | 50 / 60 Hz | Automatic<br>scanning,<br>$26 \pm 3$ KHz |
| K110881 | | | | | | |
| DONG IL<br>TECHNOLOG<br>Y LTD | Sonic Surgeon<br>300 | Class I,<br>Type B | AC 95-115 V<br>AC 210-240 V | AC Max 1A | 50 / 60 Hz | Automatic<br>scanning,<br>$26 \pm 3$ KHz |
| K080220 | | | | | | |
| ITALIA<br>MEDICA<br>S.R.L | UBS ultrasonic<br>bone surgery and<br>UDD ultrasonic<br>debridement<br>device | Class I,<br>Type CF | AC 230V±10% | 140 VA | 50 / 60 Hz | $26$ kHz $\pm$<br>10% |
Summary Comparison table
| Manufacturer | Model | Indications for Use |
|------------------------------|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K150076 | | |
| DONG IL<br>TECHNOLOGY<br>LTD | Sonic Surgeon 310L<br>Sonic Surgeon 600L<br>Sonic Surgeon 800L | Sonic Surgeon 310L, Sonic Surgeon 600L, Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied in dental use. |
| K110881 | | |
| DONG IL<br>TECHNOLOGY<br>LTD | Sonic Surgeon 300 | Ultrasonic Surgical Unit is a piezoelectric device for bone surgery that enables osteotomy and osteoplasty techniques to be applied to in dental use. |
| K080220 | | |
| ITALIA MEDICA<br>S.R.L | UBS ultrasonic bone surgery and UDD ultrasonic<br>debridement | The UBS Ultrasonic Bone Surgery and the UDD Ultrasonic Debridement Device are bone cutting instruments intended for use in oral surgery. |
{5}------------------------------------------------
### (1) Comparison between proposed Device and predicate Sonic Surgeon 300(K110881)
| Predicate Device | | Proposed Device | | remark | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------|------------------------------------|---------|
| Model Name | Sonic Surgeon 300 | Sonic Surgeon 310L | Sonic Surgeon 600L | Sonic Surgeon 800L | |
| Features | | | | | |
| LED Light | X | O | O | O | |
| LCD | LCD Display | Touch LCD | Touch LCD | Touch LCD | |
| Input Voltage | AC 95 - 115V<br>(50/60Hz)<br>AC 210 – 240 V<br>(50/60Hz) | AC100V -<br>AC240V<br>(50/60Hz) | AC100V - AC240V<br>(50/60Hz) | AC100V - AC240V<br>(50/60Hz) | |
| Pump Cover | Automatic | Manual | Manual | Manual | |
| Power Supply | Toroidal Transformer | SMPS | SMPS | SMPS | |
| Foot Switch IP class | IPX1 | IPX8 | IPX8 | IPX8 | |
| Operating Frequency | 26 ± 3 kHz | 26 ± 3 kHz | 26 ± 3 kHz | 26 ± 3 kHz | |
| Other features are same as Predicate. | | | | | |
| Dimensions | | | | | |
| Hand piece | A type:<br>$Ø$ 20×154.2mm(L) | A type:<br>$Ø$ 20×154.2mm(L)<br>B type(LED):<br>$Ø$ 20×152.7mm(L) | C type(LED):<br>$Ø$ 22.5×155.4mm(L) | C type(LED):<br>$Ø$ 22.5× | |
| Generator | H:106mm<br>L:300mm<br>W:284mm | H:106mm<br>L:300mm<br>W:298mm | H:106mm<br>L:300mm<br>W:298mm | H:106m<br>m<br>L:300m | |
| Foot Switch | H: 37 mm<br>L: 140 mm<br>W: 240 mm | H: 37 mm<br>L: 140 mm<br>W: 240 mm | H: 37 mm<br>L: 140 mm<br>W: 240 mm | H: 37 mm<br>L: 140 mm<br>W: 240 mm | |
| Tip Holder | H: 37.9 mm<br>L: 41 mm<br>W: 58 mm | H: 38 mm<br>L: 41 mm<br>W: 58 mm | H: 38 mm<br>L: 41 mm<br>W: 58 mm | H: 38 mm<br>L: 41 mm<br>W: 58 mm | |
| Torque Wrench | $Ø$ 40×31mm(L) | $Ø$ 40×31mm(L) | $Ø$ 40×31mm(L) | $Ø$ 40×31mm(L) | |
| Wire Water Pack | $Ø$ 5×477mm(L) | $Ø$ 5×470mm(L) | $Ø$ 5×470mm(L) | $Ø$ 5×470mm(L) | |
| Storage tray-not for sterilizer use | H: 61 mm<br>L: 180 mm<br>W: 200 mm | H: 61 mm<br>L: 180 mm<br>W: 200 mm | H: 61 mm<br>L: 180 mm<br>W: 200 mm | H: 61 mm<br>L: 180 mm<br>W: 200 mm | |
| Wire Hand piece | H: 128 mm<br>L: 66 mm<br>W: 83 mm | H: 128 mm<br>L: 66 mm<br>W: 83 mm | H: 128 mm<br>L: 66 mm<br>W: 83 mm | H: 128 mm<br>L: 66 mm<br>W: 83 mm | |
