K150069 · Acuitas Medical , Ltd. · LLZ · May 22, 2015 · Radiology
Device Facts
Record ID
K150069
Device Name
Fine Structure Analysis(fineSA)
Applicant
Acuitas Medical , Ltd.
Product Code
LLZ · Radiology
Decision Date
May 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
fineSA software is intended for use as a non-invasive device that processes designated MR images using the 2-D pulse sequence incorporated on Siemens Medical 3T MAGNETOM systems running syngo MR software versions B17 and D13. This includes the Aera, Skyra, Avanto and Verio models. fineSA software generates reports that provide information about the structural features in bone. The MR data, in the form of spectra or spectral images, reflect the spacing of structural features at a scale of approximately 500 microns or larger.
Device Story
fineSA is a software-only image processing system for characterizing trabecular bone microarchitecture. It processes 1D MR data acquired via standard 2D pulse sequences on Siemens 3T MAGNETOM scanners. Input data is transferred via DICOM to a networked PC running the software. The device uses Fourier and inverse Fourier transformations to analyze spectral data, identifying repetitive structural features in bone at a scale of 500 microns or larger. It generates a PDF report, which is sent back to a DICOM destination for clinician review. Used in radiology environments or remote locations with secure network access. The output assists clinicians in evaluating bone structural features; it does not provide chemical information. The device benefits patients by providing non-invasive, quantitative insights into bone microarchitecture.
Clinical Evidence
Bench testing only. Software verification and validation were performed according to FDA guidance for a 'moderate' level of concern. Testing utilized phantom and clinical data to confirm the software meets design specifications and intended use requirements.
Technological Characteristics
Software-only image processing system; operates on networked PCs. Inputs: DICOM MR data (spin echo/gradient echo). Processing: Fourier and inverse Fourier transformations, windowing. Connectivity: DICOM standard. Deployment: Radiology internal network or remote secure access. Software level of concern: Moderate.
Indications for Use
Indicated for patients undergoing MRI of bone structures on Siemens 3T MAGNETOM systems to provide information regarding structural feature spacing (≥500 microns).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 22, 2015
Acuitas Medical Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004
Re: K150069
Trade/Device Name: Fine Structure Analysis (fineSA) Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, LNH Dated: April 22, 2015 Received: April 22, 2015
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K150069
Device Name
## Fine Structure Analysis (fineSA) Software
Indications for Use (Describe)
fineSA software is intended for use as a non-invasive device that processes designated MR images using the 2-D pulse sequence incorporated on Siemens Medical 3T MAGNETOM systems running syngo MR software versions B17 and D13. This includes the Aera, Skyra, Avanto and Verio models.
fineSA software generates reports that provide information about the structural features in bone. The MR data, in the form of spectra or spectral images, reflect the spacing of structural features at a scale of approximately 500 microns or larger.
Type of Use (Select one or both, as applicable)Z Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Acuitas Medical. The word "ACUITAS" is written in purple, with a small green square on top of the "I". The word "medical" is written in green and is smaller than the word "ACUITAS". The logo is simple and modern.
Fine Structure Analysis (fineSA)
#### 510(K) SUMMARY (K150069) AS REQUIRED BY 21 CFR 807.92(c)
#### 1. Submitter
Acuitas Medical Ltd. 12 Technium One, Kings Road Swansea Waterfront SA1 8PH Wales UK
Phone: (970) 472-0613 Fax: (970) 416-8896
Contact Person: John Heinrich, Ph.D. Chief Executive Officer Phone: (970) 472-0613 Fax: (970) 416-8896 Email: jheinrich@acuitasmedical.com
# 2. Date Prepared
May 20, 2015
## 3. Device
- . Trade Name: Fine Structure Analysis (fineSA)
- Common Name: System, Image Processing, Radiological ●
- Classification Name: Picture Archiving and Communication System (21 CFR 892.2050) ●
- Regulatory Class: II ●
- . Product Code: LLZ, LNH
## 4. Predicate Device
The predicate device for the Fine Structure Analysis (fineSA) software is the syngo.MR General, which was cleared under 510(k) K130749.
{4}------------------------------------------------
# 5. Device Description
fineSA is a tool for the characterization of trabecular bone microarchitecture using magnetic resonance data. fineSA analyzes data acquired from a region of interest with a 2-D pulse sequence. The result is a spectrum that shows not chemical information, but the average spacing of the structural elements within the region of interest. fineSA provides the ability to identify repetitive structural features as small as 500 um.