| Tips | | Refer to Table 4 | | | Table.1 |
| | Refer to Table 5 & Table 6<br>The material of handpiece enclosure is different from Sonic Surgeon 300.<br>Other materials are same as predicate device. | | | | |
| Materials | | | | | |
| Performance parameters | | | | | |
| LED Light Control | X | O | O | O | Table.2 |
| Output Power | MAX 40W | 15 ~ 50W | 15 ~ 70W | 15 ~ 90W | Table.3 |
| Irrigation | 10~90 ml/min. | 7~110 ml/min. | 7~110 ml/min. | 7~110 ml/min. | Table.4 |
{6}------------------------------------------------
(2) Comparing 'output Power' between proposed Device and predicate model 'UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)'
| | Predicate Device | Proposed Device | | | remark |
|-------------------------------|---------------------------------|-------------------------|-----------------------|-----------------------|---------|
| Manufacturer | ITALIA MEDICA<br>S.R.L | DONG IL TECHNOLOGY LTD. | | | |
| Model Name | Model Ultrasonic<br>Debridement | Sonic Surgeon<br>310L | Sonic Surgeon<br>600L | Sonic Surgeon<br>800L | |
| <b>Performance parameters</b> | | | | | |
| Output<br>Power | 90 W | 30~50 W | 50~70 W | 70~90 W | Table.3 |
### Table.1 Tips (Sonic Surgeon 300/310L/600L/800L)
| No. | Sonic Surgeon 300 / 310L | | Sonic Surgeon 600L / 800L | |
|------|--------------------------|------------------|---------------------------|-----------------|
| Name | Size | Name | Size | |
| 1 | DG3-001 | Ø 4.5x28.26mm(L) | BTG-B017 | Ø 5.5x28.5mm(L) |
| 2 | DG3-002 | Ø 4.5×30.51mm(L) | BTG-R018 | Ø 5.5x29.5mm(L) |
| 3 | DM3-003 | Ø 4.5x31.86mm(L) | BTM-D019 | Ø 5.5x31.5mm(L) |
| 4 | DE3-004 | Ø 4.5×28.29mm(L) | BTE-K020 | Ø 5.5x29.0mm(L) |
| 5 | DH6-005 | Ø 4.5x30.75mm(L) | BTH-S021 | Ø 5.5x30.1mm(L) |
| 6 | DD6-016 | Ø 4.5x27.7mm(L) | BTD-F022 | Ø 5.5x27.5mm(L) |
| 7 | DD6-028i | Ø 4.5x28.2mm(L) | BTD-F023 | Ø 5.5x28.5mm(L) |
| 8 | DC6-Saw | Ø 4.5x31.5mm(L) | BTC-I024 | Ø 5.5x31.5mm(L) |
| 9 | DC6-Saw L | Ø 4.5x30.9mm(L) | BTC-I025 | Ø 5.5×31.0mm(L) |
| 10 | DC6-Saw R | Ø 4.5x30.9mm(L) | BTC-I026 | Ø 5.5x31.0mm(L) |
### Table.2 LED Brightness
| LED<br>level | LED brightness (lux) | | | Remark |
|--------------|----------------------|--------------------|--------------------|-------------------|
| | Sonic Surgeon 310L | Sonic Surgeon 600L | Sonic Surgeon 800L | |
| 1 | $1,900\pm100$ | $2,000\pm100$ | $2,000\pm100$ | |
| 2 | $2,400\pm100$ | $2,500\pm100$ | $2,500\pm100$ | Distance:<br>20mm |
| 3 | $2,600\pm100$ | $2,800\pm100$ | $2,800\pm100$ | |
{7}------------------------------------------------
## Table.3 Output Power
| [Unit: W] | | | | | | |
|--------------------|-------------|--------|--------|------------|--------|--------|
| Model Name | Normal Mode | | | Boost Mode | | |
| | Level1 | Level2 | Level3 | Level1 | Level2 | Level3 |
| Sonic Surgeon 310L | 15~17 | 17~21 | 21~27 | 30~50 | 30~50 | 30~50 |
| Sonic Surgeon 600L | 15~17 | 17~21 | 21~27 | 50~70 | 50~70 | 50~70 |
| Sonic Surgeon 800L | 15~17 | 17~21 | 21~27 | 70~90 | 70~90 | 70~90 |
## Table.4 Water flow
| Pump level | Water flow (ml/min approximately) | | Remark |
|------------|-----------------------------------|------------------------------|--------|
| | Sonic Surgeon 300 | Sonic Surgeon 310L/600L/800L | |
| 1 | 10 | 7 ~ 13 | |
| 2 | 30 | 15 ~ 40 | |
| 3 | 50 | 40 ~ 60 | |
| 4 | 70 | 60 ~ 80 | |
| 5 | 90 | 80 ~ 110 | |
# Table.5 Materials (Sonic Surgeon 300)
| No | Name | Materials | Company | CAS No. | Grade |
|----|----------------------------------|-----------------------------------------------------------------------------|----------------------------------------|-------------------------------------------|---------------------|
| 1 | Enclosure<br>(Generator<br>Unit) | PC(Polycarbonate)<br>ABS(Acrylonitrile-<br>Butadiene-Styrene<br>Terpolymer) | LG Chemical | PC<br>(103598-77-2)<br>ABS<br>(9003-56-9) | LUPOY<br>GN5001RF |
| 2 | Hose, and<br>seals | Silicone | Elkem | 7440-21-3 | Synthetic<br>rubber |
| 3 | Enclosure<br>(Hand Piece) | PES<br>(Polyethersulfone) | BASF | 25608-63-3 | Ultrason<br>E2010 |
| 4 | Tips | TrimRite stainless | CARPENTER | - | UNS S42010 |
| 5 | Tip Coating | ZrN<br>(Zirconium Nitride)<br>ZrN+Diamond | Chemetall<br>(Zirconium)<br>ProSciTech | 25658-42-8<br>Diamond<br>(77820-40-3) | CA<br>M24 |
{8}------------------------------------------------
| Table.6 Materials (Sonic Surgeon 310L/600L/800L) | | | | | |
|--------------------------------------------------|----------------------------------|-----------------------------------------------------------------------------|----------------------------------------|------------------------------------------------------------------------------------------------------|---------------------|
| No | Name | Materials | Company | CAS No. | Grade |
| 1 | Enclosure<br>(Generator<br>Unit) | PC(Polycarbonate)<br>ABS(Acrylonitrile-<br>Butadiene-Styrene<br>Terpolymer) | LG Chemical | PC<br>(103598-77-2)<br>ABS<br>(9003-56-9) | LUPOY<br>GN5001RF |
| 2 | Hose, and<br>seals | Silicone | Elkem | 7440-21-3 | Synthetic<br>rubber |
| 3 | Enclosure<br>(Hand Piece) | PPSU<br>(Polyphenylsulfone) | Solvay | Polyphenylsulfone<br>(25608-64-4)<br>Titanium dioxide<br>(13463-67-7)<br>Carbon black<br>(1333-86-4) | Duradex<br>D-3000 |
| 4 | Tips | TrimRite stainless | CARPENTER | - | UNS S42010 |
| 5 | Tip Coating | ZrN<br>(Zirconium Nitride)<br>ZrN+Diamond | Chemetall<br>(Zirconium)<br>ProSciTech | 25658-42-8<br>Diamond<br>(77820-40-3) | CA<br>M24 |
#### H. Performance Information
The Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L have been manufactured and tested to meet the safety requirements of several IEC standards. The Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L comply with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for safety and IEC 60601-1-2:2007, Medical electrical equipment - Part 1: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests.
#### l. Conclusion:
The performance tests and evidence demonstrate that Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L function in a substantially equivalent manner to the predicate device.
In terms of output power, Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are substantially equivalent to the predicate device UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220). The maximum output power of Sonic Surgeon 800L is 70~90W and that of UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)' is 90W. In terms of light power, the capacity of the device's LED light is demonstrated in 'Biocompatibility' section and an LED report was attached as evidence.
Because our proposed device has the same intended use, the same sterilization method, same materials used, same electrical input power, same frequency, and same power output as predicate devices, as well as similarity of product configuration and administration, it can be concluded the Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.