Compatible MRI scanners, acquire one-dimensional data along the length of a defined region of interest in the anatomical region using a standard pulse sequence available on the OEM scanner. The additional acquisition time required to collect the fineSA compatible data is less than 2 minutes. The resulting DICOM data object is transferred to a network personal computer loaded with fineSA software for processing and report generation.
The DICOM input data is received from the MRI scanner via standard DICOM protocols. The processing performed is driven by the received data content, and at the completion of data processing and analysis, a PDF formatted report is generated and transferred to a designated DICOM destination as a DICOM encapsulated PDF document for viewing and interpretation.
fineSA is a software only image processing system that can be installed on any network computer that meets the specified minimum hardware and operating system requirements. The system can be deployed on the radiology internal network or at a remote location with secure access to radiology internal network.
## 6. Indications for Use
fineSA software is intended for use as a non-invasive device that processes designated MR images using the 2-D pulse sequence incorporated on Siemens Medical 3T MAGNETOM systems running syngo MR software versions B17 and D13. This includes the Aera, Skyra, Avanto and Verio models.
fineSA software generates reports that provide information about the structural features in bone. The MR data, in the form of spectral images, reflect the spacing of structural features at a scale of approximately 500 microns or larger.
## 7. Predicate Device Comparison
The fineSA software is substantially equivalent to the syngo.MR General post-processing application (K130749).
A summary comparison of characteristics is provided in Table 1.
{5}------------------------------------------------
Fine Structure Analysis (fineSA)
| Characteristic | Fine Structure Analysis (fineSA) | PREDICATE<br>syngo.MR General |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance | K150069 | K130749 |
| Imaging Modality | MRI | MRI |
| Product Code | LLZ | LLZ |
| Intended Use | fineSA software is intended for use as a non-<br>invasive device that processes designated MR<br>images using the 2-D pulse sequence incorporated<br>on Siemens Medical 3T MAGNETOM systems<br>running syngo MR software versions B17 and D13.<br>This includes the Aera, Skyra, Avanto and Verio<br>models.<br><br>fineSA software generates reports that provide<br>information about the structural features in bone.<br>The MR data, in the form of spectra or spectral<br>images, reflect the spacing of structural features at a<br>scale of approximately 500 microns or larger. | Post processing software / applications to be used<br>for viewing and evaluating the designated images<br>provided by a magnetic resonance device.<br><br>syngo.MR General is a syngo based post-processing<br>software for viewing, manipulating, and evaluating<br>MR images. |
| Software Device | Yes | Yes |
| Patient Contact | No | No |
| Graphical User<br>Interface (GUI) | No | Yes |
| Data Interchange | DICOM | DICOM |
| Data Acquisition | Data acquired for fineSA processing uses spin echo<br>and gradient echo sequences. | Data acquired uses spin echo and gradient echo<br>sequences. |
| Data Analysis | Data analyzed using standard data analysis<br>techniques including Fourier and inverse Fourier<br>transformations and windowing. | Data analyzed using standard data analysis<br>techniques including Fourier and inverse Fourier<br>transformations and windowing. |
| Data Presentation | Data presented in conjunction with MR images. | Data presented as or in conjunction with MR<br>images. |
| Data Types | Spacings of anatomical elements in bone. | Various, including anatomical, relaxation time,<br>diffusion rate, spectroscopic. and chemical shift. |
#### Table 1 - fineSA & Predicates Comparative Analysis
## 8. Performance Data
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by the FDA's Guidance for Industry and FDA Staff, "Guidance for Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "moderate" level of concern, since a malfunction or latent flaw in the software could delay delivery of appropriate medical care that would likely lead to minor injury.
The fineSA processing software was fully tested, verified and validated by Acuitas as part of its internal design control requirements to ensure it meets defined specifications using phantom and clinical data. Formal test plans were executed to confirm that fineSA meets its formal requirements. This submission includes the verification and validation plan, test results, and verification report confirming the fineSA meets its intended use and performance requirements.
{6}------------------------------------------------
# 9. Risk Management
Product risk management activities were performed in accordance with ISO 14971:2007 throughout the product development process. Risk management verification consisted of both a desk audit and software testing to ensure the implementation of all risk mitigations for the device.
# 10. Substantial Equivalence Conclusion
As noted previously, the fineSA and its predicate devices have similar intended use for viewing and evaluating the designated images provided by a magnetic resonance device. While there are some differences between fineSA and its predicates, these differences are minor and do not raise different questions of safety and effectiveness. Performance data and software verification and validation demonstrate that the fineSA software should perform as intended in the specified use conditions.
Therefore, Acuitas Medical believes that the fineSA software is substantially equivalent to the predicate devices identified in this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.